K180640 · Elucent Medical, Inc. · NEU · Jun 4, 2018 · General, Plastic Surgery
Device Facts
Record ID
K180640
Device Name
SmartClip Soft Tissue Marker
Applicant
Elucent Medical, Inc.
Product Code
NEU · General, Plastic Surgery
Decision Date
Jun 4, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4300
Device Class
Class 2
Intended Use
The implanted SmartClip™ is indicated for radiographic marking of sites in soft tissue. In addition, the Marker is indicated in situations where the soft tissue site needs to be marked for future medical procedures.
Device Story
SmartClip is a sterile, single-use, permanently implantable soft tissue marker; preloaded in a 15ga stainless steel introducer needle delivery system. Marker consists of a Parylene C coated glass vial; retained in needle by biocompatible bone wax plug. Physician deploys marker into soft tissue (e.g., lung, GI, breast) via stylet-lock mechanism. Marker provides permanent radiographic, mammographic, ultrasound, and MRI visibility. Device facilitates localization of tissue sites for subsequent medical procedures. No software or algorithm involved.
Clinical Evidence
Bench testing only. Testing included simulated use, mechanical integrity, imaging assessment (X-ray, mammography, ultrasound, MRI), MR compatibility, biocompatibility, packaging, and sterilization validation. No clinical data presented.
Technological Characteristics
Permanent soft tissue marker; Parylene C coated glass vial. Delivery system: 15ga stainless steel needle (10.8 cm) with depth reference marks. Sterilization: Ethylene oxide (SAL 10^-6). Imaging compatibility: X-ray, mammography, ultrasound, MRI. Biocompatible materials. No software or electronic components.
Indications for Use
Indicated for radiographic marking of soft tissue sites for future medical procedures in patients requiring tissue localization.
Regulatory Classification
Identification
An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 4, 2018
Elucent Medical, Inc. Ms. Debra Kridner VP Regulatory and Quality 7480 Flying Cloud Drive, Suite 101 Eden Prairie, Minnesota 55344
Re: K180640
Trade/Device Name: SmartClip Soft Tissue Marker Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable Clip Regulatory Class: Class II Product Code: NEU Dated: May 21, 2018 Received: May 22, 2018
Dear Ms. Kridner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may; therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# David Krause -S
for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180640
Device Name
SmartClip™ Soft Tissue Marker
Indications for Use (Describe)
The implanted SmartClip™ is indicated for radiographic marking of sites in soft tissue. In addition, the Marker is indicated in situations where the soft tissue site needs to be marked for future medical procedures.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------|--------------------------------------------------------------|
| <div> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary
Following is a summary of 510(k) safety and effectiveness information in accordance with 21 CFR 807.92
| Date Prepared: | May 18, 2018 |
|-----------------------------|---------------------------------------------------------------------------------------------|
| Applicant: | Elucent Medical, Inc.<br>7480 Flying Cloud Drive, Suite 110<br>Eden Prairie, MN 55344 USA |
| Official Correspondent: | Debra Kridner<br>Tel: (844) 417 1700 / Fax: (952) 314-7105<br>debra.kridner@elucent.com |
| Trade Name: | SmartClip™ Soft Tissue Marker |
| Common Name: | Implanted, Radiographic, Tissue Marker |
| Classification Number: | 21 CFR 878.4300 |
| Classification Name: | Implantable Clip |
| Device Class: | Class II |
| Product Code: | NEU |
| Review Panel: | General and Plastic Surgery |
| Predicate Device (primary): | Focal Therapeutics BioZorb Marker - K143484/K152070 |
| Reference Device: | Cianna Medical Permanent Tissue Marker - K132463 |
## Indications for Use for the Device Subject to this 510(k) Notification:
The SmartClip™ is indicated for radiographic marking of sites in addition, the Marker is indicated in situations where the soft tissue site needs to be marked for future medical procedures.
## Device Description for the Device Subject to this 510(k) Notification:
The SmartClip™ is an ethylene oxide sterile, single use device composed of a soft tissue marker preloaded in a Delivery System. The marker is intended to be placed within soft tissue. The marker is visible using radiography (including mammographic imaging), ultrasound and MRI.
Image /page/3/Figure/8 description: This image shows a medical device with labels pointing to its different parts. The introducer needle length is labeled as 10.8 cm. Other labeled parts include the unlock button, deployment plunger, stylet, bone wax (3-4 mm), and SmartClip (8 mm x 1.4 mm).
Image /page/3/Figure/9 description: The image shows the word "SmartClip" with the trademark symbol next to it. The text is in a serif font and is black. The background is white.
#### Marker:
{4}------------------------------------------------
#### K180640 pg. 2 of 3
A sterile, single use device permanently implanted marker is approximately 8 mm wide. The marker is placed into the barrel of the introducer and maintained in place prior to insertion by a biocompatible bone wax plug. The marker can be implanted into various types of soft tissue (e.g., lung, gastrointestinal system, and breast) and subsequently be detected by means of radiography (including mammographic imaging), ultrasound and MRI.
## Delivery System:
The Delivery System consists of a stylet, a stylet-lock to prevent and 15ga introducer needle. The stainless steel needle is approximately 10.8 cm long with 1cm depth reference marks. The marker is preloaded inside the needle and retained by a bone wax plug. When the stylet is completely depressed the marker and bone wax plug, are deployed from the end of the needle.
| Comparison of<br>Subject Device<br>to Predicate<br>Device | Subject Device<br>(under review) | Primary Predicate Device<br>Focal Therapeutics BioZorb Marker<br>(K143484/K152070) | Reference Device:<br>Cianna Medical Permanent<br>Tissue Marker<br>(K132463) |
|-----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The SmartClip™ is indicated for<br>radiographic marking of sites in<br>soft tissue. In addition, the<br>Marker is indicated in situations<br>where the soft tissue site needs to<br>be marked for future medical<br>procedures. | The BioZorb LP Marker is indicated<br>for radiographic marking of sites in<br>soft tissue. In addition, the Marker is<br>indicated in situations where the soft<br>tissue site needs to be marked for<br>future medical procedures | The Cianna Medical<br>Permanent Tissue Marker is<br>intended to mark tissue<br>during a percutaneous<br>breast biopsy procedure and<br>be permanently visible by<br>radiography and visible for<br>up to thirty days by<br>ultrasound. |
| Intended Use | Implanted Soft Tissue Marker | Implanted Soft Tissue Marker | Implanted Soft Tissue -<br>"Breast" Marker |
| Anatomical Site | Soft Tissue | Soft Tissue | Soft Tissue - Breast |
#### Indications for Use Comparison:
# Technological and Performance Characteristics Comparison:
| Comparison of<br>Subject Device to<br>Predicate Device | Subject Device<br>(under review) | Primary Predicate Device<br>Focal Therapeutics BioZorb<br>Marker<br>(K143484/K152070) | Predicate Device:<br>Cianna Medical Permanent<br>Tissue Marker<br>(K132463) |
|--------------------------------------------------------|-------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|
| Overall Technological<br>Characteristics | Radiographically visible<br>permanent marker | Radiographically visible<br>permanent marker | Radiographically visible<br>permanent marker |
| Principle of Operation | Marker is positioned into<br>tissue site for radiographic<br>visualization of tissue site | Marker is positioned into tissue<br>site for radiographic<br>visualization of tissue site | Marker is positioned into<br>tissue site for radiographic<br>visualization of tissue site |
| Visualization<br>Compatibility | X-Ray, Mammography<br>Ultrasound, MR | X-Ray, Mammography<br>Ultrasound, CT, MR | Radiographic and<br>Ultrasound (30 days) |
| Deployment Method | Introducer | Manual, open surgical | Introducer |
| Implant Duration | Permanent | Permanent | Permanent |
| Marker Materials | Parylene C coated glass vial | Titanium (marker clip),<br>bioabsorbable polymer (spacer) | Titanium beads and Nitinol |
| Sterility | Terminal sterilization by<br>ethylene oxide, sterility<br>assurance level 10-6 | Terminal sterilization by<br>radiation, sterility assurance<br>level 10-6 | Terminal sterilization by<br>ethylene oxide, sterility<br>assurance level 10-6 |
| Biocompatibility | Passed | Passed | Passed |
# Performance Testing:
{5}------------------------------------------------
All necessary testing has been performed for the SmartClip™ to assure substantial equivalence to the predicate devices and demonstrate the device performs as intended.
#### The device design was qualified through the following tests:
- Simulated Use .
- . Mechanical Integrity
- . Imaging Assessment
- MR Compatibility
- Biocompatibility
- Packaging
- Sterilization ●
#### Conclusion:
The SmartClip™ met all specified criteria and did not raise new safety or performance questions.
#### Basis for Determination of Substantial Equivalence:
The Indications for Use, Intended Use, Fundamental Scientific Technology and Principles of Operation for the SmartClip™ are the same as those described for the predicate device. The SmartClip™ is found to have a safety and effectiveness profile that is the same as the predicate device and is determined by Elucent Medical to be substantially equivalent.
In summary, the SmartClip™ has the following similarties to the predicate device which have previously received 510(k) clearance:
- Has the same indications for use
- Has the same intended use
- Has the same implant duration
- Used in the same anatomical site
- Uses similar technological characteristics
- Uses the same principles of operation
- Uses biocompatible materials
- Has the same sterility assurance level
Therefore, the conclusions drawn from the and clinicaltests (biocompatibility) demonstrate that the devices as safe, as effective, and performs as well as the beatly marketed device predicate. The SmartClip™ is substantially equivalent to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.