← Product Code [OHS](/submissions/SU/subpart-e%E2%80%94surgical-devices/OHS) · K180856

# Neutrogena Light Therapy Aging Mask+ (K180856)

_Johnson & Johnson Consumer, Inc. · OHS · Jun 19, 2018 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHS/K180856

## Device Facts

- **Applicant:** Johnson & Johnson Consumer, Inc.
- **Product Code:** [OHS](/submissions/SU/subpart-e%E2%80%94surgical-devices/OHS.md)
- **Decision Date:** Jun 19, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

The Neutrogena Light Therapy Aging Mask+ is an over the counter device that is indicated for the treatment of full face wrinkles.

## Device Story

Light-based wrinkle reduction device; utilizes LED technology to emit red (620-640nm) and infrared (820-880nm) light. User-operated mask worn over face; powered by Nickel-Metal Hydride rechargeable batteries. Operation involves pressing 'On' button on controller module; device automatically shuts off after 10-minute treatment session. Intended for home use by consumers. Output consists of light energy (1.32 mW/cm²) delivered to facial skin to reduce wrinkles. Device provides non-invasive light therapy; benefits include potential reduction in appearance of facial wrinkles.

## Clinical Evidence

No clinical data. Evidence consists of bench testing, including performance verification, human factors evaluation, biocompatibility testing per ISO 10993-1, and electrical safety/electromagnetic compatibility testing.

## Technological Characteristics

Light-emitting diode (LED) therapy; red (620-640nm) and infrared (820-880nm) wavelengths. Intensity 1.32 mW/cm²; total energy dose 47.58 J/cm². Powered by Nickel-Metal Hydride rechargeable batteries. Form factor: wearable mask with controller module. Biocompatibility per ISO 10993-1. Home-use environment.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- LG Beauty LED MASK ([K170984](/device/K170984.md))
- Rejuvalite MD ([K133896](/device/K133896.md))
- Pro X OTC 5 Light Therapy ([K140471](/device/K140471.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 19, 2018

Johnson & Johnson Consumer, Inc. Cindy Abraham Associate Director, Regulatory Affairs 7050 Camp Hill Road Fort Washington, Pennsylvania 19034

Re: K180856

Trade/Device Name: Neutrogena Light Therapy Aging Mask+ Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHS Dated: March 21, 2018 Received: April 2, 2018

Dear Cindy Abraham:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good

{1}------------------------------------------------

manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

> Sincerely. Jennifer R. Stevenson -S3 For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director

Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K180856

### Device Name

Neutrogena Light Therapy Aging Mask+

Indications for Use (Describe)

The Neutrogena Light Therapy Aging Mask+ is an over the counter device that is indicated for the treatment of full face wrinkles.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### ട്. 510(k) SUMMARY

| 510(k) NUMBER:     | K180856                                                                              |
|--------------------|--------------------------------------------------------------------------------------|
| 510(k) SUBMITTER:  | Johnson & Johnson Consumer, Inc.<br>7050 Camp Hill Road<br>Fort Washington, PA 19034 |
| CONTACT:           | Cindy R. Abraham<br>Associate Director, Regulatory Affairs<br>cabraha3@its.jnj.com   |
| DATE PREPARED:     | March 21, 2018                                                                       |
| PROPRIETARY NAME:  | Neutrogena® Light Therapy Aging Mask+                                                |
| PANEL:             | General & Plastic Surgery                                                            |
| REGULATION NUMBER: | CFR Title 21, 878.4810                                                               |
| CLASSIFICATION:    | Class II                                                                             |
| PRODUCT CODES:     | OHS                                                                                  |
| COMMON NAME:       | Light-based Over The Counter Wrinkle Reduction                                       |

#### Predicate Devices [807.92(a)(3)] 5.1

- . LG Beauty LED MASK, K170984
- Rejuvalite MD, K133896
- . Pro X OTC 5 Light Therapy, K140471

#### 5.2 Device Description [807.92(a)(4)]

The Neutrogena Light Therapy Aging Mask+ uses known LED light therapy technology for the treatment of facial wrinkles. The device emits a combination of red light and infrared light. Users place the lightweight mask over the face and press the "On" button on the controller module to start treatment. The device will automatically turn off after each 10-minute treatment.

{4}------------------------------------------------

#### Intended Use and Indications for Use [807.92(a)(5)] 5.3

The Neutrogena Light Therapy Aging Mask+ is an over the counter device that is indicated for the treatment of full face wrinkles.

### 5.4 Comparison of Technological Characteristics with the Predicate Devices [807.92(a)(6)]

The Neutrogena Light Therapy Aging Mask+ is substantially equivalent to the predicate devices in that it has the same intended use and technological characteristics, as demonstrated in Table 5.1. Performance testing performed on the Neutrogena Light Therapy Aging Mask+ is sufficient to demonstrate that the subject device is as safe and effective as the legally marketed predicate devices. The technological and labeling differences do not raise new or different questions about safety or effectiveness. The Neutrogena Light Therapy Aging Mask+ is substantially equivalent to the predicate devices.

{5}------------------------------------------------

| Device Feature               | Johnson & Johnson<br>Neutrogena Light<br>Therapy Aging<br>Mask+<br>Subject Device                                                                   | LG Electronics<br>LG Beauty LED<br>Mask<br>Predicate 1                                                                                      | Trophy<br>Rejuvalite MD<br>Predicate 2                                                                                              | La Lumiere, LLC<br>Pro X OTC 5 Light<br>Therapy<br>Predicate 3                                                                                                                                                                    |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number                | Not yet assigned                                                                                                                                    | K170984                                                                                                                                     | K133896                                                                                                                             | K140471                                                                                                                                                                                                                           |
| Classification               | 21 CFR 878.4810                                                                                                                                     |                                                                                                                                             |                                                                                                                                     |                                                                                                                                                                                                                                   |
| Common Name                  | Light-based over-the-counter wrinkle reduction                                                                                                      |                                                                                                                                             |                                                                                                                                     |                                                                                                                                                                                                                                   |
| Intended Use                 | Light-based treatment to reduce wrinkles on the body in general or specific anatomical area                                                         |                                                                                                                                             |                                                                                                                                     |                                                                                                                                                                                                                                   |
| Product Code                 | OHS                                                                                                                                                 |                                                                                                                                             |                                                                                                                                     |                                                                                                                                                                                                                                   |
| Rx and/or OTC                | OTC                                                                                                                                                 |                                                                                                                                             |                                                                                                                                     |                                                                                                                                                                                                                                   |
| Use Environment              | Home-Use                                                                                                                                            |                                                                                                                                             |                                                                                                                                     |                                                                                                                                                                                                                                   |
| Anatomical<br>Location       | Full face                                                                                                                                           |                                                                                                                                             |                                                                                                                                     |                                                                                                                                                                                                                                   |
| Indications for<br>Use       | The Neutrogena<br>Light Therapy Aging<br>Mask+ is an over the<br>counter device that is<br>indicated for the<br>treatment of full face<br>wrinkles. | The LG BEAUTY<br>LED MASK is an<br>over the counter<br>device that is<br>intended for the use<br>in the treatment of<br>full face wrinkles. | The Rejuvalite MD is<br>an Over the Counter<br>device that is<br>intended for the use<br>in the treatment of<br>full face wrinkles. | The Pro X OTC 5 is<br>an Over-The-Counter<br>device intended for<br>use in the treatment<br>of facial wrinkles. It<br>is for people with<br>wrinkles on their face<br>and who have<br>Fitzpatrick skin types<br>I, II and/or III. |
| Power Source                 | Nickel-Metal<br>Hydride rechargeable<br>batteries                                                                                                   | Li-Ion rechargeable<br>batteries                                                                                                            | AC to DC                                                                                                                            | AA alkaline batteries                                                                                                                                                                                                             |
| Energy Source                | Light-emitting diodes                                                                                                                               |                                                                                                                                             |                                                                                                                                     |                                                                                                                                                                                                                                   |
| Wavelength (nm)              | Red 620-640nm<br>IR 820-880nm                                                                                                                       | Red 637nm<br>IR 854nm                                                                                                                       | Red 600, 622, 660nm<br>IR 860nm                                                                                                     | Red 620-640nm<br>IR 820-880nm                                                                                                                                                                                                     |
| Intensity<br>(mW/cm²)        | 1.32 mW/cm²                                                                                                                                         | 4.08 mW/cm²                                                                                                                                 | 2.35 mW/cm²                                                                                                                         | 0.76 mW/cm²                                                                                                                                                                                                                       |
| Treatment<br>Regimen         | 10 minutes/day for 60<br>sessions                                                                                                                   | 9 minutes/day, 5<br>days a week for 8<br>weeks                                                                                              | 3 minutes/day, 5 days<br>a week for 8 weeks                                                                                         | 15 minutes/day for 60<br>sessions                                                                                                                                                                                                 |
| Total Energy<br>Dose (J/cm²) | 47.58 J/cm²                                                                                                                                         | 88.21 J/cm²                                                                                                                                 | 16.94 J/cm²                                                                                                                         | 41.23 J/cm²                                                                                                                                                                                                                       |

# Table 5.1 Technological Characteristics

{6}------------------------------------------------

#### ર.5 Performance Data [807.92(b)(1), (b)(2)]

Performance bench and safety testing include

- . Performance verification
- Human Factors Evaluation .
- Biocompatibility testing in accordance with requirements put forth in . ISO 10993-1
- Electrical Safety and Electromagnetic Compatibility testing .

The verification, electrical safety testing, electromagnetic compatibility testing, and human factors data presented in this 510(k) submission demonstrate the Neutrogena Light Therapy Aging Mask+ device met the established specifications for its intended use and environment of use In addition, the testing demonstrated that the subject device does not raise new or different questions of safety or effectiveness when compared to the legally-marketed predicate devices.

#### 5.6 Conclusion [807.92(b)(3)]

The basis for substantial equivalence for the Neutrogena Light Therapy Aging Mask+ device and the predicate devices is performance data and conformity with recognized standards. Performance bench verification and validation demonstrate that the subject device performs comparably to the predicate devices that are marketed for the same intended use. Based on the performance testing and the similarities of the indications for use and the technological characteristics, it can be concluded that the Neutrogena Light Therapy Aging Mask+ device is substantially equivalent to the predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHS/K180856](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHS/K180856)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
