← Product Code [OHT](/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT) · K212697

# Leaf Smooth (K212697)

_Shenzhen Leaflife Technology Co., Ltd. · OHT · Nov 19, 2021 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT/K212697

## Device Facts

- **Applicant:** Shenzhen Leaflife Technology Co., Ltd.
- **Product Code:** [OHT](/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT.md)
- **Decision Date:** Nov 19, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

The Leaf Smooth is an over-the-counter device intended for removal of unwanted body and/or facial hair.

## Device Story

Leaf Smooth (LH-LIPLB) is personal, light-based, hair reduction device for over-the-counter use. Device utilizes Intense Pulsed Light (IPL) technology via Xenon Arc Flashlamp to reduce hair growth without cutting or pulling. User operates device via finger switch; device powered by external adapter. Safety features include capacitive sensor for skin contact detection and skin sensor for skin color detection; pulses inhibited if contact is insufficient. Device offers 6 output intensity levels. Intended for single-person use. Healthcare provider involvement is not required; device is for patient self-use at home. Output is light pulse delivery to skin surface; effect is hair reduction.

## Clinical Evidence

No clinical data. Bench testing only, including performance testing for spot size and energy output accuracy, biocompatibility (ISO 10993-5, ISO 10993-10), and electrical/photobiological safety (IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-57, IEC 62471).

## Technological Characteristics

IPL hair reduction device using Xenon Arc Flashlamp. Wavelength 475-1100 nm; spot size 3.8 cm²; fluence 4-6 J/cm²; pulse duration 2-10ms. 6 intensity levels. Capacitive skin contact sensor and skin color sensor. Powered by external adapter. Complies with ANSI/AAMI ES60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-57, IEC 62471. Biocompatible per ISO 10993-5 and 10993-10.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- IPL Home Use Hair Removal Device, Model D-1128 ([K192432](/device/K192432.md))
- IPL Salon Hair Reduction System, Model F60001 ([K181568](/device/K181568.md))
- iPulse SmoothSkin Gold Hair Removal System ([K160968](/device/K160968.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 19, 2021

Shenzhen Leaflife Technology Co., Ltd Cheng Qiang Regulatory Registration Supervisor 4 F,Bldg. C, JMD Industrial Park, No.39 Qingfeng Blvd., Baolong Industrial Area, Longgang District Shenzhen, Guangdong 518116 China

Re: K212697

Trade/Device Name: Leaf Smooth Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHT Dated: August 18, 2021 Received: August 25, 2021

Dear Cheng Qiang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K212697

Device Name Leaf Smooth

Indications for Use (Describe)

The Leaf Smooth is an over-the-counter device intended for removal of unwanted body and/or facial hair.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for LEAFLIFE. The logo is composed of the word "LEAFLIFE" in a stylized font, with the "LEAF" portion in blue and the "LIFE" portion in green. Above the word "LEAFLIFE" is a green leaf and a series of green dots. Below the word "LEAFLIFE" are two Chinese characters in gray.

# 510(k) Summary

In accordance with 21 CFR 807.92 the following summary of information is provided:

#### SUBMITTER l.

Shenzhen Leaflife Technology Co., Ltd 4F, Bldg. C JMD Industrial Park, No.39 Qingfeng Blvd., Baolong Industrial Area, Longgang Dist., Shenzhen, RP China Phone: 086-(0)17875910506 Fax: 086-0755-27215592

| Primary Contact Person: | Cheng Qiang                           |
|-------------------------|---------------------------------------|
|                         | Regulatory registration supervisor    |
|                         | Shenzhen Leaflife Technology Co., Ltd |
| Phone:                  | 086-(0)17875910506                    |
| Tel:                    | 086-0755-27216609                     |
| Fax:                    | 086-0755-27215592                     |
| Email:                  | cq@leaflife.com                       |

Date Prepared:

08/20/2021

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for LEAFLIFE. The logo has the word LEAFLIFE in blue and green, with a green leaf above the letter "E". There are also some Chinese characters below the word LEAFLIFE. The logo is simple and modern, and it conveys a sense of nature and health.

#### II. PROPOSED DEVICE

| Trade Name:          | Leaf Smooth                                                                            |
|----------------------|----------------------------------------------------------------------------------------|
| Model(s):            | LH-LIPLB                                                                               |
| Common Name:         | Light Based Over-The-Counter Hair Removal                                              |
| Classification Name: | Laser surgical instrument for use in general and plastic<br>surgery and in dermatology |
| Regulation Number:   | 21 CFR 878.4810                                                                        |
| Regulation Class:    | II                                                                                     |
| Product Code:        | OHT                                                                                    |
| Review Panel:        | General & Plastic Surgery                                                              |

III. PREDICATE DEVICE

| Predicate device 1# | IPL Home Use Hair Removal Device, Model D-1128:<br>K192432 |
|---------------------|------------------------------------------------------------|
| Predicate device 2# | IPL Salon Hair Reduction System, Model F60001:<br>K181568  |
| Predicate device 3# | iPulse SmoothSkin Gold Hair Removal System, K160968        |

The predicate has not been subject to a design-related recall.

### IV. DEVICE DESCRIPTION

The Leaf Smooth (LH-LIPLB) is a personal, light-based, hair reduction device intended to be sold over-the-counter directly to the end user. The device provides hair reduction using Intense Pulsed Light (IPL) technology. It works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain. The device is only powered by the external power adapter and its IPL emission activation is by finger switch.

The device contains a Xenon Arc Flashlamp, capacitive sensor to detect appropriate skin contact and a skin sensor to detect appropriate skin color. If the device is not properly applied to the treatment area (in full contact with the skin), the device cannot emit the treatment light pulses. The device is for single-person use only.

The Leaf Smooth has 6 levels output intensity.

{5}------------------------------------------------

Image /page/5/Picture/2 description: The image shows the logo for LEAFLIFE. The logo is composed of the word "LEAFLIFE" in blue and green, with a green leaf above the "L". There is a blue arc under the word "LEAFLIFE", and some Chinese characters to the right of the arc. The logo is simple and modern, and it conveys a sense of nature and freshness.

### V. INTENDED USE

The Leaf Smooth is an over-the-counter device intended for removal of unwanted body and/or facial hair.

## vi. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The Leaf Smooth (LH-LIPLB) has same intended use as the predicate devices. Technological characteristics such as fluence, laser wavelength, spot size, and pulse duration, of the Leaf Smooth device are similar to those of the predicate devices.

{6}------------------------------------------------

Image /page/6/Figure/0 description: The image shows the logo for LEAFLIFE. The logo is in blue and green, with the word "LEAFLIFE" in blue and green. There is a green leaf above the word "LEAFLIFE". Below the word "LEAFLIFE" are two Chinese characters in gray.

| Item                         | Proposed Device                                                                                                     | Predicate Device<br>K192432(Primary)                                                                                                | Predicate Device<br>K181568                                                                                                                                                                                                                                                                                                                                                                                                                                   | Predicate Device<br>K160968                                                                                                                                                                                                                                                                                                                                                              | Remark  |
|------------------------------|---------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Device name                  | Leaf Smooth (LH-LIPLB)                                                                                              | IPL Home Use Hair Removal<br>Device                                                                                                 | IPL Salon Hair Reduction<br>System                                                                                                                                                                                                                                                                                                                                                                                                                            | iPulse SmoothSkin Gold                                                                                                                                                                                                                                                                                                                                                                   | /       |
| K number                     | On pending                                                                                                          | K192432                                                                                                                             | K181568                                                                                                                                                                                                                                                                                                                                                                                                                                                       | K160968                                                                                                                                                                                                                                                                                                                                                                                  | /       |
| Product code                 | OHT                                                                                                                 | OHT                                                                                                                                 | OHT                                                                                                                                                                                                                                                                                                                                                                                                                                                           | OHT                                                                                                                                                                                                                                                                                                                                                                                      | Same    |
| Classification<br>regulation | 21 CFR 878.4810                                                                                                     | 21 CFR 878.4810                                                                                                                     | 21 CFR 878.4810                                                                                                                                                                                                                                                                                                                                                                                                                                               | 21 CFR 878.4810                                                                                                                                                                                                                                                                                                                                                                          | Same    |
| Indications for<br>Use       | The Leaf Smooth is an<br>over-the-counter device<br>intended for removal of<br>unwanted body and/or facial<br>hair. | IPL Home Use Hair Removal<br>Device is an over-the-counter<br>device intended for removal of<br>unwanted body and/or facial<br>hair | The IPL Salon Hair Reduction<br>System (Model: F60001) is<br>an over the Counter device<br>intended for the removal of<br>unwanted body and/or facial<br>hair in adults. It is also<br>intended for permanent<br>reduction in unwanted hair.<br>Permanent hair reduction is<br>defined as the long-term<br>stable reduction in the<br>number of hairs re-growing<br>when measured at 6, 9 and<br>12 months after the<br>completion of a treatment<br>regimen. | The iPulse SmoothSkin Gold<br>Hair Removal System is<br>indicated for the removal of<br>unwanted hair. The iPulse<br>Smoothskin Gold is also<br>indicated for the permanent<br>reduction in hair regrowth,<br>defined as the long-term,<br>stable reduction in the number<br>of hairs regrowing when<br>measured at 6, 9 and 12<br>months after the completion of<br>a treatment regime. | Same    |
| Prescription use<br>or not   | OTC                                                                                                                 | OTC                                                                                                                                 | OTC                                                                                                                                                                                                                                                                                                                                                                                                                                                           | OTC                                                                                                                                                                                                                                                                                                                                                                                      | Same    |
| Item                         | Proposed Device                                                                                                     | Predicate Device<br>K192432(Primary)                                                                                                | Predicate Device<br>K181568                                                                                                                                                                                                                                                                                                                                                                                                                                   | Predicate Device<br>K160968                                                                                                                                                                                                                                                                                                                                                              | Remark  |
| Pulsing control              | Finger switch                                                                                                       | Finger switch                                                                                                                       | Finger switch                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Finger switch                                                                                                                                                                                                                                                                                                                                                                            | Same    |
| Delivery device              | Direct illumination to tissue                                                                                       | Direct illumination to tissue                                                                                                       | Direct illumination to tissue                                                                                                                                                                                                                                                                                                                                                                                                                                 | Direct illumination to tissue                                                                                                                                                                                                                                                                                                                                                            | Same    |
| Energy medium                | Xenon Arc Flashlamp                                                                                                 | Xenon Arc Flashlamp                                                                                                                 | Xenon Arc Flashlamp                                                                                                                                                                                                                                                                                                                                                                                                                                           | Xenon Arc Flashlamp                                                                                                                                                                                                                                                                                                                                                                      | Same    |
| Wavelength<br>range          | 475-1100 nm                                                                                                         | 510-1100nm                                                                                                                          | 475-1200nm                                                                                                                                                                                                                                                                                                                                                                                                                                                    | 510-1100nm                                                                                                                                                                                                                                                                                                                                                                               | Similar |
| Spot size                    | 3.8 $cm^2$                                                                                                          | 4.5 $cm^2$ , 2.0 $cm^2$ , 3.0 $cm^2$                                                                                                | 3.02 $cm^2$<br>1.72 $cm^2$                                                                                                                                                                                                                                                                                                                                                                                                                                    | 3 $cm^2$                                                                                                                                                                                                                                                                                                                                                                                 | Similar |
| Fluence                      | 4-6 J/ $cm^2$                                                                                                       | 2.0-4.0 J/ $cm^2$ , 2.5-4.5 J/ $cm^2$                                                                                               | Max 5 J/ $cm^2$                                                                                                                                                                                                                                                                                                                                                                                                                                               | 3-6 J/ $cm^2$                                                                                                                                                                                                                                                                                                                                                                            | Similar |
| Pulse Duration               | 2-10ms                                                                                                              | 7.5-14ms                                                                                                                            | 11-12ms                                                                                                                                                                                                                                                                                                                                                                                                                                                       | 2-10ms                                                                                                                                                                                                                                                                                                                                                                                   | Similar |

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for LEAFLIFE. The logo is blue and green, with a stylized leaf above the text. The text is in a sans-serif font, and there is some Chinese text below the English text. The logo is simple and modern, and it conveys a sense of nature and life.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for LEAFLIFE. The logo is in blue and green, with the word "LEAFLIFE" in blue and green letters. There is a green leaf above the "A" in LEAFLIFE. There is a blue curve below the "L" and "E" in LEAFLIFE. Below the word LEAFLIFE are two Chinese characters in gray.

| Item                             | Proposed Device                                       | Predicate Device<br>K192432                | Predicate Device<br>K181568                | Predicate Device<br>K160968                | Remark |
|----------------------------------|-------------------------------------------------------|--------------------------------------------|--------------------------------------------|--------------------------------------------|--------|
| Biocompatibility                 |                                                       |                                            |                                            |                                            |        |
| Cytotoxicity                     | No Cytotoxicity                                       | No Cytotoxicity                            | No Cytotoxicity                            | No Cytotoxicity                            | Same   |
| Sensitization                    | No evidence of sensitization                          | No evidence of sensitization               | No evidence of sensitization               | No evidence of sensitization               | Same   |
| Irritation                       | No evidence of irritation                             | No evidence of irritation                  | No evidence of irritation                  | No evidence of irritation                  | Same   |
| EMC, Electrical and Laser Safety |                                                       |                                            |                                            |                                            |        |
| Electrical<br>Safety             | Comply with ANSI/AAMI<br>ES60601-1,<br>IEC 60601-1-11 | Comply with IEC 60601-1,<br>IEC 60601-1-11 | Comply with IEC 60601-1,<br>IEC 60601-1-11 | Comply with IEC 60601-1,<br>IEC 60601-1-11 | Same   |
| EMC                              | Comply with IEC 60601-1-2                             | Comply with IEC 60601-1-2                  | Comply with IEC 60601-1-2                  | Comply with IEC 60601-1-2                  | Same   |
| Light Safety                     | Comply with IEC 60601-2-57,<br>IEC 62471              | Comply with IEC 60601-2-57,<br>IEC 62471   | Comply with IEC 60601-2-57,<br>IEC 62471   | Comply with IEC 60601-2-57,<br>IEC 62471   | Same   |

{9}------------------------------------------------

Image /page/9/Picture/2 description: The image shows the logo for LEAFLIFE. The logo is a combination of text and a graphic. The text "LEAFLIFE" is written in a stylized font, with the "L" and "E" in blue, the "A" and "F" in green, and the "L", "I", "F", and "E" in blue. Above the text is a green leaf with a curved line. Below the text is some Chinese characters.

### VII. Non-Clinical Test Conclusion

Non-clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

ES60601-1:2005/(R) 2012 and A1:2012, C1:2009/(R)2012 and ANSI/AAMI A2:2010/(R)2012, Medical electrical equipment — Part 1: General requirements for basic safety and essential performance

IEC 60601-1-2:2014, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.

IEC 60601-1-11:2015, Medical electrical equipment - Part 1-11 General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.

IEC 60601-2-57:2011, Medical electrical equipment - Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use.

IEC 62471:2006, Photobiological safety of lamps and lamp systems.

ISO 10993-5 Third Edition 2009-06-01, Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity. (Biocompatibility)

ISO 10993-10 Third Edition 2010-08-01, Biological Evaluation of Medical Devices -Part 10: Tests for Irritation and Skin Sensitization. (Biocompatibility)

Performance Testing for Spot Size Accuracy and Energy Output Accuracy.

Software documentation consistent with moderate level of concern was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels.

VIII. CLINICAL TEST CONCLUSION

No clinical study is included in this submission.

{10}------------------------------------------------

Image /page/10/Picture/2 description: The image shows the logo for LEAFLIFE. The logo is a combination of text and a graphic element. The text "LEAFLIFE" is written in a stylized font, with the letters in blue and green. Above the text is a green leaf. Below the text are two Chinese characters in gray.

## IX. SUBSTANTIALLY EQUIVALENT (SE) CONCLUSION

The Leaf Smooth device has the same intended use and similar technology to those of the predicate devices. Performance testing conducted on the Leaf Smooth device supports that the Leaf Smooth can be used safely and effectively for the indications for use stated above.

Based on the comparison and analysis above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT/K212697](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT/K212697)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
