IPL Hair Removal Device, Model(s): T1, T2, T3, T7, T10

K232183 · Shenzhen Koli Technology Co.,Ltd · OHT · Sep 22, 2023 · General, Plastic Surgery

Device Facts

Record IDK232183
Device NameIPL Hair Removal Device, Model(s): T1, T2, T3, T7, T10
ApplicantShenzhen Koli Technology Co.,Ltd
Product CodeOHT · General, Plastic Surgery
Decision DateSep 22, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Intended Use

IPL Hair Removal Device is indicated for the removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.

Device Story

IPL Hair Removal Device (Models T1, T2, T3, T7, T10) is an OTC, home-use device for hair reduction. Device uses Intense Pulsed Light (IPL) generated by a Xenon arc flashlamp; energy is delivered to skin surface to reduce hair growth. Device is powered by external adapter; operation requires manual finger switch activation. Integrated skin sensor prevents light pulse emission unless in full contact with skin. User selects from multiple function modes (skin, face, body, underarm, bikini) and five intensity levels. Device is intended for single-person use. Output affects hair follicles to achieve long-term, stable reduction in regrowth. Benefits include non-invasive hair removal with minimal pain.

Clinical Evidence

Bench testing only. No clinical data provided. Evidence includes biocompatibility testing (ISO 10993-5, ISO 10993-10), electrical safety and EMC testing (ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-83), photobiological eye safety (IEC 62471), and software verification/validation. Usability was evaluated per FDA human factors guidance.

Technological Characteristics

Materials: ABS, PC, POM. Energy source: External power adapter (100-240V). Light source: Xenon arc flashlamp. Wavelength: 510-1100nm. Energy density: 1.03-3.08 J/cm². Pulse duration: 4-12ms. Features: Skin contact sensor, finger switch activation, 5 intensity levels. Connectivity: None. Sterilization: Not required. Software: Microprocessor-controlled.

Indications for Use

Indicated for removal of unwanted hair and permanent reduction in hair regrowth in adults.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. Underneath the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. September 22, 2023 Shenzhen Koli Technology Co.,Ltd % Yvonne Liu Registration Engineer Feiying Drug & Medical Consulting Technical Service Group Rm 2401 Zhenye International Business Center, No. 3101-90 Qianhai Road. Shenzhen, Guangdong 518052 China Re: K232183 Trade/Device Name: IPL Hair Removal Device, Model(s): T1, T2, T3, T7, T10 Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHT Dated: July 24, 2023 Received: July 24, 2023 Dear Yvonne Liu: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of {1}------------------------------------------------ Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Tanisha L. Hithe -S Tanisha L. 2023.09.22 Hithe -S 14:53:36 -04'00' Tanisha Hithe, MS, MHS Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K232183 Device Name IPL Hair Removal Device Model(s): T1, T2, T3, T7, T10 #### Indications for Use (Describe) IPL Hair Removal Device is indicated for the removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime. Type of Use (Select one or both, as applicable) | | Registration Use (Part 34 CFR 301 Subpart D) | |--|----------------------------------------------| | | On-The-Spot-Use (34 CFR 301 Subpart C) | __ Prescription Use (Part 21 CFR 801 Subpart D) |X | Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary K232183 # "510(k) Summary" as required by 21 CFR Part 807.92. #### I. Submitter Shenzhen Koli Technology Co.,Ltd Floor 4, No.145-1 Furui Road, Xintian Community, Fuhai Street, Bao'an District, Shenzhen, China Post code: 518103 Tel.: +86 15818667227 Fu Qizhen General Manager Tel: +86 15818667227 Email: fuqizhen1983@163.com # II. Device Name of Device: IPL Hair Removal Device Model(s): T1, T2, T3, T7, T10 Common or Usual Name: Light Based Over-The-Counter Hair Removal Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: OHT Regulation Number: 21 CFR 878.4810 ### III. Predicate Device and reference device Predicate device: | Manufacturer | Predicate Device | 510(k) Number | Approval Date | |-----------------------------------------------|------------------------------------------------------------------------------------------|---------------|---------------| | Shenzhen Ishine Technology<br>Company Limited | IPL Hair Removal Device,<br>Model(s):<br>skn001,skn005,skn006,skn002,<br>RoseSkinCo Lumi | K222537 | Oct 14,2022 | Reference device: | Manufacturer | Reference Device | 510(k) Number | Approval Date | |-----------------------------------------------------------------|---------------------------------------------------------------------------------------------|---------------|---------------| | Shenzhen Beauty Every<br>Moment intelligent electric<br>Co.,Ltd | IPL Home Use Hair Removal<br>Device, Models: D-1198, D-<br>1185, D-1186, D-1189, D-<br>1197 | K221001 | May 12, 2022 | {4}------------------------------------------------ # IV. Device Description IPL Hair Removal Device (Models:T1, T2, T3, T7, T10),is an over-the-counter, home-use and singleperson-use device for hair reduction by using Intense Pulsed Light (IPL). It works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain. The device is only powered by the external power adapter and its IPL emission activation is by finger switch. This product adopts irreplaceable flash window and is suitable for multiple hair removal areas. There are four function modes including skin, face,body and bikini for T1 and T7, but only body mode for T3, and five modes including skin, face,underarm,body and bikini for T2 and T10. There are five levels for each mode. The device contains a skin sensor to detect appropriate skin contact, if the device is not in full contact with the skin, the device cannot emit the treatment light pulses. IPL Hair Removal Device, models:T1, T2, T3, T7, T10 have the same indication for use, performance, structure design and operation. The main difference mainly contains product appearance, dimension , weight , number of keys and function modes, spot size and energy density . The five models are in two colors, one set is in white and the other set is in green. # V. Indications for Use IPL Hair Removal Device is indicated for the removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime. # VI. Materials | Component<br>name | Material of Component | Body Contact Category | Contact Duration | |---------------------------------------------------------------|-----------------------|-------------------------------------------|--------------------| | IPL Hair<br>Removal Device<br>(Enclosure and<br>flash window) | ABS, PC,POM | Surface-contacting<br>device: Intact skin | Less than 24 hours | We have tested the device for biocompatibility by a reliable third-party lab. For details, please refer to Section 16 "Biocompatibility Discussion". # VII.Comparison of Technological Characteristics With the Predicate Device The IPL Hair Removal Device has the same intended use, mode of action and similar operational characteristics as the predicate device. Any minor differences between the subject device and the listed predicate device and reference device do no raise any issues of safety or efficacy. Performance data supports that the device is safe and as effective as the predicate device and reference device for its intended use. Therefore, the IPL Hair Removal Device may be found substantially equivalent to its predicate device and reference device. IPL Hair Removal Device is compared with the following Predicate Device and Reference Device in terms of intended use, design, material,specifications, and performance: {5}------------------------------------------------ | Comparison<br>Elements | Subject Device | Predicate Device | Reference Device | Remark | |-------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|-----------| | 510(k)<br>Number | Pending | K222537 | K221001 | / | | Trade name | IPL Hair Removal Device<br>T1,T2,T3,T7,T10 | IPL Hair Removal Device<br>(skn001,skn005,skn006,skn002,RoseSkin Co Lumi) | IPL Home Use Hair<br>Removal Device,<br>Models: D-1198, D-<br>1185, D-1186, D-1189,<br>D-1197 | / | | Manufacturer | Shenzhen Koli Technology<br>Co.,Ltd | SHENZHEN<br>TECHNOLOGY<br>COMPANY LIMITED | Shenzhen Beauty Every<br>Moment intelligent /<br>electric Co.,Ltd | / | | Regulation<br>number | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | Same | | Product code | OHT | OHT | OHT | Same | | Device<br>classification | Class II | Class II | Class II | Same | | Indication for<br>use/ Intended<br>use | IPL Hair Removal Device<br>is indicated for the removal<br>of unwanted hair. The<br>device is also indicated for<br>the permanent reduction in<br>hair regrowth, defined as<br>the long-term, stable<br>reduction in the number of<br>hairs regrowing when<br>measured at 6, 9 and 12<br>months after the<br>completion of a treatment<br>regime. | IPL Hair Removal Device<br>is indicated for the removal<br>of unwanted hair. The<br>device is also indicated for<br>the permanent reduction in<br>hair regrowth, defined as<br>the long-term, stable<br>reduction in the number of<br>hairs regrowing when<br>measured at 6, 9 and 12<br>months after the<br>completion of a treatment<br>regime. | IPL Home Use Hair<br>Removal Device is an<br>over-the counter device<br>intended for removal of<br>unwanted body and /or<br>facial hair | Same | | Prescription<br>or OTC | OTC | OTC | OTC | Same | | Source<br>energy | Supplied by external<br>adapter | Supplied by external<br>adapter | Supplied by external<br>adapter | Same | | Power<br>Supply | 100-240V~, 50/60Hz | 100-240V~, 50/60Hz | Unknown | Same | | Dimension | T1: 176 x 62 x 35mm<br>T2: 171 x 64 x 34mm<br>T3: 214 x 63 x 64mm<br>T7: 174 x 63 x 36mm | skn001:134*83*48.5mm<br>skn005:132.5*83*48mm<br>skn006:135.5*86*52.5mm<br>skn002:134*82*40mm | Unknown | Different | | Comparison<br>Elements | Subject Device | Predicate Device | Reference Device | Remark | | | T10: 171 x 64 x 35mm | RoseSkinCo<br>134*82*40mm | Lumi: | | | Weight | T1: 200g<br>T2: 296g<br>T3: 200g<br>T7: 216g<br>T10: 216g | skn001:196g<br>skn005: 196g<br>skn006: 210g<br>skn002: 186g<br>RoseSkinCo Lumi:186g | Unknown | Different | | Sterilization | Not required | Not required | Not required | Same | | Light Source | Intense Pulsed Light | Intense Pulsed Light | Intense Pulsed Light | Same | | Energy<br>medium | Xenon Arc Flashlamp | Xenon Arc Flashlamp | Xenon Arc Flashlamp | Same | | Wavelength<br>range | 510-1100nm | 510-1100nm | 550-1100nm | Same | | Spot<br>Size<br>( cm² ) | T1: $3.0 \pm 0.25cm^2$<br>T2: $3.9 \pm 0.25cm^2$<br>T3: $3.2 \pm 0.25cm^2$<br>T7: $3.0 \pm 0.25cm^2$<br>T10: $3.9 \pm 0.25cm^2$ | $3.3cm^2$ | D-1198: $2.7 cm^2$<br>D-1185: $2.7 cm^2$<br>D-1186: $3.0 cm^2$<br>D-1189: $3.0 cm^2$<br>D-1197: $3.4 cm^2$ | Similar | | Energy<br>Density<br>(±20%) | T1,T7: 1.33~3.00 J/cm²<br>T3: 1.25-2.81 J/cm²<br>T2,T10: 1.03~ 3.08 J/cm² | For sk001,skn005,skn006<br>Level 1: 1.5 J/cm²<br>Level 2: 1.9J/cm²<br>Level 3: 2.3J/cm²<br>Level 4: 2.8J/cm²<br>Level 5: 3.5J/cm²<br>For skn002<br>and RoseSkinCo Lumi<br>Level 1: 1.6J/cm²<br>Level 2: 1.9J/cm²<br>Level 3: 2.3J/cm²<br>Level 4: 2.8J/cm²<br>Level 5: 3.2J/cm²<br>Level 6: 3.8J/cm² | 2.0~4.3 J/cm² | Similar | | Pulse<br>duration | 4-12ms | 6-8ms | 5-12 ms | Similar | | Comparison<br>Elements | Subject Device | Predicate Device | Reference Device | Remark | | Pulsing<br>control | Finger switch | Finger switch | Finger switch | Same | | Number<br>of<br>output<br>channels | One channel | One channel | One channel | Same | | Output<br>intensity<br>level | 5 Levels | Levels<br>5<br>for<br>skn001,skn005,skn006;<br>6<br>Levels for skn002<br>and<br>RoseSkinCo Lumi | Unknown | Similar | | Skin Contact<br>Sensor | Yes | Yes | Yes | Same | | Software/<br>Firmware/<br>Microprocess<br>or Control? | Yes | Yes | Yes | Same | | Electrical<br>safety | ANSI AAMI ES60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>IEC 60601-2-83 | ANSI AAMI ES60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>IEC 60601-2-83 | IEC 60601-1-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>IEC 60601-2-57 | Same | | Eye safety | IEC 62471 | IEC 62471 | IEC 62471 | Same | | Biocompatibi<br>lity | ISO 10993-5<br>ISO 10993-10 | ISO 10993-5<br>ISO 10993-10 | ISO 10993-5<br>ISO 10993-10 | Same | {6}------------------------------------------------ {7}------------------------------------------------ #### VIII. Performance Data The following performance data were provided in support of the substantial equivalence determination. ### 1) Biocompatibility Testing The biocompatibility evaluation for the body-contacting components of the IPL Hair Removal Device was conducted in accordance with the "Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices -Part 1: Evaluation and Testing Within a Risk Management Process, Document Issued on September 4, 2020", as recognized by FDA. The following testing was performed to, and passed, including: - > ISO 10993-5:2009, Biological Evaluation of Medical Devices –Par t 5: Tests for In Vitro Cytotoxicity - > ISO 10993-10:2010, Biological Evaluation of Medical Devices –Par t 10: Tests for Irritation and Skin Sensitization {8}------------------------------------------------ ## 2) Electrical Safety and EMC Electrical safety and EMC testing was performed to, and passed, as per the following standards: - > IEC 60601-1-2 Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance -Collateral standard: electromagnetic compatibility - > ANSI AAMI ES60601-1 Medical electrical equipment Part 1: General requirements for basic safety and essential performance - > IEC 60601-1-11 Medical Electrical Equipment -Part 1: General Requirements for Basic Safety and Essential Performance-Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment - > IEC 60601-2-83 Medical electrical equipment Part 2-83: Particular requirements for the basic safety and essential performance of home light therapy equipment # 3) Eye Safety - > IEC 62471 Photobiological safety of lamps and lamp systems #### 4) Software Verification and Validation Software documentation consistent with moderate level of concern was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels. ### 5) Usability The product usability has been evaluated and verified according to the following FDA guidance > Applying Human Factors and Usability Engineering to Medical Devices, issued on FEBRUARY 2016 ### IX. Conclusions Based on the above analysis and non-clinical tests performed, it can be concluded that the subject device IPL Hair Removal Device is as safe, as effective, and performs as well as the legally marketed predicate device and reference device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...