K180061 · Bard Peripheral Vascular, Inc. · NEU · Mar 9, 2018 · General, Plastic Surgery
Device Facts
Record ID
K180061
Device Name
UltraCor Twirl Breast Tissue Marker
Applicant
Bard Peripheral Vascular, Inc.
Product Code
NEU · General, Plastic Surgery
Decision Date
Mar 9, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4300
Device Class
Class 2
Indications for Use
The UltraCor® Twirl® Breast Tissue Marker is intended for use to attach to soft breast tissue, including axillary lymph nodes, to radiographically mark the location of the biopsy procedure.
Device Story
Device is a disposable, beveled 17G x 10cm needle applicator containing a nitinol radiographic ring wireform. Used by physicians to mark biopsy sites in soft breast tissue or axillary lymph nodes for long-term radiographic identification. The clinician inserts the needle into the tissue site and uses a locking plunger to deploy the nitinol ring. The marker remains in the tissue to provide a radiographic landmark, assisting in subsequent clinical decision-making and localization for follow-up procedures or surgery. No electronic or software components.
Clinical Evidence
No clinical trials were conducted. Evidence consists of a comprehensive literature review regarding the use of breast tissue markers in axillary lymph nodes for metastatic breast cancer patients, supplemented by physician consultations confirming current clinical practice.
Technological Characteristics
Implant: Nitinol radiographic ring wireform. Applicator: 17G x 10cm beveled needle with 1cm depth marks and locking plunger. Principle: Mechanical deployment of a radiopaque marker. Sterilization: Not specified, but stated as same as predicate.
Indications for Use
Indicated for patients undergoing breast biopsy procedures requiring radiographic marking of soft breast tissue or axillary lymph nodes.
Regulatory Classification
Identification
An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.
K243642 — UltraCor Twirl Breast Tissue Marker · Bard Peripheral Vascular, Inc. · Mar 24, 2025
K152510 — UltraCor Twirl Breast Tissue Marker · Bard Peripheral Vascular, Inc. · May 13, 2016
K182082 — Tumark for Eviva, Tumark for Brevera · Somatex Medical Technologies GmbH · Oct 31, 2018
K232615 — LumiMARK Biopsy Site Marker (LM0215T (Tulip Shape)); LumiMARK Biopsy Site Marker (LM0215L (Lotus Shape)); LumiMARK Biopsy Site Marker (LM0215R (Rose shape)) · Devicor Medical Products, Inc. · Jan 12, 2024
K183503 — EasyMark Tissue Marker · Inrad, Inc. · Oct 17, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.
March 9, 2018
Bard Peripheral Vascular, Inc. Meghan McKelvey Regulatory Affairs Associate 1625 W. 3rd Street Tempe, Arizona 85281
Re: K180061
Trade/Device Name: UltraCor Twirl Breast Tissue Marker Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable Clip Regulatory Class: Class II Product Code: NEU Dated: January 5, 2018 Received: January 8, 2018
Dear Ms. McKelvey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may: therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# David Krause -S
for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K180061
Device Name UltraCor® Twirl® Breast Tissue Marker
Indications for Use (Describe)
The UltraCor® Twirl® Breast Tissue Marker is intended for use to attach to soft breast tissue, including axillary lymph nodes, to radiographically mark the location of the biopsy procedure.
Type of Use (Select one or both, as applicable)
| <span style="font-size:12px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------------------------------|
| □ Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# ULTRACOR®Twirl® Breast Tissue Marker 510(k) Summary 21 CFR 807.92
As required by the Safe Medical Devices Act of 1990, coded under Section 513, Part (I)(3)(A) of the Food, Drug and Cosmetic Act, a 510(k) summary upon which substantial equivalence determination is based is as follows:
#### 1. Submitter Information:
| Applicant: | Bard Peripheral Vascular, Inc<br>1625 West 3rd Street<br>Tempe, Arizona 85281 |
|------------|-------------------------------------------------------------------------------|
| Phone: | (480) 350-6153 |
| Fax: | (480) 966-7062 |
| Contact: | Meghan McKelvey, Regulatory Affairs Associate |
| Date: | January 5, 2018 |
#### 2. Subject Device Names:
| Device Trade Name: | ULTRACOR® TWIRL® Breast Tissue Marker |
|-----------------------|----------------------------------------|
| Common or Usual Name: | Breast Tissue Marker |
| Classification: | Class II |
| Review Panel: | General & Plastic Surgery |
| Regulation Number: | 21 CFR 878.4300 (FDA Product Code NEU) |
#### 3. Predicate Devices:
| Predicate 510(k) Number | Subject Device |
|-------------------------------|------------------|
| K152510; cleared May 13, 2016 | ULTRACOR® TWIRL® |
#### Summary of Change 4.
The Indications for Use statement is being updated to include the axillary lymph nodes. Subject device labeling updates including updates to the Instructions for Use and Potential Complications, are being implemented through this submission.
{4}------------------------------------------------
#### 5. Device Description:
The ULTRACOR® Twirl® Breast Tissue Marker consists of a disposable beveled needle applicator containing a nitinol radiographic ring wireform. The wireform is intended for longterm radiographic marking of the tissue site. The applicator has a beveled 17G x 10cm needle with 1cm depth marks and a locking plunger. The ring is deployed from the beveled needle tip into the tissue site.
#### 6. Indications for Use of Device:
The ULTRACOR® Twirl® Breast Tissue Marker is intended for use to attach to soft breast tissue, including axillary lymph nodes, to radiographically mark the location of the biopsy procedure.
#### 7. Technological Comparison to Predicate Devices:
The ULTRACOR® Twirl® Breast Tissue Marker has the following similarities to the predicate device:
- . Same Intended Use
- Same Target Population .
- . Same Technological Characteristics
- Same Fundamental Scientific Technology .
- Same Operating Principal .
- . Same Implant Design and Materials
- . Same Applicator Design and Materials
- . Same Performance Specifications
- . Same Packaging Materials and Configurations
- . Same Sterility Assurance Level and Method of Sterilization
The subject device and the predicate device are different in the following manner:
- Modifications to Indications for Use .
There are no changes to the implants or delivery systems. The subject device labeling has been updated to include the updated Indications for Use language.
{5}------------------------------------------------
#### 8. Performance Data:
The change to the Indications for Use described in this submission does not affect the design of the device and no new or increased risks have been identified, therefore additional bench performance testing was not warranted.
#### 9. Clinical Analysis:
The change to the Indications for Use is clinically supported by both a comprehensive literature review and physician statements. A comprehensive literature review was conducted to collect and assess relevant clinical literature related to the use of clips (breast tissue markers) in axillary lymph nodes in metastatic breast cancer patients. To further reinforce the clinical use of a breast tissue marker in the axillary lymph node region, Bard consulted with physicians to gain further insight on current practices.
#### 10. Conclusions:
The subject device, ULTRACOR® Twirl® Breast Tissue Marker, is substantially equivalent to the legally marketed predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.