← Product Code [OHT](/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT) · K213830

# Home use hair removal device, model: Precise Touch Pro (K213830)

_Iluminage Beauty Co., Limited · OHT · Feb 1, 2022 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT/K213830

## Device Facts

- **Applicant:** Iluminage Beauty Co., Limited
- **Product Code:** [OHT](/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT.md)
- **Decision Date:** Feb 1, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

The Home use hair removal device is an over-the-counter device intended for removal of unwanted body hair and/or facial hair.

## Device Story

Personal light-based hair reduction device; uses Intense Pulsed Light (IPL) technology. Input: skin contact detection via built-in sensor. Operation: Xenon flashlamp emits optical radiation (530-1100nm) to target hair follicles; device requires full skin contact for pulse triggering. Output: light pulses for hair removal. Used in home environment by consumer. Healthcare provider role: none; device is OTC. Benefit: hair reduction/removal. Safety features: skin sensor prevents accidental firing; electrical/eye safety testing per IEC standards.

## Clinical Evidence

Bench testing only. No clinical data provided. Evidence includes biocompatibility testing (ISO 10993-5, ISO 10993-10), electrical safety (IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-83), and photobiological eye safety (IEC 62471). Software verification and validation confirmed all requirements met and hazards mitigated.

## Technological Characteristics

IPL hair removal device; Xenon arc flashlamp energy source; 530-1100nm wavelength; 2.3-4.3 J/cm² energy density; 3.6cm² spot size; 7.6-9.6ms pulse duration. Powered by external AC adapter. Includes skin contact sensor for safety interlock. Biocompatibility per ISO 10993-1. Software level of concern: moderate.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- IPL Hair removal, V501 ([K203510](/device/K203510.md))
- IPL Hair Removal Device Joy Version, CB-027 ([K173813](/device/K173813.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

February 1, 2022

Iluminage Beauty Co., Limited % Rain Yip Registered Engineer Feiying Drug & Medical Consulting Technical Service Group Rm2401, ZhenYe International Center, No.3101-90 Qianhai Road Nanshan District Shenzhen, Guangdong 518000 China

Re: K213830

Trade/Device Name: Home use hair removal device, model: Precise Touch Pro Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHT Dated: November 30, 2021 Received: December 8, 2021

Dear Rain Yip:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K213830

#### Device Name

Home use hair removal device, model: Precise Touch Pro

Indications for Use (Describe)

The Home use hair removal device is an over-the-counter device intended for removal of unwanted body hair and/or facial hair.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

#### "510(k) Summary" as required by 21 CFR Part 807.92.

#### Date: 2021-11-30

#### I. Submitter

Iluminage Beauty Co., Limited Level 54, Hopewell Centre, 183 Queen's Road East, Hong Kong. Post code: 999077 Tel.: 86 0755 29163244

Jacky Lee RD Manager Tel: +86 135 1058 4309 Email: 44069748@qq.com

#### II. Device

Name of Device: Home use hair removal device Model: Precise Touch Pro Common or Usual Name: Light Based Over-The-Counter Hair Removal Classification Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: OHT Regulation Number: 21 CFR 878.4810

#### III. Predicate Device

| Predicate Device                                                | Manufacturer                            | 510(k) Number | Approval Date |
|-----------------------------------------------------------------|-----------------------------------------|---------------|---------------|
| Primary predicate device: IPL Hair removal, V501                | Shenzhen Weikai Technology Co., Ltd.    | K203510       | Feb.26, 2021  |
| Predicate device 2: IPL Hair Removal Device Joy Version, CB-027 | Shenzhen CosBeauty Technology Co., Ltd. | K173813       | Sep. 7, 2018  |

#### IV. Device Description

The Precise Touch Pro Home use hair removal device is a personal. light-based, hair reduction device intended to be sold over-the-counter directly to the end user. The device provides hair reduction using Intense Pulsed Light (IPL) technology to perform hair removal. The device mainly consists of IPL main body, AC adapter and device stand three parts and it is only powered by the external power adapter, as well as the treatment window located in the main body which is the source of optical radiation, namely a Xenon flashlamp and the built-in skin sensor to detect

{4}------------------------------------------------

the body skin area. If the device is not properly applied to the expected treatment area (in full contact with the skin), the device cannot be automatic triggered a pulse emitting.

## V. Indications for Use

The Home use hair removal device is an over-the-counter device intended for removal of unwanted body hair and/or facial hair.

# VI. Comparison of Technological Characteristics With the Predicate Device

The Home use hair removal device has the same intended use, mode of action and similar operational characteristics as the predicate devices. Any minor differences between the subject device and the listed predicate devices do no raise any issues of safety or efficacy. Performance data supports that the device is safe and as effective as the predicate device for its intended use. Therefore, the Home use hair removal device may be found substantially equivalent to its predicate device.

Home use hair removal device is compared with the following Predicate Device in terms of intended use, design, material, specifications, and performance:

| Comparison<br>Elements                | Subject Device                                                                                                                                  | Primary Predicate<br>Device K203510                                                                                          | Predicate device 2 K173813                                                                                                                                                                                                                                                                                                        |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K Number                              | Pending                                                                                                                                         | K203510                                                                                                                      | K173813                                                                                                                                                                                                                                                                                                                           |
| Trade name                            | Home use hair<br>removal device                                                                                                                 | IPL Hair removal,<br>V501                                                                                                    | IPL Hair Removal Device Joy<br>Version, CB-027                                                                                                                                                                                                                                                                                    |
| Wavelength<br>range                   | 530-1100nm                                                                                                                                      | 530-1200nm                                                                                                                   | 510-1200nm                                                                                                                                                                                                                                                                                                                        |
| Energy<br>medium                      | Xenon Arc Flashlamp                                                                                                                             | Xenon Arc Flashlamp                                                                                                          | Xenon Arc Flashlamp                                                                                                                                                                                                                                                                                                               |
| Energy<br>density                     | $2.3~4.3 J/cm²$                                                                                                                                 | $3.0~4.0J/cm²$                                                                                                               | $1.8~5.1J/cm²$                                                                                                                                                                                                                                                                                                                    |
| Spot size                             | $3.6cm²$                                                                                                                                        | $3.7cm²$                                                                                                                     | Body: $4.2cm²$<br>Bikini and face: $2.0cm²$                                                                                                                                                                                                                                                                                       |
| Pulse duration                        | 7.6~9.6ms                                                                                                                                       | 4~13ms                                                                                                                       | 9.20~11.20ms                                                                                                                                                                                                                                                                                                                      |
| Delivery<br>device                    | Direct illumination<br>tissue                                                                                                                   | Direct illumination<br>tissue                                                                                                | Direct illumination tissue                                                                                                                                                                                                                                                                                                        |
| Indication for<br>use/Intended<br>use | The Home use hair<br>removal device is an<br>over-the-counter<br>device intended for<br>removal of unwanted<br>body hair and/or<br>facial hair. | IPL Hair removal is<br>an over-the-counter<br>device intended for<br>removal of unwanted<br>body hair and/or<br>facial hair. | The IPL Hair Removal Device<br>Joy Version is indicated for the<br>removal of unwanted hair.<br>The device is also indicated<br>for the permanent reduction in<br>hair regrowth, defined as the<br>long-term, stable reduction in the<br>number of hairs regrowing<br>when measured at 6, 9 and<br>12 months after the completion |

{5}------------------------------------------------

| Comparison<br>Elements | Subject Device | Primary Predicate<br>Device K203510 | Predicate device 2 K173813                                                                    |
|------------------------|----------------|-------------------------------------|-----------------------------------------------------------------------------------------------|
|                        |                |                                     | of a treatment regime. The<br>device is used for adults<br>with Fitzpatrick skin types I -IV. |
| Location for<br>use    | OTC            | OTC                                 | OTC                                                                                           |

## VII.Performance Data

The following performance data were provided in support of the substantial equivalence determination.

## 1) Biocompatibility Testing

The biocompatibility evaluation for the body-contacting components of the subject device was conducted in accordance with the "Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices –Part 1: Evaluation and Testing Within a Risk Management Process", as recognized by FDA. The battery of testing was performed to, and passed, including:

- > ISO 10993-5 Biological Evaluation of Medical Devices -Par t 5: Tests for In Vitro Cytotoxicity
- > ISO 10993-10 Biological Evaluation of Medical Devices –Par t 10: Tests for Irritation and Skin Sensitization

#### 2) Electrical Safety and Eye Safety

Electrical safety and Eye safety testing was performed to, and passed, the following standards:

- > IEC 60601-1-2 Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance -Collateral standard: electromagnetic compatibility
- > IEC 60601-1 Medical electrical equipment -Part 1: General requirements for basic safety and essential performance
- > IEC 60601-1-11 Medical Electrical Equipment -Part 1: General Requirements for Basic Safety and Essential Performance -Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment
- > IEC 60601-2-83 Medical Electrical Equipment Part 2-83: Particular Requirements For The Basic Safety And Essential Performance Of Home Light Therapy Equipment

## 3) Eye Safety

- > IEC 62471 Photobiological safety of lamps and lamp systems
## 4) Software Verification and Validation

{6}------------------------------------------------

Software documentation consistent with moderate level of concern was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels.

#### Summary

Based on the above performance as documented in this application, the subject device Home use hair removal device was found to have a safety and effectiveness profile that is similar to the predicate device.

#### VIII. Conclusions

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the comparison of intended use, design, materials and performance, the subject device Home use hair removal device is to be concluded substantial equivalent to its predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT/K213830](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT/K213830)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
