The IPL Hair Removal is an over the-counter device intended for removal of unwanted body and/or facial hair.
Device Story
Personal light-based hair reduction device; uses Intense Pulsed Light (IPL) technology via Xenon flashlamp. Input: manual activation via finger switch; skin sensor detects contact to enable light pulse emission. Output: optical radiation pulses for hair reduction. Used in home environment by individual users. Device includes main body, power adapter, and accessories (shaver, sunglasses). User controls intensity and mode (manual/auto flash, ice cooling) via buttons. Mechanism: light energy targets hair follicles below skin surface to reduce growth. Benefits: non-invasive hair removal with minimal pain.
Clinical Evidence
No clinical data. Bench testing only. Evidence includes electrical safety (IEC 60601-1, 60601-1-11, 60601-2-57, 60601-1-2), photobiological safety (IEC 62471), biocompatibility (ISO 10993-5, ISO 10993-10), and usability validation (IEC 62366-1, 60601-1-6).
Technological Characteristics
Light-based hair removal using Xenon arc flashlamp. Wavelength range: 470nm-1200nm. Pulse duration: 9-12ms. Energy density: 1.9-5.3 J/cm². Powered by external 12V adapter. Features skin contact sensor for safety. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-57, IEC 62471, ISO 10993-5, ISO 10993-10.
Indications for Use
Indicated for removal of unwanted body and/or facial hair in an over-the-counter setting for adult users.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Shenzhen Semlamp Intelligent Technology Co., Ltd % Cassie Lee Official Correspondent Guangzhou GLOMED Biological Technology Co., Ltd. 2231, Building 1, Rui Feng Center, Kaichuang Road Huangpu District Guangzhou, Guangdong 510530 China
Re: K240969
Trade/Device Name: IPL Hair Removal (SL-B287, SL-B329, SL-B301-1, SL-B328) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: OHT Dated: March 27, 2024 Received: April 9, 2024
Dear Cassie Lee:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{2}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for
Yan Fu -
e: 2024.05.31 09:10:04 -04'00'
Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K240969
Device Name
IPL Hair Removal (SL-B287, SL-B329, SLB301-1, SL-B328)
Indications for Use (Describe)
The IPL Hair Removal is an over the-counter device intended for removal of unwanted body and/or facial hair.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary of K240969
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
#### 1. Submitter's Information
Company Name: Shenzhen Semlamp Intelligent Technology Co., Ltd Address: Room 915, Lankun Building, Minkang Road 213, Zhangkeng Community, Minzhi Avenue, Longhua District, Shenzhen, Guangdong, China Contact Person (including title): Qi Wenjun (CEO) Tel: +86 13428733261 Fax: +86 0755-23490305 Post code: 518131 E-mail: info@semlamp.com
## Application Correspondent:
Contact Person: Ms. Cassie Lee Guangzhou GLOMED Biological Technology Co., Ltd. Address: 2231, Building 1, Rui Feng Center, Kaichuang Road, Huangpu District, Guangzhou, Guangdong, China Tel: +86 20 8266 2446 Email: regulatory@glomed-info.com
#### Subject Device Information: 2.
Classification Name: Light Based Over-The-Counter Hair Removal Trade Name: IPL Hair Removal Model Name: SL-B287, SL-B329, SL-B301-1, SL-B328 Review Panel: General & Plastic Surgery Product Code: OHT Requlation Number: 21 CFR 878.4810 Requlatory Class: II
### Predicate Device Information 3.
### Primary Predicate Device
Sponsor: Shenzhen Lescolton Electrical Appliance Co., Ltd. Trade Name: IPL Hair Removal Device (LS-T106, LS-T107, LS-T108) Classification Name: Light Based Over-The-Counter Hair Removal 510(K) Number: K232499 Review Panel: General & Plastic Surgery Product Code: OHT Regulation Number: 21 CFR 878.4810 Requlation Class: II
### Reference Device
Sponsor: Shenzhen Wochuan Electronic Co., Ltd Trade Name: IPL Hair Removal, Model: W-1095, W-1098 Classification Name: Light Based Over-The-Counter Hair Removal 510(K) Number: K232124 Review Panel: General & Plastic Surgery Product Code: OHT Regulation Number: 21 CFR 878.4810 Regulation Class: II
#### 4. Device Description
{5}------------------------------------------------
The IPL Hair Removal is a personal. light-based, hair reduction device intended to be sold over-thecounter directly to the end user. The device provides hair reduction using Intense Pulsed Light (IPL) technology. It works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain. The device is only powered by the external power adapter and its IPL emission activation is by finger switch. The device mainly contains a Xenon lamp and a skin sensor to detect appropriate skin contact. If the device is not properly applied to the treatment area (in full contact with the skin), the device cannot emit the treatment light pulses. The device is for single-person use only. For model SL-B287 and SL-B329, they consist of IPL beauty device main body and power adapter (12V3A) two parts, and one non-removable lamp head (light-emitting treatment window) located in the main body which is the source of optical radiation, namely a Xenon flashlamp. In addition, the device is equipped with a manual shaver and sunglass.
The user can control the device effectively by the buttons on the main unit. There are 3 operation buttons: Power button (turning on/off the device, select light intensity level), Flash button (to start flashing) and loe Cooling button (to select ice cooling mode and switch manual / auto flash mode).
For model SL-B301-1 and SL-B328, consisting of an IPL beauty device main body and power adapter (12V2A) two parts, and one non-removable lamp head (light-emitting treatment window) located in the main body which is the source of optical radiation, namely a Xenon flashlamp. In addition, the device is equipped with a manual shaver, sunglass and replacement lampshade.
The user can control the device effectively by the buttons on the main unit. There are 2 operation buttons: Power button (turning on/off the device and selecting the level of gears) and Mode/Flash button.
### Intended Use / Indications for Use 5.
The IPL Hair Removal is an over the-counter device intended for removal of unwanted body and/or facial hair.
| Elements of<br>Comparison | Subject Device | Primary Predicate<br>Device | Reference Device | Remark |
|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|--------|
| Company | Shenzhen Semlamp<br>Intelligent Technology<br>Co., Ltd | Shenzhen Lescolton<br>Electrical Appliance<br>Co., Ltd. | Shenzhen Wochuan<br>Electronic Co., Ltd | -- |
| Trade Name | IPL Hair Removal, Model:<br>SL-B287, SL-B329, SL-<br>B301-1, SL-B328 | IPL Hair Removal<br>Device (LS-T106,<br>LS-T107, LS-T108) | IPL Hair Removal,<br>Model W-1095, W-<br>1098 | -- |
| Classification<br>Name | Light Based Over-The-<br>Counter Hair Removal | Light Based Over-<br>The-Counter Hair<br>Removal | Light Based Over-<br>The-Counter Hair<br>Removal | -- |
| 510(k) Number | K240969 | K232499 | K232124 | -- |
| Product Code | OHT | OHT | OHT | Same |
| Class | II | II | II | Sames |
| Intended Use /<br>Indications for<br>Use | The IPL Hair Removal is<br>an over the-counter<br>device intended for<br>removal of unwanted<br>body and/or facial hair. | The IPL Hair<br>Removal Device is<br>an over-the-counter<br>device intended for<br>removal of unwanted<br>body and/ or facial<br>hair | IPL Hair Removal is<br>an over-the-counter<br>device intended for<br>removal of unwanted<br>body and/or facial<br>hair. | Same |
| Prescription or<br>OTC | OTC | OTC | OTC | Same |
| Light source | Intense Pulsed Light | Intense Pulsed Light | Intense Pulsed Light | Same |
| Energy medium | Xenon Arc Flashlamp | Xenon Arc | Xenon Arc Flashlamp | Same |
#### Comparison to predicate devices 6.
Compare with the predicate devices, the subject device is very similar in design principle, intended use, indications for use, functions, material and the applicable standards. The differences between the subject
{6}------------------------------------------------
| | | Flashlamp | | |
|-----------------------------------------|----------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------|-------------------|
| Wavelength<br>Range | 470nm-1200nm | LS-T106: 610-1200nm LS-T107:<br>560-1200nm LS-<br>T108: 470-1200nm | 470nm-1100nm | Same |
| Pulse duration | 9-12 milliseconds | LS-T106: 0.64-2.4ms<br>LS-T107: 7.2-10.8ms<br>LS-T108: 6.8-10.2ms | 4-13ms | Similar<br>Note 1 |
| Energy density | SL-B287: 2.2-5.3J/cm²<br>SL-B329: 2.2-5.3J/cm²<br>SL-B301-1: 1.9-3.4J/cm²<br>SL-B328: 1.9-3.4J/cm² | LS-T106: 2.0-<br>4.87J/cm² LS-T107:<br>2.16-5.18J/cm² LS-<br>T108: 2.0-5.62J/cm² | W-1095:1.67~4.46J/cm²<br>W-1098:1.16~2.79J/cm² | Similar<br>Note 1 |
| Spot size (size of<br>treatment window) | SL-B287: 3.6 cm²<br>SL-B329: 3.6 cm²<br>SL-B301-1:4.2 cm²/2 cm²<br>SL-B328: 4.2 cm²/2 cm² | LS-T106: 3.2 cm²<br>LS-T107: 3.7 cm²<br>LS-T108: 3.2 cm² | W-1095: 2.69cm²<br>W-1098: 4.3cm² | Similar<br>Note 2 |
| Delivery device | Direct illumination to<br>tissue | Direct illumination to<br>tissue | Direct illumination to<br>tissue | Same |
| Pulsing control | Finger switch | Finger switch | Finger switch | Same |
| Location for use | large areas (e.g. arms,<br>legs) and small areas<br>(e.g. lip) | Multiple hair removal<br>areas, including<br>small areas (e.g.<br>armpit, bikini lines)<br>and large areas (e.g.<br>arms, legs) | Face, lips, underarms,<br>bilini lines, arms, legs,<br>etc | Same |
| Safety and EMC | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>IEC 60601-2-57<br>IEC 62471 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>IEC 60601-2-83<br>IEC 62471 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>IEC 60601-2-83<br>IEC 62471 | Same |
| Biocompatibility | ISO 10993-5<br>ISO 10993-10 | ISO 10993-5<br>ISO 10993-10 | ISO 10993-5<br>ISO 10993-10 | Same |
# Comparison in Detail(s):
# Note 1:
Althouqh the "Pulse duration" and "Enerqy density" of the subject device is slightly different from that of the predicate device, the values of the pulse duration and energy density of the subject device are within the range of the reference device. So, this difference will not raise any safety or effectiveness issue.
# Note 2:
Although the "Spot size (size of treatment window)" of subject device is a little different from the predicate device, the difference in treatment window size is due to different designs. So, we believe that these differences will not affect the device's performance.
# 7. Test Summary
- 7.1 Non-Clinical Tests Performed
# 1) Electrical safety, and electromagnetic compatibility Test
Non-clinical tests were performed on the subject device in order to validate the design and to assure conformance with the following voluntary design standards in connection with medical device electrical safety, and electromagnetic compatibility:
- IEC 60601-1 2020-08 Edition 3.2 Medical electrical equipment Part 1: General requirements for basic . safety and essential performance
{7}------------------------------------------------
- � IEC 60601-1-11 Edition 2.1 2020-07 Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance --Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment.
- � IEC 60601-2-57 Edition 1.0 2011-01 Medical Electrical Equipment - Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use.
- � IEC 60601-1-2 Edition 4.1 2020-09 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests.
- � IEC 62471 First edition 2006-07 Photobiological safety of lamps and lamp systems.
# 2) Biocompatibility Test
The subject device is biocompatible for its intended use. They are complied with biocompatibility standards ISO 10993-5 (Cytotoxicity) and ISO 10993-10 (Sensitization and Skin Irritation).
### Software verification and validation 3)
Software verification and validation testing were conducted and documentation was provided as recommended by FDA'S Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level concern, since a malfunction of, or a latent design flaw in, the Software Device leads to an erroneous diagnosis or a delay in delivery of appropriate medical care that would likely lead to Minor Injury.
# 4) Usability validation
Usability testing was conducted on the IPL Hair Removal, Model: SL-B287, SL-B329, SL-B301-1, SL-B328, which complies with IEC 62366-1 and IEC 60601-1-6.
# 7.2 Summary of Clinical Performance
Clinical testing was not needed for this 510(k). The non-clinical performance testing described above is sufficient to support that the device can be used safely and effectively.
# 8. Date of the summary prepared: May 7, 2024
# 9. Final Conclusion
The subject device is as safe, as effective, and performs as well as or better than the legally marketed predicated device K232499 and K232124.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.