K052690 · Suretek Medical · NUJ · May 12, 2006 · General, Plastic Surgery
Device Facts
Record ID
K052690
Device Name
SURETEK REPROCESSED LAPAROSCOPIC INSTRUMENTS
Applicant
Suretek Medical
Product Code
NUJ · General, Plastic Surgery
Decision Date
May 12, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Suretek Reprocessed Laparoscopic Instruments are intended for use during general and laparoscopic surgery for cutting, grasping, dissection and electrocautery of tissue.
Device Story
Suretek Reprocessed Laparoscopic Instruments consist of various surgical tools, including dissectors, scissors, shears, and forceps, designed for use in general and laparoscopic surgical procedures. These devices are reprocessed single-use instruments originally manufactured by companies such as AutoSuture, ConMed, Ethicon, and Gyrus. The instruments function by providing mechanical cutting, grasping, and dissection capabilities, as well as delivering monopolar or bipolar electrosurgical energy for tissue coagulation. They are operated by surgeons in an operating room environment. The devices are intended to perform the same clinical functions as their original, non-reprocessed counterparts, facilitating surgical access and tissue management during minimally invasive procedures.
Clinical Evidence
No clinical data provided. Substantial equivalence was determined based on supplemental validation data submitted for reprocessed single-use devices.
Technological Characteristics
Devices include monopolar and bipolar electrosurgical instruments (dissectors, scissors, shears, forceps) in 5mm and 10mm diameters and various lengths (15cm to 33cm). Energy source is external electrosurgical generator. Construction materials are consistent with original predicate devices. Reprocessing involves cleaning, inspection, and sterilization protocols to restore functionality for single-use application.
Indications for Use
Indicated for use in general and laparoscopic surgery for cutting, grasping, dissection, and electrocautery of tissue in patients requiring such surgical interventions.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
AUTOSUTURE ENDO DISSECT 5mm
AUTOSUTURE ENDO SCIZ 5mm
AUTOSUTURE ENDO SHEARS 5mm
CONMED Curved Metzenbaum 5mm
CONMED Curved Metzenbaum Narrow Tip 5mm
CONMED Curved Metzenbaum Mini 5mm
CONMED Maryland Dissector 5mm
ETHICON Dissector 5mm
ETHICON Scissors 5mm
GYRUS Bipolar Cutting Forceps 5mm
GYRUS Bipolar Cutting Forceps 10mm
GYRUS Bipolar Needle Electrode 5mm
Related Devices
K012603 — REPROCESSED AUTOSUTURE UNIPOLAR LAPAROSCOPIC/ENDOSCOPIC INSTRUMENTS · Alliance Medical Corp. · Nov 7, 2001
K043013 — MILTEX LAPAROSCOPIC INSTRUMENTS · Miltex, Inc. · Mar 4, 2005
K073613 — REPROCESSED SCISSOR TIPS · Sterilmed, Inc. · Feb 12, 2008
K012700 — VANGUARD REPROCESSED ENDOSCOPIC INSTRUMENTS · Vanguard Medical Concepts, Inc. · Nov 7, 2001
K031869 — SISS INC. (D.B.A. MEDISISS) REPROCESSED ELECTROSURGICAL INSTRUMENTS AND ACCESSORIES · Surgical Instruments Service and Savings, Inc. · Dec 24, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three stripes representing the department's mission. The eagle is encircled by the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" in a circular arrangement. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## OCT 1 6 2006
SureTek Medical % Mike Sammon, BME, Ph.D. CEO/President 44 Bellwood Farms Greenville, South Carolina 29607
Re: K052690 - Supplemental Validation Submission Trade/Device Name: See Enclosed List Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: NUJ Dated: January 3, 2006 Received: January 6, 2006
Dear Dr. Sammon:
The above-referenced premarket notification (510(k)) was cleared by the Office of Device Evaluation (ODE) on May 12, 2006. We have received your supplemental validation data as required for reprocessed single-use devices by the Medical Device User Fee and Modernization Act of 2002. After reviewing your supplemental validation data, we have determined the devices listed in the enclosure accompanying this letter are substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market these devices, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your devices are classified (see above) into either class II (Special Controls) or class III (PMA) they may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your devices in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your devices comply with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's applicable requirements, including, but not limited to: registration and listing (2) CFR
{1}------------------------------------------------
Page 2 - Mike Sammon, BME, Ph.D.
Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
The FDA finding of substantial equivalence of your devices to legally marketed predicate devices results in classification for your devices and thus, permits you to legally market the devices. This letter will allow you to continue marketing the devices listed in the enclosure accompanying this letter.
If you desire specific advice for your devices on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
[signature]
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Page 3 - Mike Sammon, BME, Ph.D.
## Indications for Use
510 (k) Number (if known) : K052690
Suretek Reprocessed Laparoscopic Instruments Device Name:
Indications For Use:
Suretek Reprocessed Laparoscopic Instruments are intended for use during general and laparoscopic surgery for cutting, grasping, dissection and electrocautery of tissue.
Prescription Use AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
**(Division Sign-Off)**
Division of General, Restorative,
and Neurological Devices and Neurological Devices
Page 1 of
5100 Number
{3}------------------------------------------------
Page 4 – Mike Sammon, BME, Ph.D.
.
1. And 1998
:
Reprocessed Laparoscopic Instruments found to be substantially equivalent:
| AUTOSUTURE | ENDO DISSECT 5mm | Monopolar, 33cm Length |
|------------|------------------------------------|------------------------|
| AUTOSUTURE | ENDO SCIZ 5mm | Monopolar, 33cm Length |
| AUTOSUTURE | ENDO SHEARS 5mm | Monopolar, 33cm Length |
| CONMED | Curved Metzenbaum, 5mm | Monopolar, 32cm Length |
| CONMED | Curved Metzenbaum, Narrow Tip, 5mm | Monopolar, 32cm Length |
| CONMED | Curved Metzenbaum, Mini, 5mm | Monopolar, 32cm Length |
| CONMED | Maryland Dissector, 5mm | Monopolar, 32cm Length |
| ETHICON | Dissector, 5mm | Monopolar, 33cm Length |
| ETHICON | Scissors, 5mm | Monopolar, 33cm Length |
| GYRUS | Bipolar Cutting Forceps, 5mm | 24cm Length |
| GYRUS | Bipolar Cutting Forceps, 5mm | 33cm Length |
| GYRUS | Bipolar Cutting Forceps, 10mm | 15cm Length |
| GYRUS | Bipolar Cutting Forceps, 10mm | 33cm Length |
| GYRUS | Bipolar Needle Electrode, 5mm | 33cm Length |
:
1000 - 1000
. .
.
:
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.