← Product Code [OHT](/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT) · K250131

# IPL Hair Removal Device (M8011, M8012, M8013, M8015, M8021, M8022, M8023, M8025, M7011, M7012, M7021, M7023) (K250131)

_Mustech Electronics Co., Limited · OHT · Mar 19, 2025 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT/K250131

## Device Facts

- **Applicant:** Mustech Electronics Co., Limited
- **Product Code:** [OHT](/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT.md)
- **Decision Date:** Mar 19, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

IPL Hair Removal Device is an over-the-counter device intended for removal of unwanted body hair and/or facial hair.

## Device Story

Hand-held, over-the-counter IPL device for home use; reduces unwanted body and facial hair. Uses Xenon Arc Flashlamp to emit light pulses (530nm-1200nm) below skin surface. Includes skin contact sensor to prevent light emission unless in full contact with skin. Models M80XX include cooling function; M70XX do not. Powered by external adapter. Operated by user; device output (light pulses) targets hair follicles to inhibit growth. Benefits include hair reduction without cutting or pulling, with minimal pain.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including biocompatibility (ISO 10993-5, -10, -23), electrical safety (IEC 60601-1, -1-2, -1-11, -2-83), photobiological safety (IEC 62471), and usability (IEC 60601-1-6).

## Technological Characteristics

Xenon Arc Flashlamp light source; 530nm-1200nm wavelength; 1.8-4.5 J/cm² energy density; 3.3-3.4 cm² spot size. Materials: ABS+PC, Sapphire (Al2O3), POM, Silicone. Powered by external 100-240V adapter. Microprocessor-controlled pulse emission with skin contact sensor. Not sterile.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Hair Removal Device (R2815-G Pro, R2815-G) ([K232575](/device/K232575.md))
- Intense Pulsed Light (IPL) System ([K223928](/device/K223928.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

March 19, 2025

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".

Mustech Electronics Co., Limited % Riley Chen Registration Engineer Feiying Drug & Medical Consulting Technical Service Group Rm 2401 Zhenye International Business Center, No. 3101-90 Oianhai Road Shenzhen, Guangdong 518052 China

Re: K250131

Trade/Device Name: IPL Hair Removal Device (M8011, M8013, M8015, M8021, M8022, M8023, M8025, M7011, M7012, M7021, M7023) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHT Dated: January 17, 2025 Received: January 17, 2025

Dear Riley Chen:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/2/Picture/3 description: The image shows a digital signature. The signature is for "YAN FU -S". The date of the signature is March 19, 2025, and the time is 12:42:24. The time zone is -04'00'.

for Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K250131

Device Name

IPL Hair Removal Device (M8011, M8013, M8015, M8021, M8023, M8025, M7011, M7012, M7021, M7023)

Indications for Use (Describe)

IPL Hair Removal Device is an over-the-counter device intended for removal of unwanted body hair and/or facial hair.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# 510(k) Summary #K250131

### "510(k) Summary" as required by 21 CFR Part 807.92.

#### I. Submitter

Mustech Electronics Co., Limited Address: 3F, Building No.B2, Shanghe industry Park, NanChang Road, Xixiang Town, Baoan District, Shenzhen, Guangdong, China Post code: 518126 Fax: +86-0755-27670391

Kevin Liu Title: Project Supervisor Tel.: +86-0755-27967595-808 Email: kevin@mustech.com

Date of preparation: 2025-01-13

#### II. Device

Name of Device: IPL Hair Removal Device Model(s): M8011, M8012, M8015, M8021, M8022, M8025, M7011, M7012, M7021, M7023 Common or Usual Name: Light Based Over-The-Counter Hair Removal Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: OHT Regulation Number: 21 CFR 878.4810

#### III. Predicate Device

| Predicate Device                              | Manufacturer                                          | 510(k) Number | Product code |
|-----------------------------------------------|-------------------------------------------------------|---------------|--------------|
| Hair Removal Device<br>(R2815-G Pro, R2815-G) | Shenzhen Jianchao Intelligent<br>Technology Co., Ltd. | K232575       | OHT          |
| Intense Pulsed Light (IPL)<br>System          | Shenzhen Fansizhe Science<br>and Technology Co., Ltd  | K223928       | OHT          |

#### IV. Device Description

IPL Hair Removal Device is an over-the-counter, home use, light based device for unwanted hair reduction by using Intense Pulsed Light (IPL) technology. The device works below the skin's surface and does not involve any cutting or pulling, removing hair growth with minimal pain.

The device includes 12 models, they are M8011, M8013, M8015, M8021, M8022, M8023, M8025, M7011, M7012, M7021, M7023. Those models with model name of "M80XX" are equipped 

{5}------------------------------------------------

with cooling function for a better hair remove experience, while those of "M70XX" have no cooling function.

The device is only powered by the external power adapter and its IPL emission activation is by a switch or auto light emission, and it contains a skin sensor to detect appropriate skin contact, if the light outlet is not in full contact with the skin, the device cannot emit light pulses.

### V. Indications for Use

IPL Hair Removal Device is an over-the-counter device intended for removal of unwanted body hair and/or facial hair.

Comparison of Indications for Use: The indications for use of the subject device is comparable to the indications for use of the predicate devices.

### VI. Comparison of Technological Characteristics With the Predicate Device

The IPL Hair Removal Device has the same indications for use, mode of action and similar operational characteristics as the predicate devices. Any minor differences between the subject device and the listed predicate devices do no raise any issues of safety or efficacy. Performance data supports that the device is safe and as effective as the predicate devices for its intended use. Therefore, the IPL Hair Removal Device may be found substantially equivalent to its predicate devices.

IPL Hair Removal Device is compared with the following predicate devices in terms of intended use, design, specifications and performance:

| Comparison<br>Elements                               | Subject Device                                                                                                                                               | Primary Predicate Device                                                                                                     | Secondary Predicate Device                                                                                                          | Remark                                                      |
|------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|
| 510(k) Number                                        | Pending                                                                                                                                                      | K232575                                                                                                                      | K223928                                                                                                                             | /                                                           |
| Trade name                                           | IPL Hair Removal Device                                                                                                                                      | Hair Removal Device                                                                                                          | Intense Pulsed Light (IPL)<br>System                                                                                                | /                                                           |
| Model                                                | M8011, M8012, M8013,<br>M8015, M8021, M8022,<br>M8023, M8025, M7011,<br>M7012, M7021, M7023                                                                  | R2815-G Pro, R2815-G                                                                                                         | T023K, T023A, T023B,<br>T023C, T023D, T023E,<br>T021K, T021A, T001A,<br>T001B, T001M, T001N,<br>T011C, T016K                        | /                                                           |
| Indication<br>for<br>use/<br>Intended<br>use         | IPL Hair Removal Device is<br>an over-the-counter device<br>intended for removal of<br>unwanted body hair and/or<br>facial hair.                             | Hair Removal Device is an<br>over-the-counter device<br>intended for removal of<br>unwanted body hair and/or<br>facial hair. | The<br>Intense Pulsed Light<br>(IPL) System is an over-the-<br>counter device intended for<br>the removal of unwanted<br>body hair. | Same                                                        |
| Prescription<br>or<br>OTC                            | OTC                                                                                                                                                          | OTC                                                                                                                          | OTC                                                                                                                                 | Same                                                        |
| Design                                               | Hand-held                                                                                                                                                    | Hand-held                                                                                                                    | Hand-held                                                                                                                           | Same                                                        |
| Comparison<br>Elements                               | Subject Device                                                                                                                                               | Primary Predicate Device                                                                                                     | Secondary Predicate Device                                                                                                          | Remark                                                      |
| Power source                                         | An external power adapter                                                                                                                                    | An external power supply                                                                                                     | An external power supply                                                                                                            | Same                                                        |
| Power supply                                         | 100-240V, 50/60Hz                                                                                                                                            | 100~240V, 50/60Hz                                                                                                            | /                                                                                                                                   | Same                                                        |
| Dimensions                                           | M7011,<br>M7012,<br>M8011,<br>M8012,<br>M8013,<br>M8015:<br>179mm x 65mm x 39mm<br>M7021,<br>M7023,<br>M8021,<br>M8022, M8023, M8025:<br>180mm x 71mm x 39mm | 194*117*45mm                                                                                                                 | 90*44*225mm for T016K                                                                                                               | Different                                                   |
| Sterilization                                        | Not required                                                                                                                                                 | Not required                                                                                                                 | Not required                                                                                                                        | Same                                                        |
| Light source                                         | Intense Pulsed Light (IPL)                                                                                                                                   | Intense Pulsed Light (IPL)                                                                                                   | Intense Pulsed Light (IPL)                                                                                                          | Same                                                        |
| Energy medium                                        | Xenon Arc Flashlamp                                                                                                                                          | Xenon Arc Flashlamp                                                                                                          | Xenon Arc Flashlamp                                                                                                                 | Same                                                        |
| Wavelength<br>range                                  | 530nm-1200nm                                                                                                                                                 | 530-1200nm                                                                                                                   | 510nm~1200nm                                                                                                                        | Same                                                        |
| Energy density                                       | M8011,<br>M8012,<br>M8013,<br>M8015,<br>M8022,<br>M8021,<br>M8023, M8025: 1.8~4.0J/cm²<br>M7011, M7012, M7021,<br>M7023: 2.1~4.5J/cm²                        | Max. 5.2J/cm2<br>(range in 1.2J/cm2~5.2J/cm2)                                                                                | Max. 5.73J/cm² for T016K                                                                                                            | Similar                                                     |
| Output energy                                        | M8011,<br>M8012,<br>M8013,<br>M8015,<br>M8021,<br>M8022,<br>M8023, M8025: 6.0~13.5J<br>M7011,<br>M7012,<br>M7021,<br>M7023: 7.0~15.0J                        | 5~14.5J                                                                                                                      | 4.8J~18.9J for T016K                                                                                                                | Similar                                                     |
| Spot size                                            | M8011,<br>M8012,<br>M8013,<br>M8015,<br>M8021,<br>M8022,<br>M8023, M8025: 3.4 cm²<br>M7011,<br>M7012,<br>M7021,<br>M7023: 3.3cm²                             | 3.4±0.25 cm²                                                                                                                 | 3.3cm² for T016K                                                                                                                    | Same                                                        |
| Pulse duration                                       | 8.2~11.0ms                                                                                                                                                   | 8.5±2.5ms<br>(range in 6ms~11ms)                                                                                             | 4~12ms                                                                                                                              | Similar                                                     |
| Pulsing control                                      | Finger switch                                                                                                                                                | Finger switch                                                                                                                | Finger switch                                                                                                                       | Same                                                        |
| Delivery device                                      | Direct illumination to tissue                                                                                                                                | Direct illumination to tissue                                                                                                | Direct illumination to tissue                                                                                                       | Same                                                        |
| Software/<br>Firmware/<br>Microprocessor<br>Control? | Yes                                                                                                                                                          | Yes                                                                                                                          | Yes                                                                                                                                 | Same                                                        |
| Comparison<br>Elements                               | Subject Device                                                                                                                                               | Primary Predicate Device                                                                                                     | Secondary Predicate Device                                                                                                          | Remark                                                      |
| Material of<br>Patient contact<br>components         | ABS+PC, Sapphire (Al2O3),<br>Reflective cover bracket<br>(POM), Silicone stopper<br>(Silica gel)                                                             | PC+ABS, Sapphire<br>(Al2O3), Stainless steel 430                                                                             | T016K: PC for button, PC for<br>main housing and Probe<br>cover, Sapphire Crystal<br>Cooling Compress                               | Different,<br>but<br>solved by<br>biocompatibility<br>tests |

{6}------------------------------------------------

{7}------------------------------------------------

## VII. Non-Clinical Testing

The following performance data were provided in support of the substantial equivalence determination.

- > ISO 10993-5: 2009, Biological evaluation of medical devices -Par t 5: Tests for in vitro cytotoxicity
- > ISO 10993-10: 2021, Biological evaluation of medical devices Part 10: Tests for skin sensitization
- > ISO 10993-23: 2021, Biological evaluation of medical devices Part 23: Tests for irritation

> IEC 60601-1:2020, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance

> IEC 60601-1-2:2020. Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: electromagnetic compatibility

> IEC 60601-1-11:2020, Medical Electrical Equipment - Part 1: General Requirements for Basic

Safety and Essential Performance - Collateral Standard: Requirements for Medical Electrical

Equipment and Medical Electrical Systems Used in the Home Healthcare Environment

> IEC 60601-2-83: 2022, Medical electrical equipment - Part 2-83: Particular requirements for the basic safety and essential performance of home light therapy equipment

> IEC 62471: 2006, Photobiological safety of lamps and lamp systems

> IEC 60601-1-6: 2020, Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability

#### VIII. Clinical Testing

Not applicable.

#### IX. Conclusions

Based on the above analysis and non-clinical tests performed, it can be concluded that the subject device is as safe, as effective, and performs as well as the legally marketed predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT/K250131](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT/K250131)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
