← Product Code [OHS](/submissions/SU/subpart-e%E2%80%94surgical-devices/OHS) · K242385

# LED Therapy Mask (MN1, M226) (K242385)

_Shenzhen Borria Technology Co., Ltd. · OHS · Apr 7, 2025 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHS/K242385

## Device Facts

- **Applicant:** Shenzhen Borria Technology Co., Ltd.
- **Product Code:** [OHS](/submissions/SU/subpart-e%E2%80%94surgical-devices/OHS.md)
- **Decision Date:** Apr 7, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

Red light: Treatment of full-face wrinkles. Blue light (only suitable model M226 and MN1 LED Facial Mask): Treatment of mild to moderate inflammatory acne. Infrared light: Provide topical heating for the purpose of elevating tissue temperature; arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation. Mixed light: Treatment of full face wrinkles.

## Device Story

Wearable LED phototherapy mask for home use; treats facial wrinkles, acne, and provides topical heating. Device inputs: electrical power from rechargeable lithium battery. Principle of operation: LEDs emit narrowband light (Red 630nm, Blue 470nm, Infrared 850nm, Mixed) to target skin tissue. User operates via controller to select light mode and power on/off. Output: light energy applied to face/neck. Clinical benefit: wrinkle reduction, acne management, and muscle/joint relief via thermal effects. Used by patients in home environment.

## Clinical Evidence

Bench testing only. No clinical data provided. Evidence includes biocompatibility (ISO 10993-5, -10, -23), electrical safety (IEC 60601-1, -1-2, -1-11, -2-83, IEC 62133-2), photobiological eye safety (IEC 62471), software verification/validation, and usability engineering per FDA guidance.

## Technological Characteristics

Wearable LED mask; materials compliant with ISO 10993-5, -10, -23. Light source: LEDs (Red 630nm, Blue 470nm, NIR 850nm). Energy source: 3.7V 2500mAh rechargeable lithium battery. Connectivity: Type-C charging. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-83, IEC 62471, IEC 62133-2.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- LED light therapy mask (FM-01, FM-02, FM-03) ([K223544](/device/K223544.md))

## Reference Devices

- MyLedMask ([K223147](/device/K223147.md))
- Pro X OTC 5 Light Therapy Device ([K140471](/device/K140471.md))
- Rejuvalite MD ([K133896](/device/K133896.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

Build Correspondence

Convert to PDF

April 7, 2025

Shenzhen Borria Technology Co., Ltd
Chardy Wang
Project Leader
501, Floor 5, Building 6, Industrial Zone, No.2015, Shenyan
Road, Pengwan Community, Haishan Street, Yantian District
Shenzhen,
China

Re: K242385
Trade/Device Name: LED Therapy Mask (MN1, M226)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: OHS, ILY, OLP
Dated: August 12, 2024
Received: August 12, 2024

Dear Chardy Wang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K242385 - Chardy Wang
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.

{2}

K242385 - Chardy Wang
Page 3

See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

TANISHA
L. HITHE -S
Digitally signed by
TANISHA L. HITHE -S
Date: 2025.04.07
17:08:19 -04'00"

Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF

|  DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.  |
| --- | --- |
|  510(k) Number (if known) K242385  |   |
|  Device Name LED Therapy Mask (MN1, M226)  |   |
|  Indications for Use (Describe) Red light: Treatment of full- face wrinkles. Blue light (only suitable model M226 and MN1 LED Facial Mask): Treatment of mild to moderate inflammatory acne. Infrared light: Provide topical heating for the purpose of elevating tissue temperature; arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation. Mixed light: Treatment of full face wrinkles.  |   |
|  Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C)  |   |
|  CONTINUE ON A SEPARATE PAGE IF NEEDED.  |   |
|  This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*  |   |
|  The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov  |   |
|  "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."  |   |

{4}

SHENZHEN BORRIA TECHNOLOGY CO.,LTD
K242385 -510(k) Summary

# 510(k) Summary-K242385

"510(k) Summary" as required by 21 CFR Part 807.92.

April 2, 2025

## I. Submitter

SHENZHEN BORRIA TECHNOLOGY CO.,LTD
501, Floor 5, Building 6, Industrial Zone, No.2015, Shenyan Road, Pengwan Community, Haishan Street, Yantian District, Shenzhen
Chardy Wang
Project Leader
Tel: +86 133 1290 5505
Email: 466144494@qq.com

## II. Device

Name of Device: LED Therapy Mask
Model(s): MN1, M226
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: II
Product Code: OHS, OLP, ILY
Regulation Number: 21 CFR 878.4810

## III. Predicate Device and Reference Device

Predicate device:

|  Manufacturer | Predicate Device | 510(k) Number | Approval Date  |
| --- | --- | --- | --- |
|  Guangdong Newdermo Biotech Co., Ltd | LED light therapy mask (FM-01, FM-02, FM-03) | K223544 | Feb 23, 2023  |

Page 1 of 9

{5}

SHENZHEN BORRIA TECHNOLOGY CO.,LTD
K242385 -510(k) Summary

Reference device 1:

|  Manufacturer | Reference Device | 510(k) Number | Approval Date  |
| --- | --- | --- | --- |
|  myBlend | MyLedMask | K223147 | May 14, 2023  |

Reference device 2:

|  Manufacturer | Reference Device | 510(k) Number | Approval Date  |
| --- | --- | --- | --- |
|  La Lumiere, LLC | Pro X OTC 5 Light Therapy Device | K140471 | Jan 15, 2025  |

Reference device 3:

|  Manufacturer | Reference Device | 510(k) Number | Approval Date  |
| --- | --- | --- | --- |
|  Trophy Skin Incorporated | Rejuvalite MD | K133896 | Nov 13, 2014  |

IV. Device Description

The subject device LED Therapy Mask is a home use wearable LED phototherapy device whose purpose is to produce an even and narrowband of light for the treatment of cosmetic indications including facial wrinkles and acnes.

The subject device consists of a mask body unit that contains light emitting diodes (LEDs), a controller, straps and Type-C charging cable. And the device is powered by built-in rechargeable lithium battery on the controller.

Page 2 of 9

{6}

SHENZHEN BORRIA TECHNOLOGY CO.,LTD
K242385 -510(k) Summary

The LEDs generate 4 kinds of light which include Blue light (460nm, only suitable model M226 and MN1 LED Facial Mask), Red light (630nm), Infrared light (850nm) and Mixed light (630nm and 850nm). A controller is connect to the mask body unit to control the device, such as turn on/off the device, switch LED output. And the straps used for securing the mask unit to the body part.

## V. Indications for Use

Red light: Treatment of full-face wrinkles.

Blue light (only suitable model M226 and MN1 LED Facial Mask): Treatment of mild to moderate inflammatory acne.

Infrared light: Provide topical heating for the purpose of elevating tissue temperature; arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.

Mixed light: Treatment of full face wrinkles.

## VI. Comparison of Technological Characteristics With the Predicate Devices

|  Comparison Elements | Subject Device | Primary Predicate Device | Reference Device 1 | Reference Device 2 | Reference Device 3 | Remark  |
| --- | --- | --- | --- | --- | --- | --- |
|  510(k) Number | Pending | K223544 | K223147 | K140471 | K133896 | /  |
|  Trade name | LED Therapy Mask (MN1, M226) | LED light therapy mask (FM-01, FM-02, FM-03) | MyLedMask | Pro X OTC 5 Light Therapy Device | Rejuvalite MD | /  |

Page 3 of 9

{7}

SHENZHEN BORRIA TECHNOLOGY CO.,LTD
K242385 -510(k) Summary

|  Comparison Elements | Subject Device | Primary Predicate Device | Reference Device 1 | Reference Device 2 | Reference Device 3 | Remark  |
| --- | --- | --- | --- | --- | --- | --- |
|  Manufacturer | SHENZHEN BORRIA TECHNOLOGY CO., LTD | Guangdong Newdermo Biotech Co., Ltd | myBlend | La Lumiere, LLC | Trophy Skin Incorporated | /  |
|  Regulation number | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | Same  |
|  Product code | OHS, OLP, ILY | OHS, OLP, ILY | OHS | OHS | OHS |   |
|  Device classification | Class II | Class II | Class II | Class II | Class II | Same  |
|  Indication for use/ Intended use | Red light: Treatment of full-face wrinkles. Blue light (only suitable model M226 and MN1 LED Facial Mask): Treatment of mild to moderate inflammatory acne. Infrared light: Provide topical heating for the purpose of elevating tissue temperature; arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily | Red light: Treatment of full-face wrinkles. Blue light: Treatment of mild to moderate inflammatory acne. Infrared light: Provide topical heating for the purpose of elevating tissue temperature; arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily | MyLedMask is an over-the-counter device that is intended to emit energy in the visible and infrared region of the spectrum for use in the treatment of full-face wrinkles | The Pro XOTC 5is an Over-The-Counter device intended for use in the treatment of facial wrinkles. It is for people with wrinkles on their face and who have Fitzpatrick skin types I,II and/or III. | The Rejuvalite MD is an Over the Counter device that is intended for the use in the treatment of full face wrinkles. | Same  |

Page 4 of 9

{8}

SHENZHEN BORRIA TECHNOLOGY CO.,LTD
K242385 -510(k) Summary

|  Comparison Elements | Subject Device | Primary Predicate Device | Reference Device 1 | Reference Device 2 | Reference Device 3 | Remark  |
| --- | --- | --- | --- | --- | --- | --- |
|   | the relaxation of muscle tissue; and to temporarily increase local blood circulation. Mixed light: Treatment of full face wrinkles. | increase local blood circulation. Mixed light: Treatment of mild to moderate inflammatory acne. |  |  |  |   |
|  Location for use | Face and Neck | Face and body | Face and Neck | Full Face | Full Face | Same  |
|  Prescription or OTC | OTC | OTC | OTC | OTC | OTC | Same  |
|  Power supply | Input: DC 5V, 1A Built-in rechargeable lithium battery: DC 3.7V 2500mAh | Input: 100-240 V~, 50/60 Hz, 0.25 A Output: DC 5 V, 500 mA | Voltage: 100 to 240 volts, AC Frequency:50-60Hz Intensity: 0.35 A | Unknown | AC to DC | Same  |
|  Light source | LEDs | LEDs | LEDs | Light-emitting diodes | LEDs | Same  |
|  Wavelength | Red: 630nm±5nm
Blue: 470 nm±5nm
Near-Infrared: 850 nm±5nm | Red: 620nm
Blue: 460nm
Infrared: 850nm
Mixed: 620nm and 850nm and 460nm | Red:630 nm (OSRAM LS T67F)
NIR: 850 nm (OSRAM SFH 4253) | Red: 620-640nm
IR: 820-880nm | Red 600,622,660nm
IR 860nm | Similar Note 1  |

Page 5 of 9

{9}

SHENZHEN BORRIA TECHNOLOGY CO.,LTD
K242385 -510(k) Summary

|  Comparison Elements | Subject Device | Primary Predicate Device | Reference Device 1 | Reference Device 2 | Reference Device 3 | Remark  |
| --- | --- | --- | --- | --- | --- | --- |
|   | Mixed light: 630nm and 850nm |  |  |  |  |   |
|  LED Intensity | Red: 0.89~2.55mW/cm²
Blue: 1.44~4.09mW/cm²
Near-Infrared: 1.83~3.05mW/cm²
Mixed light: 0.95~2.64mW/cm² | Red light: 2.0~3.0 mW/cm²
Blue light: 2.0~4.0 mW/cm²
Infrared light: 2.0~4.0 mW/cm²
Mixed light: 9.0~12.0 mW/cm² | 18.7 mW/cm² (average of 6 measurement location (including one on the neck) between 10 to 27mW/cm²) | Red light and IR light: 0.76 mW/cm² | Red light and IR light: 2.35 mW/cm² | Similar Note 2  |
|  Dimensions (mm) | MN1: LED Facial Mask: 300*208.5*5.5mm
LED Neck Mask: 338.5*249.7*5.5mm
M226: LED Facial Mask: 300*208.5*5.5mm | FM-01: 207X277X43mm,
FM-02: 198X383X33.5mm,
FM-03: 237.5X108X8.1mm | LED mask: 176 x342 x 122
Storage case:365 x 225 x150 | Unknown | Unknown | Different Note 3  |
|  Treatment Time | 10 minutes each time | Manual Mode: 15minutes each time,
Automatic Mode: 10minutes each time | According to skin phototype, daily - 1 fair skin: 5 min 35 (335 s) - 2 moderately dark skin: 11 min 10 (670 s) - 3 dark | 900 seconds | 3 minutes daily, 5 days per week for 8 weeks |   |

Page 6 of 9

{10}

SHENZHEN BORRIA TECHNOLOGY CO.,LTD
K242385 -510(k) Summary

|  Comparison Elements | Subject Device | Primary Predicate Device | Reference Device 1 | Reference Device 2 | Reference Device 3 | Remark  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  |  | skin: 13 min 55 (835 s) during 6 to 8 week |  |  |   |
|  Electrical safety | IEC 60601-1
IEC 60601-1-2
IEC 60601-1-11
IEC 60601-2-83
IEC 62471
IEC 62133-2 | IEC 60601-1
IEC 60601-1-2
IEC 60601-1-11
IEC 60601-2-57 | IEC 60601-1
IEC 60601-1-2
IEC 60601-2-57 | Unknown | IEC 60601-1
IEC 60601-1-2 | Same  |
|  Biocompatibility feature | All body-contacting materials are complied with ISO 10993-5, ISO 10993-10 and ISO 10993-23 | All body-contacting materials are complied with ISO10993-5 and ISO 10993-10 | Unknown | Unknown | Unknown | Same  |

## VII. Performance Data

The following performance data were provided in support of the substantial equivalence determination.

### 1) Biocompatibility Evaluation

The biocompatibility evaluation for the body-contacting components of the Intense pulsed light device was conducted in accordance with the "Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices –Part 1: Evaluation and Testing Within a Risk Management Process, Document Issued on September 4, 2020", as recognized by FDA. The following testing was performed to, and passed, including:

{11}

SHENZHEN BORRIA TECHNOLOGY CO.,LTD
K242385 -510(k) Summary

- ISO 10993-5:2009, Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10:2021, Biological evaluation of medical devices – Part 10: Tests for skin sensitization
- ISO 10993-23:2021, Biological evaluation of medical devices – Part 23: Tests for skin irritation

## 2) Electrical Safety and EMC

Electrical safety and EMC testing was performed to, and passed, as per the following standards:

- IEC 60601-1-2 Edition 4.1 2020-09 Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral standard: electromagnetic compatibility
- ANSI AAMI ES60601-1:2005/(R)2012 &amp; A1:2012 C1:2009/(R)2012 &amp; A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD)
- IEC 60601-1-11 Edition 2.1 2020-07 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- IEC 60601-2-83:2019 Medical electrical equipment - Part 2-83: Particular requirements for the basic safety and essential performance of home light therapy equipment
- IEC 62133-2 Edition 1.0 2017-02, Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems

## 3) Eye Safety

- IEC 62471 Photobiological safety of lamps and lamp systems

## 4) Software Verification and Validation

Page 8 of 9

{12}

SHENZHEN BORRIA TECHNOLOGY CO.,LTD
K242385 -510(k) Summary

Software documentation consistent with *moderate level* of concern was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels.

## 5) Usability

The product usability has been evaluated and verified according to the following FDA guidance

- Applying Human Factors and Usability Engineering to Medical Devices, issued on FEBRUARY 2016.

## VIII. Conclusions

Based on the above analysis and non-clinical tests performed, it can be concluded that the subject device LED Therapy Mask is as safe, as effective, and performs as well as the legally marketed predicate device and reference devices.

Page 9 of 9

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHS/K242385](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHS/K242385)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
