← Product Code [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX) · K993834

# SURGICAL LIGHTSTIC, MODELS, 180 L, AND 180 C (K993834)

_Cardiofocus, Inc. · GEX · Feb 4, 2000 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K993834

## Device Facts

- **Applicant:** Cardiofocus, Inc.
- **Product Code:** [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX.md)
- **Decision Date:** Feb 4, 2000
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

The Surgical Lightstic" 180 is intended to be used as a surgical instrument for coagulation of soft tissue in conjunction with or without endoscopic equipment (including laparoscopes, hysteroscopes, bronchoscopes, cystoscopes, gastroscopes, colonoscopes), and coagulation of soft tissue in the contact mode in both open or closed surgical procedures (with or without handpiece). The Surgical Lightstic " 180 is indicated for use in medicine and surgery with 980 -1064 wavelength laser energy in the following surgical specialties: General Surgery, Plastic Surgery, and Dermatology.

## Device Story

Surgical Lightstic™ 180 is a laser fiber accessory designed to deliver 980-1064 nm laser energy for soft tissue coagulation. Used in open or closed surgical procedures, it functions with or without endoscopic equipment (e.g., laparoscopes, bronchoscopes) and optional handpieces. Operated by surgeons in clinical settings, the device acts as a conduit for laser energy to achieve hemostasis or tissue coagulation. It benefits patients by enabling precise, minimally invasive or open surgical interventions. The device is a passive accessory to existing laser systems.

## Clinical Evidence

Bench testing only.

## Technological Characteristics

Laser fiber accessory for 980-1064 nm wavelength energy delivery. Compatible with endoscopic equipment and handpieces. Passive optical component.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Rare Earth Medical Lightstic Model 180
- Domier's Lasertrode Fiber
- Xintec's Polaris™ Diode Laser System and Accessories
- CeramOptec's Ceralas Diode Laser Systems

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

#### 510(k) Summary FEB 4 2000

Surgical Lightstic™ 180

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR §807 for the Surgical Lightstic™ 180.

## Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

William A. Knape 987 University Avenue, Suite #14 Los Gatos, CA 95032-7640 Telephone: (408) 399-9347 (408) 399-9367 Facsimile:

Contact Person: same

Date Prepared: November 5, 1999

#### Name of Device and Name/Address of Sponsor

Surgical Lightstic™ 180 CardioFocus, Inc. 126B Mid-Tech Drive West Yarmouth, MA 02673

#### Common or Usual Name

Laser Fiber

#### Classification Name

Surgical Laser Instrument Accessories

#### Predicate Devices

Rare Earth Medical Lightstic " Model 180, Domier's Lasertrode Fiber, Xintec's Polaris™ Diode Laser System and Accessories, and CeramOptec's Ceralas Diode Laser Systems.

{1}------------------------------------------------

#### Intended Use

The Surgical Lightstic" 180 is intended to be used as a surgical instrument for coagulation of soft tissue in conjunction with or without endoscopic equipment (including laparoscopes, hysteroscopes, bronchoscopes, cystoscopes, gastroscopes, colonoscopes), and coagulation of soft tissue in the contact mode in both open or closed surgical procedures (with or without handpiece).

The Surgical Lightstic " 180 is indicated for use in medicine and surgery with 980 -1064 wavelength laser energy in the following surgical specialties: General Surgery, Plastic Surgery, and Dermatology.

### Technological Characteristics and Substantial Equivalence

From a clinical perspective and comparing design specifications, the CardioFocus Surgical Lightstic" 180 and the predicate device are substantially equivalent and have the same intended use.

CardioFocus, Inc. believes the minor differences of the CardioFocus Surgical Lightstic" 180 and its predicate fiber laser accessories should not raise any concerns regarding the overall safety and effectiveness.

#### Performance Data

None required.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized symbol that resembles three wavy lines or a caduceus, which is a symbol often associated with healthcare.

Public Health Service

FEB 4 2000

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

CardioFocus, Inc. c/o Mr. William A. Knape Clinical Pathways 987 University Avenue, Suite 14 Los Gatos, California 95032

K993834 Re:

Trade Name: Surgical Lightstic™ Models 180L and 180C Regulatory Class: II Product Code: GEX Dated: November 11, 1999 Received: November 12, 1999

Dear Mr. Knape:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

#### Page 2 - Mr. William A. Knape

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

fa

James E. Dillard III

Acting Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use Statement

510(k) Number (if known): K993834

Surgical Lightstic™ 180 Device Name:

Indications For Use:

The Surgical Lightstic" 180 is intended to be used as a surgical instrument for coagulation of soft tissue in conjunction with or without endoscopic equipment (including laparoscopes, hysteroscopes, bronchoscopes, cystoscopes, gastroscopes, colonoscopes), and coagulation of soft tissue in the contact mode in both open or closed surgical procedures (with or without handpiece).

The Surgical Lightstic™ 180 is indicated for use in medicine and surgery with 980 - 1064 nm wavelength laser energy in the following surgical specialties: General Surgery, Plastic Surgery, and Dermatology.

These are additional indications to the already cleared indications for market Note: release in K962068.

| Concurrence of CDRH, Office of Device Evaluation (ODE) |                                                                                                                    |    |                      |
|--------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|----|----------------------|
| <div><img alt="Signature" src="signature.png"/></div>  |                                                                                                                    |    |                      |
| (Division Sign-Off)                                    |                                                                                                                    |    |                      |
| Division of General Restorative Devices                |                                                                                                                    |    |                      |
| 510(k) Number                                          | K993834                                                                                                            |    |                      |
| Prescription Use                                       | <div style="text-align:left;"><img alt="Prescription Use" src="prescription_use.png"/></div> (Per 21 CFR §801.109) | OR | Over-The-Counter Use |

CONFIDENTIAL

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K993834](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K993834)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
