← Product Code [OHT](/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT) · K140381

# SENSILIGHT/ SENSILIGHT PLUS (K140381)

_El Global Trade, Ltd. · OHT · Aug 8, 2014 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT/K140381

## Device Facts

- **Applicant:** El Global Trade, Ltd.
- **Product Code:** [OHT](/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT.md)
- **Decision Date:** Aug 8, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

The sensiLight™ Plus device is an over-the-counter device intended for the removal of unwanted hair. The sensiLight™/sensiLight™ Plus is also intended for permanent reduction in hair regrowth, defined as a long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of treatment regimen.

## Device Story

sensiLight/sensiLight Plus are handheld, home-use pulsed light hair removal devices. They utilize selective photothermolysis; optical energy (475-1200nm) is absorbed by hair follicles to interrupt growth cycles. The device consists of a handheld applicator and external power supply. Safety features include a skin proximity sensor and a skin pigmentation sensor to prevent pulse activation on inappropriate skin tones or if not in full contact with skin. The 'Plus' model includes a movement sensor to prevent pulse overlapping. Users operate the device at home to treat unwanted hair. The device provides a 3 cm² treatment area. Clinical benefit is defined as long-term, stable reduction in hair regrowth measured at 6, 9, and 12 months post-treatment.

## Clinical Evidence

Bench testing only. No clinical data provided. Evidence includes performance testing, software validation, electrical/mechanical safety testing (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and usability engineering studies.

## Technological Characteristics

Handheld IPL device; ABS/PC plastic shell; optical long-pass glass filter. Energy source: external power supply (110-240 VAC). Wavelength: 475-1200nm; Max fluence: 5 J/cm²; Pulse duration: 500-800 µsec. Features: skin proximity sensor, skin pigmentation sensor, movement sensor (Plus model). Biocompatibility per ISO 10993-1. Software validation per IEC 62304.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Home Skinovations Ltd.'s Glide device ([K131870](/device/K131870.md))
- Home Skinovations Ltd.'s Silk'n Flash N Go ([K103184](/device/K103184.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized depiction of three human profiles facing to the right, layered on top of each other. The profiles are black and are set against a white background. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the profile graphic.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 8, 2014

EL Global Trade Ltd. Dr. Shlomit Segman Quality Assurance and Regulatory Affairs Manager Tzoran 4th Street P.O. Box 8242 Netanya, 42504 - Israel

Re: K140381

Trade/Device Name: sensiLight/sensiLight plus Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: OHT Dated: June 15, 2014 Received: July 14, 2014

Dear Dr. Segman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{1}------------------------------------------------

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# David Krause -S

for

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K140381

Device Name sensiLight/sensiLight Plus

Indications for Use (Describe)

The sensiLight™ Plus device is an over-the-counter device intended for the removal of unwanted hair. The sensiLight™/sensiLight™ Plus is also intended for permanent reduction in hair regrowth, defined as a long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of treatment regimen.

### Type of Use (Select one or both, as applicable)

- Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

# Neil R Ogden -S 2014.08.07 11:38:18 -04'00

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

sensiLight – 510K submission

RD-16044 A0

## 510(K) SUMMARY FOR EL GLOBAL TRADE LTD.'S SENSILIGHT

### DATE PREPARED: FEB 10TH, 2014

#### 1. 510(K) OWNER NAME

EL Global Trade Ltd. Tzoran 4th st, P.O.Box 8242, Netanya 42504, Israel. Phone: +972-9-7889069, Fax: +972-9-7734831.

Contact person name: Dr. Shlomit Segman, QA and RA manager Phone: +972-9-7889069, Fax: +972-9-7734831, E mail: Shlomit@elglobalt.com

#### 2. DEVICE NAME

Common/Usual Name: Light based hair removal deices Proprietary/Trade name: sensiLight (sensiLight and sensiLight Plus) Classification: EL Global Trade Ltd.'s sensiLight device has been classified as Class II device under the following classification names:

| Classification Name                           | Product<br>Code | Regulation<br>Number | Panel                          |
|-----------------------------------------------|-----------------|----------------------|--------------------------------|
| Light Based Over-The-<br>Counter Hair Removal | OHT             | 878.4810             | General and Plastic<br>Surgery |

#### 3. PREDICATE DEVICES

EL Global Trade Ltd.'s sensiLight device is substantially equivalent to the following Predicate Device:

- 3.1 Home Skinovations Ltd.'s Glide device, cleared under 510(k) number K131870 at August 14th, 2013.
- 3.2 Home Skinovations Ltd.'s Silk'n Flash N Go,

cleared under 510(k) number K103184 at November 10th, 2010.

### DEVICE DESCRIPTION

{4}------------------------------------------------

| EL Global Trade Ltd          | Tzoran 4th St, P.O.Box 8242, Netanya 42504, Israel<br>Tel: +972-9-7889069 Fax: +972-9-7734831 |
|------------------------------|-----------------------------------------------------------------------------------------------|
| sensilight - 510K submission | RD-16044-A0                                                                                   |

sensiLight - 510K submission

The sensiLight devices are pulsed light hair removal device. Light-based hair removal is based on the theory of selective photothermolysis in which optical energy is used to disable hair growth. The sensiLight devices are composed of a hand held applicator and an external power supply. The spot size (treatment area) in the sensiLight devices is 3 cm².

The device contains a lamp, a skin proximity sensor and a skin pigmentation sensor to detect appropriate skin tones. If the sensiLight is not properly applied (in full contact with the skin) or user skin tone is too dark/tanned, the sensiLight will not trigger a pulse.

(*) The sensiLight Plus model contains in addition, a movement sensor to prevent pulse overlapping.

Body contact materials were evaluated for biocompatibility with accordance to FDA's Memorandum - #G95 1, May 1, 1995 and ISO 10993-1:2009.

#### 4. INTENDED USE

The sensiLight is an over the counter devices intended for the removal of unwanted hair. The sensiLight is also intended for permanent reduction in hair regrowth, defined as a long-term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of treatment regimen.

#### TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE 5.

Table 1 presented in the next pages, compares the similarities and differences of EL Global Trade Ltd.'s sensiLight device to the predicate devices manufactured by Home Skinovations Ltd. as identified above.

{5}------------------------------------------------

## EL Global Trade Ltd

# sensiLight – 510K submission

RD-16036 A0

| Feature                             | EL Global Trade Ltd.<br><i>sensiLight &amp; sensiLight Plus</i><br>-New Devices -                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Home Skinovations Ltd.<br><i>Glide</i><br>510(k) no. K131870                                                                                                                                                                                                                                                                                                               | Home Skinovations Ltd.<br><i>Silk'n Flash N Go</i><br>510(k) no. K103184                                                                                                                                                                                  |
|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intend of use                       | The sensiLight is an over the counter device intended for the removal of unwanted hair. The sensiLight is also intended for permanent reduction in hair regrowth, defined as a long-term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of treatment regime.                                                                                                                                                                                                                                                                     | The Glide™ device is an over the counter device intended for the removal of unwanted hair. The Glide™ device is also intended for permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.                                       | The Flash N Go device is an over the counter device intended for the removal of unwanted hair. Flash N Go is also intended for permanent reduction in hair regrowth defined as a long-term, stable reduction in hair counts following a treatment regime. |
| Operation Principle                 | Photothermolysis. Light energy is absorbed, heats the hair follicles and interrupts the growth cycle.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Photothermolysis. Light energy is absorbed, heats the hair follicles and interrupts the growth cycle.                                                                                                                                                                                                                                                                      | Photothermolysis. Light energy is absorbed, heats the hair follicles and interrupts the growth cycle.                                                                                                                                                     |
| Technology                          | Home Pulsed Light                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Home Pulsed Light                                                                                                                                                                                                                                                                                                                                                          | Home Pulsed Light                                                                                                                                                                                                                                         |
| Comparison of treatment regime      | • The 1-4 treatments will be approximately two weeks apart<br>• Followed by 5-7 treatments plan four-week time intervals.<br>• Maintenance will be needed from time to time if growth is still visible                                                                                                                                                                                                                                                                                                                                                                                          | • Treatments 1-4 – plan two weeks apart<br>• Treatment 5-7 – plan four weeks apart<br>• Treatment 8+ - treat as needed, until desired results are achieved.<br>Maintenance – from time to time some upkeep may be needed if growth is still visible.                                                                                                                       | • First 3-4 hair removal sessions will be approximately two weeks apart.<br>• Sessions 5-7 will be approximately four weeks apart.<br>• After, use again from time to time if and when needed, until long-term results are achieved.                      |
| Safety                              | ○IEC 60601-1:2005/EN 60601-1:2006. Medical electric equipment-Part 1 General requirements for Basic safety and essential performance.                                                                                                                                                                                                                                                                                                                                                                                                                                                           | ○IEC 60601-1, Medical Electrical Equipment - Part 1: General Requirements for Safety, 1988; Amendment 1,                                                                                                                                                                                                                                                                   | ○IEC 60601-1, Medical Electrical Equipment - Part 1: General Requirements for Safety, 1988 (Second Edition).                                                                                                                                              |
| Feature                             | EL Global Trade Ltd.<br>sensiLight & sensiLight<br>Plus                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Home Skinovations<br>Ltd.<br>Glide                                                                                                                                                                                                                                                                                                                                         | Home Skinovations<br>Ltd.<br>Silk'n Flash N Go                                                                                                                                                                                                            |
|                                     | -New Devices -                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 510(k) no. K131870                                                                                                                                                                                                                                                                                                                                                         | 510(k) no. K103184                                                                                                                                                                                                                                        |
|                                     | oIEC 60601-1-11:2010.<br>Medical electric<br>equipment-Part 1:<br>Collateral requirements<br>for medical electrical<br>equipment and medical<br>electrical systems used in<br>the home healthcare<br>environment.<br>oIEC 60601-2-57:2011.<br>Medical electrical<br>equipment-Part 2:<br>Particular requirements<br>for basic safety and<br>essential performance of<br>non-laser light source<br>equipment intended for<br>therapeutic, diagnostic,<br>monitoring and<br>cosmetics/aesthetic use<br>oIEC 62471:2006 (First<br>Edition) Photo-biological<br>safety of lamps and lamp<br>systems | 1995.<br>oEN 6060 1-2-57: 2011<br>Medical electrical<br>equipment - Part 2-57:<br>Particular requirements<br>for the basic safety and<br>essential performance of<br>non-laser light source<br>equipment intended for<br>therapeutic, diagnostic,<br>monitoring and cosmeti<br>c/aesthetic use.<br>oEN 62471: 2008<br>Photobiological safety of<br>lamps and lamp systems. |                                                                                                                                                                                                                                                           |
| EMC                                 | IEC 60601-1-2:2013<br>(Medical electrical<br>equipment -fourth Edition).<br>Part 1-2: General<br>requirements for basic<br>safety and essential<br>performance - Collateral<br>standard: Electromagnetic<br>compatibility -<br>Requirements and tests                                                                                                                                                                                                                                                                                                                                           | IEC 60601-1-2 (2007),<br>Medical Electrical<br>Equipment - Part 1-2:<br>General<br>Requirements for Safety -<br>Collateral Standard:<br>Electromagnetic<br>Compatibility --<br>Requirements and Tests                                                                                                                                                                      | IEC 60601-1-2 (2004),<br>Medical Electrical<br>Equipment - Part 1-2:<br>General<br>Requirements for Safety -<br>Collateral Standard:<br>Electromagnetic<br>Compatibility --<br>Requirements and Tests                                                     |
| Performance Features                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                           |
| Optical<br>aperture                 | $3 [cm^2]$                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | $2.7 [cm^2]$                                                                                                                                                                                                                                                                                                                                                               | $6 [cm^2]$                                                                                                                                                                                                                                                |
| Emitted<br>Spectrum                 | $475-1200 [nm]$                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | $475-1200 [nm]$                                                                                                                                                                                                                                                                                                                                                            | $475-1200 [nm]$                                                                                                                                                                                                                                           |
| Max energy<br>level                 | $5 [joules/cm^2]$                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | $5 [joules/cm^2]$                                                                                                                                                                                                                                                                                                                                                          | $5 [joules/cm^2]$                                                                                                                                                                                                                                         |
| Feature                             | EL Global Trade Ltd.<br>sensiLight & sensiLight<br>Plus                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Home Skinovations<br>Ltd.<br>Glide                                                                                                                                                                                                                                                                                                                                         | Home Skinovations<br>Ltd.<br>Silk'n Flash N Go                                                                                                                                                                                                            |
|                                     | -New Devices -                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 510(k) no. K131870                                                                                                                                                                                                                                                                                                                                                         | 510(k) no. K103184                                                                                                                                                                                                                                        |
| (fluence)                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                           |
| User<br>selectable<br>Energy levels | 3                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | 5                                                                                                                                                                                                                                                                                                                                                                          | 5                                                                                                                                                                                                                                                         |
| Single pulse<br>duration            | 500-800 [uSec]                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 500-800 [uSec]                                                                                                                                                                                                                                                                                                                                                             | 500-800 [uSec]                                                                                                                                                                                                                                            |
| Interval<br>between<br>pulses       | ~2 [Sec]                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 1-3.4 [Sec], depending<br>on energy level                                                                                                                                                                                                                                                                                                                                  | ~ 3.5 [Sec]                                                                                                                                                                                                                                               |
| Power<br>Supply                     | External power supply<br>110-240 VAC, 50-60Hz<br>to 19VDC, 1.8Amp                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | External power supply 110-<br>240 VAC, 2Amp                                                                                                                                                                                                                                                                                                                                | 110-240 [VAC], 50-60<br>([Hz])                                                                                                                                                                                                                            |
|                                     | Safety Features                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                           |
| Close contact<br>detector           | Prevent the activation of<br>the device while not in<br>close contact with the<br>skin.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                                                                                                                                                                        | Yes                                                                                                                                                                                                                                                       |
| Pigmentation                        | Optical - does not allow                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Optical - does not allow                                                                                                                                                                                                                                                                                                                                                   | Optical - does not allow                                                                                                                                                                                                                                  |
| level detector                      | activation on dark skin                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | activation on dark skin                                                                                                                                                                                                                                                                                                                                                    | activation on dark skin                                                                                                                                                                                                                                   |
| (*)<br>Movement<br>sensor           | In the sensiLight Plus<br>(Second model)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | No                                                                                                                                                                                                                                                                                                                                                                         | No                                                                                                                                                                                                                                                        |
| Darkest skin<br>allowed             | Fitzpatrick 4                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Fitzpatrick 4                                                                                                                                                                                                                                                                                                                                                              | Fitzpatrick 4                                                                                                                                                                                                                                             |
| Materials                           | oTop, Side and Bottom<br>shell - ABS/PC plastic<br>(Acrylonitrile butadiene<br>styrene/PolyCarbonate)<br>oOptical Filter - Color<br>glass which purpose is of<br>an optical long pass filter<br>tested for<br>biocompatibility                                                                                                                                                                                                                                                                                                                                                                  | oDevice outer body -<br>ABS/PC plastic<br>(Acrylonitrile butadiene<br>styrene/PolyCarbonate)<br>oOptical Filter - Color<br>glass which purpose is of<br>an optical long pass filter<br>tested for<br>biocompatibility                                                                                                                                                      | oDevice outer body -<br>ABS/PC plastic<br>(Acrylonitrile butadiene<br>styrene/PolyCarbonate)<br>oOptical Filter - Color<br>glass which purpose is of<br>an optical long pass filter<br>tested for<br>biocompatibility                                     |
| Overall Design                      | The sensiLight device is<br>light based hair removal<br>system composed of a<br>hand-held applicator.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | The Glide device is light<br>based hair removal<br>system composed of a<br>hand held applicator.                                                                                                                                                                                                                                                                           | The Flash N Go is light<br>based hair removal<br>system composed of a<br>base unit and hand-held<br>applicator.                                                                                                                                           |

{6}------------------------------------------------

## EL Global Trade Ltd

# sensiLight – 510K submission

RD-16036 A0

......... EL Global Trade Ltd. 510(k) sensiLight

CONFIDENTIAL

{7}------------------------------------------------

## EL Global Trade Ltd

# sensiLight – 510K submission

RD-16036 A0

{8}------------------------------------------------

| EL Global Trade Ltd |
|---------------------|
|---------------------|

sensiLight - 510K submission

RD-16036 A0

| Feature | EL Global Trade Ltd.<br>sensiLight & sensiLight<br>Plus<br>-New Devices - | Home Skinovations<br>Ltd.<br>Glide<br>510(k) no. K131870 | Home Skinovations<br>Ltd.<br>Silk'n Flash N Go<br>510(k) no. K103184 |
|---------|---------------------------------------------------------------------------|----------------------------------------------------------|----------------------------------------------------------------------|
| Use     | Home Use                                                                  | Home Use                                                 | Home Use                                                             |

The sensiLight device relies on the same technology: Intense Pulsed Light (IPL) which is the same as both predicate devices. The safety and efficacy of IPL treatment for hair reduction are governed by the following parameters:

- Wavelength of the light emitted: Defines the interaction with specific chromophores (the ● part of the molecule responsible for its color) such as melanin, hemoglobin and water. The sensiLight device and the predicate devices utilize exactly the same wavelength (475 - 1200nm).
- Fluence/flux defines the energy per area (e.g. joules per cm²) for the treatment. The ● sensiLight device and the predicate devices deliver exactly the same maximum energy (5 joules/cm2).
- . Pulse duration - Provides for an efficient heating of the target molecule but not its surroundings. The sensiLight device and the predicate devices utilize exactly the same pulse.

#### PERFORMANCE DATA 6.

EL Global Trade Ltd's sensiLight has been successfully tested through bench, usability and safety tests to support the determination of substantial equivalence with predicate devices.

The sensiLight device has been tested and complies with the following voluntary recognized standards:

- 1. IEC 60601-1:2005/EN 60601-1:2006. Medical electric equipment-Part 1 General requirements for Basic safety and essential performance, 300 Ed.
- 2. IEC 60601-1-11:2010. Medical electric equipment-Part 1: Collateral requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
- 3. IEC 60601-2-57:2011. Medical electrical equipment-Part 2: Particular requirements for basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetics/aesthetic use.

{9}------------------------------------------------

sensiLight - 510K submission

RD-16036 A0

- 4. IEC 60601-1-11. Medical electric equipment-Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
- 5. FCC part 15, Subpart B. Class B.
- 6. IEC 60601-1-2:2007 Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance – Collateral standard: Electromagnetic compatibility – Requirements and tests.
- 7. Software Validation was conducted according to IEC 62304:2006 Medical device software - Software life cycle processes, and; FDA Guidance for the Content of Pre-Market Submissions for Software Contained in Medical Devices, dated May 11, 2005.
- 8. ISO 10993-1:2009 Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process and with FDA's Memorandum – #G95 1, May 1, 1995, Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing" (Blue Book Memo G95-1).
- 9. IEC 60601-1-6:2010, Medical electrical equipment Part 1-6: General requirements for basic safety and essential performance and IEC 62366:2007 - Medical devices -Application of usability engineering to medical devices - Collateral standard: Usability and FDA Guidance for Medical Device Use-Safety: Incorporating Human Factors Engineering into Risk Management (July18,2000) and ISO/IEC 14971:2007 (BS EN ISO 14971:2012) Medical devices – Application of risk management to medical devices.
- 10. IEC 62471:2006, Photo-biological safety of lamps and lamp systems.

Tests results are supporting all labeling claims in order to establish substantial equivalency.

#### 7. SUBSTANTIAL EQUIVALENCE

EL Global Trade Ltd.'s sensiLight is substantial equivalent to the predicate device selected in terms of indication for use, technology, performances, place of use, patient population and nature of body contact.

The Substantial equivalent decision was received based on the following comparison with the predicate device:

The design and components in the sensiLight, including the hand held applicator (with lamp, microcontroller, fan, skin color sensor, skin proximity sensor, indicator LEDs and operational button/s) are similar to the design and components found in the predicate devices (K131870 and K103184). The performance specifications (including light energy

{10}------------------------------------------------

sensiLight – 510K submission

RD-16036 A0

power, wavelength and pulse duration) are identical. The safety features found in the devices are the same, including the skin color sensor, skin proximity sensor, etc. These safety features in the sensiLight device are substantially equivalent to the safety features found in the predicate devices. Any minor differences in the technological characteristics do not raise new safety or effectiveness concerns. Furthermore, the new sensiLight device underwent performance testing, including software validation testing and electrical and mechanical safety testing according to IEC 60601-1 and electromagnetic compatibility testing according to IEC 60601-1-2. Usability testing was also successfully conducted.

#### 8. CONCLUSIONS

The evaluation of our device performances demonstrates that it is as safe and as effective as the predicate devices.

..............................................................................................................................................................................

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT/K140381](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT/K140381)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).