BioZorb Marker

K143484 · Focal Therapeutics · NEU · Jun 29, 2015 · General, Plastic Surgery

Device Facts

Record IDK143484
Device NameBioZorb Marker
ApplicantFocal Therapeutics
Product CodeNEU · General, Plastic Surgery
Decision DateJun 29, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4300
Device ClassClass 2

Intended Use

The BioZorb Marker is indicated for radiographic marking of sites in soft tissue. In addition, the Marker is indicated in situations where the soft tissue site needs to be marked for future medical procedures.

Device Story

BioZorb Marker is an implantable radiopaque device used to mark soft tissue sites for future medical procedures. It consists of a permanent titanium marker element held within a bioabsorbable polylactic acid (PLA) polymer spacer. The device is implanted manually during open surgery. Once in place, the marker provides radiographic visibility, allowing clinicians to visualize the site via mammography, ultrasound, X-ray, or CT imaging. The bioabsorbable component resorbs over one or more years, while the titanium element remains as a permanent marker. By providing a stable, visible reference point, the device assists healthcare providers in accurately locating tissue sites for subsequent clinical interventions, potentially improving the precision of follow-up procedures.

Clinical Evidence

Bench testing only. Evidence includes pyrogenicity (USP <85>), MR compatibility (ASTM 2052-14, F2213-06, 2182-11a, F2119-07), and shelf-life testing (ASTM D4169, F2096, F88/F88M, F1980, ISO 11607-1). Performance is identical to predicate K113202.

Technological Characteristics

Implantable radiopaque marker; materials: titanium and bioabsorbable polylactic acid (PLA) polymer. Dimensions: 2-5 cm. Sterilization: Radiation. MR compatibility tested per ASTM 2052-14, F2213-06, 2182-11a, F2119-07. Shelf life validated per ASTM D4169, F2096, F88/F88M, F1980, ISO 11607-1.

Indications for Use

Indicated for radiographic marking of soft tissue sites, including breast, requiring identification for future medical procedures.

Regulatory Classification

Identification

An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name encircling a stylized image of an eagle with three human profiles incorporated into its design. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement around the emblem. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 June 29, 2015 Focal Therapeutics Mr. George Hermann Official Correspondent 4370 Alpine Road, #101 Portola Valley, California 94028 Re: K143484 Trade/Device Name: BioZorb Marker Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable clip Regulatory Class: Class II Product Code: NEU Dated: May 22, 2015 Received: May 28, 2015 Dear Mr. Hermann: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {1}------------------------------------------------ device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # David Krause -S - Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### Indications for Use 510(k) Number (if known) K143484 Device Name BioZorb Marker Indications for Use (Describe) The BioZorb Marker is indicated for radiographic marking of sites in soft tissue. In addition, the Marker is indicated in situations where the soft tissue site needs to be marked for future medical procedures. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### 6. 510(k) Summary This 510(k) summary information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. | APPLICANT: | Focal Therapeutics | |------------------------|------------------------------------------------------------------------------------------------------------------------------------------------| | DATE PREPARED: | June 23, 2015 | | CONTACT PERSON: | George Hermann<br>Focal Therapeutics<br>4370 Alpine Rd. #101<br>Portola Valley, CA 94028<br>Phone: 650.530.2394<br>Fax: 650.530.2397 | | TRADE NAME: | BioZorb Marker | | COMMON NAME: | Implantable Radiographic Marker | | CLASSIFICATION NAME: | Implantable Clip, 21 CFR, 878.4300, 892.5050 | | DEVICE CLASSIFICATION: | Class II | | PRODUCT CODE | NEU | | PREDICATE DEVICES: | Mixed Media Marker (K102506)<br>Civco Suture-type Marker [reference predicate] (K071614)<br>Focal Therapeutics BioZorb Tissue Marker (K113202) | # Substantially Equivalent To: The BioZorb Marker is substantially equivalent in intended use, principal of operation and technological characteristics to the Mixed Media Marker (K 102506), Suture-type Marker (K071614) and the BioZorb Tissue Marker (K113202). ### Description of the Device Subject to Premarket Notification: The BioZorb Marker is an implantable radiopaque marker comprised of a bioabsorbable PLA (polylactic acid) component which resorbs completely in 1 or more years and a permanent component (titanium). The BioZorb Marker is provided sterile for single use and is implantable. ### Indication for Use: The BioZorb Marker is indicated for radiographic marking of sites in soft tissue. {4}------------------------------------------------ In addition, the Marker is indicated in situations where the soft tissue site needs to be marked for future medical procedures. ## Technical Characteristics: The BioZorb Marker has similar physical and technical characteristics to the predicate devices, as illustrated in the table below. | | Subject device | Primary | Primary | Reference | |------------------|------------------|-----------|-------------------|---------------| | | | Predicate | Predicate | Predicate | | | Focal | Focal | Cortex | Civco Medical | | | BioZorb | BioZorb | Mixed Media | Suture-type | | | Marker | Tissue | Marker | Marker | | | | Marker | (K102506) | (K071614) | | | | (K113202) | | | | Overall | Radiographically | SAME | Radiographically | SAME | | Technological | visible | | visible | | | Characteristics | permanent | | permanent | | | | marker | | marker | | | | element(s) in | | element(s) with | | | | bioabsorbable | | or without | | | | polymer spacer | | polymer | | | Principle of | Marker is | SAME | SAME | SAME | | Operation | positioned into | | | | | | tissue site for | | | | | | radiographic | | | | | | visualization of | | | | | | tissue site | | | | | Visualization | Mammography | SAME | SAME | X-Ray | | Compatibility | Ultrasound | | | CT | | | X-Ray | | | (presumed) | | | CT | | | | | Materials of | Titanium, | SAME | Gold, titanium or | Gold. | | Construction | bioabsorbable | | PEEK polymer, | bioabsorbable | | | polymer | | stainless steel | polymer | | | (spacer) | | | (suture) | | Overall Device | 2-5 cm | SAME | 1 mm x ~ 20 mm | >5 cm (with | | Length | | | | suture) | | Typical | Soft tissue | | | | | Anatomical | including breast | SAME | SAME | breast | | Treatment Site | | | | | | Method of | Manual, | SAME | Manual, | SAME | | Marker | open surgical | | percutaneous | | | Deployment | | | | | | Marker | Sutured in place | Tissue | Tissue retention | SAME | | Stability | | retention | | | | Provided sterile | Yes | Yes | Yes | Yes | | Steriliz. method | Radiation | SAME | Unknown | Unknown | {5}------------------------------------------------ ## Performance Data: Performance data included: - Pyrogenicity, LAL (USP Pharmacopeia <85>) . - MR compatibility (ASTM 2052-14, ASTM F2213-06, ASTM 2182-11a, . ASTM F2119-07) and, - Shelf life testing (ASTM D4169, ASTM F2096, ASTM F88/F88M, . ASTM F1980, ISO 11607-1). The Focal BioZorb Marker performance is identical to the predicate device (K113202). # Basis for Determination of Substantial Equivalence: Upon reviewing the safety and efficacy information provided in this submission and comparing intended use, principle of operation and overall technological characteristics, the BioZorb Marker is determined by Focal Therapeutics to be substantially equivalent to existing legally marketed devices.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...