← Product Code [OHS](/submissions/SU/subpart-e%E2%80%94surgical-devices/OHS) · K252093

# Light Therapy Mask (1019055) (K252093)

_Merchsource, LLC · OHS · Oct 16, 2025 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHS/K252093

## Device Facts

- **Applicant:** Merchsource, LLC
- **Product Code:** [OHS](/submissions/SU/subpart-e%E2%80%94surgical-devices/OHS.md)
- **Decision Date:** Oct 16, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

The Light Therapy Mask is an over-the-counter device intended for the treatment of full-face wrinkles and mild to moderate inflammatory acne using LED light therapy. - Red light: Treatment of full-face wrinkles. - Blue light: Treatment of mild to moderate inflammatory acne. - Red+Infrared Light: Treatment of full-face wrinkles. - Mixed light (Red+Blue+Infrared): Treatment of mild to moderate inflammatory acne.

## Device Story

Light Therapy Mask is a wearable, home-use LED phototherapy device. Input: user-selected light mode (Red, Blue, Infrared, or Mixed). Operation: mask body contains LEDs emitting specific wavelengths (635nm, 465nm, 850nm) to treat facial skin; controller manages power and vibration intensity (for relaxation). Output: narrow-band LED light delivered to face. Operated by patient at home. Healthcare providers do not interact with device output; device is intended for aesthetic self-care. Benefits include non-invasive treatment of wrinkles and acne.

## Clinical Evidence

Bench testing only. No clinical data provided. Evidence includes biocompatibility (ISO 10993-1, -5, -10, -23), electrical safety (IEC 60601-1, -1-2, -1-11, -2-83), battery safety (IEC 62133-2), photobiological eye safety (IEC 62471), and usability engineering per FDA guidance.

## Technological Characteristics

Wearable mask with LEDs (Red 635nm, Blue 465nm, IR 850nm). Powered by 3.7V 2000mAh Li-ion battery. Materials: body-contacting components tested per ISO 10993. Connectivity: USB-C for charging. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-83, IEC 62471, IEC 62133-2. Software: basic level of concern.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- LED Light Therapy Device ([K241857](/device/K241857.md))

## Reference Devices

- LED Light Therapy Mask/Model: SKB-1818P,SKB-1918,SKB-1918P,SKB-1918PLUS,SKB-2318L,S KB-2318P,SKB-2318PRO,IN-M002,SKB-2418 ([K243040](/device/K243040.md))
- Infrared Red Blue LED Light Heat Beauty Machine ([K210545](/device/K210545.md))
- Pulsaderm Wrinkle Mask 28, Pulsaderm Wrinkle Mask 72 ([K163329](/device/K163329.md))
- LED light therapy mask (FM-01, FM-02, FM-03) ([K223544](/device/K223544.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

October 16, 2025

Merchsource LLC
% Candice Qiu
Registration Specialist
Feiying Drug &amp; Medical Consulting Technical Service Group
Rm 2401 Zhenye International Business Center, No. 3101-90
Qianhai Road
Shenzhen, Guangdong 518052
China

Re: K252093
Trade/Device Name: Light Therapy Mask (1019055)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: OHS, OLP, ISA
Dated: June 19, 2025
Received: July 3, 2025

Dear Candice Qiu:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K252093 - Candice Qiu
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

{2}

K252093 - Candice Qiu
Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

YAN FU -S
Digitally signed by YAN FU
Date: 2025.10.16 09:34:09
-04'00"

for Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF

|  DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.  |
| --- | --- |
|  510(k) Number (if known) K252093  |   |
|  Device Name Light Therapy Mask Model: 1019055  |   |
|  Indications for Use (Describe) The Light Therapy Mask is an over-the-counter device intended for the treatment of full-face wrinkles and mild to moderate inflammatory acne using LED light therapy. - Red light: Treatment of full-face wrinkles. - Blue light: Treatment of mild to moderate inflammatory acne. - Red+Infrared Light: Treatment of full-face wrinkles. - Mixed light (Red+Blue+Infrared): Treatment of mild to moderate inflammatory acne.  |   |
|  Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C)  |   |
|  CONTINUE ON A SEPARATE PAGE IF NEEDED.  |   |
|  This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*  |   |
|  The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov  |   |
|  "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."  |   |

{4}

MerchSource LLC

510(k)s -510(k) Summary

# 510(k) Summary-K252093

"510(k) Summary" as required by 21 CFR Part 807.92.

September 10, 2025

## I. Submitter

MerchSource LLC

7755 Irvine Center Drive Suite 100, Irvine, CA 92618, USA

Shirley Luo

Senior Compliance Manager

Tel: 1 949 900 6598

Email: shirley.luo@merchsource.com

## II. Device

Name of Device: Light Therapy Mask

Model: 1019055

Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology

Regulatory Class: II

Product Code: OHS, OLP, ISA

Regulation Number: 21 CFR 878.4810

## III. Predicate Device and Reference Device

Predicate device:

|  Manufacturer | Predicate Device | 510(k) Number | Approval Date  |
| --- | --- | --- | --- |

Page 1 of 11

{5}

MerchSource LLC

510(k)s -510(k) Summary

|  Manufacturer | Predicate Device | 510(k) Number | Approval Date  |
| --- | --- | --- | --- |
|  Dongguan Boyuan Intelligent Technology Co.,Ltd | LED Light Therapy Device (Model(s): KFB290, KFB291, KFB265, KFB293) | K241857 | October 11, 2024  |

Reference device 1:

|  Manufacturer | Reference Device | 510(k) Number | Approval Date  |
| --- | --- | --- | --- |
|  Shenzhen Siken 3D Technology Development Co., Ltd. | LED Light Therapy Mask/Model: SKB-1818P,SKB-1918,SKB-1918P,SKB-1918PLUS,SKB-2318L,S KB-2318P,SKB-2318PRO,IN-M002,SKB-2418 | K243040 | December 20, 2024  |

Reference device 2:

|  Manufacturer | Reference Device | 510(k) Number | Approval Date  |
| --- | --- | --- | --- |
|  Marci Beauty Inc | Infrared Red Blue LED Light Heat Beauty Machine | K210545 | May 20, 2022  |

Reference device 3:

|  Manufacturer | Reference Device | 510(k) Number | Approval Date  |
| --- | --- | --- | --- |
|  Pulsaderm LLC | Pulsaderm Wrinkle Mask 28, Pulsaderm Wrinkle Mask 72 | K163329 | April 14, 2017  |

Reference device 4:

|  Manufacturer | Reference Device | 510(k) Number | Approval Date  |
| --- | --- | --- | --- |
|  Guangdong Newdermo Biotech Co.,Ltd | LED light therapy mask (FM-01, FM-02, FM-03) | K223544 | February 23, 2023  |

Page 2 of 11

{6}

MerchSource LLC
510(k)s -510(k) Summary

## IV. Device Description

The subject device Light Therapy Mask is a home use wearable LED phototherapy device whose purpose is to produce an even and narrow-band of light for the treatment of aesthetic indications including facial wrinkles and acnes.

The subject device consists of a mask body unit that contains light emitting diodes (LEDs), a controller, straps and USB-C charging cable. And the device is powered by built-in rechargeable lithium battery on the controller.

The LEDs generate 3 kinds of light which include Red light (wavelength 635nm±5nm), Blue light (wavelength 465nm±5nm), Infrared light (wavelength 850nm±5nm). A controller is connect to the mask body unit to control the device, such as turn on/off the device. And the straps used for securing the mask unit to the body part.

The device have vibration function, and user can adjust the vibration intensity. Vibration is for general relaxation purposes.

## V. Indications for Use

The Light Therapy Mask is an over-the-counter device intended for the treatment of full-face wrinkles and mild to moderate inflammatory acne using LED light therapy.

- Red light: Treatment of full-face wrinkles.
- Blue light: Treatment of mild to moderate inflammatory acne.
- Red+Infrared Light: Treatment of full-face wrinkles.
- Mixed light (Red+Blue+Infrared): Treatment of mild to moderate inflammatory acne.

## VI. Comparison of Technological Characteristics With the Predicate Devices

Page 3 of 11

{7}

MerchSource LLC

510(k)s -510(k) Summary

|  Comparison Elements | Subject Device | Predicate Device | Reference Device 1 | Reference Device 2 | Reference Device 3 | Reference Device 4 | Remark  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  510(k) Number | K252093 | K241857 | K243040 | K210545 | K163329 | K223544 | /  |
|  Trade name | Light Therapy Mask (Model:1019055) | LED Light Therapy Device (Models: KFB290, KFB291) | LED light therapy mask | Infrared Red Blue LED Light Heat Beauty Machine | Pulsaderm Wrinkle Mask 28, Pulsaderm Wrinkle Mask 72 | LED light therapy mask, Models: FM-01, FM-02, FM-03 | /  |
|  Manufacturer | MerchSource LLC | Dongguan Boyuan Intelligent Technology Co.,Ltd | Shenzhen Siken 3D Technology Development Co., Ltd. | Marci Beauty Inc | Pulsaderm LLC | Guangdong Newdermo Biotech Co.,Ltd | /  |
|  Regulation number | 21 CFR 878.4810 | 21 CFR 878.4810
21 CFR 890.5500 | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810
21 CFR 890.5500 | Same  |
|  Classification Name | Light Based Over The Counter Wrinkle Reduction(OHS), Over-The-Counter Powered Light Based Laser For Acne(OLP) | Light Based Over The Counter Wrinkle Reduction(OHS), Over-The-Counter Powered Light Based Laser For Acne(OLP), Infrared Therapeutic Heating(ILY) | Light Based Over The Counter Wrinkle Reduction (OHS): Over-The-Counter Powered Light Based Laser For Acne(OLP) | Light Based Over The Counter Wrinkle Reduction(OHS), Over-The-Counter Powered Light Based Laser For Acne(OLP) | Light Based Over the Counter Wrinkle Reduction | Light Based Over The Counter Wrinkle Reduction(OHS), Over-The-Counter Powered Light Based Laser For Acne(OLP), Infrared Therapeutic Heating(ILY) | Same  |
|  Product code | OHS, OLP, ISA | OHS, OLP, ILY | OHS, OLP | OHS, OLP | OHS | OHS, OLP, ILY | Same  |
|  Device classification | Class II | Class II | Class II | Class II | Class II | Class II | Same  |
|  Indication for use/ Intended | The Light Therapy Mask is an over-the- | - Red light: Treatment of full-face wrinkles. | LED Light Therapy Mask is an over the | The Infrared Red Blue LED Light Heat | The Pulsaderm Wrinkle Masks 28 | Red light: Treatment of full-face wrinkles. Blue light: Treatment | Same  |

Page 4 of 11

{8}

MerchSource LLC

510(k)s -510(k) Summary

|  Comparison Elements | Subject Device | Predicate Device | Reference Device 1 | Reference Device 2 | Reference Device 3 | Reference Device 4 | Remark  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  510(k) Number | K252093 | K241857 | K243040 | K210545 | K163329 | K223544 | /  |
|  use | counter device intended for the treatment of full-face wrinkles and mild to moderate inflammatory acne using LED light therapy.
- Red light: Treatment of full-face wrinkles.
- Blue light: Treatment of mild to moderate inflammatory acne.
- Red+Infrared Light: Treatment of full-face wrinkles.
- Mixed light (Red+Blue+Infrared): Treatment of mild to moderate inflammatory acne. | - Infrared light: Provide topical heating for the purpose of elevating tissue temperature; arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.
- Red+Infrared Light: Treatment of full-face wrinkles.
- Amber light: Treatment of full-face wrinkles.
- Blue light: Treatment of mild to moderate inflammatory acne.
- Mixed light(Red+Blue+Infrared): Treatment of mild to moderate inflammatory acne. | counter device that is intended to use LED light for the treatment of wrinkles and mild to moderate acne. | Beauty Machine is an hand-held over-the-counter phototherapy device, the red and infrared light is intended for the use in treating wrinkles on the face and the blue light is intended for the treatment of the mild to moderate inflammatory acne. This device is indicated for adults only. | and 72 are intended for the use in the treatment of facial wrinkles and for people with Fitzpatrick Skin Types I, II and III. | of mild to moderate inflammatory acne. Infrared light: Provide topical heating for the purpose of elevating tissue temperature; arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation. Mixed light:Treatment of mild to moderate inflammatory acne. |   |

Page 5 of 11

{9}

MerchSource LLC

510(k)s -510(k) Summary

|  Comparison Elements | Subject Device | Predicate Device | Reference Device 1 | Reference Device 2 | Reference Device 3 | Reference Device 4 | Remark  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  510(k) Number | K252093 | K241857 | K243040 | K210545 | K163329 | K223544 | /  |
|  Prescription or OTC | OTC | OTC | OTC | OTC | OTC | OTC | Same  |
|  Dimension | LED Mask
294mm*226.7*18.59mm
Controller
130.1mm*37mm*25.7mm | KFB290:
Approximately 302 mm x 197 mm x 22 mm

KFB291:
Approximately 412 mm x 195 mm x 22 mm | Model SKB-1818P, SKB-1918, SKB-1918P:
194*170*131mm
Model SKB-1918PLUS:
217*152*173mm
SKB-2318L, SKB-2318P, SKB-2318PRO, IN-FM002, SKB-2418:
411*195*4mm | Not publicly available | Not publicly available | FM-01: 207 * 277 * 43 mm,
FM-02: 198 * 383 * 33.5 mm,
FM-03: 237.5 * 108 * 8.1 mm | Different
**Note 1**  |
|  Power supply | Input: DC5V, 1A
Battery specifications:
Li-ion rechargeable battery
18650 - 2000MAh
3.7V | Input: 100 -240 V, 50 / 60 Hz
Output: 5V, 1A
Lithium ion battery:
1300mAh | Input: AC 100-240V 50-60Hz 0.3A
Output: DC 5V 1A 3.7V 650mAh
Li-ion Battery 3.7V1000mAh Li-ion Battery | 900mAh,
Rechargeable Li-Ion batteries
100~240V AC 50/60HZ 0.5A | Rechargeable Ni-MH battery | Input: 100-240V, 50/60Hz, 0.25 A
Output: DC 5 V, 500 mA | Different
**Note 2**  |
|  Sterilization | Not applicable | Not applicable | Not applicable | Not applicable | Not applicable | Not applicable | Same  |
|  Light source | Light Emitting Diodes | Light Emitting Diodes | Light Emitting Diodes | Light Emitting Diodes | Light Emitting Diodes | Light Emitting Diodes | Same  |

Page 6 of 11

{10}

MerchSource LLC

510(k)s -510(k) Summary

|  Comparison Elements | Subject Device | Predicate Device | Reference Device 1 | Reference Device 2 | Reference Device 3 | Reference Device 4 | Remark  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  510(k) Number | K252093 | K241857 | K243040 | K210545 | K163329 | K223544 | /  |
|  Location for use | Face | Face | Face | Entire Face | Face | Face and body | Same  |
|  Wavelength | 635nm ±5nm visible red light;
850nm±5nm Invisible red light;
465±5nm blue light | 635nm ±5nm visible red light;
850nm±5nm Invisible red light;
465±5nm blue light;
605±5nm amber light | SKB-1818P,SKB-1918,SKB-1918P,SKB-1918PLUS:
Blue: 460nm±10nm
Red: 620nm±10nm
SKB-2318L,SKB-2318P,SKB-2318PRO,IN-FM002,SKB-2418:
Blue: 460nm±10nm
Red: 620nm±10nm
Infrared: 850nm±10nm | Blue: 465nm
Red: 620-630nm
IR: 845-855nm | 620-630 nm red and 850 nm Infrared | Red: 620nm
Blue: 460nm
Infrared: 850nm
Mixed: 620nm and 850nm and 460nm | Same  |
|  Irradiance (mW/cm²) | Red: 5~20mW/cm²;
Blue: 5~20mW/cm²;
Red+IR: 5~20mW/cm²;
Mixed light: 10mW/cm²; | Red: 25mW/cm²;
IR: 3mW/cm²;
Red+IR: 30mW/cm²;
Blue: 18mW/cm²;
Amber: 20mW/cm²;
Mixed light: 9mW/cm²; | SKB-1818P:
Mode 1: Red: 5.5mW/cm²
Mode 2: Blue: 9mW/cm²
SKB-1918:
Mode 1: Red: 5mW/cm²
Mode 2: Blue: 6mW/cm²
SKB-1918P: | Blue Light Mode: 5.4
Red+IR Light Mode: 7.2 | Red: 15.94
Infrared: 5.24
Total: 21.18 | Red light: 2.0~3.0mW/cm²
Blue light: 2.0~4.0mW/cm²
Infrared light: 2.0~4.0 mW/cm²
Mixed light: 9.0~12.0 mW/cm² | Similar Note 3  |

Page 7 of 11

{11}

MerchSource LLC

510(k)s -510(k) Summary

|  Comparison Elements | Subject Device | Predicate Device | Reference Device 1 | Reference Device 2 | Reference Device 3 | Reference Device 4 | Remark  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  510(k) Number | K252093 | K241857 | K243040 | K210545 | K163329 | K223544 | /  |
|   |  |  | Mode 1: Red: 6mW/cm²
Mode 2: Blue: 6.5mW/cm²
SKB-1918PLUS: Mode 1: Red: 3.5mW/cm²
Mode 2: Blue: 4.0mW/cm²
SKB-2318L, SKB-2318P, SKB-2318PRO, IN-FM002, SKB-2418: Mode 1: Red: 3.5~10mW/cm²
Mode 2: Blue: 2.5~8mW/cm² |  |  |  |   |
|  Treatment time | 5, 10, 15, 20 minutes | For red, blue and red+infrared: 10, 20, 30 minutes
For infrared, amber light and mixed light: 10, 20 minutes. | It is recommended to use it for 10 minutes a day, 3 times per week. | 8 minutes | 15 minutes | Manual Mode: 15 minutes each time, Automatic Mode: 10 minutes each time. | Similar  |
|  Compliance with voluntary | IEC 60601-1; | IEC 60601-1; | IEC 60601-1; | IEC 60601-1; | IEC 60601-1; | IEC 60601-1; | Same  |

Page 8 of 11

{12}

MerchSource LLC

510(k)s -510(k) Summary

|  Comparison Elements | Subject Device | Predicate Device | Reference Device 1 | Reference Device 2 | Reference Device 3 | Reference Device 4 | Remark  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  510(k) Number | K252093 | K241857 | K243040 | K210545 | K163329 | K223544 | /  |
|  standards | IEC 60601-1-2;
IEC 60601-1-11;
IEC 60601-2-83;
IEC 62471 | IEC 60601-1-2;
IEC 60601-1-11;
IEC 60601-2-83;
IEC 62471 | IEC 60601-1-2;
IEC 60601-1-11;
IEC 60601-2-57;
IEC 62471
IEC 62133-2 | IEC 60601-1-2;
IEC 60601-1-11;
IEC 62471 | IEC 60601-1-2;
IEC 62471 | IEC 60601-1-2 |   |
|  Biocompatibility | ISO 10993-1
ISO 10993-5
ISO 10993-10
ISO 10993-23 | ISO 10993-5
ISO 10993-10
ISO 10993-23 | All body-contacting materials are complied with ISO10993-5, ISO 10993-10 and ISO 10993-23 | ISO 10993-1
ISO 10993-5
ISO 10993-10 | ISO 10993 | All body-contacting materials are complied with ISO10993-5 and ISO 10993-10 | Same  |

## VII. Performance Data

The following performance data were provided in support of the substantial equivalence determination.

1) Biocompatibility Evaluation

{13}

MerchSource LLC
510(k)s -510(k) Summary

The biocompatibility evaluation for the body-contacting components of the Intense pulsed light device was conducted in accordance with the "Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices –Part 1: Evaluation and Testing Within a Risk Management Process, Document Issued on September 4, 2020", as recommended by FDA. The following testing was performed to, and passed, including:

- ISO 10993-5:2009, Biological evaluation of medical devices –Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10:2021, Biological evaluation of medical devices –Part 10: Tests for skin sensitization
- ISO 10993-23:2021, Biological evaluation of medical devices –Part 23: Tests for skin irritation

## 2) Electrical Safety and EMC

Electrical safety and EMC testing was performed to, and passed, as per the following standards:

- IEC 60601-1-2:2014+A1:2020 Medical electrical equipment –Part 1-2: General requirements for basic safety and essential performance – Collateral standard: electromagnetic compatibility
- IEC 60601-1:2005 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-11:2015 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- IEC 60601-2-83:2019 Medical electrical equipment - Part 2-83: Particular requirements for the basic safety and essential performance of home light therapy equipment
- IEC 62133-2:2017 Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems

## 3) Eye Safety

- IEC 62471:2006 Photobiological safety of lamps and lamp systems

Page 10 of 11

{14}

MerchSource LLC
510(k)s –510(k) Summary

## 4) Software Verification and Validation

Software documentation consistent with basic documentation level of concern was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels.

## 5) Usability

The product usability has been evaluated and verified according to the following FDA guidance

- Applying Human Factors and Usability Engineering to Medical Devices, issued on FEBRUARY 2016.

## VIII. Conclusions

Based on the above analysis and non-clinical tests performed, it can be concluded that the subject device Light Therapy Mask is as safe, as effective, and performs as well as the legally marketed predicate device and reference devices.

Page 11 of 11

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHS/K252093](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHS/K252093)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
