← Product Code [NEU](/submissions/SU/subpart-e%E2%80%94surgical-devices/NEU) · K121113

# BIOPSY SCIENCES, INC. HYDROMARK BREST BIOSY SITE MARKER (K121113)

_Biopsy Sciences, LLC · NEU · Jul 12, 2012 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/NEU/K121113

## Device Facts

- **Applicant:** Biopsy Sciences, LLC
- **Product Code:** [NEU](/submissions/SU/subpart-e%E2%80%94surgical-devices/NEU.md)
- **Decision Date:** Jul 12, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4300
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Indications for Use

To mark tissue during a percutaneous breast biopsy procedure, be visible under ultrasound for at least 6 weeks, and be permanently visible by x-ray and MRI.

## Device Story

The HydroMark Breast Biopsy Site Marker is a sterile, disposable implantable device used to mark tissue during percutaneous breast biopsy procedures. It consists of a resorbable hydrogel component and a permanent metallic (stainless steel or titanium) coil. The device is pre-loaded in an applicator compatible with commercially available biopsy devices. Upon deployment into the biopsy tract, the hydrophilic hydrogel expands on fluid contact to anchor the marker at the biopsy site. The hydrogel provides temporary ultrasound visibility (at least 6 weeks) while it degrades via hydrolysis. The metallic coil remains permanently in place to provide visibility under x-ray and MRI. The device is used by clinicians in a clinical setting to assist in the long-term localization of biopsy sites, aiding in subsequent clinical decision-making and patient follow-up.

## Clinical Evidence

No clinical data. Substantial equivalence is supported by non-clinical performance testing, including physical, functional, and biocompatibility evaluations of the finished product using hydrogel from the new supplier, confirming performance meets original specifications.

## Technological Characteristics

Implantable marker consisting of a resorbable hydrophilic hydrogel and a permanent metallic (stainless steel or titanium) coil. Hydrogel expands via fluid contact for ultrasound visibility; degrades via hydrolysis. Metallic coil provides permanent x-ray and MRI visibility. Provided in a sterile, disposable applicator.

## Regulatory Identification

An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.

## Predicate Devices

- Biopsy Sciences HydroMark Breast Biopsy Site Marker ([K060769](/device/K060769.md))
- Biopsy Sciences HydroMark Breast Biopsy Site Marker ([K083006](/device/K083006.md))
- Biopsy Sciences HydroMark Breast Biopsy Site Marker ([K090501](/device/K090501.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for Biopsy Sciences. The word "Biopsy" is split into two parts, with "Bio" in white letters on a black background and "psy" in black letters on a white background. Below the word "Biopsy" is the word "Sciences" in a smaller font.

JUL 12 2012

## 510(K) SUMMARY (21 CFR 807.92)

## HYDROMARK BIOPSY SITE MARKER

510(k) Owner: Biopsy Sciences, Inc. 4900 Creekside Drive, Suite C Clearwater, FL 33760 Tel: 727-290-9825 Fax: 727-290-9828

- Sharon Rockwell Contact Person: Tel: 714-695-9269 E-mail: srockwell(@writeme.com
- July, 2012 Date Prepared:
- HydroMark Biopsy Site Marker Trade Name:
- Biopsy site marker Common Name:
- Classification Name: Implantable clip per 21 CFR 878.4300, NEU
- Predicate Devices: Biopsy Sciences HydroMark Breast Biopsy Site Marker, K060769 Biopsy Sciences HydroMark Breast Biopsy Site Marker, K083006 Biopsy Sciences HydroMark Breast Biopsy Site Marker, K090501
- The HydroMark Breast Biopsy Site Marker contains a resorbable Device Description: hydrogel component and a metallic component for permanent marking. The hydrogel has features that are unique and highly desirable for breast tissue marking.

The HydroMark Site Marker is provided pre-loaded in a sterile, disposable applicator that is compatible with specified commercially available biopsy devices. The HydroMark is deployed by the delivery system and is left in the tract created during the biopsy procedure.

The Biopsy Sciences HydroMark Breast Biopsy Site Markers will have a change in vendor for the supplier of the hydrogel material. The change in the supplier of the hydrogel has no effect on the finished product specifications.

{1}------------------------------------------------

- To mark tissue during a percutaneous breast biopsy procedure, be Intended Use: visible under ultrasound for at least 6 weeks, and be permanently visible by x-ray and MRI.
## Technological Characteristics:

The hydrogel component expands on fluid contact to fill the track of the biopsy needle anchoring the HydroMark at the exact location of biopsy. Because the hydrogel is hydrophilic, it is clearly distinct from normal breast structure under ultrasound imaging. The hydrogel material degrades via hydrolysis over time leaving the internal stainless steel or titanium coil which provides permanent visibility under x-ray and MRI.

Non-Clinical

- Performance Data: Non-clinical testing included physical, functional and biocompatibility of the finished products manufactured in a newly qualified facility, using hydrogel material from a new supplier. The devices performed as intended according to the original specifications established for the finished device.
The testing supports a determination of substantial equivalence to Conclusions: predicate devices cleared by FDA.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's mission to protect the health of all Americans and provide essential human services. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the left side of the logo.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Biopsy Sciences, LLC % Ms. Sharon Rockwell Consultant 5582 Chalon Road Yorba Linda, Florida 33760

JUL 12 2012

Re: K121113

Trade/Device Name: Biopsy Sciences HydroMark Breast Biopsy Site Marker Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable clip Regulatory Class: II Product Code: NEU Dated: April 09, 2012 Received: April 12, 2012

Dear Ms. Rockwell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{3}------------------------------------------------

Page 2 - Ms. Sharon Rockwell

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical Or ice-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## INDICATIONS FOR USE STATEMENT

510(k) Number (if known): _

Device Name: Biopsy Sciences HydroMark Breast Biopsy Site Marker

Indications for Use:

To mark tissue during a percutaneous breast biopsy procedure, be visible under ultrasound for at least 6 weeks, and be permanently visible by x-ray and MRI.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Daniel Kronefu MXM
(Division Sign-Off)

Page of

Division of Surgical. Orthopedic. and Restorative Devices

510(k) Number K121113

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/NEU/K121113](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/NEU/K121113)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
