Cosmedico Ultraviolet Lamps of Various Models

K152273 · Cosmedico Light, Inc. · LEJ · Oct 30, 2015 · General, Plastic Surgery

Device Facts

Record IDK152273
Device NameCosmedico Ultraviolet Lamps of Various Models
ApplicantCosmedico Light, Inc.
Product CodeLEJ · General, Plastic Surgery
Decision DateOct 30, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4635
Device ClassClass 2

Intended Use

This ultraviolet lamp is intended for use in sunlamp products for tanning of the human skin.

Device Story

Low-pressure, mercury-rare gas discharge lamps (fluorescent-type) used in sunlamp products. Device consists of tubular glass envelope coated with fluorescent phosphor, containing inert gas and mercury drop, with electrodes at each end. When connected to electrical source and ballast, free electrons accelerate through mercury vapor, causing mercury atom excitation and photon emission. Photons strike phosphor coating, which emits radiation primarily in UVA range (320-400nm) with small proportion of UVB (260-320nm). Used in tanning beds/booths; operated by consumers. Output is UV radiation for skin tanning. Benefits include cosmetic tanning. No software or AI/ML components.

Clinical Evidence

Bench testing only. No clinical data provided. Performance testing included functionality, electrical characteristics, mechanical safety, dimensional integrity, and UV radiation characteristics (UVA irradiance, erythemal effective irradiance, and UVC/UVB ratio per 21 CFR 1040.20).

Technological Characteristics

Low-pressure mercury-rare gas discharge lamp. Components: tubular glass envelope, fluorescent phosphor coating, electrodes/cathodes, inert fill gas, mercury, and electrical bases. Operates via ballast-controlled electrical discharge. Spectral output: primarily UVA (320-400nm) with minor UVB (260-320nm). Conforms to 21 CFR 1040.20.

Indications for Use

Indicated for tanning of human skin in sunlamp products. No specific age or gender restrictions provided; intended for over-the-counter use.

Regulatory Classification

Identification

A sunlamp product is any device designed to incorporate one or more ultraviolet (UV) lamps intended for irradiation of any part of the living human body, by UV radiation with wavelengths in air between 200 and 400 nanometers, to induce skin tanning. This definition includes tanning beds and tanning booths. A UV lamp intended for use in sunlamp products is any lamp that produces UV radiation in the wavelength interval of 200 to 400 nanometers in air.

Special Controls

*Classification.* Class II (special controls). The special controls for sunlamp products and UV lamps intended for use in sunlamp products are:(1) Conduct performance testing that demonstrates the following: (i) Device meets appropriate output performance specifications such as wavelengths, energy density, and lamp life; and (ii) Device's safety features, such as timers to limit UV exposure and alarms, function properly. (2) Demonstrate that device is mechanically safe to prevent user injury. (3) Demonstrate software verification, validation, and hazard analysis. (4) Demonstrate that device is biocompatible. (5) Demonstrate that device is electrically safe and electromagnetically compatible in its intended use environment. (6) *Labeling* —(i)*Sunlamp products.* (A) The warning statement below must appear on all sunlamp products and must be placed in a black box. This statement must be permanently affixed or inscribed on the product when fully assembled for use so as to be legible and readily accessible to view by the person who will be exposed to UV radiation immediately before the use of the product. It shall be of sufficient durability to remain legible throughout the expected lifetime of the product. It shall appear on a part or panel displayed prominently under normal conditions of use so that it is readily accessible to view whether the tanning bed canopy (or tanning booth door) is open or closed when the person who will be exposed approaches the equipment and the text shall be at least 10 millimeters (height). Labeling on the device must include the following statement: Attention: This sunlamp product should not be used on persons under the age of 18 years. (B) Manufacturers shall provide validated instructions on cleaning and disinfection of sunlamp products between uses in the user instructions. (ii) *Sunlamp products and UV lamps intended for use in sunlamp products.* Manufacturers of sunlamp products and UV lamps intended for use in sunlamp products shall provide or cause to be provided in the user instructions, as well as all consumer-directed catalogs, specification sheets, descriptive brochures, and Web pages in which sunlamp products or UV lamps intended for use in sunlamp products are offered for sale, the following contraindication and warning statements:(A) “Contraindication: This product is contraindicated for use on persons under the age of 18 years.” (B) “Contraindication: This product must not be used if skin lesions or open wounds are present.” (C) “Warning: This product should not be used on individuals who have had skin cancer or have a family history of skin cancer.” (D) “Warning: Persons repeatedly exposed to UV radiation should be regularly evaluated for skin cancer.” (c) *Performance standard.* Sunlamp products and UV lamps intended for use in sunlamp products are subject to the electronic product performance standard at § 1040.20 of this chapter.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, with flowing lines suggesting hair or clothing. The profiles are arranged in a way that creates a sense of depth and unity. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 October 30, 2015 Cosmedico Light Incorporated Mr. Steven C. Schlitt Director Engineering & Quality 233 Libbey Industrial Parkway Weymouth, Massachusetts 02189 Re: K152273 Trade/Device Name: Cosmedico Ultraviolet Lamps of Various Models Regulation Number: 21 CFR 878.4635 Regulation Name: Sunlamp products and ultraviolet lamps intended for use in sunlamp products Regulatory Class: Class II Product Code: LEJ Dated: August 15, 2015 Received: August 19, 2015 Dear Mr. Schlitt: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR [SELECT ONE: Part 801 [or, for IVDs only] Parts 801 and 809]); medical device reporting (reporting of medical device-related adverse events) (21 CFR {1}------------------------------------------------ 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Joshua C. Nipper -S For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K152273 Not used. #### Device Name Cosmedico Ultraviolet Lamps of Various Models (intended for use in Sunlamp Products and commonly referred to as "sunlamps") Indications for Use (Describe) INTENDED USE: This ultraviolet lamp is intended for use in sunlamp products for tanning of the human skin. | Type of Use (Select one or both, as applicable) | | | |---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------|------------------------------------------------------------------------| | <table style="border:none;"><tr><td style="border:none;"><input type="checkbox"/> Research Use Only (Does not require IRB review)</td><td style="border:none;"><input checked="" type="checkbox"/> Gene Therapy (Requires IRB review)</td></tr></table> | <input type="checkbox"/> Research Use Only (Does not require IRB review) | <input checked="" type="checkbox"/> Gene Therapy (Requires IRB review) | | <input type="checkbox"/> Research Use Only (Does not require IRB review) | <input checked="" type="checkbox"/> Gene Therapy (Requires IRB review) | | | | Prescription Use (Part 21 CFR 801 Subpart D) |X | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary Per Sec. 807.92(c) ## 807.92(a)(1) | Date Prepared: | August 10, 2015/ Revised 10-30-15 | |-------------------------------|-----------------------------------------------------| | 510(k) Company / Holder Name: | Cosmedico Light, Inc. | | Contact Name: | Steven C. Schlitt | | Title: | Director Engineering & Quality | | Address: | 233 Libbey Industrial Parkway<br>Weymouth, MA 02189 | | Phone Number: | 781-331-0949 (Ext. 107) | | Fax Number: | 781-331-4766 | ## 807.92(a)(2) | Device Proprietary Name: | Cosmedico Ultraviolet Lamps of Various<br>Models<br>See attachment F-1 for list of models and part numbers. | |-----------------------------------|------------------------------------------------------------------------------------------------------------------| | Device Common or Usual Name: | Ultraviolet Sunlamps | | Indication for Use / Intended Use | INTENDED USE: This ultraviolet lamp is<br>intended for use in sunlamp products for<br>tanning of the human skin. | | Classification Name: | Sunlamp products and ultraviolet lamps<br>intended for use in sunlamp products | | Classification Code: | LEJ | | Regulation Number: | 21 CFR Part 878.4635 | | Device Classification | II | {4}------------------------------------------------ #### 807.92(a)(3) #### Predicate Devices: Prior to September 2, 2014. the basic ultraviolet sunlamps named in this 510(k) existed and were offered for sale as legally marketed, Class I, 510(k) exempt, medical devices. These devices are historically Over-the-Counter devices. For the basis of this 510(k), Cosmedico Light claims substantial equivalence to these legally marketed OTC devices. Since none of the ultraviolet lamps named in this 510(k) have changed in 1) intended use or 2) technological design characteristics after September 2, 2014, essentially the contemporary ultraviolet lamps are identical to the predicate devices (i.e., those basic sunlamps offered for sale prior to the FDA's cutoff date of Sept. 2, 2014). Such an approach of using the legally marketed Class I medical devices as predicate devices is validated in the Final Reclassification Order published on June 2, 2014 in the Federal Register Vo. 79, No. 105, Page 31212 whereby it is stated that: " FDA cleared several 510(k)s for sunlamp products prior to exempting the devices from premarket notification submission. At least one 510(k) for a sunlamp product has been cleared since then under product code LEJ. These cleared sunlamp products, as well as any 510(k)-exempt sunlamp product or UV lamp intended for use in a sunlamp product legally offered for sale on or before September 2, 2014, can serve as predicates for substantial equivalence purposes." lt will be demonstrated in this 510(k) that the named ultraviolet lamps: - 1) have the same intended use as the predicate OTC devices - 2) have the same technological characteristics as the predicate devices and - 3) conform to the special controls required by the reclassification order In short, the contemporary ultraviolet sunlamps of this 510(k) are as safe and as effective as the predicate devices. #### Description of the Devices: This 510(k) applies to more than one hundred and sixty individual devices that fall into a generic class of ultraviolet lamps that have the same basic technological features and exactly the same intended use. The ultraviolet sunlamps of this 510(k) are classified scientifically as low pressure, mercury-rare gas discharge devices. The general lamp construction and technological principle of operation of all of the devices of this 510(k) is the same as that of a common fluorescent lamp. The main differentiators for the various sunlamps of this 510(k) include: lamp length, lamp diameter, lamp wattage, spectral {5}------------------------------------------------ characteristics and private labeling thereof. The ultraviolet lamps named herein comprise a tubular glass envelope, the internal surface of which is coated with a fluorescent phosphor. To each end of the tube, a glass mount is sealed. The main purpose of the mount is to provide 1) for a means of hermetically sealing the internal atmosphere of the tube, 2) a means of supporting an electrode and a means of 3) electrically accessing the internal atmosphere of the device via electrically conducting lead-wires. Once assembled as described, the envelope is evacuated of air, the air is replaced by an inert gas at a pressure significantly lower than that of the normal atmosphere, and a drop of mercury is injected into the enclosed envelope. Finally, the envelope is sealed. In operation, the device, now called a "lamp", is connected to an electrical source typically in series with a ballast that provides for 1) ignition of a discharge through the inert-gas-mercury vapor mixture and 2) its subsequent stabilization and sustainment of the electrical behavior of the lamp. Once connected to the applied electric field, free electrons are accelerated through the mercury vapor wherein collisions take place with the mercury atoms. These collisions cause excitation of the electrons in the mercury (Hq) atom and the production of photons at various wavelengths. Certain of these photons travel to the bulb wall and excite the phosphor powder. The phosphor, in turn, emits radiation that is transmitted through the transparent glass envelope. The radiation that is emitted from a standard fluorescent general lighting lamp is primarily in the visible region of the spectrum (380nm-760nm). The radiation that is emitted from an ultraviolet sunlamp is primarily in the UV range of the electromagnetic spectrum (300nm-400nm). Modern day, low pressure ultraviolet sunlamps typically emit most of their energy in the wavelength range that is closest to the visible: that is, in the UVA range (320nm-400nm) range. The emission spectrum of sunlamps usually contains a relatively small proportion of UVB radiation. (260nm-320nm) Some critical design components and aspects of the sunlamps include: Glass Bulb (tubular envelope): Must be properly dimensioned and provides for transmission of ultraviolet energy in the range required for tanning. Phosphor Powder: In addition to proper physical properties such as particle size and aqueous suspension compatibility, the phosphor must emit in the wavelength range required for a particular tanning application. It is the phosphor that primarily determines the spectral irradiance distribution of the radiation from the lamp. Fill Gas and Pressure: The fill gas (in conjunction with the electrical control gear) determines the electrical characteristics of the lamp and has a large influence on the intensity of the radiation and the temporal depreciation of light output over the usage range. {6}------------------------------------------------ Coated Electrodes (Cathodes / Coils): Provide a surface area for cathode coating and to a great extent influence starting of the lamp and its physical life. Mercury: Mercury provides the basis for the operation of the lamp. It is through the excitation of mercury atoms and the subsequent quenching of the excited atoms in the vapor state that "light" is produced. It is the shortwave UV photons that resulting from the excitations that excite phosphor to produce light in a specific spectral range. Bases: Allow the lamp to be installed in appropriate sockets for electrical connectivity to the electrodes. Image /page/6/Figure/4 description: This image shows a diagram of a fluorescent lamp and its components. The diagram labels the different parts of the lamp, including the bulb, phosphor, cathode, exhaust tube, base, mercury, gas, stem press, and lead-in wires. The text provides a brief description of each component and its function within the lamp. For example, the phosphor coating inside the bulb transforms ultraviolet radiation into visible light. For illustrative purposes: **Figure 2.** Basic parts of a typical hot cathode fluorescent lamp. ### Intended Use/Indications for Use INTENDED USE: This ultraviolet lamp is intended for use in sunlamp products for tanning of the human skin. TYPE OF USE: The predicate devices named in this 510(k) are historically used as Over the Counter (OTC) devices. #### Technological Characteristics: The technological characteristics of the ultraviolet sunlamps which are the subject of this 510(k) are identically the same as the technological characteristics of the predicate devices. The tables and data that are reported in Section 10 and the information summarized in Section 12, "Substantial Equivalence Discussion" of this 510(k), document that there are no significant differences between the subject and predicate devices. In fact, the subject and the predicate devices are "one in the same". {7}------------------------------------------------ ### Performance Testing Summary: - . All lamps are 100% tested for functionality at the end of the manufacturing process. Lamps that do not meet critical characteristics for functionality and safety are rejected. - . According to a Quality Sampling Plan, lamps are tested in-process and / or post-production for: - Electrical Characteristics and Safety A - Mechanical Safety A - > Dimensional Integrity - > Physical Design Attributes - > UV radiation Characteristics, including: - UVA Irradiance - -Erythemal Effective Irradiance - . UVC/UVB ratio per 21 CFR 1040.20 As noted, the subject lamps of this 510(k) are identical to the predicate devices. Therefore, the subject lamps and the predicate devices have the same performance requirements. | Test (Performance Criteria | Lamps meet criteria of predicate<br>devices (Yes/No) | |-----------------------------------------|------------------------------------------------------| | Functionality / Light up | Yes | | Electrical Characteristics | Yes | | Burn-in Behavior | Yes | | Dimensional Characteristics | Yes | | Mechanical Safety | Yes | | Electrical Safety | Yes | | Physical Design (e.g., reflector angle) | Yes | | UV Performance | Yes | | UVA Irradiance | Yes | | Erythemal Effective Irradiance | Yes | | UVC/UVB Ratio (per 21 CFR1040.20) | Yes | | | | Lamps that do not meet Acceptance Criteria are either re-worked or scrapped. {8}------------------------------------------------ | COSMEDICO<br>PART NUMBER | LAMP DESCRIPTION | |--------------------------|------------------------------------------------------| | 10018 | FR18T8 / COSMOLUX HRF | | 10036 | F25WT5 / COSMOLUX LS | | 10044 | F15WT5 / COSMOLUX X | | 10180 | F59T12 / HI TAN / 80W | | 10186 | F71T12 / Bermuda Gold Premium Plus / 105W | | 10187 | F73T12 / Bermuda Gold / RDC / 105W | | 10198 | F71T12 / HI TAN / 100W | | 10199 | F73T12 / HI TAN / 100W / RDC | | 10201 | F73T12 / EXPRESS TAN HO / 100W / RDC | | 10202 | F73T12 / Genesis / 120W / RDC | | 10204 | F71T12 / Genesis Hyperbronze / 105W | | 10205 | F73T12 / Genesis Hyperbronze / 105W | | 10206 | F72T12 / Genesis Plus / 105W | | 10212 | F71T12 / Genesis / 120W | | 10217 | F71T12 / Eternal Sun XPFT / 100W | | 10218 | F73T12 / Eternal Sun XPFT / 100W / RDC | | 10219 | F71T12 / Eternal Sun XP Gold / 100W | | 10220 | F73T12 / Eternal Sun XP Gold / 100W / RDC | | 10222 | F59T12 / Eternal Sun XP Gold / 80W | | 10224 | F71T12 / Eternal Sun XP Gold / 120W | | 10318 | FR18T8 / Sundash HRF | | 10330 | F25T5 / UVA Plus | | 10331 | F25T5 / COSMOLUX LS | | 10332 | FR25WT5 / LSR | | 10990 | F71T12 / COSMOLUX GSP 9K90 / 100W | | 10994 | F71T12 / Eternal Sun XP Gold 9K90 / 100W | | 11282 | F59T12 / GOLDARIUM SP / 80W | | 11305 | F71T12 / GOLDARIUM SP / 100W | | 11306 | F73T12 / GOLDARIUM SP / 100W | | 12482 | F59T12 / COSMOSTAR LONGLIFE / 80W | | 12503 | F71T12 / COSMOSTAR LONGLIFE / 100W | | 12600 | F71T12 / COSMOLUX ESP / 100W | | 12601 | F73T12 / COSMOLUX ESP / 100W / RDC | | 12607 | F73T12 / Genesis GS 120W / RDC | | 15163 | FR59T12 / Eternal Sun XP*R / 80W | | 15165 | FR71T12 / Palm Beach Mega Tan S1-S2 / 100W | | 15211 | FR59T12 / COSMOLUX VLR-TT / 80W / 220° | | 15212 | FR71T12 / COSMOLUX VLR TT / 100 W | | 15214 | FR71T12 / Bermuda Gold PLUS / 100W | | 15215 | FR71T12 / COSMOSUN R / 100W | | 15216 | FR71T12 / Eternal Sun XP*R / 100W | | 15220 | FR79T12 / COSMOLUX VLR / 120W | | 15602 | FR71T12 / COSMOLUX ESP R / 100 W | | COSMEDICO<br>PART NUMBER | LAMP DESCRIPTION | | 15603 | FR73T12 / COSMOLUX ESP / 100W / RDC | | 15640 | FR71T12 / Genesis Super Power VHP15 Plus / 120W | | 15642 | FR71T12 / Genesis Super Power VHP20 Plus / 120W | | 15649 | FR71T12 / Genesis Hyperbronze / 105W | | 15656 | FR71T12 / Bermuda Gold / Supreme 20 / 120W | | 15691 | FR71T12 / /COSMOLUX VLR 9K-90 / 100W | | 15700 | FR71T12 / COSMOLUX RA PLUS / 100W | | 15704 | FR24T12 / COSMOLUX RA PLUS / 40W | | 15710 | FR71T12 / COSMOLUX RCS / 100W | | 16077 | FR73T12 / HI TAN / VHO / RDC | | 16107 | FR73T12 / COSMOLUX VHR-TT / 160W / RDC | | 16112 | FR59T12 / COSMOLUX VHR-TT / 140W | | 16121 | FR71T12 / Genesis 170WRS-E / 160W | | 16161 | FR71T12 / COSMOLUX / VHO | | 16169 | FR71T12 / COSMOLUX VHR-TT / 160W | | 16172 | F72T12 / HI TAN / VHO / RDC | | 16180 | FR71T12 / COSMOLUX VHR / 160W | | 16202 | FR75T12 / COSMOLUX VHR 200 / 1.9 M | | 16211 | FR75T12 / COSMOLUX VHR / HP / 180W | | 16218 | FR71T12 / COSMOLUX VHR-200 | | 16220 | FR79T12 / COSMOLUX VHR 200 / 2.0 M | | 16223 | FR79T12 / COSMOLUX 9K90 / 180W-200W | | 16234 | FR79T12 / COSMOLUX 9K90 / 225W | | 16236 | FR71T12 / COSMOSUN R / 160W | | 16238 | FR79T12 / Ergoline Turbo Power / 200W | | 16239 | FR74T12 / COSMOLUX VHR-TT / 160-200W | | 16250 | FR79T12 / Suncapsule 250 MAX / 180-250W | | 16258 | FR79T12 / Eternal Sun XP*r / 2.0M / 180-200W | | 16259 | FR79T12 / Eternal Sun XP*R / 9K90 / 2.0M / 180-200W | | 16286 | FR79T12 / COSMOLUX VHR HP 250 MAX / 180-250W | | 16307 | FR72T12 / Genesis 170WRS / RDC | | 16317 | FR73T12 / Express Tan VR 160W / RDC | | 16332 | FR71T12 / COSMOLUX VHR / 200 / HP | | 16341 | FR71T12 / Genesis Turbo Power VHP20 PLUS / 160W | | 16344 | FR71T12 / Genesis Turbo Power VHP15 PLUS / 160W | | 16367 | FR71T12 / Bermuda Gold Premium Plus / 160W | | 16385 | FR71T12 / Ergoline Turbo Power VHP12 / 160W | | 16386 | FR72T12 / Genesis GS 170W / RDC | | 16388 | FR71T12 / HI-TAN / VHR / 160W | | 16391 | FR71T12 / COSMOLUX VHR / 9K90 / 160W | | 16392 | FR71T12 / Genesis 9K90 / 160WR | | 16393 | FR71T12 / COSMOLUX VHR FLEX / 160-200W | | 16394 | FR71T12 / COSMOLUX VHR-XT 160W | | COSMEDICO<br>PART NUMBER | LAMP DESCRIPTION | | 16395 | FR71T12 / COSMOLUX VHR / HP / 160W | | 16398 | FR71T12 / Genesis Hyperbronze / 170-200W | | 16399 | FR71T12 / Suncapsule SCR 160 / 160W | | 16445 | FR71T12 / Eternal Sun XP*R / 160W | | 16447 | FR71T12 / Eternal Sun XP*R 9K90 /160W | | 16459 | FR59T12 / Eternal Sun XP*R / 140W | | 16481 | FR71T12 / Eternal Sun XP / 160W | | 16542 | FR71T12 Eternal Sun XP HR 160W | | 18810 | FR71T12 / COSMOLUX VLR.T / 100 W | | 21069 | F25T5 / Radiance / 25W (07040) | | 21248 | F72T12 / Radiance 8000 7.5 160W RDC (03017) | | 21249 | F74T12 / Radiance 8.5 / 100-120W (07034) | | 21250 | F72T12 / Radiance 7000 6.5 / 100-120W (03016) | | 21626 | FR18T8 / Radiance HRF (07041) | | 22010 | FR79T12 / Genesis VHP 20 / 120W | | 22015 | FR79T12 / Genesis VHP 15 / 120W | | 22300 | F73T12 / Radiance 120 / 120W / RDC (07044) | | 22302 | F73T12 / Eternal Sun XP Gold / 120W | | 26111 | F73T12 / FABUTAN /100W / RDC | | 26113 | FR71T12 / Radiance 71i / 120W | | 26114 | FR75T12 / Radiance 75i / 120W | | 26180 | F71T12 / Radiance JAGUAR 10.0 / 100W (61747) | | 26189 | FR71T12 / Silver Sun / 100W R | | 26190 | F71T12 / Radiance 9000 8.5 / 100W (03011) | | 26191 | F71T12 / Radiance 5K 5.0 / 100W (03001) | | 26192 | F73T12 / Radiance 7000 RDC 6.5 / 100W (03006) | | 26193 | F71T12 / Radiance 8000 7.5 / 100W (03009) | | 26194 | F71T12 / Radiance PLUS 7000 6.5 / 100-120W (07011) | | 26195 | F73T12 / Radiance 9000 8.5 / 100-120W / RDC (03012) | | 26196 | FR73T12 / Radiance 7000 6.5 / 100W / RDC (03008) | | 26197 | F73T12 / Millennium 7000 6.5 / 100W (03027) | | 26198 | F73T12 / Radiance Plus 7000 6.5 / 100-120W (07013) | | 26199 | F59T12 / Radiance 5000 / 5.0 / 80W (03000) | | 26202 | F73T12 / Radiance +5000 5.0 / 100W (07003) | | 26203 | FR73T12 / Radiance+5000 5.0 / 100-120W / RDC (07006) | | 26267 | F73T12 / Radiance 8000 7.5 / 100-120W | | 26303 | F71T12 / California Sun Ultra / 105W | | 26304 | FR71T12 / California Sun Mega / 120W | | 26305 | FR71T12 / California Sun Premier / 170W | | 26308 | F71T12 / Radiance Black / 100W (03076) | | 26309 | F73T12 / Radiance Black / 100-120W / RDC (03076) | | 26673 | F73T12 / Radiance 8000 7.5 / 160W RDC (03018) | | 26682 | FR71T12 / Radiance R2 XLP 1.7 / 100-120W (03100) | | COSMEDICO<br>PART NUMBER | LAMP DESCRIPTION | | 26774 | FR71T12 / SilverSun / 160WR | | 27355 | FR79T12 / Maxlight e-power XL / 200W-R | | 27450 | FR71T12 / Radiance R2 / 100-120W / HOR 3.7 (07043) | | 27451 | FR71T12 / Radiance R2 VHO R / 160-200W (07051) | | 27483 | FR72T12 / Radiance R2 VHOR / 160-200W (07052) | | 27524 | F73T12 / Radiance 5000 5.0 / RDC / 100W (03002) | | 27525 | FR71T12 / Radiance 5000 5.0 / 100-120W (03003) | | 27526 | F71T12 / Radiance 7000 6.5 / 100-120W (03005) | | 27527 | FR71T12 / Radiance 7000 6.5 / 100-120W (03007) | | 27528 | FR71T12 / Radiance Plus+ 7000 6.5 / 100-120W (07015) | | 27529 | FR59T12 / Radiance R2 / 140W (07050) | | 27530 | FR73T12 / Radiance R2 / 160-200W / RDC (07053) | | 27531 | F59T12 / Radiance 7000 6.5 / 80W (03014) | | 27532 | FR59T12 / Radiance Plus+ 6.5 / 80W (07014) | | 27533 | FR59T12 / Radiance R2 3.7 / 80W (07042) | | 27534 | FR71T12 / Palm Beach Mega Tan / 160W | | 27535 | FR59T12 / Premium Plus Bermuda Gold Plus / 140W | | 27538 | FR71T12 / Sundazzler Premium / 160W | | 27540 | FR2.0M / Genesis GSXR 200 / 180W | | 27541 | FR71T12 / Radiance R2 XLP 7.0 / 160W R (03098) | | 27548 | FR79T12 / Radiance R2 / 2M / 120W (03081) | | 27574 | FR74T12 / Eternal Sun XP*R / 200W / RDC | | 27712 | FR73T12 / Radiance R2 3.7 / 100-120W (07045) | | 27728 | F71T12 / Millennium 5000 5.0 /100W (03021) | | 27733 | FR72T12 / RadianceR2 VIP / FR71-RDC / 160W (07066) | | 27736 | FR71T12 / GOLDARIUM SR / 100W | | 27748 | FR79T12 / Eternal Sun BronzeR / 225W | | 27749 | FR71T12 / Eternal Sun XP20R / 120W | | 27751 | FR71T12 / Radiance Black / 100-120W (07075) | | 27756 | FR71T12 / Radiance Black / 160W (07076) | | 27757 | FR71T12 / Bermuda Gold / 100 W | | 28018 | FR79T12 / Radiance R2 2M / 200W (03080) | | 28019 | FR74T12 Radiance R3 VHOR / RDC / 200W (07048) | | 28021 | F71T12/ Millennium 7000 6.5 / 100W (03026) | | 28022 | FR71T12 / Radiance R2 Plus / 200W (07059) | | 28996 | FR71T12 Millennium M2 3.15 / 160-180W (03087) | | 901935140 | FR71T12 / Ergoline Super Power SR / 100W | | 926295040 | FR71T12 / CLEO ADVANTAGE XPT / 160W-R | | 971036940 | FR71T12 / CLEO ADVANTAGE / 100W-R | {9}------------------------------------------------ {10}------------------------------------------------ {11}------------------------------------------------
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...