← Product Code [OHT](/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT) · K243036

# IPL Hair Removal Device (BHRL-01, BHRL-02, BHRL-03, BHRL-04, BHRL-05, BHRL-06, BHRL-06B, BHRL-07, BHRL-08, BHRL-09, BHRL-10, BHRL-11, BHRL-12, BHRL-13, BHRL-14, BHRL-15, BHRL-16, BHRL-17, BHRL-18, BHRL-19, BHRL-20, BHRL-21, BHRL-22, BHRL-23, BHRL-24, BHRL-25, BHRL-26, BHRL-27, BHRL-28, BHRL-29) (K243036)

_Bestway Plastic and Metal Products , Ltd. · OHT · Jun 15, 2025 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT/K243036

## Device Facts

- **Applicant:** Bestway Plastic and Metal Products , Ltd.
- **Product Code:** [OHT](/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT.md)
- **Decision Date:** Jun 15, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

IPL Hair Removal Device is indicated for the removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.

## Device Story

Over-the-counter, home-use IPL hair removal device; 23 models (BHRL-01 through BHRL-22). Device uses Xenon Arc lamp to emit intense pulsed light; energy density 1.2-4.1 J/cm²; wavelength 510-1200nm. User operates via finger switch; device requires full skin contact for pulse emission via integrated skin sensor. Most models include cooling function to reduce treatment area temperature. Powered by external AC adapter. Output affects hair follicles below skin surface to reduce regrowth; non-invasive. Healthcare provider does not operate; patient self-use. Benefits include hair reduction with minimal pain.

## Clinical Evidence

Bench testing only. No clinical data provided. Evidence includes biocompatibility (ISO 10993-1), electrical safety/EMC (IEC 60601-1, -1-2, -1-11, -2-83), photobiological eye safety (IEC 62471), software verification/validation, and human factors/usability engineering.

## Technological Characteristics

Materials: ABS, aluminum alloy. Energy: Xenon Arc lamp (IPL). Wavelength: 510-1200±15nm. Energy density: 1.2-4.1 J/cm². Spot size: 2.7-3.6 cm². Pulse duration: 6.4±2.0ms. Connectivity: None (standalone). Power: 100-240V AC adapter. Safety: Integrated skin contact sensor, cooling function (except BHRL-06B). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-83, IEC 62471.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Hand-held IPL device (IPL Home Use Hair Removal Device) Model: FZ-608, FZ-608G, FZ-100, FZ-200 ([K230739](/device/K230739.md))

## Reference Devices

- IPL Hair Removal Device Model(s): BSXT101, BSXT102, BSXT103, BSXT105, BSXT106, BSXT108 ([K230097](/device/K230097.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

June 15, 2025

Bestway Plastic And Metal Products Ltd.
% Candice Qiu
Registration Specialist
Feiying Drug &amp; Medical Consulting Technical Service Group
Rm 2401 Zhenye International Business Center, No. 3101-90,
Qianhai Road
Shenzhen, Guangdong 518052
China

Re: K243036

Trade/Device Name: IPL Hair Removal Device (BHRL-01, BHRL-02, BHRL-03, BHRL-04, BHRL-05, BHRL-06, BHRL-06B, BHRL-07, BHRL-08, BHRL-09, BHRL-10, BHRL-11, BHRL-12, BHRL-13, BHRL-14, BHRL-15, BHRL-16, BHRL-17, BHRL-18, BHRL-19, BHRL-20, BHRL-21, BHRL-22.

Regulation Number: 21 CFR 878.4810

Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology

Regulatory Class: Class II

Product Code: OHT

Dated: May 23, 2025

Received: May 23, 2025

Dear Candice Qiu:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K243036 - Candice Qiu
Page 2

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

{2}

K243036 - Candice Qiu
Page 3

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

TANISHA L. HITHE -S
Digitally signed by
TANISHA L. HITHE -S
Date: 2025.06.15
19:58:49 -04'00'

Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Submission Number (if known)

K243036

Device Name

IPL Hair Removal Device (BHRL-01, BHRL-02, BHRL-03, BHRL-04, BHRL-05, BHRL-06, BHRL-06B, BHRL-07, BHRL-08, BHRL-09, BHRL-10, BHRL-11, BHRL-12, BHRL-13, BHRL-14, BHRL-15, BHRL-16, BHRL-17, BHRL-18, BHRL-19, BHRL-20, BHRL-21, BHRL-22)

Indications for Use (Describe)

IPL Hair Removal Device is indicated for the removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.

Type of Use (Select one or both, as applicable)

☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

&gt; "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}

Bestway Plastic And Metal Products Ltd.
510(k)s -510(k) Summary

# 510(k) Summary

# K243036

"510(k) Summary" as required by 21 CFR Part 807.92.

June 6, 2025

# I. Submitter

Bestway Plastic And Metal Products Ltd.
Room 701, Building 2, No.1 Chuangye San Road, Mapu Ao, Fenggang Town, Dongguan City, Guangdong Province, China
Jonah Ci
General Manager
Tel: +86 13412459592
Email: Jonah@pm-bestways.com

# II. Device

Name of Device: IPL Hair Removal Device
Model(s): BHRL-01, BHRL-02, BHRL-03, BHRL-04, BHRL-05, BHRL-06, BHRL-06B, BHRL-07, BHRL-08, BHRL-09, BHRL-10, BHRL-11, BHRL-12, BHRL-13, BHRL-14, BHRL-15, BHRL-16, BHRL-17, BHRL-18, BHRL-19, BHRL-20, BHRL-21, BHRL-22.
Common or Usual Name: Light Based Over-The-Counter Hair Removal
Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology
Regulatory Class: II
Product Code: OHT
Regulation Number: 21 CFR 878.4810

# III. Predicate Device and Reference Device

Predicate device:

|  Manufacturer | Predicate Device | 510(k) Number | Approval Date  |
| --- | --- | --- | --- |
|  Shenzhen IONKA Medical Technology Co., Ltd. | Hand-held IPL device (IPL Home Use Hair Removal Device) Model: FZ-608, FZ-608G, FZ-100, FZ-200 | K230739 | May 26, 2023  |

Reference device:

|  Manufacturer | Reference Device | 510(k) Number | Approval Date  |
| --- | --- | --- | --- |
|  Shenzhen BSX Technology Electronics Co., Ltd. | IPL Hair Removal Device Model(s): BSXT101, BSXT102, BSXT103, | K230097 | April 6, 2023  |

{5}

Bestway Plastic And Metal Products Ltd. 510(k)s -510(k) Summary

|  Manufacturer | Reference Device | 510(k) Number | Approval Date  |
| --- | --- | --- | --- |
|   | BSXT105, BSXT106, BSXT108 |  |   |

## IV. Device Description

IPL Hair Removal Device, is an over-the-counter, home-use device for unwanted hair reduction by using Intense Pulsed Light (IPL), and it has been designed twenty-three models with the same IPL technology for hair removal, which is model BHRL-01, BHRL-02, BHRL-03, BHRL-04, BHRL-05, BHRL-06, BHRL-06B, BHRL-07, BHRL-08, BHRL-09, BHRL-10, BHRL-11, BHRL-12, BHRL-13, BHRL-14, BHRL-15, BHRL-16, BHRL-17, BHRL-18, BHRL-19, BHRL-20, BHRL-21, BHRL-22. The device works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain.

The device is only powered by the external power adapter and its IPL emission activation is by finger switch.

The IPL Hair Removal Device has an irreplaceable light exit and it can cover an area of  $3.6 \pm 0.25 \mathrm{~cm}^2$  (Model BHRL-01, BHRL-02, BHRL-03, BHRL-04, BHRL-07, BHRL-08, BHRL-09, BHRL-10) and  $2.7 \pm 0.25 \mathrm{~cm}^2$  (Model BHRL-05, BHRL-11, BHRL-12, BHRL-13, BHRL-14, BHRL-15, BHRL-16, BHRL-17, BHRL-06, BHRL-06B, BHRL-18, BHRL-19, BHRL-20, BHRL-21, BHRL-22) that is suitable for multiple hair removal areas, such as large areas (e.g. face, arms, legs) and small areas (e.g. armpits, bikini line).

The device contains a skin sensor to detect appropriate skin contact, if the light exit is not in full contact with the skin, the device cannot emit the treatment light pulses. Besides, the IPL Hair Removal Device has the cooling function (except model: BHRL-06B), which can be activated throughout the whole hair removal process to cool down the treatment area's temperature and provide the user with a better using experience.

## V. Indications for Use

IPL Hair Removal Device is indicated for the removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.

## VI. Materials

|  Component name | Material of Component | Body Contact Category | Contact Duration  |
| --- | --- | --- | --- |
|  IPL Hair Removal Device (Enclosure) | ABS, aluminum alloy | Surface-contacting device: Intact skin | Less than 24 hours  |

{6}

Bestway Plastic And Metal Products Ltd.
510(k)s -510(k) Summary

# VII. Comparison of Technological Characteristics With the Predicate Device

The IPL Hair Removal Device has the same intended use, mode of action and similar operational characteristics as the predicate device. Any minor differences between the subject device and the listed predicate device and reference device do not raise any issues of safety or efficacy. Performance data supports that the device is safe and as effective as the predicate device and reference device for its intended use. Therefore, the IPL Hair Removal Device may be found substantially equivalent to its predicate device and reference device.

IPL Hair Removal Device is compared with the following Predicate Device and Reference Device in terms of intended use, design, material, specifications, and performance:

|  Comparison Elements | Subject Device | Predicate Device | Reference Device 1 | Remark  |
| --- | --- | --- | --- | --- |
|  510(k) Number | Pending | K230739 | K230097 | /  |
|  Trade name | IPL Hair Removal Device Model: BHRL-01, BHRL-02, BHRL-03, BHRL-04, BHRL-05, BHRL-06, BHRL-06B,BHRL-07, BHRL-08, BHRL-09, BHRL-10, BHRL-11, BHRL-12, BHRL-13, BHRL-14, BHRL-15, BHRL-16, BHRL-17, BHRL-18, BHRL-19, BHRL-20, BHRL-21, BHRL-22 | Hand-held IPL device (IPL Home Use Hair Removal Device) Model: FZ-608, FZ-608G, FZ-100, FZ-200 | IPL Hair Removal Device Model(s): BSXT101, BSXT102, BSXT103, BSXT105, BSXT106, BSXT108 | /  |
|  Manufacturer | Bestway Plastic And Metal Products Ltd. | Shenzhen IONKA Medical Technology Co., Ltd. | Shenzhen BSX Technology Electronics Co., Ltd. | /  |
|  Regulation number | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | Same  |
|  Product code | OHT | OHT | OHT | Same  |
|  Device classification | Class II | Class II | Class II | Same  |
|  Indication for use/ Intended use | IPL Hair Removal Device is indicated for the removal of unwanted hair. The device is also indicated for the permanent | IPL Home Use Hair Removal Device is indicated for the removal of unwanted hair. The device is also indicated for the | The IPL Hair Removal Device is an over-the-counter device intended for removal of unwanted body and/ or facial | Same  |

{7}

Bestway Plastic And Metal Products Ltd.
510(k)s -510(k) Summary

|  Comparison Elements | Subject Device | Predicate Device | Reference Device 1 | Remark  |
| --- | --- | --- | --- | --- |
|   | reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime. | permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime. | hair. |   |
|  Prescription or OTC | OTC | OTC | OTC | Same  |
|  Applicable skin | Fitzpatrick skin types I-V | Unknown | Fitzpatrick Skin Types I-V | Same  |
|  Dimension | BHRL-01, BHRL-02, BHRL-03, BHRL-04, BHRL-07, BHRL-08, BHRL-09, BHRL-10: 80*180*170mm
BHRL-05, BHRL-11, BHRL-12, BHRL-13, BHRL-14, BHRL-15, BHRL-16, BHRL-17: 70*170*40mm
BHRL-06, BHRL-06B, BHRL-18, BHRL-19, BHRL-20, BHRL-21, BHRL-22: 60*180*30mm | FZ-608, FZ-608G: 98*147*60(mm)
FZ-100: 198*71*44(mm)
FZ-200: 216*68*52(mm) | 190x70x45mm | Difference  |
|  Power supply | 100-240V, 50/60Hz, 0.8AMax | Unknown | 100-240V, 50/60Hz | Same  |
|  Sterilization | Not required | Not required | Not required | Same  |
|  Light source | Intense Pulsed Light | Intense Pulsed Light | Intense Pulsed Light | Same  |
|  Energy medium | Xenon Arc lamp | Xenon Arc lamp | Xenon Arc lamp | Same  |
|  Wavelength range | 510-1200±15nm | 550-1200nm | 470nm-1200nm | Similar  |
|  Energy density | 1.2-4.1 J/cm² | FZ-608, FZ-608G: 3.33 J/cm²
FZ-100: 5.43 J/cm²
FZ-200: 4.5 J/cm² | Max 5.0 J/cm² | Similar  |

Page 4 of 7

{8}

Bestway Plastic And Metal Products Ltd.
510(k)s -510(k) Summary

|  Comparison Elements | Subject Device | Predicate Device | Reference Device 1 | Remark  |
| --- | --- | --- | --- | --- |
|  Output energy | 4.5 ~ 10J (±20%) | FZ-608, FZ-608G:
Level 1: 4.16J
Level 2: 4.36J
Level 3: 5.1J
Level 4: 6.1J
Level 5: 6.96J
Level 6: 7.96J
Level 7: 8.63J
Level 8: 9.13J
Level 9: 10.0J | Pure mode: 5~10J (±20%)
Armpit mode: 6~10J (±20%)
Body mode: 6.5~11J (±20%)
Bikini mode: 8~12.5J (±20%) | Similar  |
|  Spot size | (1) Model BHRL-01, BHRL-02, BHRL-03, BHRL-04, BHRL-07, BHRL-08, BHRL-09, BHRL-10: 3.6±0.25cm²
(2) Model BHRL-05, BHRL-11, BHRL-12, BHRL-13, BHRL-14, BHRL-15, BHRL-16, BHRL-17, BHRL-06, BHRL-06B, BHRL-18, BHRL-19, BHRL-20, BHRL-21, BHRL-22: 2.7±0.25cm² | 3cm² | 3.0cm²± 0.5cm² | Similar  |
|  Pulse duration | 6.4±2.0ms | 0.5-0.8ms | 4-10ms | Similar  |
|  Pulsing control | Finger switch | Finger switch | Finger switch | Same  |
|  Output intensity level | 5 Levels |  | 5 Levels | Different  |
|  Software/ Firmware/ Microprocessor Control? | Yes | Yes | Yes | Same  |
|  Electrical safety | IEC 60601-1
IEC 60601-1-2
IEC 60601-1-11
IEC 60601-2-83 | IEC 60601-1
IEC 60601-1-2
IEC 60601-1-11
IEC 60601-2-83 | ANSI AAMI ES60601-1
IEC 60601-1-2
IEC 60601-1-11
IEC 60601-2-83 | Same  |
|  Eye safety | IEC 62471 | IEC 62471 | IEC 62471 | Same  |

Page 5 of 7

{9}

Bestway Plastic And Metal Products Ltd.
510(k)s –510(k) Summary

## VIII. Performance Data

The following performance data were provided in support of the substantial equivalence determination.

### 1) Biocompatibility Evaluation

- The biocompatibility evaluation for the body-contacting components of the IPL Hair Removal Device was conducted in accordance with the "Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices –Part 1: Evaluation and Testing Within a Risk Management Process, Document Issued on September 4, 2020", as recognized by FDA.

### 2) Electrical Safety and EMC

Electrical safety and EMC testing was performed to, and passed, as per the following standards:

- IEC 60601-1-2 Medical electrical equipment –Part 1-2: General requirements for basic safety and essential performance –Collateral standard: electromagnetic compatibility.
- IEC 60601-1 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.
- IEC 60601-1-11 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
- IEC 60601-2-83 Medical electrical equipment - Part 2-83: Particular requirements for the basic safety and essential performance of home light therapy equipment.

### 3) Eye Safety

- IEC 62471 Photobiological safety of lamps and lamp systems

### 4) Software Verification and Validation

Software documentation consistent with **basic documentation level** of concern was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels.

### 5) Usability

The product usability has been evaluated and verified according to the following FDA guidance

- Applying Human Factors and Usability Engineering to Medical Devices, issued on FEBRUARY 2016.

## IX. Conclusions

{10}

Bestway Plastic And Metal Products Ltd.
510(k)s –510(k) Summary

Based on the above analysis and non-clinical tests performed, it can be concluded that the subject device IPL Hair Removal Device is as safe, as effective, and performs as well as the legally marketed predicate device and reference device.

Page 7 of 7

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT/K243036](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT/K243036)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
