← Product Code [GEY](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEY) · K012738

# SYMPHONY GRAFT DELIVERY SYSTEM (GDS) (K012738)

_Depuyacromed · GEY · Nov 14, 2001 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEY/K012738

## Device Facts

- **Applicant:** Depuyacromed
- **Product Code:** [GEY](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEY.md)
- **Decision Date:** Nov 14, 2001
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4820
- **Device Class:** Class 1
- **Review Panel:** General, Plastic Surgery

## Indications for Use

The Symphony Graft Delivery System is indicated for the delivery of allograft, autograft, or synthetic bone graft materials to an orthopedic surgical site. In addition, it is designed to facilitate pre-mixing of bone graft materials with I.V. fluids, blood, plasma, platelet rich plasma, bone marrow or other specific blood component(s) as deemed necessary by the clinical use requirements.

## Device Story

Symphony Graft Delivery System is a piston syringe designed for orthopedic surgical use. It collects, mixes, and delivers bone graft materials (allograft, autograft, or synthetic) to surgical sites. The device facilitates pre-mixing of graft materials with I.V. fluids, blood, plasma, platelet-rich plasma, or bone marrow. The graft chamber can be provided empty or pre-filled with human allograft. Operated by surgeons in an orthopedic surgical setting to improve graft handling and delivery.

## Clinical Evidence

Bench testing only. Biocompatibility testing performed per Tripartite Biocompatibility Guidance, Pharmacopeia XXII Class VI, and FDA modified ISO 10993-1 for human body fluid contact of 30 days or less.

## Technological Characteristics

Piston syringe design. Materials: medical grade polycarbonate resin and medical grade high density polyethylene. Biocompatibility: meets Pharmacopeia XXII Class VI and ISO 10993-1 standards.

## Regulatory Identification

Surgical instrument motors and accessories are AC-powered, battery-powered, or air-powered devices intended for use during surgical procedures to provide power to operate various accessories or attachments to cut hard tissue or bone and soft tissue. Accessories or attachments may include a bur, chisel (osteotome), dermabrasion brush, dermatome, drill bit, hammerhead, pin driver, and saw blade.

## Predicate Devices

- Symphony Graft Delivery System ([K003286](/device/K003286.md))
- Symphony Graft Delivery System ([K010320](/device/K010320.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

| VIII. 510(k) Summary |                                                                                                                                                                                                                                                                                                                                                                                                                          | K012738      |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| SUBMITTER:           | DePuy AcroMed, Inc.<br>325 Paramount Drive<br>Raynham, MA 02780                                                                                                                                                                                                                                                                                                                                                          | NOV 1 4 2001 |
| CONTACT PERSON:      | Karen F. Jurczak                                                                                                                                                                                                                                                                                                                                                                                                         |              |
| DATE PREPARED:       | August 15, 2001                                                                                                                                                                                                                                                                                                                                                                                                          |              |
| CLASSIFICATION NAME: | Piston Syringe                                                                                                                                                                                                                                                                                                                                                                                                           |              |
| PROPRIETARY NAME:    | Symphony Graft Delivery System                                                                                                                                                                                                                                                                                                                                                                                           |              |
| PREDICATE DEVICES:   | Symphony Graft Delivery System (K003286, K010320)                                                                                                                                                                                                                                                                                                                                                                        |              |
| INTENDED USE:        | The Symphony Graft Delivery System is indicated for the<br>delivery of allograft, autograft, or synthetic bone graft<br>materials to an orthopedic surgical site. In addition, it is<br>designed to facilitate pre-mixing of bone graft materials<br>with I.V. fluids, blood, plasma, platelet rich plasma, bone<br>marrow or other specific blood component(s) as deemed<br>necessary by the clinical use requirements. |              |
| MATERIALS:           | Medical grade polycarbonate resin<br>Medical grade High Density Polyethylene                                                                                                                                                                                                                                                                                                                                             |              |
| PERFORMANCE DATA:    | The resins, from which the components of the Symphony<br>Graft Delivery System are manufactured, meet the<br>requirements set forth in the Tripartite Biocompatibility<br>Guidance for Medical Devices. This Guidance includes<br>testing requirements for Pharmacopeia XXII, Class VI as<br>well as the FDA modified ISO 10993-1 tests for<br>biocompatibility for human body fluid contact of 30 days or<br>less.      |              |
| DESIGN:              | The Symphony Graft Delivery System Syringe is designed<br>specifically to collect, mix and deliver bone graft materials<br>to a surgical site. The Graft Chamber component of the<br>Symphony GDS can be provided either empty or pre-filled<br>with human allograft                                                                                                                                                     |              |

... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

!

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular emblem with the department's name encircling an abstract symbol. The symbol resembles a stylized caduceus, featuring a staff with a serpent entwined around it, representing health and medicine. The logo is rendered in black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV 1 4 2001

Ms. Karen F. Jurczak Regulatory Affairs Associate DePuy AcroMed, Inc. 325 Paramount Drive Raynham, Massachusetts 02767

Re: K012738

Trade/Device Name: Symphony Graft Delivery System (GDS) Regulation Number: 878.4820 Regulation Name: Surgical instrument motors and accessories/attachments Regulatory Class: I Product Code: GEY Dated: August 15, 2001 Received: August 16, 2001

Dear Ms. Jurczak:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

Page 2 - Ms. Karen F. Jurczak

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Sincerely yours,

lo Mark n Mellum

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## IV. Indications for Use

| 510(k) Number (if known): | K012738 |
|---------------------------|---------|
|---------------------------|---------|

|  | NOV 1 4 2001 |
|--|--------------|
|--|--------------|

Device Name: Symphony Graft Delivery System

The Symphony Graft Delivery System is indicated for the delivery of allograft, or synthetic bone graft materials to an orthopedic surgical site. In addition, it is designed to facilitate pre-mixing of bone graft materials with I.V. fluids, blood, plasma, platelet rich plasma, bone marrow or other specific blood component(s) as deemed necessary by the clinical use requirements.

(Please do not write below this line - continue on another page if needed)

Concurrence of CDRH, Office of Device Evaluation (ODE)

| Prescription Use: Use:<br>(Per 21 CFR 801.109) | OR Over-The-Counter |
|------------------------------------------------|---------------------|
|------------------------------------------------|---------------------|

for (Division Sign-Off)
Division of General, Restorative
and Neurological Devices

510(k) Number: K012738

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEY/K012738](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEY/K012738)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
