← Product Code [OHT](/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT) · K140527

# SENSILIGHT MINI (K140527)

_El Global Trade, Ltd. · OHT · Jul 31, 2014 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT/K140527

## Device Facts

- **Applicant:** El Global Trade, Ltd.
- **Product Code:** [OHT](/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT.md)
- **Decision Date:** Jul 31, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

The sensiLight Mini is an over the counter devices intended for the removal of unwanted hair. The sensiLight Mini is also intended for permanent reduction in hair regrowth, defined as a long-term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of treatment regimen.

## Device Story

The sensiLight Mini is a handheld, over-the-counter pulsed light device for hair removal. It utilizes selective photothermolysis, emitting optical energy (475–1200nm) to disable hair growth. The device consists of an applicator and external power supply. It features integrated skin proximity and skin pigmentation sensors; these sensors prevent pulse triggering if the device is not in full contact with skin or if the user's skin tone is too dark/tanned. Operated by the user in a home environment, the device provides a 3 cm2 treatment area. The user applies the device to the skin; the system automatically verifies safety conditions before enabling light emission. This process aims to provide long-term, stable reduction in hair regrowth. The device is designed for intuitive use without professional training, relying on provided instructions and labeling.

## Clinical Evidence

No clinical trials were conducted. Evidence consists of bench testing and a usability study. The usability study enrolled 20 potential end users with varying skin tones and educational backgrounds in a simulated home environment. All 20 subjects (100%) successfully completed all tasks, including device operation and self-selection based on labeling, demonstrating the device can be used safely and effectively by lay users.

## Technological Characteristics

Intense Pulsed Light (IPL) device. Wavelength: 475–1200nm. Max energy: 5 J/cm2. Features skin proximity and pigmentation sensors. Handheld applicator with lamp, microcontroller, fan, and indicator LEDs. Biocompatibility per ISO 10993-1:2009. Electrical safety per IEC 60601-1, IEC 60601-1-11, and IEC 60601-2-57. Electromagnetic compatibility per IEC 60601-1-2. Software validation per IEC 62304.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Home Skinovations Ltd.'s Glide device ([K131870](/device/K131870.md))
- Home Skinovations Ltd.'s Silk'n Flash N Go ([K103184](/device/K103184.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Tzoran 4th St, P.O.Box 8242, Netanya 42504, Israel EL Global Trade Ltd Tel: +972-9-7889069 Fax: +972-9-7734831 sensiLight Mini – 510K Hold Letter Respond RD-17025 A0

# 510(K) SUMMARY FOR EL GLOBAL TRADE LTD.'S SENSILIGHT MINI

DATE PREPARED: JUNE 19TH, 2014

- 1. 510(K) OWNER NAME
EL Global Trade Ltd. Tzoran 4th st, P.O.Box 8242, Netanya 42504, Israel. Phone: +972-9-7889069, Fax: +972-9-7734831.

Contact person name: Dr. Shlomit Segman, QA and RA manager Phone: +972-9-7889069, Fax: +972-9-7734831, E mail: Shlomit@elglobalt.com

#### 2. DEVICE NAME

Common/Usual Name: Light based hair removal deices Proprietary/Trade name: sensiLight Mini

Classification: EL Global Trade Ltd.'s sensiLight Mini device has been classified as Class II device under the following classification names:

| Classification Name                           | Product<br>Code | Regulation<br>Number | Panel                          |
|-----------------------------------------------|-----------------|----------------------|--------------------------------|
| Light Based Over-The-<br>Counter Hair Removal | OHT             | 878.4810             | General and Plastic<br>Surgery |

#### 3. PREDICATE DEVICES

EL Global Trade Ltd.'s sensiLight Mini device is substantially equivalent to the following Predicate Device:

- 3.1 Home Skinovations Ltd.'s Glide device, cleared under 510(k) number K131870 at August 14th, 2013.
- 3.2 Home Skinovations Ltd.'s Silk'n Flash N Go,

cleared under 510(k) number K103184 at November 10th, 2010.

3 1 2014

{1}------------------------------------------------

| EL Global Trade Ltd                        |  | Tzoran 4th St, P.O.Box 8242, Netanya 42504, Israel<br>Tel: +972-9-7889069       Fax: +972-9-7734831 |  |
|--------------------------------------------|--|-----------------------------------------------------------------------------------------------------|--|
| sensiLight Mini - 510K Hold Letter Respond |  | RD-17025 A0                                                                                         |  |

### DEVICE DESCRIPTION

The sensiLight Mini device is pulsed light hair removal device. Light-based hair removal is based on the theory of selective photothermolysis in which optical energy is used to disable hair growth. The sensiLight Mini device is composed of a hand held applicator and an external power supply. The spot size (treatment area) in the sensiLight Mini device is 3 cm2.

The device contains a lamp, a skin proximity sensor and a skin pigmentation sensor to detect appropriate skin tones. If the sensiLight Mini is not properly applied (in full contact with the skin) or user skin tone is too dark/tanned, the sensiLight Mini will not trigger a pulse.

Body contact materials were evaluated for biocompatibility with accordance to FDA 's Memorandum - #G95 1, May 1, 1995 and ISO 10993-1:2009.

#### 4. INTENDED USE

The sensiLight Mini is an over the counter devices intended for the removal of unwanted hair. The sensiLight Mini is also intended for permanent reduction in hair regrowth, defined as a long-term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of treatment regimen.

#### ਨ. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE

The sensiLight Mini device relies on the same technology: Intense Pulsed Light (IPL) which is the same as both predicate devices. The safety and efficacy of IPL treatment for hair reduction are governed by the following parameters:

- . Wavelength of the light emitted: Defines the interaction with specific chromophores (the part of the molecule responsible for its color) such as melanin, hemoglobin and water. The sensiLight Mini device and the predicate devices utilize exactly the same wavelength (475 - 1200nm).
- . Fluence/flux - defines the energy per area (e.g. joules per cm³) for the treatment. The sensiLight Mini device and the predicate devices deliver exactly the same maximum energy (5 joules/cm2).
- Pulse duration Provides for an efficient heating of the target molecule but not its . surroundings. The sensiLight Mini device and the predicate devices utilize exactly the same pulse.

{2}------------------------------------------------

| EL Global Trade Ltd                        | Tzoran 4th St, P.O.Box 8242, Netanya 42504, Israel<br>Tel: +972-9-7889069        Fax: +972-9-7734831 |
|--------------------------------------------|------------------------------------------------------------------------------------------------------|
| sensiLight Mini – 510K Hold Letter Respond | RD-17025 A0                                                                                          |

#### 6. PERFORMANCE DATA

EL Global Trade Ltd's sensiLight has been successfully tested through bench, usability and safety tests to support the determination of substantial equivalence with predicate devices.

The sensiLight Mini device has been tested and complies with the following voluntary recognized standards:

- 1. IEC 60601-1:2005/EN 60601-1:2006. Medical electric equipment-Part 1 General requirements for Basic safety and essential performance, 3th Ed.
- 2. IEC 60601-1-11:2010. Medical electric equipment-Part 1: Collateral requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
- 3. IEC 60601-2-57:2011. Medical electrical equipment-Part 2: Particular requirements for basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetics/aesthetic use.
- 4. FCC part 15. Subpart B, Class B.
- 5. IEC 60601-1-2:2007 Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance – Collateral standard: Electromagnetic compatibility - Requirements and tests.
- 6. Software Validation was conducted according to IEC 62304:2006 Medical device software - Software life cycle processes, and; FDA Guidance for the Content of Pre-Market Submissions for Software Contained in Medical Devices, dated May 11, 2005.
- 7. ISO 10993-1:2009 Biological evaluation of medical devices Part 1; Evaluation and testing within a risk management process and with FDA's Memorandum ~ #G95 1, May 1, 1995, Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing" (Blue Book Memo G95-1).
- 8. IEC 60601-1-6:2010, Medical electrical equipment Part 1-6: General requirements for basic safety and essential performance and IEC 62366:2007 - Medical devices -Application of usability engineering to medical devices - Collateral stundard: Usability and FDA Guidance for Medical Device Use-Safety: Incorporating Human Factors Engineering into Risk Management (July18,2000) and ISO/IEC 14971:2007 (BS EN ISO 14971:2012) Medical devices - Application of risk management to medical devices.

Tests results are supporting all labeling claims in order to establish substantial equivalency.

{3}------------------------------------------------

| EL Global Trade Ltd                        | Tzoran 4th St, P.O.Box 8242, Netanya 42504, Israel<br>Tel: +972-9-7889069 Fax: +972-9-7734831 |
|--------------------------------------------|-----------------------------------------------------------------------------------------------|
| sensiLight Mini – 510K Hold Letter Respond | RD-17025 A0                                                                                   |

#### 7. NON-CLINICAL (BENCH) PERFORMANCE DATA

The objective of the Usability Study was to test the sensiLight Mini device self-selection and usability, i.e., the safe and effective device use, by potential end users, under actual use conditions.

Twenty (20) potential device end users with different Fitzpatrick skin tone and with different educational level were enrolled in the Usability Study. To strengthen the assessment that the sensiLight Mini device is intuitive and no specific training is required besides reading the Instructions for Use, self-selection was conducted in order to measure the ability of lay users to determine whether they can use the devices (based on the box labeling and instruction for use). Three participants were not enrolled to this study after self-selection screening.

The sensiLight Mini device, along with box labelling and instruction for use was provided to the participants in a simulated home use environment. The participants labeling was in the content intended for distribution. All subjects were provided with a list of tasks to complete, including applying and operating the device.

All of the 20 enrolled subjects (100%) completed all tasks successfully. The sensiLight Mini device, along with box labelling and instruction for use. can be used safely and effectively by potential end users, under actual use conditions.

#### 8. SUBSTANTIAL EQUIVALENCE

EL Global Trade Ltd.'s sensiLight Mini is substantial equivalent to the predicate device selected in terms of indication for use, technology, performances, place of use, patient population and nature of body contact.

The Substantial equivalent decision was received based on the following comparison with the predicate device:

The design and components in the sensiLight Mini, including the hand held applicator (with lamp, microcontroller, fan, skin color sensor, skin proximity sensor, indicator LEDs and operational button/s) are similar to the design and components found in the predicate devices (K131870 and K103184). The performance specifications (including light energy power, wavelength and pulse duration) are identical. The safety features found in the devices are the same, including the skin color sensor, skin proximity sensor, etc. These safety features in the sensiLight Mini device are substantially equivalent to the safety features found in the predicate devices. Any minor differences in the technological characteristics do not raise new safety or effectiveness concerns. Furthermore, the new sensiLight Mini device underwent performance testing, including software validation testing and electrical and mechanical safety testing according to IEC 60601-1 and

..............................................................................................................................................................................

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ EL Global Trade Ltd. 510(k) sensiLight Mini

CONFIDENTIAL

{4}------------------------------------------------

| EL Global Trade Ltd<br>Tzoran 4th St, P.O.Box 8242, Netanya 42504, Israel<br>Tel: +972-9-7889069 Fax: +972-9-7734831 |             |
|----------------------------------------------------------------------------------------------------------------------|-------------|
| sensiLight Mini – 510K Hold Letter Respond                                                                           | RD-17025 A0 |

electromagnetic compatibility testing according to IEC 60601-1-2. Usability testing was also successfully conducted.

#### 9. CONCLUSIONS

The evaluation of our device performances demonstrates that it is as safe and as effective as the predicate devices.

... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

y == = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = =

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 31, 2014

El Global Trade Ltd. Dr. Shlomit Segman Quality Assurance and Regulatory Affairs Manager Tzoran 4th Street, P.O. Box 8242 Netanya 42504. Israel

Re: K140527

Trade/Device Name: sensiLight Mini Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: OHT Dated: June 25, 2014 Received: July 3, 2014

Dear Dr. Segman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{6}------------------------------------------------

Page 2 - Dr. Shlomit Segman

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# David Krause -S

- for
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

## Indications for Use

510(k) Number (if known) K140527

Device Name sensiLight Mini

Indications for Use (Describe)

The sensiLight Mini is an over the counter device intended for the removal of unwanted hair. The sensiLight Mini is also intended for permanent reduction in hair regrowth, defined as a long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of treatment regimen.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

## FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

# Neil R Ogden -S 2014.07.31 15:45:01 -04'00'

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT/K140527](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT/K140527)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).