K190151 · AtriCure, Inc. · OCL · Feb 25, 2019 · General, Plastic Surgery
Device Facts
Record ID
K190151
Device Name
COBRA Fusion Ablation System
Applicant
AtriCure, Inc.
Product Code
OCL · General, Plastic Surgery
Decision Date
Feb 25, 2019
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The COBRA Fusion® ablation system is intended to ablate cardiac tissue during cardiac surgerv using radiofrequency (RF) energy when connected directly to the Electrosurgical unit (ESU). The COBRA Fusion Pacing/Recording Adapter Cable may be used for temporary cardiac pacing, sensing, recording, and stimulation during the evaluation of cardiac arrhythmias during surgery when connected to a temporary external cardiac pacemaker or recording device.
Device Story
COBRA Fusion ablation system; single-use sterile ablation probe with integrated suction; magnetic retriever system; temporary external pacing cable; electrosurgical unit (ESU). Used during cardiac surgery by surgeons. Probe engages cardiac tissue via suction; RF energy delivered via ESU to ablate tissue. Pacing/recording adapter cable enables temporary pacing, sensing, and stimulation for arrhythmia evaluation when connected to external pacemaker/recording devices. System facilitates surgical treatment of cardiac arrhythmias.
Clinical Evidence
No clinical data. Bench testing only. Previous non-clinical testing (reliability, transit, shelf-life, packaging, sterilization, mechanical, biocompatibility, and electrical testing) conducted under K113475 remains applicable.
Technological Characteristics
Single-use sterile ablation probe with integrated suction; magnetic retriever system; temporary external pacing cable; electrosurgical unit (ESU). RF energy source. No changes to materials, biocompatibility, or sterilization methods from predicate.
Indications for Use
Indicated for patients undergoing cardiac surgery requiring cardiac tissue ablation or temporary cardiac pacing, sensing, recording, and stimulation for arrhythmia evaluation.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.
February 25, 2019
AtriCure, Inc. Melissa Smallwood Regulatory Affairs Specialist 7555 Innovation Way Mason, Ohio 45040
Re: K190151
Trade/Device Name: COBRA Fusion Ablation System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: OCL Dated: January 28, 2019 Received: January 30, 2019
Dear Melissa Smallwood:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Minde Jellman
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190151
Device Name COBRA Fusion® ablation system
Indications for Use (Describe)
The COBRA Fusion® ablation system is intended to ablate cardiac surgery using radiofrequency (RF) energy when connected directly to the Electrosurgical unit (ESU).
The COBRA Fusion Pacing/Recording Adapter Cable may be used for temporary cardiac pacing, recording, and stimulation during the evaluation of cardiac arrhythmias during surgery when connected to a temporary external cardiac pacemaker or recording device.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for AtriCure. The word "Atri" is in blue, and the word "Cure" is in orange. There is a registered trademark symbol to the right of the word "Cure". The logo is simple and modern.
# 510(k) Summary
### l. Applicant Information
| Manufacturer: | AtriCure®, Inc.<br>7555 Innovation Way<br>Mason, Ohio 45040<br>P: 513-644-4736<br>F: 513-895-9013 |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Melissa Smallwood<br>Regulatory Affairs Specialist |
| Date Prepared: | 01/28/2018 |
| II. Device Information | |
| Proprietary Name: | COBRA Fusion® ablation system |
| Common Name: | Cardiac Ablation System |
| Classification: | Electrosurgical Cutting and Coagulation Device and Accessories<br>Regulatory Class: Class II; per 21 CFR 878.4400<br>Product Code: OCL<br>Classification Panel: |
| Predicate Device: | COBRA Fusion ablation system |
|-------------------|------------------------------|
| | (K113475, March 20, 2012) |
### III. Device Description
The COBRA Fusion ablation system consists of a single use, sterile, ablation probe with integrated suction to engage the tissue during the ablation procedures; single use, sterile, magnetic retriever system to facilitate introduction and advancement of the probe, a temporary external pacing cable and an electrosurgical unit with accessories.
This CBE Special 510(k) contains labeling modifications that do not change the intended use, or the Indications for Use of the device as previously cleared. There are no changes to the labels affixed to the device or the packaging.
{4}------------------------------------------------
### IV. Intended Use/ Indications for Use
The COBRA Fusion® ablation system is intended to ablate cardiac tissue during cardiac surgerv using radiofrequency (RF) energy when connected directly to the Electrosurgical unit (ESU).
The COBRA Fusion Pacing/Recording Adapter Cable may be used for temporary cardiac pacing, sensing, recording, and stimulation during the evaluation of cardiac arrhythmias during surgery when connected to a temporary external cardiac pacemaker or recording device.
### V. Comparison of Technological Characteristics
- The devices have the same intended use, and indications for use;
- The fundamental scientific technology and operating principles remain the same; ●
- The device has the same design and configuration: .
- . There is no change to device materials or biocompatibility;
- . No changes were made to the packaging and sterilization;
- . No changes were made to performance specifications; and
- Verification and validation testing remain applicable to the device. .
Modifications included in the COBRA Fusion ablation system Special 510(k) were added for clarification to the existing device instructions for use regarding anticoagulation; the Indications for Use and Intended Use remain the same as previously cleared per K113475.
### VI. Performance Data
Appropriate testing has been performed to ensure that device meets its product specifications. Mechanical and reliability testing for the device remain applicable as there are no changes to the device specifications. There are no changes to the patient contact materials so biocompatibility testing as previously established remains applicable. Packaging, sterilization, pyrogen and transit testing performed remain representative of the device in packaged form. There are no changes to the ESU and prior validation and electrical testing remain valid.
## Non-clinical Bench Testing previously conducted per K113475 remains applicable as follows:
- Reliability Testing ●
- Transit ●
- Shelf-life ●
- . Packaging
- Sterilization
- Mechanical testing .
- Biocompatibility ●
- Electrical Testing (ESU) ●
### VII. Conclusions
The modified COBRA Fusion ablation system is substantially equivalent to the current legally marketed COBRA Fusion ablation system cleared per K113475.