K193162 · AtriCure, Inc. · OCL · Dec 13, 2019 · General, Plastic Surgery
Device Facts
Record ID
K193162
Device Name
EPi-Sense Guided Coagulation System with VisiTrax
Applicant
AtriCure, Inc.
Product Code
OCL · General, Plastic Surgery
Decision Date
Dec 13, 2019
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The EPi-Sense Guided Coagulation System with VisiTrax is intended for the coagulation of cardiac tissue using Radiofrequency (RF) energy using thoracoscopic, and laparoscopic surgical techniques. The EPi-Sense Guided Coagulation System with VisiTrax may be used for temporary cardiac signal sensing and recording during surgery when connected to an external recording device.
Device Story
EPi-Sense Guided Coagulation System uses RF energy to coagulate cardiac tissue; features flexible, cooled electrode with suction stabilizer. System includes Subtle Cannula for probe/laparoscope access and Sensing Cable for connection to external electrogram recording systems. Used in OR by surgeons; device transmits RF energy from external generator to produce local tissue heating. Sensing electrodes allow temporary cardiac signal monitoring during surgery. Modifications include updated strain reliefs, yoke, and receptacle on sensing cable; shortened shaft, lengthened tip, and added NiTi wires on cannula. System provides surgical coagulation and diagnostic signal sensing; benefits include minimally invasive access and integrated visualization/sensing capabilities.
Clinical Evidence
Bench testing only. Mechanical design verification testing performed per Quality System to verify conformance to design controls and specifications. Testing confirmed modified devices met predetermined acceptance criteria and demonstrated equivalence to previously cleared devices.
Technological Characteristics
RF energy-based coagulation system. Components: flexible cooled electrode probe, suction stabilizer, Subtle Cannula (access tool), Sensing Cable. Materials include NiTi wires and stainless steel. Connectivity: connects to external RF generator and external electrogram recording systems. Sterilization: sterile (cannula/probe) and non-sterile (cables) components. Compliance: IEC 60601-2-2.
Indications for Use
Indicated for coagulation of cardiac tissue via thoracoscopic or laparoscopic surgical techniques and temporary cardiac signal sensing/recording during surgery.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
EPi-Sense Guided Coagulation System with VisiTrax (K142084)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the agency's name, "U.S. Food & Drug Administration," in blue text.
December 16, 2019
AtriCure, Inc. Caitlin Wunderlin Regulatory Affairs Specialist 7555 Innovation Wav Mason, Ohio 45040
Re: K193162
Trade/Device Name: EPi-Sense Guided Coagulation System with VisiTrax Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: OCL Dated: November 14, 2019 Received: November 15, 2019
Dear Caitlin Wunderlin:
This letter corrects our substantially equivalent letter sent on December 13, 2019.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
## Nicole M. Gillette -S
Nicole Gillette Acting Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K193162
Device Name
EPi-Sense Guided Coagulation System with VisiTrax
Indications for Use (Describe)
The EPi-Sense Guided Coagulation System with VisiTrax is intended for the coagulation of cardiac tissue using Radiofrequency (RF) energy using thoracoscopic, and laparoscopic surgical techniques.
The EPi-Sense Guided Coagulation System with VisiTrax may be used for temporary cardiac signal sensing and recording during surgery when connected to an external recording device.
Type of Use (Select one or both, as applicable)
| <span style="font-size: large;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------------------------------------------------------------------------------|
| <span style="font-size: large;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the AtriCure logo. The logo is in two colors, blue and orange. The word "Atri" is in blue, while the word "Cure" is in orange. There is a blue dot in the middle of the "C" in "Cure". There is a registered trademark symbol to the right of the word "Cure".
## 510(k) Summary
### l. Applicant Information
| Manufacturer: | AtriCure®, Inc.<br>7555 Innovation Way<br>Mason, Ohio 45040<br>P: 513-755-4100 |
|--------------------|--------------------------------------------------------------------------------|
| Contact Person: | Caitlin Wunderlin<br>Regulatory Affairs Specialist |
| Alternate Contact: | Jonathan McElwee<br>Sr. Manager, Regulatory Affairs |
| Date Prepared: | 14 November 2019 |
#### II. Device Information
| Proprietary Name: | EPi-Sense Guided Coagulation System with VisiTrax® |
|-------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Guided Coagulation System |
| Classification: | Surgical device for cutting, coagulation, and/or ablation of tissue,<br>including cardiac tissue<br>Regulatory Class: Class II; per 21 CFR 878.4400<br>Product Code: OCL<br>Classification Panel: Cardiovascular |
| Predicate Device: | EPi-Sense Guided Coagulation System with VisiTrax<br>(K142084, 28 October 2014) |
#### lll. Device Description
The EPi-Sense Guided Coagulation System flexible, cooled electrode device, with a suction stabilizer feature, transmits radiofrequency (RF) energy from an Electrosurgical RF Generator (nonsterile, re-useable) connected through an Instrument Cable (sterile). The device is designed to produce local tissue heating which results in coagulation of cardiac tissue. A temporary sensing electrode feature may be used with an off-the-shelf electrogram recording system with the use of an additional instrument cable (non-sterile).
The Subtle Cannula is a sterile, single-use access tool used to introduce the EPi-Sense probe into the chest cavity. The cannula is a 30 cm or 40 cm long device with a large central lumen to accommodate both the EPi-Sense probe and a commercially available laparoscope for visualization. The cannula is comprised of a distal tip, a shaft with a textured grip at the proximal end, a vacuum line and an integrated quidewire.
{4}------------------------------------------------
## Atril.
The Sensing Cable is a reusable, non-sterile cable that provides connection for the sensing electrodes of the EPi-Sense probe to a commercially available external electrogram recording system.
The Subtle Cannula and Sensing Cable are accessories to the EPi-Sense Guided Coagulation System.
### IV. Indications for Use
The EPi-Sense Guided Coaqulation System with VisiTrax is intended for the coagulation of cardiac tissue using Radiofrequency (RF) energy using thoracoscopic, and laparoscopic surgical techniques.
The EPi-Sense Guided Coaqulation System with VisiTrax may be used for temporary cardiac signal sensing and recording during surgery when connected to an external recording device.
### V. A. Device Changes From Predicate
The changes for the CSK-2030 Sensing Cable include:
- Overmolded strain reliefs
- o Overmolded yoke
- In-line receptacle ●
- In-line receptacle strain relief .
The changes for the Subtle Cannula include:
- Shortened shaft and internal spring ●
- Lengthened molded tip ●
- Added lumens and NiTi wires ●
- . Shortened stainless steel wires
## B. Comparison of Technological Characteristics
- . The devices include the same intended use, and;
- No changes were made in operating principle, or specifications of performance, and; .
- The biocompatibility remains unchanged, and;
- The results of the verification and validation testing: ●
- Demonstrated equivalency in mechanical performance and system compatibility o
- Compliance to IEC 60601-2-2 was met o
- Did not raise any new issues of safety or effectiveness ೧
### VI. Performance Data
Mechanical design verification testing was completed per AtriCure's Quality System to verify the device's conformance to appropriate design controls and specifications to demonstrate equivalence to the previously cleared Sensing Cable and Subtle Cannula devices. The modified devices met the new and existing predetermined acceptance criteria ensuring substantial equivalence to the previously cleared K142084 cannula device. No new safety or performance issues were raising during testing.
{5}------------------------------------------------
# AtriCure®
### VII. Conclusions
AtriCure demonstrated that the EPi-Sense Guided Coagulation System with modified Sensing Cable and Subtle Cannula is substantially equivalent in fundamental design, technology, function, device materials, packaging, sterilization, operating principle, and intended use/ indication for use as the previously cleared EPi-Sense Guided Coagulation System via K142084 on 28 October 2014.