Finessis Zero Flexylon Powder Free Sterile White Surgical Glove Tested for use with Chemotherapy Drugs

K173386 · Lucenxia Prescience AG · KGO · Apr 9, 2018 · General, Plastic Surgery

Device Facts

Record IDK173386
Device NameFinessis Zero Flexylon Powder Free Sterile White Surgical Glove Tested for use with Chemotherapy Drugs
ApplicantLucenxia Prescience AG
Product CodeKGO · General, Plastic Surgery
Decision DateApr 9, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4460
Device ClassClass 1

Indications for Use

This surgeon's glove is a device made of Flexylon intended to be worn by operating room personnel to protect a surgical wound from contamination. These Gloves are tested for use with Chemotherapy Drugs DO NOT USE with Carmustine or Thiotepa.

Device Story

Sterile, powder-free surgical glove composed of Flexylon; intended for use by operating room personnel to provide barrier protection for surgical wounds. Device provides resistance against permeation by various chemotherapy drugs. Tested per ASTM D6978-05; breakthrough times established for 32 chemotherapy agents. Contraindicated for use with Carmustine and Thiotepa due to insufficient breakthrough times. Used in clinical/surgical settings to protect patient and clinician from contamination.

Clinical Evidence

Bench testing only. Permeation testing performed per ASTM D6978-05 for 32 chemotherapy drugs. Results demonstrate no breakthrough up to 240 minutes for 30 drugs; breakthrough times for Carmustine and Thiotepa were identified and labeled as contraindications.

Technological Characteristics

Material: Flexylon. Form factor: Sterile, powder-free surgical glove. Testing standard: ASTM D6978-05 for chemotherapy drug permeation. No electronic, software, or energy-based components.

Indications for Use

Indicated for use by operating room personnel as a protective barrier for surgical wounds. Tested for resistance to specific chemotherapy drugs per ASTM D6978-05. Contraindicated for use with Carmustine or Thiotepa.

Regulatory Classification

Identification

A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo, which features a stylized human figure. To the right of that is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. April 9, 2018 Lucenxia Prescience AG Robert Hill Regulatory Director Rathausstrasse 7 Baar, CH6341, Switzerland Re: K173386 Trade/Device Name: Finessis Zero Flexylon Powder Free Sterile White Surgical Gloves, Tested for use with Chemotherapy Drugs Regulation Number: 21 CFR 878.4460 Regulation Name: Surgeon's Glove Regulatory Class: Class I Product Code: KGO, LZC Dated: March 1, 2018 Received: March 5, 2018 Dear Robert Hill: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). > Sincerely, Geeta K. Pamidimukkala -5 for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration #### Indications for Use 510(k) Number (if known) K173386 Device Name Finessis Zero Flexylon Powder Free Sterile White Surgical gloves Tested for use with Chemotherapy Drugs #### Indications for Use (Describe) This surgeon's glove is a device made of Flexylon intended to be worn by operating room personnel to protect a surgical wound from contamination. These Gloves are tested for use with Chemotherapy Drugs DO NOT USE with Carmustine or Thiotepa. | Tested Chemotherapy Drug<br>and Concentration per ASTM D6978-05 | Minimum BREAKTHROUGH DETECTION<br>(Specimen 1/2/3) (Minutes) | |-----------------------------------------------------------------|--------------------------------------------------------------| | 1 Carmustine (BCNU), 3.3mg/ml (3,300 ppm) | 6.9 (9.0, 12.3, 6.9) | | 2 Cisplatin, 1.0 mg/ml (1,000 ppm) | No Breakthrough up to 240 min. | | 3 Cyclophosphamide (Cytoxan), 20 mg/ml (20,000 ppm) | No Breakthrough up to 240 min. | | 4 Dacarbazine (DTIC) 10.0 mg/ml (10,000 ppm) | No Breakthrough up to 240 min. | | 5 Doxorubicin Hydrochloride, 2.0mg/ml (2,000 ppm) | No Breakthrough up to 240 min. | | 6 Etoposide (Toposar), 20.0 mg/ml (20,000 ppm) | No Breakthrough up to 240 min. | | 7 Fluorouracil, 50.0 mg/ml (50,000 ppm) | No Breakthrough up to 240 min. | | 8 Ifosfamide, 50.0 mg/ml (50,000 ppm) | No Breakthrough up to 240 min. | | 9 Methotrexate, 25 mg/ml (25,000 ppm) | No Breakthrough up to 240 min. | | 10 Mitomycin C, 0.5 mg/ml (500 ppm) | No Breakthrough up to 240 min. | | 11 Paclitaxel (Taxol), 6.0 mg/ml (6,000 ppm) | No Breakthrough up to 240 min. | | 12 Thiotepa, 10.0 mg/ml (10,000 ppm) | 7.6 (9.2, 7.6, 10.0) | | 13 Vincristine Sulfate, 1.0 mg/ml (1,000 ppm) | No Breakthrough up to 240 min. | | 14 Bleomycin, 15 mg/ml (15,000 ppm) | No Breakthrough up to 240 min. | | 15 Busulfan, 6 mg/ml (6,000 ppm) | No Breakthrough up to 240 min. | | 16 Carboplatin, 10 mg/ml (10,000 ppm) | No Breakthrough up to 240 min. | | 17 Cytarabine, 100 mg/ml (100,000 ppm) | No Breakthrough up to 240 min. | | 18 Daunorubicin, 5 mg/ml (5,000 ppm) | No Breakthrough up to 240 min. | | 19 Docetaxel, 10.0 mg/ml (10,000 ppm) | No Breakthrough up to 240 min. | | 20 Epirubicin, 2.0 mg/ml (2,000 ppm) | No Breakthrough up to 240 min. | | 21 Fludarabine, 25 mg/ml (25,000 ppm) | No Breakthrough up to 240 min. | | 22 Ganciclovir, 10 mg/ml (10,000 ppm) | No Breakthrough up to 240 min. | | 23 Gemcitabine (Gemzar), 38 mg/ml (38,000 ppm) | No Breakthrough up to 240 min. | | 24 Idarubicin, 1.0 mg/ml (1,000 ppm) | No Breakthrough up to 240 min. | | 25 Irinotecan, 20.0 mg/ml (20,000 ppm) | No Breakthrough up to 240 min. | | 26 Mechlorethamine HCI, 1.0 mg/ml (1,000 ppm) | No Breakthrough up to 240 min. | | 27 Melphalan, 5.0 mg/ml (5,000 ppm) | No Breakthrough up to 240 min. | | 28 Mitoxantrone, 2.0 mg/ml (2,000 ppm) | No Breakthrough up to 240 min. | | 29 Oxaliplatin, 2.0 mg/ml (2,000 ppm) | No Breakthrough up to 240 min. | | 30 Rituximab, 10 mg/ml (10,000 ppm) | No Breakthrough up to 240 min. | | 31 Trisenox, 0.1 mg/ml (100 ppm) | No Breakthrough up to 240 min. | | 32 Vinorelbine, 10 mg/ml (10,000 ppm) | No Breakthrough up to 240 min. | ### TIME Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below. {3}------------------------------------------------ Type of Use (Select one or both, as applicable) □ Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. # *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the The burden time for this collection of information is connation the data needed and complete time to review instructions, Search existing data Sources, gather and his hurden estimate or any other aspect and review the concollorion, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...