← Product Code [NEU](/submissions/SU/subpart-e%E2%80%94surgical-devices/NEU) · K142198

# BNX Fine Needle Aspiration System (K142198)

_Covidien, LLC · NEU · Oct 8, 2014 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/NEU/K142198

## Device Facts

- **Applicant:** Covidien, LLC
- **Product Code:** [NEU](/submissions/SU/subpart-e%E2%80%94surgical-devices/NEU.md)
- **Decision Date:** Oct 8, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4300
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Indications for Use

The BNX FNA System is used to sample targeted sub-mucosal and extramural gastrointestinal lesions through the accessory channel of an ultrasound endoscope. The needle is designed with a passive (i.e., automatic) safety shielding feature to aid in the prevention of needle stick injury. The 19Ga. and 22Ga. BNX Aspiration Needles are also intended to implant fiducial markers under endoscopic ultrasound to radiographically mark soft tissue for future therapeutic procedures.

## Device Story

BNX Fine Needle Aspiration (FNA) System is a sterile, single-use endoscopic ultrasound aspiration needle; consists of Beacon Endoscopic Ultrasound Delivery System and BNX Aspiration Needle. Device inserted through accessory channel of ultrasound endoscope; used by clinicians to sample gastrointestinal lesions or implant fiducial markers. Needle tip etched for enhanced ultrasonic visualization. Features integrated passive safety shield that automatically covers needle tip upon removal to prevent needle stick injuries. For fiducial delivery, marker is loaded into needle and secured with bone wax; deployed into target tissue by advancing stylet. Modular design allows needle exchange without disconnecting handle from endoscope. Benefits include reduced risk of sharps injury and ability to radiographically mark soft tissue for subsequent therapeutic procedures.

## Clinical Evidence

Bench testing and simulated use testing performed. No clinical data provided. Testing demonstrated that the subject device is substantially equivalent to the predicate devices for the proposed intended use.

## Technological Characteristics

Modular design; metallic needle with stylet housed in sheath with handle assembly. Needle sizes: 19Ga, 22Ga, 25Ga. Tips etched for ultrasonic visualization. Integrated passive needle protection shield. Sterile, single-patient use. Luer-lock connection to endoscope. Manual operation via handle piston.

## Regulatory Identification

An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.

## Predicate Devices

- BNX Fine Needle Aspiration System ([K133008](/device/K133008.md))
- Wilson-Cook Medical EchoTip Ultra Fiducial Needle ([K111895](/device/K111895.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is simple and recognizable, and it is used to represent the U.S. Department of Health & Human Services.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 8, 2014

Covidien, LLC Saket Bhatt Regulatory Affairs Manager 2000 Commonwealth Ave., Suite 110 Auburndale, MA 02466

Re: K142198 Trade/Device Name: BNX Fine Needle Aspiration System Regulation Number: 21 CFR§ 878.4300 Regulation Name: Implantable Clip Regulatory Class: II Product Code: NEU, FCG Dated: August 8, 2014 Received: August 11, 2014

Dear Saket Bhatt,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Benjamin R. Fisher -A

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Food and Drug Administration Indications for Use

Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known)

#### Device Name

BNX Fine Needle Aspiration System

### Indications for Use (Describe)

The BNX FNA System is used to sample targeted sub-mucosal and extramural gastrointestinal lesions through the accessory channel of an ultrasound endoscope. The needle is designed with a passive (i.e., automatic) safety shielding feature to aid in the prevention of needle stick injury. The 19Ga. and 22Ga. BNX Aspiration Needles are also intended to implant fiducial markers under endoscopic ultrasound to radiographically mark soft tissue for future therapeutic procedures.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

| FOR FDA USE ONLY                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| <p>This section applies only to requirements of the Paperwork Reduction Act of 1995.</p> <p><b>*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*</b></p> <p>The burden time for this collection of information is estimated to average 79 hours per response, including the             time to review instructions, search existing data sources, gather and maintain the data needed and complete             and review the collection of information. Send comments regarding this burden estimate or any other aspect             of this information collection, including suggestions for reducing this burden, to:</p> <p>Department of Health and Human Services</p> <p>Food and Drug Administration</p> |

Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (1/14)

Page 1 of 1

PSC Publishing Services (301) 443-6740 EF

{3}------------------------------------------------

## 510(k) SUMMARY 5.0 Page 1 of 2

This 510(k) Summary for the BNX Fine Needle Aspiration System is being submitted in accordance with 21 CFR 807.92.

| Submitter's<br>Name and<br>Address: | Covidien IIc<br>2000 Commonwealth Ave. Suite 110<br>Auburndale, MA 02466-2008                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|-------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact<br>Person:                  | Saket Bhatt, Regulatory Affairs Manager<br>540 Oakmead Parkway<br>Sunnyvale, CA 94085<br>Phone & Fax: 408-328-7357<br>Email: Saket.Bhatt@covidien.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Date:                               | August 8, 2014                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Name of<br>Medical Device:          | Device Regulation: 21 CFR 878.4300 and 21 CFR 876.1075, Class II<br>Common/Usual Name: Implantable clip and Gastroenterology-urology biopsy instrument<br>Product Code: NEU (Marker, Radiographic, Implantable) and FCG (Kit, Needle, Biopsy)<br>Proprietary Name: BNX Fine Needle Aspiration System<br>Classification Panel: Gastroenterology-Urology Devices Panel                                                                                                                                                                                                                                                                                                                                                                                                             |
| Predicate<br>Devices:               | The subject device is substantially equivalent to the BNX Fine Needle Aspiration<br>System (K133008, cleared November 20, 2013).<br><br>The expanded indication for delivery of fiducial markers is substantially equivalent to the<br>Wilson-Cook Medical EchoTip Ultra Fiducial Needle cleared under K111895 (cleared<br>April 27, 2012).                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Device<br>Description:              | The BNX Fine Needle Aspiration (FNA) System is a sterile, single patient use endoscopic<br>ultrasound aspiration needle. The device consists of the BeaconTM Endoscopic<br>Ultrasound Delivery System and BNXTM Fine Needle Aspiration Needle which are<br>assembled before insertion through the accessory channel of an ultrasound endoscope.<br>The device is offered with needle sizes of 19, 22 and 25 ga (however only the 19 and 22<br>gauge needles will be indicated for fiducial delivery). The BNX FNA System has an<br>integrated needle protection shield that automatically engages over the distal end of the<br>needle during removal to cover the needle sharp. In this manner, the needle tip is<br>covered to help protect against inadvertent needle sticks. |
| Indication For<br>Use:              | The BNX FNA System is used to sample targeted sub-mucosal and extramural<br>gastrointestinal lesions through the accessory channel of an ultrasound endoscope. The<br>needle is designed with a passive (i.e., automatic) safety                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |

{4}------------------------------------------------

# 510(k) SUMMARY Page 2 of 2

shielding feature to aid in the prevention of needle stick injury. The 19Ga. and 22Ga. BNX Fine Needle Aspiration Needles are also intended to implant fiducial markers under endoscopic ultrasound to radiographically mark soft tissue for future therapeutic procedures.

Technological The proposed BNX FNA System's indications for use and technological characteristics are identical when compared with the predicate BNX FNA System cleared in K133008. Characteristics: With respect to fiducial delivery, the BNX FNA System indication for use statement proposed is identical to the predicate Wilson-Cook Medical EchoTip Ultra Fiducial Needle cleared under K111895. There have been no changes made to the design of the BNX FNA System to allow for expanded indication.

> The referenced BNX FNA device incorporates a long stiff metallic needle with stylet housed in a sheath with handle assembly. The handle is screwed onto the luer-lock connection of the endoscope. The needle is manipulated by a handle piston which is locked and unlocked by means of a screw to avoid advancement of the needle during introduction and withdrawal of the biopsy assembly. The tips of the Aspiration Needles are etched for enhanced ultrasonic needle visualization. The BNX FNA System is modular in design, i.e., the sheath and handle assembly are incorporated in a Delivery System as a separate component from the Aspiration Needle/stylet assembly. The modular design facilitates exchange of any size aspiration needle as the needle can be removed from the scope without requiring that the handle be disconnected. Additionally, the BNX FNA System has an integrated needle protection shield as a sharps injury protection feature that automatically engages over the distal end of the needle during removal to cover the needle sharp. In this manner, the needle tip is covered to help protect against inadvertent needle sticks. For delivery of fiducial markers the needle is loaded with the fiducial marker and fixated in place using bone wax. The fiducial marker is deployed into the target tissue by advancing the stylet.

Performance This premarket notification is supported with bench and simulated use testing. Testing Data: performed demonstrates that the subject device is substantially equivalent to the predicate devices for the proposed intended use.

Conclusion: Covidien has demonstrated that the proposed BNX FNA System is substantially equivalent to the predicate BNX FNA System and Wilson-Cook Medical EchoTip Ultra Fiducial Needle when used to implant fiducial markers under endoscopic ultrasound to radiographically mark soft tissue for future therapeutic procedures.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/NEU/K142198](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/NEU/K142198)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
