← Product Code [NBY](/submissions/SU/subpart-e%E2%80%94surgical-devices/NBY) · K173922

# C-PTFETM Surgical Suture (K173922)

_Cp Medical, Inc. · NBY · Mar 26, 2018 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/NBY/K173922

## Device Facts

- **Applicant:** Cp Medical, Inc.
- **Product Code:** [NBY](/submissions/SU/subpart-e%E2%80%94surgical-devices/NBY.md)
- **Decision Date:** Mar 26, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.5035
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

CP Medical nonabsorbable C-PTFETM Surgical Suture is a removable, nonabsorbable surgical suture intended for use in general soft tissue approximation and/or ligation, including dental surgical procedures. CP Medical C-PTFETM sutures are contra indicated for use in microsurgery, ophthalmic procedures, or peripheral neural tissue.

## Device Story

C-PTFE™ Surgical Suture is a sterile, nonabsorbable, monofilament polytetrafluoroethylene (PTFE) suture; provided with or without needles. Used in clinical settings by healthcare professionals for soft tissue approximation and ligation. Sutures are supplied in figure-8 or racetrack configurations within Tyvek/film pouches. Device is single-use and disposable. Sterilized via ethylene oxide (EO). No electrical, software, or mechanical components. Clinical benefit derived from providing secure tissue approximation during surgical procedures.

## Clinical Evidence

Bench testing only. No clinical data or animal studies were conducted. Performance supported by comparative bench testing against predicate device, including USP-40-NF35:2017 standards for tensile strength, diameter, and needle attachment, and biocompatibility testing per ISO 10993.

## Technological Characteristics

Material: Monofilament Polytetrafluoroethylene (PTFE). Form factor: Suture with/without needles, figure-8 or racetrack configuration. Sterilization: Ethylene Oxide (EO) per EN ISO-11135:2014. Biocompatibility: Per ISO 10993. No electrical components, software, or active energy sources.

## Regulatory Identification

Nonabsorbable expanded polytetrafluoroethylene (ePTFE) surgical suture is a monofilament, nonabsorbable, sterile, flexible thread prepared from ePTFE and is intended for use in soft tissue approximation and ligation, including cardiovascular surgery. It may be undyed or dyed with an approved color additive and may be provided with or without an attached needle(s).

## Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.

## Predicate Devices

- MonoTex ([K140415](/device/K140415.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 26, 2018

CP Medical Inc. Mr. John Hartigan Director QA/RA 1775 Corporate Dr. Suite 150 Norcross, Georgia 30093

Re: K173922

Trade/Device Name: C-PTFFE™ Surgical Suture Regulation Number: 21 CFR 878.5035 Regulation Name: Nonabsorbable Expanded Polytetrafluoroethylene Surgical Suture Regulatory Class: Class II Product Code: NBY Dated: December 21, 2017 Received: December 26, 2017

Dear Mr. Hartigan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application(PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820);

{1}------------------------------------------------

#### Page 2 - John Hartigan

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

> Sincerely. Jennifer R. Stevenson -S For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K173922

Device Name C-PTFETM Surgical Suture

#### Indications for Use (Describe)

The CP Medical nonabsorbable C-PTFETM Surgical Suture is a removable, nonabsorbable surgical suture intended for use in general soft tissue approximation and/or ligation, including dental surgical procedures. CP Medical C-PTFETM sutures are contra indicated for use in microsurgery, ophthalmic procedures, or peripheral neural tissue.

Type of Use (Select one or both, as applicable)

| <div> <span> <svg height="12" width="12"> <rect fill="none" height="12" stroke="black" width="12"></rect> <line stroke="black" x1="0" x2="12" y1="0" y2="12"></line> <line stroke="black" x1="0" x2="12" y1="12" y2="0"></line> </svg> </span>    Prescription Use (Part 21 CFR 801 Subpart D)   </div> |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg height="12" width="12"> <rect fill="none" height="12" stroke="black" width="12"></rect> </svg> </span>     Over-The-Counter Use (21 CFR 801 Subpart C)    </div>                                                                                                                      |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 1. 510(k) Summary

In accordance with 21 CFR 807.87 (h) and 21 CRF 807.92, the 510(k) summary for the C-PTFE® Surgical Suture is provided below.

| Date Summary<br>Prepared                                                                                              | December 14, 2017                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|-----------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer/Distribu<br>tor/ Sponsor                                                                                 | CP Medical Inc.<br>1775 Corporate Drive, Ste. 150<br>Norcross, GA 30093                                                                                                                                                                                                                                                                                                                                                                                                                           |
| 510(k) Contact                                                                                                        | CP Medical<br>John Hartigan<br>1775 Corporate Drive<br>Norcross, GA 30093<br>470-273-6005 (phone)<br>hartiganj@cpmedical.com (email)                                                                                                                                                                                                                                                                                                                                                              |
| Trade Name                                                                                                            | C-PTFETM Surgical Suture                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Common Name                                                                                                           | Surgical Suture, Nonabsorbable                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Code -Name -<br>Classification                                                                                        | NBY – Surgical Suture Class II: 21 CFR §878.5035                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Predicate Devices                                                                                                     | MonoTex K140415<br>This predicate has not been subject to a design-related recall<br>(FDA's TPLC website,. no filings from 2000 to date 2017)                                                                                                                                                                                                                                                                                                                                                     |
| Device Description                                                                                                    | Device Description: The CP Medical C-PTFETM Surgical<br>Sutures are Nonabsorbable, Polytetraflouroethylene,<br>single use, sterile, disposable, monofilament sutures and<br>are provided with or without needles.<br><br>Individual sutures are available in a "racetrack"<br>configuration, attached around a HDPE delivery system,<br>and/or in a figure 8 in a Tyvek/film pouch.<br><br>The CP Medical C-PTFETM Surgical Sutures are sterilized<br>via EO in accordance with EN ISO-11135:2014 |
| Intended Use                                                                                                          | CP Medical nonabsorbable C-PTFETM Surgical Suture is<br>a removable, nonabsorbable surgical suture intended for<br>use in general soft tissue approximation and/or ligation,<br>including dental surgical procedures. CP Medical C-                                                                                                                                                                                                                                                               |
| PTFE™ sutures are contra indicated for use in<br>microsurgery, ophthalmic procedures, or peripheral<br>neural tissue. |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| <b>Non-Clinical<br/>Performance Testing</b>                                                                           | The CP Medical nonabsorbable C-PTFE™ Surgical Suture<br>was tested recognized standard 6-393: USP-40-NF35:2017,<br>Nonabsorbable Surgical Suture, 6-394: USP-40-NF35:2017,<br><881>, Tensile Strength, 6-395: USP-40-NF35:2017 <861>,<br>Diameter, 6-396 USP-40-NF35:2017 <871>, Sutures –<br>Needle Attachment.                                                                                                                                                                                  |
| <b>Substantial<br/>Equivalence<br/>Summary<br/>(Conclusion)</b>                                                       | Based on the technological; characteristics and non-clinical<br>performance testing the CP Medical C-PTFE™ Surgical<br>Suture was shown to be substantially equivalent to the<br>predicate device, the MonoTex K140415.                                                                                                                                                                                                                                                                           |

{4}------------------------------------------------

{5}------------------------------------------------

Additionally, a comparison between the CP Medical C-PTFETM Surgical Suture and the predicate device is summarized below:

| Feature                                          | CP Medical C-PTFETM<br>Surgical Suture                                                                                                                                                                                                                                                                                                                                                         | MonoTex K140415                                                                                                                                                                                                                                                                                                                                                                 | Substantially<br>Equivalent                              |
|--------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|
| Device<br>Classification                         | Class II: 21 CFR<br>§878.5035                                                                                                                                                                                                                                                                                                                                                                  | Class II: 21 CFR<br>§878.5035                                                                                                                                                                                                                                                                                                                                                   | Yes                                                      |
| Product Code                                     | NBY – Suture Surgical                                                                                                                                                                                                                                                                                                                                                                          | NBY – Suture Surgical                                                                                                                                                                                                                                                                                                                                                           | Yes                                                      |
| Device<br>Common<br>Name as<br>Cleared by<br>FDA | Suture, Surgical<br>Nonabsorbable,<br>Polytetraflouroethylene                                                                                                                                                                                                                                                                                                                                  | Suture, Surgical<br>Nonabsorbable,<br>Polytetraflouroethylene                                                                                                                                                                                                                                                                                                                   | Yes                                                      |
| Indications for<br>use                           | CP Medical<br>nonabsorbable C-<br>PTFETM Surgical<br>Suture is a removable,<br>nonabsorbable surgical<br>suture intended for use<br>in general soft tissue<br>approximation and/or<br>ligation, including<br>dental surgical<br>procedures. CP<br>Medical C-PTFETM<br>sutures are contra<br>indicated for use in<br>microsurgery,<br>ophthalmic procedures,<br>or peripheral neural<br>tissue. | Riverpoint MonoTex<br>surgical suture is<br>indicated for use in<br>general soft tissue<br>approximation and/or<br>ligation, including<br>cardiovascular, dental,<br>general surgical<br>procedures and repair<br>of dura matter.<br>Riverpoint MonoTex<br>sutures are not<br>indicated for use in<br>microsurgery,<br>ophthalmic procedures,<br>or peripheral neural<br>tissue | Yes                                                      |
| Patient<br>Population                            | Single use only                                                                                                                                                                                                                                                                                                                                                                                | Single use only                                                                                                                                                                                                                                                                                                                                                                 | Yes                                                      |
| Environment<br>of Use                            | Point of procedure –<br>clinical settings.                                                                                                                                                                                                                                                                                                                                                     | Point of procedure –<br>clinical settings.                                                                                                                                                                                                                                                                                                                                      | Yes                                                      |
| Feature                                          | CP Medical C-PTFE™<br>Surgical Suture                                                                                                                                                                                                                                                                                                                                                          | MonoTex K140415                                                                                                                                                                                                                                                                                                                                                                 | Substantially<br>Equivalent                              |
| Device<br>Description                            | CP Medical C-PTFE™<br>Surgical Suture are<br>sterile, nonabsorbable,<br>monofilament sutures<br>available with or<br>without needles, The<br>sutures are delivered in<br>a figure 8 or racetrack<br>configuration, which is<br>contained within a<br>Tyvek/film package.<br>The sutures are<br>sterilized by EO.                                                                               | Riverpoint PTFE<br>MonoTex Surgical<br>Suture are sterile,<br>nonabsorbable,<br>monofilament sutures<br>available with or<br>without needles, The<br>sutures are delivered in<br>a figure 8 or racetrack<br>configuration, which is<br>contained within a<br>Tyvek/film package.<br>The sutures are<br>sterilized by EO.                                                        | Yes                                                      |
| Principles of<br>Operation                       | CP Medical C-PTFE™<br>Surgical Suture are for<br>use in clinical settings<br>and carry a prescription.<br>The sutures are<br>removed for device<br>packing following<br>aseptic protocols.                                                                                                                                                                                                     | Riverpoint PTFE<br>MonoTex Surgical<br>Suture are for use in<br>clinical settings and<br>carry a prescription.<br>The sutures are<br>removed for device<br>packing following<br>aseptic protocols.                                                                                                                                                                              | Yes                                                      |
| Sizes                                            | Various diameters<br>based on USP (unless<br>stated on labeling),<br>various lengths.                                                                                                                                                                                                                                                                                                          | Various diameters<br>based on USP (unless<br>stated on labeling),<br>various lengths.                                                                                                                                                                                                                                                                                           | Yes                                                      |
| Needles                                          | Available with or<br>without standard<br>needles attached.                                                                                                                                                                                                                                                                                                                                     | Available with or<br>without standard<br>needles attached.                                                                                                                                                                                                                                                                                                                      | Yes                                                      |
| Colors<br>available                              | White/Undyed                                                                                                                                                                                                                                                                                                                                                                                   | White/Undyed                                                                                                                                                                                                                                                                                                                                                                    | Yes                                                      |
| Feature                                          | CP Medical C-PTFE™ Surgical Suture                                                                                                                                                                                                                                                                                                                                                             | MonoTex K140415                                                                                                                                                                                                                                                                                                                                                                 | Substantially Equivalent                                 |
| Non-Clinical<br>Testing<br>Summary               | The CP Medical<br>nonabsorbable C-PTFE™ Surgical<br>Suture was tested per<br>recognized standard:<br>6-393: USP-40-<br>NF35:2017,<br>Nonabsorbable<br>Surgical Suture, 6-394: USP-40-<br>NF35:2017, <881>,<br>Tensile Strength. 6-395: USP-40-<br>NF35:2017 <861>,<br>Diameter 6-396 USP-40-<br>NF35:2017 <871>,<br>Sutures – Needle<br>Attachment.                                            | Riverpoint PTFE<br>MonoTex are tested<br>per USP <861>, <871><br>and <881>.                                                                                                                                                                                                                                                                                                     | Yes – Test<br>report 17-033<br>confirmed<br>equivalency. |
| Sterile                                          | Yes                                                                                                                                                                                                                                                                                                                                                                                            | Yes                                                                                                                                                                                                                                                                                                                                                                             | Yes                                                      |
| Sterilization<br>method                          | EO                                                                                                                                                                                                                                                                                                                                                                                             | EO                                                                                                                                                                                                                                                                                                                                                                              | Yes                                                      |
| Biocompatible<br>materials                       | Per ISO 10993                                                                                                                                                                                                                                                                                                                                                                                  | Per ISO 10993                                                                                                                                                                                                                                                                                                                                                                   | Yes                                                      |
| Instructions<br>for Use                          | Yes                                                                                                                                                                                                                                                                                                                                                                                            | Yes                                                                                                                                                                                                                                                                                                                                                                             | Yes                                                      |
| Materials                                        | Monofilament<br>Polytetraflouroethylene                                                                                                                                                                                                                                                                                                                                                        | Monofilament<br>Polytetraflouroethylene                                                                                                                                                                                                                                                                                                                                         | Yes                                                      |
| RX                                               | Yes                                                                                                                                                                                                                                                                                                                                                                                            | Yes                                                                                                                                                                                                                                                                                                                                                                             | Yes                                                      |

{6}------------------------------------------------

{7}------------------------------------------------

# PERFORMANCE DATA

The following performance data were provided in support of the substantial equivalence determination.

{8}------------------------------------------------

### Biocompatibility testing

The biocompatibility evaluation for CP Medical C-PTFE™ was conducted in accordance with FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO 10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing," May 1, 1995, and International Standard ISO 10993-1 "biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk management Process," as recognized by FDA. The battery of testing included the following tests:

- . Muscle Implantation Study in Rabbits - 4 weeks
- . Maximization Sensitization
- . Bacterial Reverse Mutation
- Systemic Toxicity in Mice
- USP Pyrogen Study .
- Cytotoxicity
- Intracutaneous .
- Chemical Characterization ●

The C-PTFE™ device is considered a >30 implantable device

#### Electrical safety and electromagnetic compatibility (EMC)

There are no electrical components in the C-PTFE™. No Electrical safety information is required for the safe functioning of this device.

#### Software Verification and Validation Testing

There is no software in the C-PTFE™ and the C-PTFE™ sutures requires no software to function. This section does not apply to this device.

#### Mechanical and acoustical testing

This section does not apply to this device.

{9}------------------------------------------------

### Animal study

This section does not apply to this device.

### Clinical Studies

This section does not apply to this device.

### Summary

Based on the comparative testing between post sterile, aged, C-PTFE™ and MonoTex (predicate) currently in interstate commerce, the C-PTFE™ was found to have a safety and effectiveness profile that is similar to the predicate device.

## Conclusions

The non-clinical data support the safety of the device and the sterilization validation demonstrate that the C-PTFE™ device should perform as intended in the specific use conditions. Based on the technological characteristics and non-clinical performance testing, the CP Medical C-PTFE™ Surgical Suture performs comparably the predicate device, MonoTex K140415, which is currently marketed for the same intended use.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/NBY/K173922](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/NBY/K173922)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
