← Product Code [OHS](/submissions/SU/subpart-e%E2%80%94surgical-devices/OHS) · K112362

# C100 WRINKLE REDUCTION DEVICE (K112362)

_Silver Bay, LLC · OHS · Aug 14, 2012 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHS/K112362

## Device Facts

- **Applicant:** Silver Bay, LLC
- **Product Code:** [OHS](/submissions/SU/subpart-e%E2%80%94surgical-devices/OHS.md)
- **Decision Date:** Aug 14, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

The Quasar C100 Wrinkle Reduction Device is intended to emit energy in the red and IR region of the spectrum, specifically indicated for the treatment of periorbital wrinkles.

## Device Story

Quasar Calypso C100 is a handheld light therapy device for home use. It utilizes an array of 20 red and 20 infrared LEDs (610nm to 850nm) to deliver light energy to the skin. Powered by a 12V wall-mount supply, the device emits a constant waveform with a total output of 65 mW/cm². The user operates the device by placing the light head directly on or slightly above the skin in the periorbital area for 3 minutes daily, at least 5 days per week. The device is intended to reduce periorbital wrinkles. It is constructed of ABS plastic with polycarbonate lenses. No complex software or AI algorithms are involved; operation is limited to an on/off switch.

## Clinical Evidence

Bench testing only. The device complies with IEC/EN 60601-1 (electrical safety), IEC/EN 60601-1-2 (EMC), and IEC 62471 (photobiological safety). No clinical data provided.

## Technological Characteristics

Handheld LED light therapy system. 20 red and 20 infrared LEDs (610-850nm). Constant waveform. 65 mW/cm² output. Materials: ABS plastic body, polycarbonate lenses. Power: 12V wall-mount supply. Standards: IEC/EN 60601-1, IEC/EN 60601-1-2, IEC 62471. No software/algorithm.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Light for Wrinkles ([K101190](/device/K101190.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K112362 P. 1 of 9

1 of 5

# AUG 1 4 2012

# 510(k) Summary

K

This summary of 510(k) summary information is being submitted in accordance with the requirements of 21 CFR § 878.4810.

Submission Date: August 1, 2011

1. Submitter Information:

AEGIS Regulatory, Inc. - Robert T. Wagner 1131 Anthem View Lane Knoxville, TN 37922 Tel.: 865-982-5552 Email: bob@fdalistingconsultants.com Silver Bay, LLC d/b/a Quasar Bio-Technologies Attn: Peter Nesbitt 1431 Tallevast Rd. Sarasota, FL 34231 Tel.: 941-306-5812 Email: peter@quasarbiotech.com

# For Manufacturer:

2. General Information

2.1 Classification Name: Wrinkle Reduction Device

2.2 Common/Usual Name: Light Therapy System, Quasar Calypso C100

2.3 Proprietary Names: Quasar Calypso C100 Wrinkle Reduction Device

2.4 Classification: Class II

2.5 Classification Number: 878.4810

2.6 Product Codes: OHS

#### 3. Device Description:

The Silver Bay, Quasar Calypso, C100 Anti-Wrinkle device consists of a collection of red and near infrared diodes (LEDs), packaged in a compact hand held device. The device has a head containing the LED array and an on/off switch.

The device is made of ABS plastic with clear polycarbonate lenses covering the LED light source. The Quasar Calypso C100 uses a 12 volt wall mount power supply.

{1}------------------------------------------------

K112362. p. 2.0+5

The C100 light delivery system used for applying therapy for the use in the treatment of periorbital wrinkles, by emitting at least 65 mW/cm² of red and near infrared (610nm to 850nm) light via an electic light emitting diodes [LEDs] energy source. There are 20 red LEDs and 20 infrared LEDs in the head. The device is not intended for ocular applications or direct eye exposure.

### 4. Indications / Intended Use:

The Quasar Calypso C100 is an Over-The-Counter handheld device intended to emit energy in the red/IR spectrum, specifically indicated for use in the treatment of periorbital wrinkles.

Rx or OTC:

The C100 is an Over the Counter (OTC) device. The labeling, instructions, and User Operations (21 CFR § 801.60 and 61), are designed for layman understanding and use. The predicate device is OTC.

#### 5. Predicate Device:

This device is substantially equivalent to the following predicate devices, which are currently in safe and effective commerce:

1. K101190 - Light for Wrinkles (Led Intellectual Properties, LLC)

#### Predicate Chart

| Device                | Light for Wrinkles                                                                                                                                                                                                                                                         | C100                                                                                                                                                                                                                                                                                                                            |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                       | AAL OCT (K101190)<br>LED Intellectual<br>Properties LLC                                                                                                                                                                                                                    | Quasar Bio-Tech Inc                                                                                                                                                                                                                                                                                                             |
|                       | Properties LLC                                                                                                                                                                                                                                                             | K112362                                                                                                                                                                                                                                                                                                                         |
|                       | A Predicate Device                                                                                                                                                                                                                                                         | This Submission                                                                                                                                                                                                                                                                                                                 |
| Indications           | The Light for Wrinkles<br>is an Over-The-Counter<br>handheld device<br>intended for use in the<br>treatment of<br>periorbital wrinkles.<br>The target patient<br>population for the AAL<br>OCT is persons with<br>periorbital wrinkles. It<br>is designed for home<br>use. | The C100 is intended<br>to emit energy in the<br>red and IR region of<br>the spectrum,<br>specifically indicated<br>for use in the<br>treatment of<br>periorbital wrinkles.<br>The target patient<br>population for the<br>C100 is the same as the<br>predicate devices. Like<br>the predicate devices,<br>the C100 is designed |
| Device                | Light for Wrinkles                                                                                                                                                                                                                                                         | C100                                                                                                                                                                                                                                                                                                                            |
|                       | AAL OCT (K101190)<br>LED Intellectual<br>Properties LLC                                                                                                                                                                                                                    | Quasar Bio-Tech Inc<br>K112362                                                                                                                                                                                                                                                                                                  |
|                       | A Predicate Device                                                                                                                                                                                                                                                         | This Submission                                                                                                                                                                                                                                                                                                                 |
|                       |                                                                                                                                                                                                                                                                            | for home use.                                                                                                                                                                                                                                                                                                                   |
| Handheld              | Yes                                                                                                                                                                                                                                                                        | Yes                                                                                                                                                                                                                                                                                                                             |
| Wavelength<br>s       | 605nm, 630nm,<br>660nm, 855nm                                                                                                                                                                                                                                              | 610, 630 and 660,<br>850nm                                                                                                                                                                                                                                                                                                      |
| Modes                 | On/Off                                                                                                                                                                                                                                                                     | On/Off                                                                                                                                                                                                                                                                                                                          |
| IR power<br>source    | LEDs                                                                                                                                                                                                                                                                       | LEDs                                                                                                                                                                                                                                                                                                                            |
| Visible light<br>LEDs | Yes                                                                                                                                                                                                                                                                        | Yes                                                                                                                                                                                                                                                                                                                             |
| Waveform              | Constant                                                                                                                                                                                                                                                                   | Constant                                                                                                                                                                                                                                                                                                                        |
| Energy<br>Source      | 70 LED's over 27 sq.<br>cm.                                                                                                                                                                                                                                                | 40 LEDs. Over 20 sq.<br>cm                                                                                                                                                                                                                                                                                                      |
| Energy<br>Level       | 65 mW total                                                                                                                                                                                                                                                                | 65 mW total                                                                                                                                                                                                                                                                                                                     |
| Power<br>Supply       | 115VAC Electric Outlet<br>Power Supply                                                                                                                                                                                                                                     | 115VAC Electric Outlet<br>Power Supply                                                                                                                                                                                                                                                                                          |
| Treatment<br>Time     | 3 minutes daily,<br>minimum 5 days per<br>week                                                                                                                                                                                                                             | 3 minutes daily,<br>minimum 5 days per<br>week                                                                                                                                                                                                                                                                                  |
| Target<br>Population  | Individuals with<br>periorbital lines and<br>wrinkles.                                                                                                                                                                                                                     | Individuals with<br>periorbital lines and<br>wrinkles.                                                                                                                                                                                                                                                                          |
| Location for<br>Use   | OTC                                                                                                                                                                                                                                                                        | OTC                                                                                                                                                                                                                                                                                                                             |

{2}------------------------------------------------

Summary of the technological characteristics of the device compared to predicate device:

·

.

{3}------------------------------------------------

U of 5

The C100 and the above referenced predicate. Light for Wrinkles device, are Over the Counter Devices used to treat wrinkles as defined in 21 CFR § 878.4810. These devices utilize red and infrared diodes from 610 to 850 nm to provide narrow bands of light energy to treat wrinkles. The performance achieved by these devices is similar with equal power output. The devices are handheld, and intended to be placed directly on the skin or held just over the skin. They are manufactured out of similar materials. Based upon comparison to the predicate device, the C100 has the same intended uses, with similar technological characteristics as the predicate device. The system performs as intended and does not raise any new safety or effectiveness issues.

#### 6. Biocompatibility:

The only patient contact material in the C100 is the light head and body of the device.

The light head in contact with the face is Polycarbonate and the body is constructed of ABS plastic. the same materials used in the predicate device. The biocompatibility of these materials are well known and considered safe when in contact with heatthy skin. A review of the Biocompatibility decision is shown on the "General Program Memorandum- #G95-1. Attachment C, Biocompatibility Flow Chart for the Selection of Toxicity Tests for 510(k)s."

The conclusion is that the C100 m does not raise any new safety issues.

#### 7. Performance Testing and Standards:

These devices have been tested under and are in compliance with performance standards that have been established for such devices under Section 878 of the Federal Food, Drug, and Cosmetics Act. All electrical and radiological products made by the applicant have been OSHA/NRTL listed, and have received constituent marks.

The device has been tested and is in conformity with IEC/EN 60601-1, IEC/ EN 60601-1-2, IEC 62471 Standards.

#### 8. Statement of Safety and Effectiveness:

The information in this 510(k) submission was used to support the safety and effectiveness of this device with respect to its cited predicates.

#### 9. Substantial Equivalence Discussion

After an analysis of the safety, indications, intended uses,

performance, features, design materials, power output, technological properties, treatment areas, treatment regimes and methods of operation, the manufacturer believes that no

{4}------------------------------------------------

significant differences exist between the device and the predicates listed in Section 5 Therefore substantial equivalency is requested.

:

·

K112362 2.5

{5}------------------------------------------------

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/5/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features the department's symbol, which is a stylized caduceus, a symbol often associated with medicine and healthcare. The symbol is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The text is in all capital letters and is evenly spaced around the symbol.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Silver Bay, LLC % Aegis Regulatory, Incorporated Mr. Robert T. Wagner CEO 1131 Anthem View Lane Knoxville, Tennessee 37922

AUG 1 4 2012

Re: K112362

Trade/Device Name: Quasar C100 Wrinkle Reduction Device Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: OHS Dated: August 13, 2012 Received: August 13, 2012

Dear Mr. Wagner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{6}------------------------------------------------

#### Page 2 - Mr. Robert T. Wagner

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Mark A. Millman

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Ki12362

# INDICATIONS FOR USE STATEMENT

510(k) Number: K112362

Device Name: Quasar C100 Wrinkle Reduction Device

Indications for Use:

The Quasar C100 Wrinkle Reduction Device is intended to emit energy in the red and IR region of the spectrum, specifically indicated for the treatment of periorbital wrinkles.

## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

#### Concurrence of CDRH, Office of Device Evaluation (ODE)

OR

Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Per 21 CFR 801.109)

Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________ (Optional Format 1-2-96)

Page __ of __

(Division Sign-Off) (Division Sign-Off)
Division of Surgical, Orthopedic,
Division of Surgices Division of Surgh - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -

K112362

510(k) Number

Silver Bay LLC Quasar C100

Premarket Notification

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHS/K112362](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHS/K112362)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
