K132613 · Biorenew Labs, LLC · OHS · Feb 5, 2014 · General, Plastic Surgery
Device Facts
Record ID
K132613
Device Name
SONILASE LIGHT THERAPY DEVICE
Applicant
Biorenew Labs, LLC
Product Code
OHS · General, Plastic Surgery
Decision Date
Feb 5, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Sonilase™ Light Device Red is used to treat periorbital wrinkles and fine lines around the eyes (i.e., crow's feet).
Device Story
Sonilase Light Device is a handheld, battery-operated, OTC red light therapy system for periorbital wrinkles and fine lines. Device utilizes 660nm LED light source to deliver 50 mW/cm2 energy directly to skin. System includes handheld unit, lens cover, cleansing brush, recharging stand, and UV sanitizing cradle. Operated by consumer; applied directly to periorbital region for 160-second treatment sessions, twice weekly for six weeks. Device provides non-invasive light therapy; intended to improve appearance of fine lines. No user-serviceable components.
Clinical Evidence
No clinical efficacy data provided. Bench testing included functional performance, electrical safety (IEC 60601-1-1), and electromagnetic compatibility (IEC 60601-1-2). Usability/label comprehension study conducted with 38 participants; 100% comprehension rate and 97% successful device usage rate reported.
Technological Characteristics
Handheld, cordless, battery-powered (7.4V Li-ion). Light source: 660nm red LEDs. Output: 50 mW/cm2. Includes recharging stand and UV sanitizing cradle. Software validated per FDA guidance. Complies with IEC 60601-1-1 and IEC 60601-1-2.
Indications for Use
Indicated for individuals with periorbital wrinkles and fine lines (crow's feet). OTC use.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
# 510(k) Summary
## K132613
This summary of 510(k) summary information is being submitted in accordance with the requirements of 21 CFR § 878.4810.
Submission Date: August 8, 2013
#### 1. Submitter Information:
AEGIS Regulatory, Inc. - Susan Anthoney-DeWet 2424 Dempster Drive Coralville, IA 52241 Tel.: 865-982-5552 Email: sue@fdalistingconsultants.com
For Specifications Developer: Biorenew Labs, LLC
Attn: Jon David 415 Pier Avenue Hermosa Beach, CA 90254 Tel.: 310-400-0631 Email: jon@biorenewlabs.com
#### 2. General Information
2.1 Classification Name: Light Based Over-The-Counter Wrinkle Reduction Device
2.2 Common/Usual Name: Sonilase Light Device Red
2.3 Proprietary Names: Sonilase Light Device
#### 2.4 Classification: Class II
2.5 Classification Number: 878.4810
2.6 Product Codes: OHS
#### 3. Device Description:
The Sonilase Light Deviceis a modular system that offers red light therapy for the treatment of wrinkles, rhytides, and fine lines in the periorbital region.
{1}------------------------------------------------
The Sonilase system components include the handheld unit containing the LED module , attachable clear plastic lens cover, an adjunct attachable cleansing brush head, recharging stand, 7.4V LI-ION BATTERY, 9V UL approved power supply , UV Sanitizing cradle and goggles.
The unit is applied directly to the skin to ensure consistent administration of light during each treatment. The device does not contain any user serviceable components. The device is sold as Over the Counter (OTC).
### 4. Indications / Intended Use:
The Sonilase™ Light Device is used to treat periorbital wrinkles and fine lines around the eyes
(i.e., crow's feet).
Rx or OTC:
The Sonilase Light Device is an Over the Counter (OTC) device. The labeling, instructions, and User Operations (21 CFR § 801.60 and 61), are designed for layman understanding and use. The predicate device is OTC.
#### 5. Predicate Device:
This device is substantially equivalent to the following predicate device, which is currently in safe and effective commerce:
- 1. K110735 Tanda Max OTC System (Pharos Life Corporation)
#### Predicate Chart
| Device | Tanda Max OTC<br>System | Sonilase Red OTC<br>System |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Pharos Life Corp. | Biorenew Labs, LLC |
| | K110735 | K132613 |
| | A Predicate Device | This Submission |
| Indications | The Tanda Max OTC<br>System is intended to be<br>used for the treatment of<br>wrinkles, rhytides, and<br>fine lines in the<br>periorbital region. | The Sonilase Red OTC<br>System is intended to be<br>used for the treatment of<br>wrinkles, rhytides, and<br>fine lines in the<br>periorbital region. |
| Handheld | Yes | Yes |
| Wavelength | 660 nm | 660 nm |
| Modes | On/Off | On/Off |
| Red power<br>source | LEDs | LEDs |
{2}------------------------------------------------
| Device | Tanda Max OTC<br>System | Sonilase Red OTC<br>System |
|-----------------------|--------------------------------------------------------|--------------------------------------------------------|
| | Pharos Life Corp. | Biorenew Labs, LLC |
| | K110735 | K132613 |
| | A Predicate Device | This Submission |
| Visible light<br>LEDs | Yes | Yes |
| Energy Level | 50 mW total | 50 mW total |
| Power Supply | One 9v rechargeable<br>alkaline battery | One 9v rechargeable<br>alkaline battery |
| Treatment<br>Time | 160 seconds2 days per<br>week for 6 weeks | 160 seconds2 days per<br>week for 6 weeks |
| Target<br>Population | Individuals with<br>periorbital lines and<br>wrinkles. | Individuals with<br>periorbital lines and<br>wrinkles. |
| Location for<br>Use | OTC | OTC |
Summary of the technological characteristics of the device compared to predicate device:
1. Has the same intended use as the predicate (i.e., Treatment of wrinkles, rhytides and fine lines in the periorbital region;
2. Has the same output (i.e., 50 mW/cm2 ) as the predicate;
3. Utilizes the same wavelength (i.e., 660 nm) as the predicate device;
4. Utilizes the same treatment duration (i.e., 160 seconds) as the predicate device;
5. Utilizes the same treatment regimen of twice a week for six weeks.
The Sonilase Light Device and the above referenced predicate, Tanda Max OTC device, are Over the Counter Devices used to treat wrinkles as defined in 21 CFR § 878.4810. These devices utilize red diodes at 660nm to provide narrow bands of light energy to treat periorbital wrinkles. The performance achieved by these devices is similar with equal power output. The devices are handheld, and intended to be placed directly on the skin. They are manufactured out of similar materials. Based upon comparison to the predicate device, the Sonilase Light Device has the same intended uses, with similar
{3}------------------------------------------------
technological characteristics as the predicate device. The system performs as intended and does not raise any new safety or effectiveness issues.
#### 6. Technological Characteristics
The Sonilase Light Device operates by a cordless system drawing upon its 7.4V rechargeable battery to deliver the treatment. The Sonilase Light Device delivery system used for applying therapy for the use in the treatment of periorbital wrinkles, is by emitting at least 50 mW/cm² of red (660nm) light via an electric light emitting diodes [LEDs] energy source. The device is not intended for ocular applications or direct eye exposure.
### 7. Performance Testing and Standards:
Testing of the Sonilase included functional performance testing, software validation testing and user safety testing.
The results of this testing are as follows:
Complies with IEC 60601-1-2 Electromagnetic Compatibility Complies with IEC 60601-1-1 Electrical Safety
The Sonilase software was tested and validated in accordance with FDA's "Guidance for the content of Premarket Submissions for Software Contained in Medical Devices"
A Usability/Label Comprehension Study was conducted with 38 participants. The results of the study found that: 100% of the participants were able to comprehend the labeling. 97% of the participants were able to use the device successfully.
## 9. Substantial Equivalence Conclusion
After an analysis of the safety, indications, intended uses, performance, design materials, power output, technological properties, treatment areas, and treatment regimes the manufacturer believes that no significant differences exist between the device and the predicate device. Therefore substantial equivalency is requested.
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the words "Public Health Service" in a simple, sans-serif font. The text is black and appears to be on a white background. The words are stacked on top of each other, with "Public Health" on the top line and "Service" on the bottom line.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Biorenew Labs, LLC % Ms. Susan Anthoney AEGIS Regulatory Incorporated 2424 Dempster Drive Coralville, Iowa 52241
Re: K132613
Trade/Device Name: Sonilase Light Device Red Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: OHS Dated: January 5, 2013 Received: January 8, 2013
Dear Ms. Anthoney:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
February 5, 2014
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{5}------------------------------------------------
Page 2 - Ms. Susan Anthoney
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
For
Sincerely yours.
Joshua F.C.Nipper -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K132613
Device Name Sonilase Light Device Red
#### Indications for Use (Describe)
The Sonilase™ Light Device Red is used to treat periorbital wrinkles and fine lines around the eyes (i.e., crow's feet).
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FOR FDA USE ONLY ( ್ನು ಇತ್ತಿದ್ದರೆ ಕೆ ಬ Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Image /page/6/Picture/11 description: The image shows the name "Neil R Ogden" followed by "-S". Below that is the date "2014.02.04" and the time "14:46:55". To the right of the name is a logo that appears to be a stylized design. The time also includes "05:00".
FORM FDA 3881 (1/14) for BSA Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.