← Product Code [NEU](/submissions/SU/subpart-e%E2%80%94surgical-devices/NEU) · K062528

# TISSUE SITE MARKING SYSTEM (K062528)

_Suros Surgical Systems, Inc. · NEU · Jan 26, 2007 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/NEU/K062528

## Device Facts

- **Applicant:** Suros Surgical Systems, Inc.
- **Product Code:** [NEU](/submissions/SU/subpart-e%E2%80%94surgical-devices/NEU.md)
- **Decision Date:** Jan 26, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4300
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Intended Use

The Tissue Site Marking System is indicated for the permanent radiographic marking of sites in soft tissue.

## Device Story

Tissue Site Marking System consists of an implantable marker (bioabsorbable and permanent components) and a manual deployment device. Used to mark soft tissue sites for identification under ultrasound, x-ray, MRI, nuclear imaging, CT, or direct visualization. Operated by clinicians in a clinical setting. Marker remains in situ to provide a permanent radiographic reference point, assisting in follow-up localization or surgical planning.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

System comprises a sterile, single-use deployment device and an implantable marker containing both bioabsorbable and permanent components. Designed for manual deployment under various imaging modalities (ultrasound, x-ray, MRI, nuclear, CT).

## Regulatory Identification

An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# SECTION I - 510(K) Summa FOI Re leasable

JAN 26 2007

Pursuant to § 513(i)(3)(A) of the Food, Drug, and Cosmetic Act, Suros Surgical Systems, Inc. is required to submit with this premarket notification " ... adequate summary of any information respecting safety and effectiveness or state that such information will be made available upon request of any person." Suros Surgical Systems, Inc. chooses to submit a summary of information respecting safety and effectiveness.

| Classification Name:   | Marker, Radiographic Implantable |            |
|------------------------|----------------------------------|------------|
| Common/Usual Name:     | Tissue Site Marking System       |            |
| Proprietary Name:      | None at this time                |            |
| Device Classification: | Class II                         |            |
| Description            | Number                           | 21 CFR Ref |
| Implantable Clip       | NEU                              | § 878.4300 |

Owner/Operator: Suros Surgical Systems, Inc. - FDA Registration # 3003862400 Northwest Technology Center 6100 Technology Center Drive Indianapolis, IN 46278 317-344-7500

Contact: Heidi Strunk Director, Quality & Regulatory Suros Surgical Systems, Inc. Northwest Technology Center 6100 Technology Center Drive Indianapolis, IN 46278 317-344-7500

#### DESCRIPTION OF DEVICE

The Tissue Site Marking System is supplied as a sterile device and is intended for single patient use only. The Tissue Site Marking System is composed of two elements:

1) an implantable device (marker) including a bioabsorable component and a permanent component and

{1}------------------------------------------------

### SECTION I - 510(K) Summary

2) a deployment device.

The Tissue Site Marking System is intended for use with the manual method of deployment under the following imaging modalities: ultrasound, x-ray, magnetic resonance, nuclear, computed tomography, and direct visualization.

### INDICATIONS FOR USE

The Tissue Site Marking System is indicated for the permanent radiographic marking of sites in soft tissue.

## CONTRAINDICATIONS

None known

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows a logo for the Ministry of Health & Human Services. The logo is circular and contains the text "MINISTRY OF HEALTH & HUMAN" around the left side of the circle. To the right of the text is a stylized image of an eagle.

Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850

Ms. Heidi H. Strunk Director, Quality and Regulatory Suros Surgical Systems. Inc. 6100 Technology Center Drive INDIANAPOLIS IN 46278

Re: K062528

Trade/Device Name: Tissue Site Marking System Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable clip Regulatory Class: II Product Code: NEU Dated: December 22, 2006 Received: December 26, 2006

JAN 2 6 7007

#### Dear Ms. Strunk:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indiver for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cometic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. TTo general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulter of de

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Image /page/2/Picture/10 description: The image shows a circular logo with the letters "FDA" in bold, stylized font in the center. Above the letters, the years "1906 - 2006" are printed, indicating a centennial celebration. Below the letters, the word "Centennial" is written in a cursive font, and there are three stars underneath. The logo is surrounded by a dotted circle, and there are small star shapes placed around the perimeter of the circle.

Promoting and Promoting Public Health

{3}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology)                      | 240-276-0120 |
| Other          |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150

or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.

Sincerely yours.

Nancy C. Hogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number (if known):

Device Name: Tissue Site Marking System

Indications for Use: The Tissue Site Marking System is indicated for the permanent radiographic marking of sites in soft tissue.

2528

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

David A. Ingram

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________

Prescription Use: 
(Per 21 CFE 801.109)

OR

Over-the Counter Use:

:

(Optional Format 1-2-96)

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/NEU/K062528](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/NEU/K062528)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
