CAPSULAR SHRINKAGE ELECTRODE

K991830 · Linvatec Corp. · JOS · Aug 20, 1999 · General, Plastic Surgery

Device Facts

Record IDK991830
Device NameCAPSULAR SHRINKAGE ELECTRODE
ApplicantLinvatec Corp.
Product CodeJOS · General, Plastic Surgery
Decision DateAug 20, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The HeatWave Electrode is intended to be used for electro-coagulation of soft tissue in shoulder, ankle, wrist, elbow and knee arthroscopic procedures.

Device Story

HeatWave Electrode is a sterile, single-use, monopolar electrosurgical device. It connects to an external electrosurgical generator via an electrosurgical pencil and dispersive pad. Used by surgeons during arthroscopic procedures to perform soft tissue electro-coagulation. The device delivers electrical energy to the target tissue to achieve coagulation. It is provided individually or in a kit containing the electrode, pencil, and dispersive pad.

Clinical Evidence

Bench testing only.

Technological Characteristics

Monopolar electrode. Materials: 303 Stainless Steel (tip/shaft), Kynar (shrink tube), Polystyrene (hub). Dimensions: 2.3mm, 3.0mm, 4.0mm diameters; 4-8.5 inch working length. Sterilization: ETO. Energy source: External electrosurgical generator.

Indications for Use

Indicated for electro-coagulation of soft tissue during arthroscopic procedures of the shoulder, ankle, wrist, elbow, and knee.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K991830 、元 20 103 August 5, 1999 # SUMMARY OF SAFETY AND EFFECTIVENESS In accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CFR 807.92, Linvatec Corporation is hereby submitting the 510(k) Summary of Safety and Effectiveness for the HeatWave™ Electrode, 510(k) Number K991830. #### A. Submitter Linvatec Corporation 11311 Concept Boulevard Largo, Florida 33773-4908 ### Company Contact B. Laura Seneff Manager, Regulatory Affairs ### C. Device Name | Trade Name: | HeatWave™ Electrode | |----------------------------------------|-----------------------------------------------------------------------------| | Common Name: | Electrode | | Classification Names: | Electrosurgical cutting and coagulation device<br>and accessories, 878.4400 | | Proposed Class/Device:<br>Product Code | Class II, 79 JOS, Electrode,<br>Electrosurgical | ### D. Predicate/Legally Marketed Devices TAC™ Probes ORATEC Interventions, Inc. ArthoWand CAPS ArthroCare Corporation {1}------------------------------------------------ Summary of Safety and Effectiveness HeatWave Electrode 510(k) # K991830 August 5, 1999 Page 2 of 3 ### E. Device Description The HeatWave™ Electrode is a sterile, single-use, monopolar electrode designed to be used in conjunction with an electrosurgical generator via an electrosurgical pencil and a dispersive pad. The HeatWave Electrode will be sold individually and/or in a kit containing an electrode, electrosurgical pencil and an electrosurgical dispersive pad. ### F. Intended Use The HeatWave Electrode is intended to be used for electro-coagulation of soft tissue in shoulder, ankle, wrist, elbow and knee arthroscopic procedures. #### G. Substantial Equivalence The HeatWave Electrode is substantially equivalent in design, function and intended use to the TAC™ Probes (ORATEC Interventions, Inc.)and the ArthoWand CAPS (ArthroCare Corporation). Testing has been done to prove safety and effectiveness of the devices. The similarities/dissimilarities to the predicates are shown in the attached table. {2}------------------------------------------------ # Summary of Safety and Effectiveness HeatWave Electrode 510(k) # K991830 August 5, 1999 Page 3 of 3 # CHART OF SIMILARITIES AND DISSIMILARITIES | Company | Device Name | Intended Use | Material | Single-Use<br>Reusable<br>Method of<br>Sterilization | Design/Dimensions | |------------------------------------------------------------------|---------------------------------------|---------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|------------------------------------------------------|----------------------------------------------------------------------------------| | NEW PRODUCT<br>Linvatec | HeatWaveTM<br>Electrode | Electro-coagulation of soft<br>tissue in shoulder, ankle,<br>wrist, elbow and knee<br>arthroscopic procedures | Tip/Shaft: 303 Stainless Steel<br>Shrink Tube: Kynar<br>Hub: Polystyrene | Single-Use<br>ETO | Diameter: 2.3mm<br>3.0mm<br>4.0mm<br>Working length: 4-8½in.<br>Monopolar | | PREDICATE<br>ORATEC<br>Interventions, Inc.<br>510(k)#<br>K984185 | TAC-STM<br>Monopolar<br>Cautery Probe | Electro-coagulation of soft<br>tissue in shoulder, ankle,<br>wrist, elbow and knee<br>arthroscopic procedures | Tip/Shaft: 303 Stainless Steel<br>Shrink Tube: Teflon<br>Hub/Connector: ABS Plastic | Single Use<br>Sterilization<br>Method Unknown | Diameter: 2.0mm<br>Working Length: 6 in.<br>Monopolar | | PREDICATE<br>ArthroCare<br>Corporation<br>510(K) #<br>Unknown | ARTHROWAND<br>CAPS | Electro-coagulation of soft<br>tissue during arthroscopic<br>procedures | Shaft: 300 Series Stainless Steel<br>Tip: Ceramic<br>Hub/Connector: ABS Plastic | Single-Use<br>Sterilization<br>Method Unknown | Degree/Diameter:<br>25°/3.0mm<br>35°/3.0mm<br>Working length: 5.0 in.<br>Bipolar | {3}------------------------------------------------ Image /page/3/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with text around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 AUG 20 1999 Ms. Laura Seneff Manager, Regulatory Affairs Linvatec Corporation 11311 Concept Boulevard Largo, Florida 33773 K991830 Re: Trade Name: HeatWave™ Electrode Regulatory Class: II Product Code: JOS Dated: August 5, 1999 Received: August 10, 1999 Dear Ms. Seneff: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {4}------------------------------------------------ Page 2 - Ms. Laura Seneff This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely vours. Ser Plumell J. Rayer Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ August 5, 1999 Page 1 of 1 510(k) Number (if known): K991830 Device Name: HeatWave™ Electrode Indications for Use: The HeatWave Electrode is intended to be used for electro-coagulation of soft tissue in shoulder, ankle, wrist, elbow and knee arthroscopic procedures. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use OR Over-the-Counter Use (Per 21 CFR 801.109) Thunell V. Hagen for Jas (Division Sign-Off Division of General Res 510lk) Number (Optional Format 1-2-96)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...