OTOSILK OTOLOGICAL GRAFT DRESSING

K964534 · Boston Medical Products, Inc. · NAD · Apr 9, 1997 · General, Plastic Surgery

Device Facts

Record IDK964534
Device NameOTOSILK OTOLOGICAL GRAFT DRESSING
ApplicantBoston Medical Products, Inc.
Product CodeNAD · General, Plastic Surgery
Decision DateApr 9, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4020
Device ClassClass 1
AttributesTherapeutic

Indications for Use

To be used as a temporary dressing in the external auditory canal following tympanoplasty or stapedectomy procedures.

Device Story

OTOSILK™ Otological Graft Dressing; sterile, disposable, non-adherent material. Used as carrier for split-thickness skin grafts in ear canal procedures (tympanoplasty, canalplasty, mastoid tympanoplasty, congenital aural atresia correction, canal exostoses removal). Graft placed dermis-side up on dressing; trimmed; applied to canal. Blue color aids visualization for complete removal 1-2 weeks post-operatively. Used by surgeons in clinical/OR settings. Benefits: facilitates graft placement; ensures total dressing removal without disturbing graft.

Clinical Evidence

Bench testing only. Sterile finished device samples passed biocompatibility testing; determined to be non-reactive and non-cytotoxic.

Technological Characteristics

Sterile, disposable, non-adherent material. Blue-colored for visualization. Form factor designed for ear canal application.

Indications for Use

Indicated for use as a temporary dressing in the external auditory canal for patients undergoing tympanoplasty or stapedectomy procedures.

Regulatory Classification

Identification

An occlusive wound dressing is a nonresorbable, sterile or non-sterile device intended to cover a wound, to provide or support a moist wound environment, and to allow the exchange of gases such as oxygen and water vapor through the device. It consists of a piece of synthetic polymeric material, such as polyurethane, with or without an adhesive backing. This classification does not include an occlusive wound dressing that contains added drugs such as antimicrobial agents, added biologics such as growth factors, or is composed of materials derived from animal sources.

Predicate Devices

Submission Summary (Full Text)

{0} APR - 9 1997 K964534 # SMDA Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. ## Product: OTOSILK™ Otological Graft Dressing ## Description: The OTOSILK™ Graft Dressing is a sterile, disposable graft dressing constructed of non-adherent material, designed for use with split thickness skin grafts used in procedures such as tympanoplasty, canalplasty, mastoid tympanoplasty, correction of congenital aural atresia, and removal of canal exostoses. Split thickness skin grafts may be taken from several sites, using either a razor blade or dermatome. The OTOSILK dressing will be used as a carrier for this skin graft to the ear canal. The graft is placed dermis side up on the dressing and the uncovered material is trimmed away. The skin graft and dressing are then applied to the canal. The dressing is removed from the canal one to two weeks post-operatively. The blue color of the OTOSILK Graft Dressing increases visualization of the dressing, ensuring total removal from the ear canal without disturbing the graft. ## Indication For Use: To be used as a temporary dressing in the external auditory canal following tympanoplasty or stapedectomy procedures. ## Predicate Device: Product No. XO-2904 originally marketed by Xomed (8641 Baypine Rd., Jacksonville, FL 32216) and subsequently marketed as Product No. 14-28010 by Xomed-Treace (6743 Southpoint Drive North, Jacksonville, FL 32216). ## Testing: Sterile finished device samples passed biocompatibility testing and were determined to be non-reactive and non-cytotoxic. Submitted by: Stuart K. Montgomery, President Date:
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...