← Product Code [OHS](/submissions/SU/subpart-e%E2%80%94surgical-devices/OHS) · K250125

# LMNT O2 (K250125)

_Xplr , Ltd. · OHS · Mar 24, 2025 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHS/K250125

## Device Facts

- **Applicant:** Xplr , Ltd.
- **Product Code:** [OHS](/submissions/SU/subpart-e%E2%80%94surgical-devices/OHS.md)
- **Decision Date:** Mar 24, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

LMNT O2 is an over-the-counter device indicated to emit energy in the red and IR region of the spectrum for use in dermatology for the treatment of periorbital wrinkles. In addition, LMNT O2 indicated to emit energy in the blue region of the spectrum for use in dermatology for the treatment of mild to moderate acne.

## Device Story

Hand-held, home-use phototherapy device; utilizes 36 LEDs (12 red 633nm, 12 IR 830nm, 12 blue 415nm) to emit light energy; constant waveform; uniform distribution without hot spots. User operates device on face/periorbital area. Red/IR mode treats periorbital wrinkles; blue mode treats mild to moderate acne. Output affects skin via photobiomodulation/phototherapy. Benefits include reduction of wrinkles and acne lesions. Powered by lithium battery.

## Clinical Evidence

Bench testing only. No clinical data provided. Testing included electrical safety (ANSI AAMI ES60601-1), EMC (IEC 60601-1-2), photobiological safety (IEC 62471), biocompatibility (ISO 10993-5, -10, -11, -23), and performance testing (skin temperature, wavelength/power density, and hot spot verification).

## Technological Characteristics

Materials: Rigid ABS and stainless steel. Light source: 36 LEDs (12x 633nm, 12x 830nm, 12x 415nm). Energy density: 69mW/cm2 (633nm), 53mW/cm2 (830nm), 49mW/cm2 (415nm). Power: Lithium battery. Form factor: Hand-held. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 62471, ISO 10993 series.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- BLU TOTALE ([K232656](/device/K232656.md))
- Avologi ENEO ([K181650](/device/K181650.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

March 24, 2025

XPLR Ltd. % Doris Dong Shanghai CV Technology Co., Ltd. Room 805, No. 19 Dongbao Road, Songjiang Area, Shanghai 201613 China Shanghai, 201613 China

Re: K250125

Trade/Device Name: Lmnt 02 Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHS, OLP Dated: January 15, 2025 Received: January 17, 2025

Dear Doris Dong:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

{1}------------------------------------------------

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4. Subpart A) for combination products: and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical

{2}------------------------------------------------

devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

TANISHA L.
Digitally signed by
TANISHA L. HITHE -S
HITHE -S
Date: 2025.03.24 23:53:11
-04'00'

Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K250125

Device Name LMNT 02

#### Indications for Use (Describe)

LMNT O2 is an over-the-counter device indicated to emit energy in the red and IR region of the spectrum for use in dermatology for the treatment of periorbital wrinkles. In addition, LMNT O2 indicated to emit energy in the blue region of the spectrum for use in dermatology for the treatment of mild to moderate acne.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# 510(k) Summary #K250125.

[As required by 21 CFR 807.92]

| [As required by 21 CFR 807.92] |                                                                                                                                                                                                                                                                                                            |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1. Submission Information      |                                                                                                                                                                                                                                                                                                            |
| 510(k) Number:                 |                                                                                                                                                                                                                                                                                                            |
| Date:                          | January 15, 2025                                                                                                                                                                                                                                                                                           |
| Type of 510(k) Submission:     | Traditional 510(k)                                                                                                                                                                                                                                                                                         |
| Basis for 510(k) Submission:   | New device                                                                                                                                                                                                                                                                                                 |
| Owner:                         | XPLR ltd                                                                                                                                                                                                                                                                                                   |
|                                | Ben ami 6 st, Tel Aviv, Israel                                                                                                                                                                                                                                                                             |
|                                | Tel: +972-52-4200400                                                                                                                                                                                                                                                                                       |
|                                | Email:Info@xplr-med.com                                                                                                                                                                                                                                                                                    |
|                                | Website:www.Xplr-med.com                                                                                                                                                                                                                                                                                   |
| Contact:                       | Doris Dong                                                                                                                                                                                                                                                                                                 |
|                                | [Consultant, from Shanghai CV Technology Co., Ltd.]<br>Add: Room 805, No. 19 Dongbao Road, Songjiang Area, Shanghai, 201613 China                                                                                                                                                                          |
|                                | E-mail: doris.d@ceve.org.cn                                                                                                                                                                                                                                                                                |
|                                | Tel: 86 21-31261348 / Fax: 86 21-57712250                                                                                                                                                                                                                                                                  |
|                                |                                                                                                                                                                                                                                                                                                            |
| 2. Device Description          |                                                                                                                                                                                                                                                                                                            |
| Proprietary Name:              | LMNT O2                                                                                                                                                                                                                                                                                                    |
| Classification Name:           | Light Based Over The Counter Wrinkle Reduction,                                                                                                                                                                                                                                                            |
|                                | Over-The-Counter Powered Light Based Laser For Acne                                                                                                                                                                                                                                                        |
| Regulation Number:             | 21 CFR 878.4810                                                                                                                                                                                                                                                                                            |
| Product Code:                  | OHS,OLP                                                                                                                                                                                                                                                                                                    |
| Device Class:                  | II                                                                                                                                                                                                                                                                                                         |
| Review Panel:                  | General & Plastic Surgery                                                                                                                                                                                                                                                                                  |
| Device Description:            | LMNT O2 is a hand held device that uses low power light spectrum at red                                                                                                                                                                                                                                    |
|                                | and infrared LED, at wavelength of 633 ±15nm, 830 ±15nm distribution<br>with no hot spots for the treatment of periorbital wrinkles. The device also<br>uses blue LED, at wavelength of 415 ±15nm emitting optical power in a<br>uniform distribution with no hot spots for the treatment of acne.Designed |
|                                | for home-use.                                                                                                                                                                                                                                                                                              |
| Indications for use:           | LMNT O2 is an over-the-counter device indicated to emit energy in the red                                                                                                                                                                                                                                  |
|                                | and IR region of the spectrum for use in dermatology for the treatment of                                                                                                                                                                                                                                  |
|                                | periorbital wrinkles. In addition, LMNT O2 indicated to emit energy in the                                                                                                                                                                                                                                 |
|                                | blue region of the spectrum for use in dermatology for the treatment of mild                                                                                                                                                                                                                               |
|                                | to moderate acne.                                                                                                                                                                                                                                                                                          |

# 3. Predicate Device Identification

| Predicate 510(k) Number:  | K232656                      |
|---------------------------|------------------------------|
| Marketing clearance date: | 11/14/2023                   |
| Product name:             | BLU TOTALE (Model: ENEOB852) |
| Manufacturer:             | Premier North America Inc.   |
| Predicate 510(k) Number:  | K181650                      |

| Predicate 510(k) Number:  | K181039    |
|---------------------------|------------|
| Marketing clearance date: | 07/26/2018 |

{5}------------------------------------------------

Product name: Manufacturer:

Avologi ENEO Premier North America Inc.

{6}------------------------------------------------

|                          | New Device                                                                                                                                                                                                                                                                                                                    | Predicate Device                                                                                                         | Predicate Device                                                                                                                                                               | Remark  |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| 510(k) Number            | To be assigned                                                                                                                                                                                                                                                                                                                | K232656                                                                                                                  | K181659                                                                                                                                                                        | --      |
| Proprietary Name         | LMNT O2                                                                                                                                                                                                                                                                                                                       | BLU TOTALE                                                                                                               | Avologi ENEO                                                                                                                                                                   | --      |
| Model                    | --                                                                                                                                                                                                                                                                                                                            | ENEOB852                                                                                                                 | --                                                                                                                                                                             | --      |
| Owner                    | XPLR ltd.                                                                                                                                                                                                                                                                                                                     | Premier North America Inc.                                                                                               | Premier North America Inc.                                                                                                                                                     | --      |
| Product Code             | OHS,OLP                                                                                                                                                                                                                                                                                                                       | OLP                                                                                                                      | OHS                                                                                                                                                                            | Same    |
| Classification           | II                                                                                                                                                                                                                                                                                                                            | II                                                                                                                       | II                                                                                                                                                                             | Same    |
| Indications for use      | LMNT O2 is an over-the-counter device indicated to emit energy in the red and IR region of the spectrum for use in dermatology for the treatment of periorbital wrinkles. In addition, LMNT O2 indicated to emit energy in the blue region of the spectrum for use in dermatology for the treatment of mild to moderate acne. | BLU TOTALE is an over the counter phototherapy device for use in dermatology for the treatment of mild to moderate acne. | Avologi ENEO is an over the counter device indicated to emit energy in the red and IR region of the spectrum for use in dermatology for the treatment of periorbital wrinkles. | Same    |
| Type                     | Hand-held                                                                                                                                                                                                                                                                                                                     | Hand-held                                                                                                                | Hand-held                                                                                                                                                                      | Same    |
| Materials                | Rigid ABS and stainless steel                                                                                                                                                                                                                                                                                                 | Rigid ABS and stainless steel                                                                                            | Rigid ABS and stainless steel                                                                                                                                                  | Same    |
| Target Population        | Individuals with periorbita wrinkles and mild to moderate acne on the face.                                                                                                                                                                                                                                                   | Individuals with mild to moderate acne on the face.                                                                      | Individuals with periorbita wrinkles                                                                                                                                           | Same    |
| Anatomical Sites         | Entire face and periorbital area                                                                                                                                                                                                                                                                                              | Entire Face                                                                                                              | Periorbital Area                                                                                                                                                               | Same    |
| Light source             | Light emitting diode (LED)                                                                                                                                                                                                                                                                                                    | Light emitting diode (LED)                                                                                               | Light emitting diode (LED)                                                                                                                                                     | Same    |
| Light spectrum region    | Blue light,red and IR region of the spectrum                                                                                                                                                                                                                                                                                  | Blue light                                                                                                               | red and IR region of the spectrum                                                                                                                                              | Same    |
| Wavelengths              | 633 $             \pm         $ 15nm, 830 $             \pm         $ 15nm,415 $             \pm         $ 15nm                                                                                                                                                                                                               | 415 $             \pm         $ 15nm                                                                                     | 633 $             \pm         $ 5nm, 830 $             \pm         $ 5nm                                                                                                       | Same    |
| Waveform                 | Constant                                                                                                                                                                                                                                                                                                                      | Constant                                                                                                                 | Constant                                                                                                                                                                       | Same    |
| Visible light LEDs       | Yes                                                                                                                                                                                                                                                                                                                           | Yes                                                                                                                      | Yes                                                                                                                                                                            | Same    |
| Energy source            | 36 pcs LEDs(12pcs 633nm LEDs +<br>12pcs 830nm LEDs+ 12pcs 415nm<br>LEDs)over 10cm2                                                                                                                                                                                                                                            | 24 pcs LEDs over 15cm2                                                                                                   | 24 pcs LEDs (12pcs 633nm LEDs + 12pcs<br>830nm LEDs ) over 15cm2                                                                                                               | Similar |
| Energy density           | 69mW/cm2 for 633nm<br>53mW/cm2 for 830nm<br>49mW/cm2 for 415nm                                                                                                                                                                                                                                                                | 50mW/cm2                                                                                                                 | 69mW/cm2 for 633nm<br>55mW/cm2 for 830nm                                                                                                                                       | Similar |
| Initial treatment course | Red light mode:<br>For the first month (4 weeks), treatment<br>should be performed 3 times a week for<br>15-20 minutes each time (5-7 minutes<br>on each treatment zone).<br>Blue light mode:<br>4 minutes per area, twice per week for 4<br>weeks (total of 8 treatments).                                                   | 4 minutes per area, twice per week for 4<br>weeks (total of 8 treatments)                                                | For the first month (4 weeks), treatment<br>should be performed 3 times a week for<br>15-20 minutes each time (5-7 minutes on<br>each treatment zone).                         | Same    |
| Power supply             | Lithium battery                                                                                                                                                                                                                                                                                                               | Lithium battery                                                                                                          | Lithium battery                                                                                                                                                                | Same    |

### 4. Substantial Equivalence to Predicate device

{7}------------------------------------------------

### 5. Non-clinical Testing

The LMNT O2 has been evaluated the safety and conforms to the following international consensus standards:

Electrical safety:

- ANSI AAMI ES66601-1:2005/(R)2012 & A1:2010(R)2012 & A2:2010(R)2012 (Cons. Text) [Incl. AMD2:2021] Medical electrical equipment Part 1 : Geneal . requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) [Including Amendment 2 (2021)];
# EMC:

- IEC 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION Medical equipments for basic and exceptivements for basices and performance-. Collateral Standard: Electromagnetic disturbances - Requirements and tests;
# Additional safety testing:

- . IEC 60601-1-11 Edition 2.1 2020-07 CONSOLIDATED VERSION Medical electrical equirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical systems used in the home healthcare environment;
- . IEC TR 60601-4-2 Edition 1.0 2016-05 Medical equipment - Part 4-2: Guidane and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems;

{8}------------------------------------------------

- IEC 62471 First edition 2006-07 Photobiological safety of lamps and lamp systems; .
#### Biocompatibility testing:

- ISO 10993-5 Third edition 2009-06-01 Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity; ●
- ISO 10993-10 Fourth edition 2021-11 Biological evaluation of medical devices Part 10: Tests for skin sensitization; ●
- ISO 10993-11 Third edition 2017-09 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity;
- ISO 10993-23 First edition 2021-01 Biological evaluation of medical devices Part 23: Tests for irritation; ●

# Performance testing:

- Skin Temperature Testing ●
- LED Wavelength And Power Density Testing ●
- No hot spots Testing ●

# 6. Conclusions

The onclusion from the non-clinical tests demonstrate that the subject device is as sell as the legally markets predicate device K22156 and K181659.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHS/K250125](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHS/K250125)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
