K123133 · Advanced Medical Solutions (Plymouth), Ltd. · MPN · May 22, 2013 · General, Plastic Surgery
Device Facts
Record ID
K123133
Device Name
BARLE TISSUE ADHESIVE
Applicant
Advanced Medical Solutions (Plymouth), Ltd.
Product Code
MPN · General, Plastic Surgery
Decision Date
May 22, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4010
Device Class
Class 2
Attributes
Therapeutic
Intended Use
Barle Tissue Adhesive topical skin adhesive is intended for topical applications only, to hold closed easily approximated skin edges of wounds from surgical incisions, including punctures from minimally invasive surgery and simple, thoroughly cleansed, trauma induced lacerations. Barle Tissue Adhesive topical skin adhesive may be used in conjunction with, but not in place of, deep dermal stitches.
Device Story
Barle Tissue Adhesive is a sterile, single-patient-use topical skin adhesive containing 2-octyl cyanoacrylate. The device consists of a crushable glass ampoule housed within a polypropylene applicator. Clinicians apply the adhesive by crushing the ampoule via the applicator 'wings' and dispensing the liquid onto approximated skin edges. The adhesive polymerizes within minutes to bond skin edges, maintaining wound closure. Used in clinical settings for surgical incisions and trauma-induced lacerations; intended for topical application only. The device provides a mechanical barrier to hold wound edges together, facilitating healing. It is not a replacement for deep dermal sutures.
Clinical Evidence
Bench testing only. Performance established via comparative testing against predicates: tensile strength (ASTM F2255-05, F2258-05, F2458-05, F2256-05), degradation by hydrolysis, heat of polymerization, ease of actuation, polymerization setting time, purity analysis, moisture content, and porcine wound healing. Biocompatibility testing (cytotoxicity, irritation, sensitization, acute dermal toxicity, intramuscular implantation) performed per ISO 10993 and FDA Class II Special Controls guidance.
Technological Characteristics
Liquid 2-octyl cyanoacrylate adhesive in a polypropylene applicator with a crushable glass ampoule. Sterile, single-use. Sterilization: dry heat (adhesive) and ethylene oxide (applicator/packaging) to SAL 10^-6. Complies with ISO 10993 for biocompatibility and ASTM standards for tensile strength (F2255-05, F2258-05, F2458-05, F2256-05).
Indications for Use
Indicated for topical closure of easily approximated skin edges of surgical incisions (including minimally invasive surgery punctures) and simple, cleansed, trauma-induced lacerations. For use in conjunction with, but not replacing, deep dermal stitches.
Regulatory Classification
Identification
A tissue adhesive for the topical approximation of skin is a device intended for topical closure of surgical incisions, including laparoscopic incisions, and simple traumatic lacerations that have easily approximated skin edges. Tissue adhesives for the topical approximation of skin may be used in conjunction with, but not in place of, deep dermal stitches.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: “Tissue Adhesive for the Topical Approximation of Skin.” See § 878.1(e) of this chapter for the availability of this guidance document.(b)
*Tissue adhesive for non-topical use* —(1)*Identification.* A tissue adhesive for non-topical use, including adhesives intended for use in the embolization of brain arteriovenous malformation or for use in ophthalmic surgery, is a device used for adhesion of internal tissues and vessels.(2)
*Classification.* Class III (premarket approval). As of May 28, 1976, an approval under section 515 of the act is required before this device may be commercially distributed. See § 878.3 of this chapter.
{0}------------------------------------------------
K123/33 page 1 of 4
Barle Tissue Adhesive Traditional 510(k) Premarket Notification Page 5-1
#### TRADITIONAL 510(K) SUMMARY 5.0
Submitted by: Advanced Medical Solutions (Plymouth) Ltd. Western Wood Way Langage Science Park Plymouth, Devon, UK PL7 5BG Tel: United Kingdom + 44 (0)1752 209955 Contact Person: Michael Browne Quality and Regulatory Affairs Manager Advanced Medical Solutions (Plymouth) Ltd 20th May 2013 Date of Summary: Device Trade Name: Barle Tissue Adhesive MPN Product Codes: Common or Usual Name: Topical Skin Adhesive Tissue Adhesive (21 CFR 878.4010) Classification Name: LiquiBand® Ultima (K100284) Predicate Device(s): Dermabond Nx (K100423) Dermabond (P960052) SurgiSeal (K082993)
Derma+Flex QS (K101276)
-
Device Description:
Indication for Use:
Barle Tissue Adhesive is a sterile, topical tissue adhesive containing 2-octyl cyanoacrylate for wound closure. Barle Tissue Adhesive is supplied in a single patient use configuration. The applicator is composed of a crushable glass ampoule contained within a plastic polypropylene applicator. The ampoule is crushed through force applied by the clinician to the 'wings' of the applicator body. It is applied to easily approximated skin edges and polymerizes within minutes. The device is contained within a PET/tyvek blister.
Barle Tissue Adhesive topical skin adhesive is intended for topical applications only, to hold closed easily approximated skin edges of wounds from surgical incisions, including punctures from minimally invasive surgery and simple, thoroughly cleansed, trauma induced lacerations. Barle Tissue Adhesive topical skin adhesive may be used in conjunction with, but not in place of, deep dermal stitches.
## MAY 2 2 2013
{1}------------------------------------------------
K123133 page 2 of 4
#### Technological Characteristics
The technological characteristics of Barle Tissue Adhesive are substantially equivalent to the predicate devices.
Barle Tissue Adhesive consists of a liquid topical skin adhesive formulation packaged within a dispensing applicator. The device is supplied in a sterile single use package for use in wound closure procedures. Barle Tissue Adhesive design allows for precise application of the adhesive to the wound area. The topical skin adhesive is designed to bond to the skin to provide wound closure maintaining wound approximation.
The main differences between Barle Tissue Adhesive and the LiquiBand Ultima predicate are
1 - change in formulation from n-butyl cyanoacrylate to 2-octyl cyanoacrylate
2 - Modifications to the LiquiBand® Ultima applicator design to use only one tip
{2}------------------------------------------------
Page 5-3
## Substantial Equivalence:
Barle Tissue Adhesive is substantially equivalent to Dermabond Nx (K100423) with regard to Indication For Use, formulation, technology, target population, intended application, mechanism of action and performance at achieving its intended use and LiquiBand Ultima Topical Skin Adhesive (K100284) with regard to Indication For Use, applicator technology, target population, intended application, mechanism of action and performance at achieving its intended use.
#### Biocompatibility
The biological evaluation of Barle Tissue Adhesive has been performed in accordance with ISO 10993 and the FDA-modified 'Use of International Standard ISO 10993, Biological Evaluation of Medical Devices Part-1: Evaluation and Testing, for breached or compromised surface with blood contact for the wound closure and subsequent layer adhesives' and the Special Controls document 'Guidance for Industry and FDA Staff: Class II Special Controls Guidance Document: Tissue Adhesive for the Topical Approximation of Skin' May 30th 2008
The biocompatibility tests were conducted for a "breached or compromised surface with blood contact device with prolonged contact duration of greater than 24 hours but less than 30 days" since the adhesive is applied to a wound and allowed to dry. All of the testing was performed in accordance with ISO 10993 and using Good Laboratory Practices (GLP).
The results provide evidence that Barle Tissue Adhesive is safe and biocompatible for its intended use and therefore substantially equivalent to the predicate devices.
### Sterilisation and Shelf Life
Sterilisation of Barle Tissue Adhesive is the same as the predicate LiquiBand® Ultima (K100284). Sterilisation is carried out to a SAL 10° by dry heat for the adhesive and ethylene oxide for the applicator and packaging.
Real time and accelerated stability testing data has been generated to support this submission.
{3}------------------------------------------------
Page 5-4
Substantial Equivalence Testing Summary:
The following comparative testing demonstrated substantially equivalent performance between Barle Tissue Adhesive, Dermabond Nx and LiquiBand Ultima;
- Tensile strength (ASTM F2255-05, F2258-05, F2458-05, . F2256-05)
- Degradation by hydrolysis I
- . Heat of polymerization
- 트 Ease of actuation
- Polymerization setting time l
- Purity analysis .
- Moisture content I
- Porcine wound healing l
- Biocompatibility testing; cytotoxicity, irritation, sensitization, 트 acute dermal toxicity and intramuscular implantation
Conclusion
Based on the nonclinical testing carried out Barle Tissue Adhesive is considered as safe, as effective and performs as well or better than the legally marketed predicate devices -Dermabond Nx and LiquiBand Ultima.
Barle Tissue Adhesive was evaluated in tests to establish a performance and safety profile in accordance with the Class II Special Controls Guidance Document: Tissue Adhesive for Topical Approximation of Skin.
Advanced Medical Solutions (Plymouth), Ltd.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three stylized, curved lines.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
Advanced Medical Solutions (Plymouth) Ltd. % Mr. Michael Browne Western Wood Way Langage Science Park Plymouth, Devon United Kingdom PL7 5BG
Re: K123133
Trade/Device Name: Barle Tissue Adhesive Regulation Number: 21 CFR 878.4010 Regulation Name: Tissue adhesive Regulatory Class: Class II Product Code: MPN Dated: April 25, 2013 Received: May 02, 2013
#### Dear Mr. Browne:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations; Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination.does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
May 22, 2013
{5}------------------------------------------------
Page 2 - Mr. Michael Browne
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours, For
# Peter DERRumm -S
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Page 4-1
0244
#### INDICATIONS FOR USE STATEMENT 4.0
K123133 510(k) Number:
Device Name: Barle Tissue Adhesive
Model Number: 000-579
Indications For Use: Barle Tissue Adhesive topical skin adhesive is intended for topical applications only, to hold closed easily approximated skin edges of wounds from surgical incisions, including punctures from minimally invasive surgery and simple, thoroughly cleansed, trauma induced lacerations. Barle Tissue Adhesive topical skin adhesive may be used in conjunction with, but not in place of, deep dermal stitches.
Prescription Use: YES (Part 21 CFR 801 Subpart D) AND/OR
Over-the-Counter Use: NO (21 CFR 801 Subpart C)
K123133
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
JiyoungDang -S
(Division Sign-Off) Division.of Surgical Devices 510(k) Number: K123133
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.