← Product Code [OHT](/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT) · K180383

# Light based hair removal device (K180383)

_Shenzhen Gsd Tech Co., Ltd. · OHT · May 11, 2018 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT/K180383

## Device Facts

- **Applicant:** Shenzhen Gsd Tech Co., Ltd.
- **Product Code:** [OHT](/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT.md)
- **Decision Date:** May 11, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

Light based hair removal device is an over-the-counter device intended for removal of unwanted hair such as but not limited to small areas such as underarm and facial hair below the chin line and large areas such as legs, excluding patients with Fitzpatrick Skin Phototypes VI.

## Device Story

Personal, over-the-counter light-based hair reduction device; uses Xenon flashlamp technology to deliver optical radiation to skin. Device consists of main body, detachable lamp cap, and external power adapter. User operates device via finger switch to deliver direct illumination to tissue. Principle of operation involves selective photothermolysis for hair reduction. Intended for home use by consumers. Output is light pulse; healthcare providers do not interact with device output; device enables self-administered hair removal, potentially reducing need for professional clinical treatments.

## Clinical Evidence

No clinical data. Evidence consists of bench testing, including biocompatibility (ISO 10993-5, ISO 10993-10), electrical safety (IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-57), photobiological safety (IEC 62471), and software verification and validation.

## Technological Characteristics

Xenon arc flashlamp source; energy density 4.2 J/cm2 max; pulse duration 2-3 ms; spot size 1.05cm x 2.95cm. Powered by external adapter. Biocompatibility per ISO 10993-1. Electrical safety per IEC 60601 series. Photobiological safety per IEC 62471. Software level of concern: moderate.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Perfectsmooth ([K161428](/device/K161428.md))
- iPulse SmoothSkin Gold Hair Removal System ([K160968](/device/K160968.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 11, 2018

Shenzhen GSD Tech Co., Ltd % Rain Yip Registered Engineer Feiying Drug & Medical Consulting Technical Service Group Rm. 3005, Area B, Bldg.1, Southward Ruifeng Business Center Guimiao Road, S Shenzhen, 518000 Cn

Re: K180383

Trade/Device Name: Light based hair removal device Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHT Dated: January 31, 2018 Received: February 12, 2018

Dear Rain Yip:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm_for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

# Sincerely. Jennifer R. Stevenson -53

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K180383

Device Name Light based hair removal device

Indications for Use (Describe)

Light based hair removal device is an over-the-counter device intended for removal of unwanted hair such as but not limited to small areas such as underarm and facial hair below the chin line and large areas such as legs, excluding patients with Fitzpatrick Skin Phototypes VI.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

### "510(k) Summary" as required by 21 CFR Part 807.92.

### Date: 2018-01-31

#### I. Submitter

SHENZHEN GSD TECH CO., LTD BUILDING A JUNSD HI-TECH PARK, WEST OF BAO'AN RD., WATCH & CLOCK BASE, GUANGMING DISTRICT, SHENZHEN, GUANGDONG, CHINA Tel.: +86 755 2916 9431 ext.8040 Tax: +86 755 2910 9786

Ziyi Gui Regulatory Engineer Tel: +86 130 7784 1845 Email: gzy@gsd.com.cn

### II. Device

Name of Device: Light based hair removal device Common or Usual Name: Light Based Over-The-Counter Hair Removal Classification Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: OHT Regulation Number: 21 CFR 878.4810

# III. Predicate Device

The predicate devices are listed as below:

| Manufacturer                    | Predicate Device                              | 510(k) Number | Approval Date  |
|---------------------------------|-----------------------------------------------|---------------|----------------|
| Shen Zhen CosBeauty<br>Co., Ltd | Perfectsmooth                                 | K161428       | March 23, 2017 |
| CyDen Limited                   | iPulse SmoothSkin Gold Hair<br>Removal System | K160968       | April 04, 2016 |

# IV. Device Description

The Light based hair removal device is a personal, light-based, hair reduction device intended to be sold over-the-counter directly to the end user. The device provides hair reduction using IPL

{4}------------------------------------------------

technology. The device consists of IPL main body and adapter two parts, and a detachable lamp cap located in the main body which is the source of optical radiation, namely a Xenon flashlamp. Meanwhile, the device is power from adapter via an external power.

# V. Indications for Use

Light based hair removal device is an over-the-counter device intended for removal of unwanted hair such as but not limited to small areas such as underarm and facial hair below the chin line and large areas such as legs, excluding patients with Fitzpatrick Skin Phototypes VI.

# VI. Comparison of Technological Characteristics With the Predicate Device

The Light based hair removal device has the same intended use, mode of action and similar operational characteristics as the predicate device. Any minor differences between the subject device and the listed predicate devices do no raise any issues of safety or efficacy. Performance data supports that the device is safe and as effective as the predicate device for its intended use. Therefore, the Light based hair removal device may be found substantially equivalent to its predicate device.

Light based hair removal device is compared with the following Predicate Devices in terms of intended use, design, material, specifications, and performance.

1) K161428, "PerfectSmooth", manufactured by "Shen Zhen CosBeauty Co., Ltd" in Guangdong, China

| Comparison<br>Elements | Subject Device                                | Predicate Device 1                            | Predicate Device 2                               |
|------------------------|-----------------------------------------------|-----------------------------------------------|--------------------------------------------------|
| K Number               | Pending                                       | K161428                                       | K160968                                          |
| Common name            | Light Based Over-The-<br>Counter Hair Removal | Light Based Over-The-<br>Counter Hair Removal | Light Based Over-The-<br>Counter Hair Removal    |
| Trade name             | Light based hair<br>removal device            | PerfectSmooth                                 | iPulse SmoothSkin<br>Gold Hair Removal<br>System |
| Energy medium          | Xenon Arc Flashlamp                           | Xenon Arc Flashlamp                           | Xenon Arc Flashlamp                              |
| Pulse duration         | 2-3 milliseconds                              | 11-13 milliseconds                            | 2-10 milliseconds                                |
| Energy density         | 4.2 J/cm2Max.                                 | 4.7J/cm2Max.                                  | 3~6 J/cm2                                        |
| Spot size              | 1.05cm×2.95cm<br>(3.1 cm3)                    | 4.5 cm2                                       | 3 (3cm by 1cm)                                   |
| Delivery device        | Direct illumination to                        | Direct illumination                           | Direct illumination                              |

2) K160968. "iPulse SmoothSkin Gold Hair Removal System ", manufactured by "CyDen Limited" in Wales, UK

{5}------------------------------------------------

| Comparison<br>Elements                | Subject Device                                                                                                                                                                                                                                                                                                            | Predicate Device 1                                                                                                                                                                                                                            | Predicate Device 2                                                                                                                                                                                                                                                                                                                                                                                |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                       | tissue                                                                                                                                                                                                                                                                                                                    | tissue                                                                                                                                                                                                                                        | tissue                                                                                                                                                                                                                                                                                                                                                                                            |
| Pulsing control                       | Finger switch                                                                                                                                                                                                                                                                                                             | Finger switch                                                                                                                                                                                                                                 | Finger switch                                                                                                                                                                                                                                                                                                                                                                                     |
| Indication<br>for<br>use/Intended use | Light based hair<br>removal device is an<br>over-the-counter device<br>intended for removal of<br>unwanted hair such as<br>but not limited to small<br>areas such as underarm<br>and facial hair below<br>the chin line and large<br>areas such as legs,<br>excluding patients with<br>Fitzpatrick Skin<br>Phototypes VI. | The PerfectSmooth is<br>an<br>over-the-counter<br>device intended for<br>removal of unwanted<br>hair such as but not<br>limited to small areas<br>such as underarm and<br>facial hair below the<br>chin line and large areas<br>such as legs. | The iPulse SmoothSkin<br>Gold Hair Removal<br>System is indicated for<br>the removal of<br>unwanted hair. The<br>iPulse Smoothskin Gold<br>is also indicated for the<br>permanent reduction in<br>hair regrowth, defined<br>as the long-term, stable<br>reduction in the number<br>of hairs regrowing<br>when measured at 6, 9<br>and 12 months after the<br>completion of a<br>treatment regime. |
| Location for use                      | OTC                                                                                                                                                                                                                                                                                                                       | OTC                                                                                                                                                                                                                                           | OTC                                                                                                                                                                                                                                                                                                                                                                                               |

# VII.Performance Data

The following performance data were provided in support of the substantial equivalence determination.

#### 1) Biocompatibility Testing

The biocompatibility evaluation for the body-contacting components of the Light based hair removal device was conducted in accordance with the "Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices -Part 1: Evaluation and Testing Within a Risk Management Process, Document Issued on June 16, 2016", as recognized by FDA. The battery of testing was performed to, and passed, including:

- > ISO 10993-5:2009/(R)2014, Biological Evaluation of Medical Devices –Par t 5: Tests for In Vitro Cytotoxicity
- > ISO 10993-10:2010/(R)2014, Biological Evaluation of Medical Devices –Par t 10: Tests for Irritation and Skin Sensitization

# 2) Electrical Safety and Eye Safety

{6}------------------------------------------------

Electrical safety and Eye safety testing was performed to, and passed, the following standards:

- > IEC 60601-1-2 Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance -Collateral standard: electromagnetic compatibility
- > IEC 60601-1 Medical electrical equipment -Part 1: General requirements for basic safety and essential performance
- > IEC 60601-1-11 Medical Electrical Equipment -Part 1: General Requirements for Basic Safety and Essential Performance -Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment
- > IEC 60601-2-57 Medical electrical equipment -Part 2-57: Particular requirements for the basic safety and essential performance of non-laser source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use

# 3) Eye Safety

- > IEC 62471 Photobiological safety of lamps and lamp systems
### 4) Software Verification and Validation

Software documentation consistent with moderate level of concern was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels.

#### Summary

Based on the above performance as documented in this application, Light based hair removal device was found to have a safety and effectiveness profile that is similar to the predicate device.

#### VIII. Conclusions

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the comparison of intended use, design, materials and performance, the Light based hair removal device is to be concluded substantial equivalent to its predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT/K180383](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT/K180383)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
