K231579 · Molli Surgical, Inc. · NEU · Oct 26, 2023 · General, Plastic Surgery
Device Facts
Record ID
K231579
Device Name
MOLLI 2
Applicant
Molli Surgical, Inc.
Product Code
NEU · General, Plastic Surgery
Decision Date
Oct 26, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4300
Device Class
Class 2
Indications for Use
The MOLLI Marker is intended to be placed percutaneously in soft tissue to temporarily mark a surgical site intended for surgical removal. The MOLLI Marker can only be implanted for less than 30 days. Using imaging guidance (such as ultrasound or radiography) or aided by non-imaging guidance (MOLLI 2 System), the MOLLI Marker is located and surgically removed with the target tissue. The MOLLI 2 System is intended only for the non-imaging detection of the MOLLI Marker that has been implanted in a site intended for surgical removal.
Device Story
MOLLI 2 is a surgical guidance system for locating non-palpable lesions. System components include a temporary implantable marker, a marker delivery introducer, a detection wand, and a visualization tablet. The marker is placed percutaneously in soft tissue using imaging guidance. During surgery, the surgeon uses the detection wand to locate the marker; the system provides non-imaging detection and localization, with output displayed on the tablet. This assists the surgeon in identifying the target tissue for excision. The device is used in a clinical/surgical setting by physicians. The system facilitates precise surgical removal of lesions, potentially improving surgical outcomes and efficiency.
Clinical Evidence
Bench testing only. Biocompatibility testing performed for the modified MOLLI Introducer. Effectiveness demonstrated via bench testing of the modified introducer. Justification provided for updating the cardiac device contraindication to a warning.
Technological Characteristics
System consists of an implantable marker, delivery introducer, detection wand, and visualization tablet. Introducer needle cannula is siliconized. Operates via non-imaging detection of the marker. Designed for temporary implantation (<30 days).
Indications for Use
Indicated for patients requiring temporary percutaneous marking of soft tissue surgical sites for removal. Marker implantation duration is limited to <30 days. Contraindicated/Warning: Use caution with patients having magnetically susceptible implanted devices (e.g., pacemakers, ICDs, cochlear implants, nerve stimulators, infusion pumps); maintain minimum 40 mm distance.
Regulatory Classification
Identification
An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.
K253888 — MOLLI 2 System · Stryker Endoscopy · Dec 31, 2025
K234149 — MOLLI 2 System · Molli Surgical, Inc. · Sep 26, 2024
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" stacked on top of each other.
October 26, 2023
MOLLI Surgical, Inc. Joseph De Croos Director. Ouality Assurance & Regulatory Affairs 50 Wellington Street East Suite 400 Toronto, ON M5E 1C8 Canada
Re: K231579
Trade/Device Name: Molli 2 Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable Clip Regulatory Class: Class II Product Code: NEU Dated: September 29, 2023 Received: September 29, 2023
Dear Joseph De Croos:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).
{2}------------------------------------------------
Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Digitally signed by Tek Tek N. N. Lamichhane -S Lamichhane -S Pate: 2023.10.26 16:08:14 -04'00' Tek N. Lamichhane, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231579
Device Name MOLLI 2
Indications for Use (Describe)
The MOLLI Marker is intended to be placed percutaneously in soft tissue to temporarily mark a surgical site intended for surgical removal. The MOLLI Marker can only be implanted for less than 30 days. Using imaging guidance (such as ultrasound or radiography) or aided by non-imaging guidance (MOLLI 2 System), the MOLLI Marker is located and surgically removed with the target tissue.
The MOLLI 2 System is intended only for the non-imaging detection of the MOLLI Marker that has been implanted in a site intended for surgical removal.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: Arial, sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Molli. The word "molli" is written in a dark blue font, with a light blue dot above the second "i". Below the word "molli" is the phrase "precision surgery made simpler" in a smaller font, with the word "simpler" in light blue and the rest of the phrase in dark blue.
## K231579 - 510(k) Summary
#### DATE PREPARED
October 25, 2023
## MANUFACTURER AND 510(k) OWNER
MOLLI Surgical, Inc. 50 Wellington Street East Suite 400 Toronto ON M5E 1C8 Canada Telephone: +1-833-665-5463 Dr. Joseph De Croos, Director of Quality Assurance and Regulatory Affairs Official Contact: Email: jdecroos@mollisurgical.com
## DEVICE INFORMATION
| Proprietary Name/Trade Name: | MOLLI 2 |
|------------------------------|--------------------------------------------------------------------|
| Common Name: | Implantable radiographic marker |
| Regulation Number: | 21 CFR 878.4300 |
| Class: | II |
| Product Code: | NEU |
| Premarket Review: | OPEQ/OHT4/Infection Control and Plastic Surgery Devices<br>(DHT4B) |
| Review Panel: | General & Plastic Surgery |
#### PREDICATE DEVICE IDENTIFICATION
MOLLI 2 is substantially equivalent to the following predicate:
| 510(k) Number | Predicate Device Name / Manufacturer | Primary Predicate |
|---------------|--------------------------------------|-------------------|
| K223107 | MOLLI 2 / MOLLI Surgical, Inc. | ✓ |
The predicate device has not been subject to a design related recall.
## DEVICE DESCRIPTION
MOLLI 2 is a precision surgical marking and guidance system for locating non-palpable lesions during surgery. MOLLI 2 consists of a temporary marker (MOLLI Marker), a marker delivery system (MOLLI Introducer), a detection wand (MOLLI Wand 2), and a visualization tablet (MOLLI Tablet 2). The MOLLI 2 Wand and MOLLI 2 Tablet constitute the MOLLI 2 system. The MOLLI 2 is intended for the non-imaging detection and localization of the MOLLI Marker that has been implanted in a site intended for surgical removal.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Molli, a company that specializes in precision surgery. The logo consists of the word "molli" in a dark blue sans-serif font, with a small light blue circle above the "i". Below the word "molli" is the tagline "precision surgery made simpler" in a smaller font, with the word "simpler" in light blue and the rest of the tagline in dark blue.
The purpose of this 510(k) premarket notification is to introduce the following changes to the previously cleared MOLLI device:
- . Modification to the MOLLI Introducer
- Update to contraindications and warning in the MOLLI Marker & Introducer user guide
## INDICATIONS FOR USE
The MOLLI Marker is intended to be placed percutaneously in soft tissue to temporarily mark a surgical site intended for surgical removal. The MOLLI Marker can only be implanted for less than 30 days. Using imaging guidance (such as ultrasound or radiography) or aided by nonimaging guidance (MOLLI 2 System), the MOLLI Marker is located and surgically removed with the target tissue.
The MOLLI 2 System is intended only for the non-imaging detection and localization of the MOLLI Marker that has been implanted in a site intended for surgical removal.
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
MOLLI Surgical, Inc. believes that the subject device is substantially equivalent to the predicate device based on the information summarized here:
The subject device has the same intended use, design, and technological characteristics as the predicate device cleared in K223107. Minor modifications to the subject device as compared to the predicate device include:
- MOLLI Introducer needle cannula is siliconized
- . Update to contraindication and warning in the MOLLI Marker & Introducer user guide
- o Current Contraindication: "The MOLLI Marker must not be placed in the same breast as an implanted cardiac device. If placing the MOLLI Marker in the contralateral breast, it must be a minimum of 20 mm away from an implanted cardiac device".
- Change cardiac device contraindication to a warning: "Caution should be o exercised when using the MOLLI Marker on patients with magnetically susceptible implanted devices, including active implantable and wearable medical devices, and devices using a permanent maqnet (e.q., pacemakers, implantable defibrillators, cochlear implants, nerve stimulators, deep brain stimulators, infusion pumps, or any implantable or body worn permanent magnet). The MOLLI Marker must be implanted at a minimum of 40 mm away from magnetically susceptible implanted devices."
790 Estate Drive · 2nd Floor Deerfield · Illinois · 60015 United States
50 Wellington Street E. · Suite 400 Toronto · Ontario · M5E 1C8 Canada
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Molli, a company that specializes in precision surgery. The word "molli" is written in a bold, sans-serif font, with a small blue circle above the "i". Below the word "molli" is the tagline "precision surgery made simpler®", with the word "simpler" in a lighter shade of blue. The logo is clean and modern, and the tagline suggests that Molli's products and services are designed to make surgery more efficient and effective.
## SUMMARY OF NON-CLINICAL TESTING
Due to the changes in the processing of the patient-contacting components of the MOLLI 2 (i.e., the MOLLI Introducer) and the updates to the MOLLI Marker & Introducer user guide, additional non-clinical testing was performed.
The following reports are included in this submission demonstrate safety and effectiveness based on current industry standards:
- . Biocompatibility testing to demonstrate safety of MOLLI Introducer
- . Bench testing to demonstrate effectiveness of MOLLI Introducer
- . Justification for updating cardiac device contraindication to a warning
The results of these reports indicate that the subject device is substantially equivalent to the predicate device.
## CONCLUSION
Based on the testing performed, it can be concluded that the subject device does not raise new issues of safety or effectiveness compared to the predicate device. The same indications for use, technological characteristics, and performance characteristics for the proposed MOLLI 2 are assessed to be substantially equivalent to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.