QUASAR C100, BABY QUASAR PLUS, QUASAR MD PLUS; WRINKLE REDUCTION DEVICE

K130225 · Silverbay, LLC, D/B/A Quasar Bio-Tech · OHS · Jul 11, 2013 · General, Plastic Surgery

Device Facts

Record IDK130225
Device NameQUASAR C100, BABY QUASAR PLUS, QUASAR MD PLUS; WRINKLE REDUCTION DEVICE
ApplicantSilverbay, LLC, D/B/A Quasar Bio-Tech
Product CodeOHS · General, Plastic Surgery
Decision DateJul 11, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Quasar C100 Wrinkle Reduction Device is intended to emit energy in the red and IR region of the spectrum, specifically indicated for the treatment of full-face wrinkles. The Baby Quasar Plus Wrinkle Reduction Device is intended to emit energy in the red and iR region of the spectrum, specifically indicated for the treatment of fullface wrinkles. The Quasar MD Plus Wrinkle Reduction Device is intended to emit energy in the red and IR region of the spectrum, specifically indicated for the treatment of fullface wrinkles.

Device Story

Handheld light therapy devices (Quasar C100, Baby Quasar Plus, Quasar MD Plus) emit red and near-infrared light (610nm to 850nm) to treat full-face wrinkles. Devices consist of LED arrays (12-20 red/infrared pairs) powered by 12V wall-mount supplies. Operated by consumers at home; users apply light to facial skin for 3-minute sessions, minimum 5 days per week. Output is constant-wave light energy (at least 65 mW/cm²). No complex processing or AI; simple on/off functionality. Benefits include non-invasive wrinkle reduction. Safety ensured by biocompatible materials (ABS plastic, polycarbonate, anodized aluminum) and compliance with electrical/radiological safety standards.

Clinical Evidence

Bench testing only. Compliance with IEC/EN 60601-1 (electrical safety), IEC/EN 60601-1-2 (EMC), and IEC 62471 (photobiological safety) demonstrated. Human Factors Study conducted for label comprehension and device use confirmed safety and effectiveness for intended users and environments.

Technological Characteristics

Handheld LED arrays; wavelengths 610-850nm; constant waveform; 12V wall-mount power supply. Materials: ABS plastic, polycarbonate lenses, anodized aluminum. Head sizes: 8.7 cm² to 20 cm². Compliance: IEC/EN 60601-1, IEC/EN 60601-1-2, IEC 62471. No software/algorithm.

Indications for Use

Indicated for the treatment of full-face wrinkles in individuals. Intended for over-the-counter home use. Not intended for ocular applications or direct eye exposure.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(k) Summary #### K130225 This summary of 510(k) summary information is being submitted in accordance with the requirements of 21 CFR § 878.4810. Submission Date: January 15, 2013 | 1. Submitter Information: | AEGIS Regulatory, Inc. - Robert T. Wagner | |---------------------------|-----------------------------------------------| | | 2424 Dempster Drive | | | Coralville, IA 52241 | | | Tel.: 865-982-5552 | | | Email: bob@fdalistingconsultants.com | | For Manufacturer: | Silver Bay, LLC d/b/a Quasar Bio-Technologies | | | Attn: Peter Nesbitt | | | 1431 Tallevast Rd. | | | Sarasota, FL 34231 | | | Tel.: 941-306-5812 | | | Email: peter@quasarbiotech.com | #### 2. General Information - 2.1 2.2 Common/Usual Name: Light Therapy System, Quasar C100, Baby Quasar Plus, Quasar MD Plus 2.3 Proprietary Names: Quasar Calypso C100 Wrinkle Reduction Device, Baby Quasar Plus, Quasar MD Plus - 2.4 Classification: Class II - 2.5 Classification Number: 878.4810 - 2.6 Product Codes: OHS {1}------------------------------------------------ #### 3. Device Description: The Silver Bay, Quasar C100. Anti-Wrinkle device consists of a collection of red and near infrared diodes (LEDs), packaged in a compact hand held device. The device has a head size of 20 cm² containing the LED array and an on/off switch. The device is made of ABS plastic with clear polycarbonate lenses covering the LED light source. The Quasar C100 uses a 12 volt wall mount power supply. The C100 light delivery system used for applying therapy for the treatment of full-face wrinkles, by emitting at least 65 mWcm² of red and near infrared (610nm to 850nm) light via an electric light emitting diodes [LEDs] energy source. There are 20 red LEDs and 20 infrared LEDs in the head. The device is not intended for ocular applications or direct eye exposure. The Silver Bay, Baby Quasar Plus Anti-Wrinkle device consists of a collection of red and near infrared diodes (LEDs), packaged in a compact hand held device. The device has a head size of 8.7 cm 2 containing the LED array and an on/off switch. The device is made of anodized aluminum. The Baby Quasar Plus uses a 12 volt wall mount power supply. The Baby Quasar Plus light delivery system used for applying therapy for the use in the treatment of full-face wrinkles, by emitting at least 65 mWcm² of red and near infrared (610nm to 850nm) light via an electric light emitting diodes [LEDs] energy source. There are 12 red LEDs and 12 infrared LEDs in the head. The device is not intended for ocular applications or direct eye exposure. The Silver Bay, Quasar MD Plus Anti-Wrinkle device consists of a collection of red and near infrared diodes (LEDs), packaged in a compact hand held device. The device has a head size of 15.2 cm 2 containing the LED array and an on/off switch. The device is made of anodized aluminum. The Baby Quasar Plus uses a 12 volt wall mount power supply. The Quasar MD Plus light delivery system used for applying therapy for the use in the treatment of full-face wrinkles, by emitting at least 65 mWcm² of red and near infrared (610nm to 850m), light via an electric light emitting diodes [LEDs] energy source. There are 12 red LEDs and 12 infrared LEDs in the head. The device is not intended for ocular applications or direct eve exposure. #### 4. Indications / Intended Use: The Quasar C100 is an Over-The-Counter handheld device intended to emit energy in the red/IR spectrum, specifically indicated for use in the treatment of full-face wrinkles. The Baby Quasar Plus is an Over-The-Counter handheld device intended to emit energy in the red/IR spectrum, specifically indicated for use in the treatment of full-face wrinkles. The Quasar MD Plus is an Over-The-Counter handheld device intended to emit energy in the red/IR spectrum, specifically indicated for use in the treatment of full-face wrinkles {2}------------------------------------------------ Rx or OTC: he C100, Baby Quasar Plus, and Quasar MD Plus are Over the Counter . The labeling, instructions, and # Predicate Device ખં This device is substantially equivalent to the following predicate devices, which are currently in safe and effective commerce - K120775 Light for Wrinkles (Led Intellectual Properties, LLC : し - K112362 Quasar C100 Anti-Wrinkle Device (Quasar Bio-Tech Inc ਪ # Predicate Cha | Device | | | | | | |-------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Light for Wrinkles<br>AAL OCT (K120775)<br>LED Intellectual<br>Properties LLC | C100<br>Quasar Bio-Tech Inc<br>K112362<br>A Predicate Device | C100<br>Quasar Bio-Tech Inc<br>KXXXXXX<br>This Submission | Baby Quasar Plus<br>Quasar Bio-Tech Inc<br>KXXXXXX<br>This Submission | Quasar MD Plus<br>Quasar Bio-Tech Inc<br>KXXXXXX<br>This Submission | | | Indications | The Light for Wrinkles is<br>an Over-The-Counter<br>handheld device<br>intended for use in the<br>treatment of full-face<br>wrinkles. The target<br>patient population for<br>the AAL OCT is persons<br>with facial wrinkles. It<br>is designed for home<br>use. | The C100 is intended to<br>emit energy in the red<br>and IR region of the<br>spectrum, specifically<br>indicated for use in the<br>treatment of periorbital<br>wrinkles. The target<br>patient population for the<br>C100 is periorbital<br>wrinkles. The C100 is<br>designed for home use. | The C100 is intended to<br>emit energy in the red<br>and IR region of the<br>spectrum, specifically<br>indicated for use in the<br>treatment of full-face<br>wrinkles. The target | The Baby Quasar Plus is<br>intended to emit energy in<br>the red and IR region of<br>the spectrum, specifically<br>indicated for use in the<br>treatment of full-face<br>wrinkles. The target<br>patient population for the<br>Baby Quasar Plus is the<br>same as the predicate<br>devices. Like the<br>predicate devices, the<br>Baby Quasar Plus is<br>designed for home use. | The Quasar MD Plus is<br>intended to emit energy in<br>the red and IR region of<br>the spectrum, specifically<br>indicated for use in the<br>treatment of full-face<br>wrinkles. The target<br>patient population for the<br>Quasar MD Plus is the<br>same as the predicate<br>devices. Like the predicate<br>devices, the Quasar MD<br>Plus is designed for home<br>use. | {3}------------------------------------------------ | Device | Light for Wrinkles<br>AAL OCT (K120775)<br>LED Intellectual<br>Properties LLC | C100<br>Quasar Bio-Tech Inc | C100<br>Quasar Bio-Tech Inc | Baby Quasar Plus<br>Quasar Bio-Tech Inc | Quasar MD Plus<br>Quasar Bio-Tech Inc | |-----------------------|-------------------------------------------------------------------------------|------------------------------------------------|------------------------------------------------|------------------------------------------------|------------------------------------------------| | | A Predicate Device | K112362 | KXXXXXX | KXXXXXX | KXXXXXX | | | A Predicate Device | A Predicate Device | This Submission | This Submission | This Submission | | Handheld | Yes | Yes | Yes | Yes | Yes | | Wavelengths | 605nm, 630nm, 660nm,<br>855nm | 610, 630 and 660, 850nm | 610, 630 and 660, 850nm | 610, 630 and 660, 850nm | 610, 630 and 660, 850nm | | Modes | On/Off | On/Off | On/Off | On/Off | On/Off | | IR power<br>source | LEDs | LEDs | LEDs | LEDs | LEDs | | Visible light<br>LEDs | Yes | Yes | Yes | Yes | Yes | | Waveform | Constant | Constant | Constant | Constant | Constant | | Energy Source | 70 LEDs over 27 sq. cm. | 40 LEDs. Over 20 sq. cm | 40 LEDs. Over 20 sq. cm | 24 LEDs. Over 8.7 sq. cm | 24 LEDs. Over 15.2 sq. cm | | Energy Level | 65 mW total | 65 mW total | 65 mW total | 65 mW total | 65 mW total | | Power Supply | 115VAC Electric Outlet<br>Power Supply | 115VAC Electric Outlet<br>Power Supply | 115VAC Electric Outlet<br>Power Supply | 115VAC Electric Outlet<br>Power Supply | 115VAC Electric Outlet<br>Power Supply | | Treatment<br>Time | 3 minutes daily,<br>minimum 5 days per<br>week | 3 minutes daily,<br>minimum 5 days per<br>week | 3 minutes daily,<br>minimum 5 days per<br>week | 3 minutes daily,<br>minimum 5 days per<br>week | 3 minutes daily,<br>minimum 5 days per<br>week | | Target<br>Population | Individuals with facial<br>lines and wrinkles. | Individuals with<br>periorbital lines and | Individuals with facial<br>lines and wrinkles. | Individuals with facial<br>lines and wrinkles. | Individuals with facial<br>lines and wrinkles. | , " {4}------------------------------------------------ | Device | | | | | Location for Use | | |--------------------|---------------------------------------------------------|--------------------|--------------------|-----------|------------------|--| | Light for Wrinkles | AAL OCT (K120775)<br>LED Intellectual<br>Properties LLC | A Predicate Device | | wrinkles. | OTC | | | C100 | Quasar Bio-Tech Inc | K112362 | A Predicate Device | | OTC | | | C100 | Quasar Bio-Tech Inc | KXXXXXX | This Submission | | OTC | | | Baby Quasar Plus | Quasar Bio-Tech Inc | KXXXXXX | This Submission | | OTC | | | Quasar MD Plus | Quasar Bio-Tech Inc | KXXXXXX | This Submission | | OTC | | mmary of the technological characteristics of the device compared to predicate dev are abovent grouped {5}------------------------------------------------ #### 6. Biocompatibility: The only patient contact material in the C100 is the light head and body of the device. The light head in contact with the face is Polycarbonate and the body is constructed of ABS plastic, the same materials used in the predicate device. The biocompatibility of these materials are well known and considered safe when in contact with healthy skin. A review of the Biocompatibility decision is shown on the "General Program Memorandum- #G95-1, Attachment C, Biocompatibility Flow Chart for the Selection of Toxicity Tests for 510(k)s." The conclusion is that the C100 m does not raise any new safety issues. #### 7. Performance Testing and Standards: These devices have been tested under and are in compliance with performance standards that have been established for such devices under Section 878 of the Federal Food, Drug, and Cosmetics Act. All electrical and radiological products made by the applicant have been OSHA/NRTL listed, and have received constituent marks. The device has been tested and is in conformity with IEC/EN 60601-1, IEC/ EN 60601-1-2, IEC 62471 Standards. A Human Factors Study for label comprehension and device use was conducted. The results of the HF Study found that the C100, Baby Quasar + and Quasar MD + have been found to be adequately safe and effective for the intended users, its intended uses, and use environments. #### 8. Statement of Safety and Effectiveness: The information in this 510(k) submission was used to support the safety and effectiveness of this device with respect to its cited predicates. #### 9. Substantial Equivalence Discussion After an analysis of the safety, indications, intended uses, dose rates, performance, features, design materials, power output, technological properties, treatment areas, treatment regimes and methods of operation, the manufacturer believes that no significant differences exist between the device and the predicates listed in Section 5 Therefore substantial equivalency is requested. {6}------------------------------------------------ Image /page/6/Picture/0 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized abstract symbol on the right, consisting of three curved lines that resemble a person with outstretched arms. To the left of the symbol, the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol. The text is in all caps and is smaller than the symbol. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002 Silverbay LLC, d/b/a Quasar Biotech % AEGIS Regulatory, Inc. Ms. Susan Anthoney 2424 Dempster Drive Coralville, Iowa 52241 July 11, 2013 Re: K130225 Trade/Device Name: Quasar C100, Baby Quasar Plus, Quasar MD Plus; Wrinkle Reduction Device Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: OHS Dated: June 11, 2013 Received: June 18, 2013 Dear Ms. Anthoney: We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must {7}------------------------------------------------ Page 2 - Ms. Susan Anthoney comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safetv/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. # Mark N. Melkerson -S Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {8}------------------------------------------------ ### INDICATIONS FOR USE STATEMENT 510(k) Number: K130225 Device Name: Quasar C100, Baby Quasar Plus, Quasar MD Plus; Wrinkle Reduction Device Indications for Use: The Quasar C100 Wrinkle Reduction Device is intended to emit energy in the red and IR region of the spectrum, specifically indicated for the treatment of full-face wrinkles. The Baby Quasar Plus Wrinkle Reduction Device is intended to emit energy in the red and iR region of the spectrum, specifically indicated for the treatment of fullface wrinkles. The Quasar MD Plus Wrinkle Reduction Device is intended to emit energy in the red and IR region of the spectrum, specifically indicated for the treatment of fullface wrinkles. ## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) OR Prescription Use (Per 21 CFR 801.109) Over-The-Counter Use__X (Optional Format 1-2-96) زنرین Neil R Ogden 2013.07.11 09:24:16 -04'00' (Division Sign-Off) for MXM Division of Surgical Devices 510(k) Number _K130225 Silver Bay LLC .
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...