← Product Code [OHS](/submissions/SU/subpart-e%E2%80%94surgical-devices/OHS) · K170984

# LG BEAUTY LED MASK (K170984)

_Lg Electronics · OHS · Aug 23, 2017 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHS/K170984

## Device Facts

- **Applicant:** Lg Electronics
- **Product Code:** [OHS](/submissions/SU/subpart-e%E2%80%94surgical-devices/OHS.md)
- **Decision Date:** Aug 23, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

The LG BEAUTY LED MASK is an over the counter device that is intended for the use in the treatment of full face wrinkles.

## Device Story

LG Beauty LED Mask is an over-the-counter, wearable facial mask for wrinkle reduction. Device emits red (637 nm) and infrared (854 nm) LED light to induce photobiological effects on facial skin. System comprises a mask with internal LEDs, a controller for operation, and a power supply unit for charging. User wears mask over face; controller manages treatment duration; device features automatic shut-off. Protective eyecups prevent ocular exposure to LED light. Intended for home use by the patient. Output is light energy; clinical benefit is reduction of facial wrinkles.

## Clinical Evidence

Bench testing only. Performance data includes measurement of wavelength, average output power, and total irradiance. Compliance with consensus standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 62471-1 (optical hazard), ISO 14971 (risk management), and IEC 62366-1 (usability). Biocompatibility supported by use of materials (polypropylene, polycarbonate, silicone) common in other medical devices.

## Technological Characteristics

LED-based light therapy device. Wavelengths: 637 nm (Red) and 854 nm (IR). Power density: 25 mW/cm2. Materials: Polypropylene, polycarbonate, silicone. Standards: IEC 60601-1, IEC 60601-1-2, IEC 62471-1, ISO 14971, IEC 62366-1. Form factor: Wearable mask with controller and power supply. Battery-operated.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Pulsaderm Wrinkle Mask 28 and Wrinkle Mask 72 ([K163329](/device/K163329.md))
- Rejuvalite MD ([K133896](/device/K133896.md))

## Reference Devices

- illuMask Acne Light Therapy Mask ([K123999](/device/K123999.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 23, 2017

LG Electronics. Inc. % Minusk Kim BT Solutions, Inc. Unit 502, 148 Yuksamro Seoul, Gangnam-gu 06249 Republic of Korea

Re: K170984

Trade/Device Name: LG Beauty LED Mask Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: OHS Dated: July 19, 2017 Received: July 21, 2017

Dear Minusk Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply

{1}------------------------------------------------

with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely, Jennifer R. Stevenson -

# S3

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K170984

Device Name LG BEAUTY LED MASK

Indications for Use (Describe)

The LG BEAUTY LED MASK is an over the counter device that is intended for the use in the treatment of full face wrinkles.

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

# 6. 510(k) Summary

## 1. General Information

Date of preparation of 510(k) summary: July 16, 2017

Submitter: LG Electronics, Inc. Address: 222 LG-ro Jinwi-myeon, Pyeongtaek-si, Gyeonggi-do, 17709, Rep. of Korea (South Korea)

Contact Person: Minsuk Kim, BT Solutions Telephone: +82.2.538.9140 Email: cto@btsolutions.co.kr

### 2. Device Name and Code

| Device Name:         | LG BEAUTY LED MASK                                                                                                                      |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name:          | LG BEAUTY LED MASK                                                                                                                      |
| Classification Name: | Light Based Over The Counter Wrinkle Reduction                                                                                          |
| Product Code:        | OHS                                                                                                                                     |
| Regulation Number:   | 878.4810                                                                                                                                |
| Classification:      | Class II                                                                                                                                |
| Review Panel:        | General & Plastic Surgery (ODE)<br>Division of Surgical Devices (DSD)<br>General Surgery Devices Branch One - Light Based/Laser (GSDB1) |

## 3. Predicate Devices

LG BEAUTY LED MASK is substantially equivalent to the following devices

| Applicant         | Device Name                                      | 510(k) Number |
|-------------------|--------------------------------------------------|---------------|
| Pulsaderm LLC     | Pulsaderm Wrinkle Mask 28 and<br>Wrinkle Mask 72 | K163329       |
| Trophy Skin, Inc. | Rejuvalite MD                                    | K133896       |

Table 6.1 Predicate devices

Table 6.2 Reference device

| Applicant               | Device Name                      | 510(k) Number |
|-------------------------|----------------------------------|---------------|
| Photo Therapeutics, Ltd | illuMask Acne Light Therapy Mask | K123999*      |

*This device is considered as reference device, of which only limited aspects are equivalent to those of the LG BEAUTY LED MASK.

### 4. Device Description

The LG BEAUTY LED MASK is a device which allows emission of LED light in the RED (637 nm) and IR (854 nm) spectrum on to the face, which induces photobiological effect to the face for reduction of wrinkle. The LG BEAUTY LED MASK include the mask that contains LEDs on the inner surface, the controller which turns on and off the device, and the

{4}------------------------------------------------

### 510(k) Summary

power supply unit that delivers electrical power to the controller for operation or for charging the battery contained in the controller.

Users place the lightweight mask over the face, and use the controller to operate the LG BEAUTY LED MASK. The device will automatically turn off after each treatment. Emitted light from LEDs are not intended for ocular or ophthalmic treatment. To prevent irradiation of LED lights to eyes during the treatment, LG BEAUTY LED MASK has protective eyecup which blocks light energy from LEDs.

## 5. Indications / Intended Use

The LG BEAUTY LED MASK is an over the counter device that is intended for the use in the treatment of full face wrinkles.

## 6. Technical Characteristics in Comparison to Predicate Devices

|                   | Predicate Device                                                                                                                                                              | Predicate Device                                                                                                        | Proposed Device                                                                                                                             |
|-------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant         | Pulsaderm LLC                                                                                                                                                                 | Trophy Skin, Inc.                                                                                                       | LG Electronics                                                                                                                              |
| Device Name       | Pulsaderm Wrinkle<br>Mask 28 and Wrinkle<br>Mask 72                                                                                                                           | Rejuvalite MD                                                                                                           | LG BEAUTY LED<br>MASK                                                                                                                       |
| 510(k)<br>number  | K163329                                                                                                                                                                       | K133896                                                                                                                 | N/A                                                                                                                                         |
| Product Code      | OHS                                                                                                                                                                           | OHS                                                                                                                     | OHS                                                                                                                                         |
| Intended Use      | The Pulsaderm Wrinkle<br>Masks 28 and 72 are<br>intended for the use in<br>the treatment of facial<br>wrinkles and for people<br>with Fitzpatrick Skin<br>Types I, II and III | The Rejuvalite MD is<br>an over-the-counter<br>device intended for the<br>use in the treatment of<br>full-face wrinkles | The LG BEAUTY<br>LED MASK is an<br>over the counter<br>device that is<br>intended for the use<br>in the treatment of<br>full face wrinkles. |
| LED<br>emission   | Emission of red (620-<br>630 nm) and IR (850<br>nm)                                                                                                                           | Emission of red (600,<br>622, 660 nm) and IR<br>(820 nm)                                                                | RED (637 nm) and IR<br>(854 nm)                                                                                                             |
| LED power         | 21.18 - 25.35 mW/cm $^{2}$ total                                                                                                                                              | 62 mW/cm $^{2}$ total                                                                                                   | 25 mW/cm $^{2}$ total                                                                                                                       |
| Treatment<br>time | 15 minutes everyday                                                                                                                                                           | 3 minutes daily<br>5 days per week for 8<br>weeks                                                                       | 9 minutes daily<br>5 days per week for 8<br>weeks                                                                                           |

LG BEAUTY LED MASK is substantially equivalent to the following devices

## 7. Performance Data

Non-clinical tests: Measurement of wavelength, average output power, and total irradiance (power density, in units of mW/cm2) of treatment LEDs were performed. Other performance, such as optical hazard, electromagnetic compliance, etc, were tested using following consensus standards:

- Basic safety and essential performance of the LG BEAUTY LED MASK is tested and evaluated according to the FDA-recognized consensus standard, IEC 60601-1.

{5}------------------------------------------------

### 510(k) Summary

- Effect to the device by electromagnetic disturbances were tested and evaluated using -IEC 60601-1-2.
- Optical hazard was evaluated using IEC 62471-1. -
- -Risk management was recorded by referring to ISO 14971.
- -Usability was documented using IEC 62366-1.

The portion of the device that touches patient body is made of polypropylene, polycarbonate plastic, and silicone, which have been used for other medical devices without any biocompatibility risk.

## 8. Conclusion

On the basis of the information provided in this Summary, LG Electronics believes LG BEAUTY LED MASK is substantially equivalent to legally commercialized predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHS/K170984](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHS/K170984)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
