ENCORE Latex Ortho, Powder Free Latex Surgical Gloves with Protein Content Labeling Claim (<50µg/dm2 per glove of total aqueous extractable protein) and Tested for use with Chemotherapy drugs, GAMMEX Latex Ortho, Powder Free Latex Surgical Gloves with Protein Content Labeling Claim (<50 µg/dm2 per glove of total aqueous extractable protein) and Tested for use with Chemotherapy drugs

K210656 · Careplus (M) Sdn Bhd · KGO · Jun 24, 2021 · General, Plastic Surgery

Device Facts

Record IDK210656
Device NameENCORE Latex Ortho, Powder Free Latex Surgical Gloves with Protein Content Labeling Claim (<50µg/dm2 per glove of total aqueous extractable protein) and Tested for use with Chemotherapy drugs, GAMMEX Latex Ortho, Powder Free Latex Surgical Gloves with Protein Content Labeling Claim (<50 µg/dm2 per glove of total aqueous extractable protein) and Tested for use with Chemotherapy drugs
ApplicantCareplus (M) Sdn Bhd
Product CodeKGO · General, Plastic Surgery
Decision DateJun 24, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4460
Device ClassClass 1

Indications for Use

Powder Free Surgical gloves are sterile disposable devices intended to be worn by operating room personnel to protect a surgical wound from contamination.

Device Story

Sterile, powder-free latex surgical gloves; intended for use by operating room personnel to protect surgical wounds from contamination. Gloves provide a barrier against patient fluids and microorganisms. Tested for permeation resistance against specific chemotherapy drugs per ASTM D6978. Output is a physical barrier; used in surgical settings to prevent cross-contamination between patient and clinician. Benefits include reduced risk of surgical site infection and protection of healthcare workers from hazardous chemotherapy agents.

Clinical Evidence

Bench testing only. Testing included ASTM D6978 standard practice for assessment of medical gloves to permeation by chemotherapy drugs, measuring breakthrough detection times for various agents.

Technological Characteristics

Material: Latex. Form factor: Powder-free surgical gloves. Energy source: None. Connectivity: None. Sterilization: Sterile. Standards: ASTM D6978 for chemotherapy drug permeation.

Indications for Use

Indicated for operating room personnel to protect surgical wounds from contamination. Tested for use with specific chemotherapy drugs.

Regulatory Classification

Identification

A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. June 24, 2021 Careplus (M) SDN BHD Lim Shyan CEO/Managing Director Lot 120 & 121, Jalan Senawang 3, Senawang Industrial Estate Seremban, Negeri Sembilan 70450 Malaysia Re: K210656 Trade/Device Name: ENCORE Latex Ortho, Powder Free Latex Surgical Gloves with Protein Content Labeling Claim (<50ug/dm2 per glove of total aqueous extractable protein) and Tested for use with Chemotherapy drugs GAMMEX Latex Ortho, Powder Free Latex Surgical Gloves with Protein Content Labeling Claim (<50 ug/dm2 per glove of total aqueous extractable protein) and Tested for use with Chemotherapy drugs Regulation Number: 21 CFR 878.4460 Regulation Name: Non-Powdered Surgeon's Glove Regulatory Class: Class I, reserved Product Code: KGO Dated: May 25, 2021 Received: May 27, 2021 Dear Lim Shyan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. {1}------------------------------------------------ If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809 ); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products): good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. For Clarence W. Murray III. Ph.D. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health {2}------------------------------------------------ ## Indications for Use 510(k) Number k210656 ## Device Name GAMMEX® Latex Ortho, Powder free Latex Surgical Gloves with Protein Content Labeling Claim (< 50 ug/dm2 per glove of total aqueous extractable protein) and Tested for use with Chemotherapy drugs. ENCORE® Latex Ortho, Powder free Latex Surgical Gloves with Protein Content Labeling Claim (< 50 ug/dm2 per glove of total aqueous extractable protein) and Tested for use with Chemotherapy drugs Indications for Use (Describe) Powder Free Surgical gloves are sterile disposable devices intended to be worn by operating room personnel to protect a surgical wound from contamination. These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy: | Test Chemotherapy Drug (Concentration) | Minimum Breakthrough Detection Time in<br>minutes | |----------------------------------------|---------------------------------------------------| | Fluorouracil (50.0mg/ml) | >240 | | Etoposide (20.0mg/ml) | >240 | | Cyclophosphamide (20.0mg/ml) | >240 | | Carmustine (3.3mg/ml) | 22.0 | | Thiotepa (10.0mg/ml) | 24.2 | | Paclitaxel (6.0mg/ml) | >240 | | Doxorubicin Hydrochloride (2.0mg/ml) | >240 | | Methotrexate (25.0mg/ml) | >240 | | Vincristine Sulfate (1.0mg/ml) | >24 | "WARNING: Do not use with Carmustine and Thiotepa" Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: {3}------------------------------------------------ Department of Health and Human Services Food and Drug Administration
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...