INUMI™ Flex Needle

K233884 · Galvanize Therapeutics, Inc. · OAB · May 22, 2024 · General, Plastic Surgery

Device Facts

Record IDK233884
Device NameINUMI™ Flex Needle
ApplicantGalvanize Therapeutics, Inc.
Product CodeOAB · General, Plastic Surgery
Decision DateMay 22, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Intended Use

The INUMI™ Flex Needle is intended for use as part of the Aliya™ System for surgical ablation of soft tissue.

Device Story

INUMI™ Flex Needle is an endoscopic soft tissue ablation device used with the Aliya™ System (K212871). It features a 160 cm working length, compatible with endoscopes having ≥2.0 mm working channels. The device delivers pulsed electric field energy from the Aliya™ Generator to target tissue via a monopolar electrode. Operated by physicians in clinical settings, the device enables endoscopic access to soft tissue for ablation, contrasting with the predicate's percutaneous approach. The system transforms electrical energy into thermal effects to achieve tissue destruction. Clinical benefit includes minimally invasive surgical ablation capabilities. Healthcare providers use the device under endoscopic visualization to navigate to and ablate target tissues, with clinical decision-making guided by the intended therapeutic ablation outcome.

Clinical Evidence

No human clinical trials were conducted. Evidence includes bench testing (biocompatibility per ISO 10993-1; sterilization per ISO 11135:2014; electrical safety/EMC per IEC 60601-1/1-2; capacitive coupling per IEC 60601-2-18; mechanical/functional testing) and animal studies. Animal testing included a comparative ablation study in porcine liver, kidney, and muscle to characterize tissue destruction, and a chronic porcine GLP study to evaluate endoscopic navigation, access, and ablation performance.

Technological Characteristics

160 cm length, endoscopic-compatible (≥2.0 mm channel). Monopolar electrode. Energy source: Pulsed electric field (via Aliya™ Generator). Materials: Biocompatible per ISO 10993-1. Sterilization: Ethylene oxide (SAL 10^-6). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18, ASTM F1980-21, ISTA 3A 2018, ASTM F2096-11, ASTM F1929-15, ASTM F88/F88M-21.

Indications for Use

Indicated for surgical ablation of soft tissue in patients requiring such procedures, using the Aliya™ System.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of the symbol, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on the first line and "ADMINISTRATION" on the second line. May 22, 2024 Galvanize Therapeutics, Inc. Dongbo Wang Senior Director of Regulatory Affairs 3200 Bridge Parkway Redwood City, California 94065 ## Re: K233884 Trade/Device Name: INUMI™ Flex Needle Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: OAB Dated: May 20, 2024 Received: May 20, 2024 Dear Dongbo Wang: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). {1}------------------------------------------------ Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. # Long H. Chen =S Digitally signed by Long H.Chen -S Date: 2024.05.22 Date: 2024.05.22 08:59:39 -04'00' Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health {2}------------------------------------------------ # Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. Submission Number (if known) K233884 Device Name INUMI™ Flex Needle Indications for Use (Describe) The INUMI™ Flex Needle is intended for use as part of the Aliya™ System for surgical ablation of soft tissue. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## Date of 510(k) Summary Preparation: 11 April 2023 #### Submitter Information | Manufacturer: | Galvanize Therapeutics, Inc.<br>3200 Bridge Pkwy,<br>Redwood City, CA 94065<br>Phone: 650-268-4252<br>Fax: N/A | |---------------|----------------------------------------------------------------------------------------------------------------| | Contact: | Dongbo Wang<br>Senior Director of Regulatory Affairs<br>dwang@galvanizetx.com<br>Phone: 650-268-4252 | #### Subject Device | Device Trade Name: | INUMIT™ Flex Needle | |--------------------|----------------------------------------------------------------| | 510(k) Number: | K233884 | | Common Name: | Low energy soft tissue ablation device | | Regulation Number: | 21 CFR §878.4400 | | Regulation Name: | Electrosurgical cutting and coagulation device and accessories | | Product Code: | OAB low energy direct current thermal ablation system | | Review Panel: | General & Plastic Surgery (SU) | #### Primary Predicate Device | Device Trade Name: | Aliya™ System | |--------------------|------------------------| | 510(k) Number | K212871 | | Manufacturer: | Galvanize Therapeutics | #### Device Description The INUMI™ Flex Needle (subject device) is an endoscopically compatible soft tissue ablation device that is used as part of the 510(k)-cleared Aliya™ System (K212871). The subject device has a 160 cm working length, and is compatible with endoscope working channels with a minimum of 2.0 mm diameter. #### Proposed Indications for Use statement The INUMI™ Flex Needle is intended for use as part of the Aliya™ System for surgical ablation of soft tissue. {4}------------------------------------------------ ## Comparison to Primary Predicate Device | Characteristic | INUMI™ Flex Needle<br>Subject Device | Aliya™ System<br>Primary Predicate<br>Device | Comparison | |----------------------------------|------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------| | 510(k) number | TBD (K233884) | K212871 | N/A | | Classification<br>Regulation | 21 CFR §878.4400<br>Electrosurgical cutting<br>and coagulation device<br>and accessories. | 21 CFR §878.4400<br>Electrosurgical cutting<br>and coagulation device<br>and accessories. | Identical | | Product Code | OAB low energy direct<br>current thermal<br>ablation system | OAB low energy direct<br>current thermal ablation<br>system | Identical | | Indications for Use<br>statement | The INUMI™ Flex<br>Needle is intended for<br>use as part of the<br>Aliya™ System for<br>surgical ablation of soft<br>tissue. | The Aliya™ System is<br>indicated for the surgical<br>ablation of soft tissue. | Identical, for surgical ablation<br>of soft tissue, subject device<br>specifies compatibility with the<br>Aliya System | | Intended use | Ablation of soft tissue | Ablation of soft tissue | Identical | | Type of Use | Prescription | Prescription | Identical | | Energy Source | Galvanize Aliya™<br>Generator (K212871) –<br>Pulsed Electric Field | Galvanize Aliya™<br>Generator (K212871) –<br>Pulsed Electric Field | Identical | | Energy Polarity | Monopolar | Monopolar | Identical | | Electrode | Aliya™ Electrode<br>(K212871) | Aliya™ Electrode<br>(K212871) | Identical | | Access | Endoscopic | Percutaneous | Different | #### Comparison of Indications for Use statement with the Primary Predicate Device The Indications for Use statements are identical in that both devices (predicate and subject devices) are indicated for the surgical ablation of soft tissue. The Indication for Use statement for the subject device adds the specificity for the compatibility with the Aliya System. The additional language defining compatibility does not alter or expand the intended use, it is a technical limitation on compatible devices. #### Comparison of Technological Characteristics with the Primary Predicate Device The subject device and the primary predicate device utilize the same generator and electrode to achieve soft tissue ablation. The main difference in technological characteristics is the form of access - endoscopic for the subject device and percutaneous for the predicate device. #### Performance Testing (bench) Performance testing has been completed to demonstrate (in part) substantial equivalence of the subject device to the predicate device. The testing performed is summarized, below: {5}------------------------------------------------ ## Biocompatibility Biocompatibility testing was performed for patient-contacting components of the INUMI™ Flex Needle in accordance with ISO 10993-1 requirements, as well as the FDA Guidance Document Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a Risk Management Process," September 2023. # Sterilization The INUMI™ Flex Needle was validated per ISO 11135:2014 to achieve a sterility assurance level of 10-6 using ethylene oxide sterilization. # Shelf-Life/Package Validation Shelf-Life and Package Validation were performed to ensure that the devices maintain package intecrity and sterility. Applicable standards were used: ASTM F1980-21, ISTA 3A 2018, ASTM F2096-11, ASTM F1929-15, ASTM F88/F88M-21. # Electrical Safety and Electromagnetic Compatibility (EMC) Electrical safety and EMC testing was completed for the applicable components of the INUMI™ Flex Needle. The results demonstrated compliance to all applicable standards, including IEC 60601-1 and IEC 60601-1-2. # Capacitive Couple Testing Capacitive coupling testing was done per FDA Guidance on Premarket Notifications (510(k)) Submissions for Electrosurgical Devices for General Surgery according to the applicable clauses of IEC 60601-2-18. # Mechanical and Electrical Performance Testing Mechanical and Functional testing was completed to confirm that the performance of the INUMI™ Flex Needle conforms to the predefined mechanical and electrical specifications after sterilization and accelerated aging. # Ex Vivo Characterization of Thermal Effects on Tissue The temperature-time history was measured consistent with the recommendations of the FDA Guidance on Premarket Notification (510(k)) Submission for Electrosurgical Devices for General Surgery in three types of ex-vivo porcine tissue - liver, kidney, and muscle. Comparative testing was performed with the subject device and the primary predicate device. # Performance Testing (animal) # Comparative Ablation Performance Porcine model was utilized to characterize tissue destruction with subject device and the primary predicate device in the liver, kidney, and longissimus dorsi skeletal muscle in accordance with the FDA quidance on Premarket Notification [510(k)] for Electrosurgical Devices for General Surgery. {6}------------------------------------------------ ## Endoscopic Performance A chronic porcine GLP study was conducted to evaluate the performance of the endoscopic use of subject device regarding navigation, access, and ablation of target tissues. ## Conclusion The subject INUMI™ Flex Needle and the legally marketed predicate Aliya™ System (K212871) share the same Intended Use and consistent Indications for Use. Both performance testing (bench) and performance testing (animal) support a determination of substantial equivalence to the legally marketed predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...