The Red Head Collection Device, The Jordy Connection System

K161372 · Miami Fat Supply · MUU · Feb 24, 2017 · General, Plastic Surgery

Device Facts

Record IDK161372
Device NameThe Red Head Collection Device, The Jordy Connection System
ApplicantMiami Fat Supply
Product CodeMUU · General, Plastic Surgery
Decision DateFeb 24, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.5040
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Red Head and Jordy Connection System is indicated for use as a liposuction adipose (fatty) collection system used in the aesthetic body contouring and collection of autologous adipose tissue. The Red Head and Jordy Connection System is intended to be used in the following surgical procedures: Aspiration of adipose (fatty) tissue Harvesting adipose (fatty) tissue Filtering adipose (fatty) tissue Extraction of autologous adipose (fatty) tissue

Device Story

The Red Head and Jordy Connection System is a single-use, closed-loop liposuction collection device. It consists of a polycarbonate canister, funnel, and lid, with polypropylene evacuation components and silicone tubing. During aesthetic body contouring, the device connects to a standard liposuction aspiration pump. Adipose tissue is harvested from the patient and enters the canister through a collection port. An internal funnel with perforations separates fatty tissue from fluids; waste is evacuated by closing a valve on the lid. The collected adipose tissue is then withdrawn via a bottom port using a Toomey syringe for autologous re-injection. The device is operated by physicians in a clinical setting. It benefits patients by providing a sterile, closed-loop method for collecting and filtering autologous fat for transfer, minimizing tissue manipulation and maintaining tissue viability.

Clinical Evidence

Bench testing only. Performance testing included assembly verification, leak testing, canister implosion, tubing collapse, and pull-off force testing. Biocompatibility testing (cytotoxicity, intracutaneous reactivity, sensitization, and acute systemic toxicity) was performed per ISO 10993. Tissue viability testing confirmed no adverse effects on adipose tissue compared to the predicate.

Technological Characteristics

Materials: Polycarbonate (canister, funnel, lid), Polypropylene (spout, clamps), Silicone (tubing). Principle: Vacuum-assisted aspiration and gravity/perforation-based filtration. Form factor: Canister-based collection system, up to 2500 mL capacity. Connectivity: None. Sterilization: User-sterilized (SAL 10^-6). No software or electronic components.

Indications for Use

Indicated for patients undergoing aesthetic body contouring requiring aspiration, harvesting, filtering, and extraction of autologous adipose tissue.

Regulatory Classification

Identification

A suction lipoplasty system is a device intended for aesthetic body contouring. The device consists of a powered suction pump (containing a microbial filter on the exhaust and a microbial in-line filter in the connecting tubing between the collection bottle and the safety trap), collection bottle, cannula, and connecting tube. The microbial filters, tubing, collection bottle, and cannula must be capable of being changed between patients. The powered suction pump has a motor with a minimum of 1/3 horsepower, a variable vacuum range from 0 to 29.9 inches of mercury, vacuum control valves to regulate the vacuum with accompanying vacuum gauges, a single or double rotary vane (with or without oil), a single or double diaphragm, a single or double piston, and a safety trap.

Special Controls

*Classification.* Class II (special controls). Consensus standards and labeling restrictions.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of a caduceus, which is a symbol often associated with healthcare and medicine. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 24, 2017 Miami Fat Supply % Ms. Ayanna Brown Bgf Consulting 325 South Mcgee Ave. Apopka. Florida 32703 Re: K161372 Trade/Device Name: The Red Head Collection Device. The Jordy Connection System Regulation Number: 21 CFR 878.5040 Regulation Name: Suction Lipoplasty System Regulatory Class: Class II Product Code: MUU Dated: May 9, 2016 Received: May 17, 2016 Dear Ms. Brown: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # David Krause -S for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ## Indications for Use 510(k) Number (if known) K161372 Device Name Red Head Collection Device and Jordy Connection System #### Indications for Use (Describe) The Red Head and Jordy Connection System is indicated for use as a liposuction adipose (fatty) collection system used in the aesthetic body contouring and collection of autologous adipose tissue. The Red Head and Jordy Connection System is intended to be used in the following surgical procedures: - Aspiration of adipose (fatty) tissue - Harvesting adipose (fatty) tissue - Filtering adipose (fatty) tissue - Extraction of autologous adipose (fatty) tissue Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995, #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. {3}------------------------------------------------ # 510(k) Summary | Device Name: | The Red Head and Jordy Connection System | |------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) Submission: | K161372 | | Date of Submission: | January 25, 2017 | | 510(k) Owner & Manufacturer: | Donnell Mark Jordan<br>Miami Fat Supply Inc.<br>1510 Max Hooks Rd. #F<br>Groveland, FL 34736<br>Telephone: 352-557-6979<br>Fax: 352-388-9447<br>miamifatsupply@gmail.com | | 510(k) Submitter and Contact: | Ayanna Brown<br>BGF Consulting LLC<br>325 South McGee Avenue<br>Apopka, FL 32703<br>Phone: 407-574-7195<br>Fax: 407-814-0127<br>ayanna@bgfconsulting.com | | FDA Classification Name: | 21 CFR Part 878- General and Plastic Surgery<br>Subpart E - Surgical Devices<br>Sec. 878.5040 Suction Lipoplasty System<br>Class II Device<br>Product Code MUU | | Classification Panel:<br>Common Name:<br>Predicate Device: | 21 CFR Part 880- General Hospital and Personal Use<br>Subpart G - General Hospital and Personal Use<br>Sec. 880.6960 Irrigation Syringe<br>Class I (Sterile) Device<br>Product Code KYZ<br>General and Plastic Surgery<br>Lipoplasty System<br>K092284 Lipisystems AquaVage<br>(Product Code MUU)<br>21CFR 880.6960 Irrigation Syringe Classification<br>(Product Code KYZ) | {4}------------------------------------------------ ## Indications for Use: The Red Head and Jordy Connection System is indicated for use as a liposuction adipose (fatty) collection system used in the aesthetic body contouring and collection of autologous adipose tissue. The Red Head and Jordy Connection System is intended to be used in the following surgical procedures: - Aspiration of adipose (fatty) tissue - Harvesting adipose (fatty) tissue - Filtering adipose (fatty) tissue - Extraction of autologous adipose (fatty) tissue # Device Description: Device Functions: - 1. Single Use closed loop system collection device (same) - 2. Canister contains ports on lids for interface between: Jordy Connection System to Canister and aspirator pump. - 3. Connection system to connect tubing to interfaces - 4. Funnel to enable fat to separate from fluids - 5. Channel for fluid evacuation. - Toomey Syringe to extract fat from device. 6. ## Device Design: - 1. Owner holds patent to predicate device and has improved performance and safety through this design. - 2. Non-Sterile, Single Use - 3. Canister able to withstand 30 in/Hg (18 in/Hg maximum vacuum used for collection of adipose (fatty tissue) for harvesting. - 4. Tissue enters canister through collection port on canister lid. - 5. Funnel perforation large enough for only adipose (fatty) tissue to be collected - 6. Waste material is removed from canister by closing valve on center of lid. - 7. Remaining adipose (fatty) tissue can be withdrawn through tissue port on bottom of center of canister for autologous adipose (fatty) tissue extraction. # Physical Properties and Materials Used: - 1. Canister, Funnel and Outer Lid made from Polycarbonate - 2. Evacuation spout and pinch clamps made from Polypropylene - 3. Tubing made from silicone # Intended use of Device: For use in aspirating subcutaneous fatty tissue including autologous fat collection. {5}------------------------------------------------ # Predicate and Reference device comparison table | Device Name | The Red Head<br>Miami Fat Supply | Medical Device ResourceCorp.<br>Lipisystems Aquavage<br>Predicate Device | |---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device<br>Description | The Red Head is a single use, closed loop<br>tissue collection device comprised of a<br>medical grade canister, vacuum port,<br>collection port, tissue port and cap<br>intended to be used with standard<br>liposuction aspiration to collect fatty<br>tissue for aesthetic body contouring. As<br>the tissue is harvested from the patient it<br>enters the canister through the collection<br>port on the canister lid. The physician<br>removes waste materials from the canister<br>by closing the valve on the center of the lid.<br>The fatty tissue remains that can be<br>withdrawn through the tissue port on the<br>bottom center of the canister for<br>autologous fat re-injection. | The Lipisystems Aquavage consists of a<br>plastic canister, silicone tubing, with a<br>vacuum port, collection port and bottom<br>tissue port and lid, intended to be used<br>with a standard liposuction pump to<br>collect fatty tissue for aesthetic body<br>contouring. As the tissue is harvested<br>from the patient it enters the canister via<br>the port in the canister lid. The physician<br>removes unwanted waste materials from<br>the collection system via the vacuum port<br>by closing the valve on the lid. This<br>process leaves fatty tissue that can be<br>transferred to syringes via the tissue port<br>for autologous fat reinjection. | | Picture of<br>Product | Image: The Red Head product | Image: Lipisystems Aquavage product | | Intended Use: | The Red Head is used in the aspiration,<br>harvesting, filtering and extraction of<br>autologous adipose tissue for aesthetic<br>body contouring | The Aquavage is used in the aspiration,<br>harvesting, filtering and extraction of<br>autologous adipose tissue for aesthetic<br>body contouring | | Technology<br>Comparison | The Red Head employs the same<br>technological characteristics as the<br>predicate device. | The Lipisystems Aquavage employs the<br>same technological characteristics as the<br>predicate device. | | Suction<br>Source | Aspiration Device | Aspiration Device | | Volume | Up to 2500 mL | 2000 cc or 1200 cc | | Range | | | | Shipped Sterile | No, sterilized by user prior to use | Yes | | Sterility Assurance Level | $10^{-6}$ | $10^{-6}$ | | Disposable or Reusable | Single Use, Disposable | Single Use Disposable | | Resterilization Method | Not Applicable | Not Applicable | {6}------------------------------------------------ ### Substantial Equivalence Discussion: The Subject and Predicate device are identical in terms of function and performance. Performance testing has demonstrated that the subject device operates as designed and functions as intended. Performance tests include Assembly Verification, Leak Testing, Canister Implosion, Tubing Collapse and Pull Off Force Testing. Biocompatibility was evaluated in accordance with ISO 10993 - Biological Evaluation of Medical Devices. Tests for cytotoxicity, intracutaneous reactivity, sensitization, and acute systemic toxicity were completed and passed. The subject device of this submission is indicated for collection and transfer of autologous adipose tissue. Tissue viability testing found no adverse effect on adipose tissue with use of the subject device. The minimally manipulated nature of such tissue in the subject device and the predicate device form the basis of substantial equivalence. ### Conclusion for the Substantial Equivalence of the Device: Based on the performance and comparison data contained within this submission, Miami Fat Supply deems that the Red Head and Jordy Connection System is substantially equivalent to the identified predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...