← Product Code [NUJ](/submissions/SU/subpart-e%E2%80%94surgical-devices/NUJ) · K191073

# Vein360 Endovenous Radiofrequency Ablation (RFA) Catheter (K191073)

_Vein 360, LLC · NUJ · Oct 22, 2019 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/NUJ/K191073

## Device Facts

- **Applicant:** Vein 360, LLC
- **Product Code:** [NUJ](/submissions/SU/subpart-e%E2%80%94surgical-devices/NUJ.md)
- **Decision Date:** Oct 22, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Vein360 Endovenous RFA Catheter is intended for endovascular coagulation of blood vessels in patients with superficial vein reflux.

## Device Story

Reprocessed single-use endovenous radiofrequency ablation (RFA) catheter; consists of molded handle with actuation switch, integrated cable, and flexible shaft with distal RF heating element. Device delivers thermal energy to treatment site via RF heating; relays temperature data to external RF generator. Used by clinicians in clinical settings for superficial vein reflux treatment. Reprocessing involves cleaning, inspection, packaging, and sterilization of OEM devices. Output is thermal coagulation of target vein; facilitates vein closure. Benefits include minimally invasive treatment of venous reflux.

## Clinical Evidence

No clinical data. Evidence consists of bench testing and ex vivo studies. Bench testing included cleaning validation (AAMI TIR30), sterilization (ISO 11135), biocompatibility (ISO 10993), pyrogenicity (ST72), electrical safety (IEC 60601), and mechanical integrity. Ex vivo study compared ablative performance and thermal injury zones across three tissue types (muscle, liver, kidney) between reprocessed and OEM devices, demonstrating equivalent performance.

## Technological Characteristics

Reprocessed catheter with RF heating element. Materials validated for biocompatibility per ANSI AAMI ISO 10993. Sterilization via EN ISO 11135:2014. Electrical safety per IEC 60601. Connectivity via integrated cable to external RF generator. No software algorithm; device functions as a passive delivery tool for RF energy.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- VNUS ClosureFAST Catheter ([K061373](/device/K061373.md))
- Covidien ClosureFAST Radiofrequency Catheter ([K111887](/device/K111887.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The seal features a stylized depiction of a human figure, while the text is in a clear, sans-serif font, with "FDA" in a blue square.

October 22, 2019

Vein360, LLC Suzanne Meyer CEO 4460 Lake Forest Drive, Suite 230 Blue Ash, Ohio 45242-3741

Re: K191073

Trade/Device Name: Vein360 Endovenous Radiofrequency Ablation (RFA) Catheter Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories, Reprocessed Regulatory Class: Class II Product Code: NUJ Dated: April 17, 2019 Received: April 22, 2019

Dear Suzanne Meyer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

Reprocessed Single-Use Device Models Subject to Clearance:

{2}------------------------------------------------

| OEM<br>Model<br>Number | Device Name/Description                                                                          | Original<br>Manufacturer           | Reprocessed<br>Model<br>Number |
|------------------------|--------------------------------------------------------------------------------------------------|------------------------------------|--------------------------------|
| CF7-3-60               | CovidienTM ClosureFASTTM Radiofrequency Catheter<br>(3cm heating length, 60cm insertable length) | Covidien                           | VEN-3-60                       |
| CF7-7-60               | VNUS® ClosureFASTTM Catheter (7cm heating length,<br>60cm insertable length)                     | VNUS Medical<br>Technologies, Inc. | VEN-7-60                       |
| CF7-7-100              | VNUS® ClosureFASTTM Catheter (7cm heating length,<br>100cm insertable length)                    | VNUS Medical<br>Technologies, Inc. | VEN-7-100                      |

{3}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known)

K191073

Device Name

Vein360 Endovenous Radiofrequency Ablation (RFA) Catheter

Indications for Use (Describe)

The Vein360 Endovenous Radiofrequency Ablation (RFA) Catheter is intended for endovascular coagulation of blood vessels in patients with superficial vein reflux.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# SECTION 5: 510(k) SUMMARY

Image /page/4/Picture/3 description: The image shows the logo for Vein 360. The word "VEIN" is in large, bold, black letters at the top of the logo. Below the word "VEIN" is the number "360" in a smaller, but still bold, black font. A blue arrow that transitions to green is below the number 360. The arrow is curved in a semi-circle.

April 17, 2019 (revised October 22, 2019) Date of Preparation:

#### Company Name / Contact:

- Company: Vein360. LLC 4460 Lake Forest Drive Suite 230 Blue Ash, OH 45242
Contact:

Suzanne Meyer CEO Phone: (513) 554-1300

### Device Identification:

| Proprietary Name:          | Vein360 Endovenous Radiofrequency<br>Ablation (RFA) Catheter                      |
|----------------------------|-----------------------------------------------------------------------------------|
| Common Name:               | Electrosurgical Device                                                            |
| Classification Name:       | Electrosurgical Cutting and<br>Coagulation Device and Accessories,<br>Reprocessed |
| Classification Reference:  | 21 CFR § 878.4400                                                                 |
| Classification Panel:      | 79 - General & Plastic Surgery                                                    |
| Device Product Code:       | NUJ                                                                               |
| Proposed Regulatory Class: | Class II                                                                          |

### Device Description:

The Vein360 Endovenous RFA Catheter is a reprocessed single-use device (SUD) consisting of a molded handle with actuation switch, an integrated instrument cable, and a flexible catheter shaft with a radiofrequency (RF) heating element at the distal end. The catheter and integrated connection cable are provided sterile and meant for single patient use. The catheter's function is to provide thermal energy to the desired treatment site via RF heating of the heating element and to relay temperature back to the RF generator.

The Vein360 Endovenous RFA Catheter is subjected to reprocessing

{5}------------------------------------------------

operations following the initial clinical use of the OEM predicate. These operations include cleaning, inspection, packaging and sterilization. After reprocessing, the Vein360 reprocessed Endovenous RFA Catheter retains substantially equivalent performance to that of the OEM predicate. The Vein360 reprocessed Endovenous RFA Catheter is reprocessed one (1) time.

### Indications for Use:

The Vein360 Endovenous RFA Catheter is intended for endovascular coagulation of blood vessels in patients with superficial vein reflux.

## Substantial Equivalence Information:

The Vein360 Endovenous RFA Catheter is a reprocessed SUD that is substantially equivalent to the new, unused device of the same product currently marketed by the device's original equipment manufacturer (OEM) and described herein with respect to indications for use, device design, materials, performance and method of sterilization. As a reprocessed SUD, there are no changes to the clinical applications, patient population, or method of operation. Within the proposed class, the following OEM devices are used as a predicate device for comparison: VNUS® ClosureFAST™ Catheters (K061373) Models CF7-7-60 and CF7-7-100, and Covidien™ ClosureFAST™ Radiofrequency Catheter (K111887) Model CF7-3-60. The following table summarizes the Vein360 reprocessed Endovenous RFA Catheters vs. the predicate OEM devices:

| OEM Model<br>Number | OEM Name                           | Predicate 510(k) | Device Name/Description                                                                              | Vein360<br>Model<br>Number |
|---------------------|------------------------------------|------------------|------------------------------------------------------------------------------------------------------|----------------------------|
| CF7-3-60            | Covidien                           | K111887          | Covidien™ ClosureFAST™<br>Radiofrequency Catheter<br>(3cm heating length, 60cm<br>insertable length) | VEN-3-60                   |
| CF7-7-60            | VNUS Medical<br>Technologies, Inc. | K061373          | VNUS® ClosureFAST™<br>Catheter (7cm heating length,<br>60cm insertable length)                       | VEN-7-60                   |
| CF7-7-100           | VNUS Medical<br>Technologies, Inc. | K061373          | VNUS® ClosureFAST™<br>Catheter (7cm heating length,<br>100cm insertable length)                      | VEN-7-100                  |

Table 1. OEM vs. Vein360 Models

### Performance Data:

Results of performance testing demonstrate the Vein360 reprocessed Endovenous RFA Catheters are substantially equivalent to the OEM predicate devices and effective for their intended function. Substantial equivalence determination was concluded through successful completion of a battery of testing, which included:

- Cleaning validation per AAMI TIR30:2011/(R)2016 0

{6}------------------------------------------------

- Sterilization validation per EN ISO 11135:2014 ●
- Biocompatibility per ANSI AAMI ISO 10993 o
- Pyrogenicity per ANSI AAMI ST72:2011/(R)2016 ●
- Physical and mechanical integrity testing ●
- Electrical safety testing per IEC 60601 ●
- o Comparative ex vivo tissue testinq in three (3) tissue types

With respect to SUD reprocessing, comprehensive cleaning validation studies included enumeration of clinical soil levels to establish a worst-case basis for artificial test soil development. Using both clinical and artificial test soils, the cleaning operation was validated with a high degree of confidence by objectively demonstrating removal of all physical soil under minimum operating conditions. The body of this submission includes all data related to the cleaning process and validation.

Electro-mechanical performance testing was performed to demonstrate that the reprocessing operations did not adversely affect the predicate device's form, fit or function. The following list summarizes the extensive scope of performance tests executed to fully demonstrate substantial equivalence of the Vein360 reprocessed Endovenous RFA catheters to their OEM predicate, all of which are detailed within the body of this submission:

- Guide wire testing ●
- Physical dimensions ●
- Impact resistance ●
- Intearity testing ●
- Thermal accuracy ●
- Continuity and resistance
- Connector testing
- Leakage current ●
- Radiated emissions ●
- RF immunity ●

Finally, the Vein360 reprocessed Endovenous RFA Catheters were compared to the predicate device in a comprehensive ex vivo study which was performed by an independent laboratory. The study measured ablative performance across three (3) tissue types (muscle, liver and kidney). The Vein360 reprocessed Endovenous RFA Catheters performed in a substantially equivalent manner to their OEM predicates, indicating that the reprocessing of the catheters does not impact ablation performance and that the thermal injury zones between the Vein360 reprocessed Endovenous RFA catheter and the OEM predicate are equivalent.

The totality of data collected through comprehensive performance testing has successfully demonstrated that the Vein360 reprocessed Endovenous RFA Catheters are equivalent to their OEM predicates.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/NUJ/K191073](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/NUJ/K191073)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
