OMNITIP SIDE FIRING SWITCHABLE TIP WITH SUCTION/IRRIGATION, MODEL 20371-HP

K992230 · Trimedyne, Inc. · GEX · Jul 12, 1999 · General, Plastic Surgery

Device Facts

Record IDK992230
Device NameOMNITIP SIDE FIRING SWITCHABLE TIP WITH SUCTION/IRRIGATION, MODEL 20371-HP
ApplicantTrimedyne, Inc.
Product CodeGEX · General, Plastic Surgery
Decision DateJul 12, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Intended Use

The OmniTip™ Side Firing Switchable Tip with Suction/Irrigation is intended for use with any pulsed Holmium: YAG 2.1 micrometer laser with a compatible connector for incision, excision, ablation, vaporization, and coagulation.

Device Story

Fiber optic energy delivery device; consists of near-contact laser fiber enclosed in needle assembly (capillary tube and stainless steel shaft). Attaches to Omni Multiuse Handpiece via quick-connect mechanism. Proximal luer connector links suction/irrigation channel to external pump. Used by physicians in clinical settings to deliver laser energy for tissue interaction. Suction/irrigation feature allows simultaneous fluid management during laser procedures. Benefits include enhanced visualization and debris removal during surgery.

Clinical Evidence

Simulated clinical testing performed using heart tissue. Physician feedback utilized to validate clinical utility. No human clinical trial data provided.

Technological Characteristics

Fiber optic laser delivery system; stainless steel shaft; capillary tube; luer connector for suction/irrigation. Compatible with pulsed Holmium: YAG 2.1 micrometer lasers. Mechanical, biological, and performance testing conducted.

Indications for Use

Indicated for use with pulsed Holmium: YAG 2.1 micrometer lasers for incision, excision, ablation, vaporization, and coagulation of tissue.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # K992230 Attachment 4 # 510(k) Summary of Safety and Effectiveness Information Trimedyne® OmniTip™ Side Firing Switchable Tips with Suction/Irrigation This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. | Submitter Information: | Trimedyne, Inc.<br>P.O. Box 57001<br>Irvine, CA 92619-7001<br>(949) 559-5300<br>(949) 559-1330 | |------------------------|------------------------------------------------------------------------------------------------| | Contact person: | Susan H. Gamble<br>Vice President, Regulatory Affairs and Quality | | Summary Date: | July 1, 1999 | ### II. Device Name OmniTip™ Side Firing Switchable Tip with Suction/Irrigation Proprietary: Common: Laser Fiber Classification: Accessories to Laser-Powered Instrument (unclassified) ## Predicate Device III. The predicate devices are the Trimedyne OmniTip™ Switchable Tips and Tapertip™ Holmium Arthroscopic Handpieces. ## IV. Device Description The OmniTip Side Firing Switchable Tip with Suction/Irrigation is a fiber optic energy delivery device consisting of a near-contact laser fiber enclosed in a needle assembly of a capillary tube and an external stainless steel shaft. The device is attached to the Omni Multiuse Handpiece through a quick-connect mechanism. A luer connector assembly installed at the proximal end of the device connects the suction/irrigation channel to a pump via an external line. {1}------------------------------------------------ ### V. Intended Use This device is intended for use with any Holmium laser (with a compatible connector) for its cleared applications. ### VI. Technological Characteristics The device differs from the predicate devices due mainly only to the incorporation of suction/irrigation channel and minor configuration changes. #### VII. Nonclinical Tests The device was subjected to a series of tests, including mechanical, biological, and performance studies. ### VIII. Clinical Tests The device was subjected to a simulated clinical testing using heart tissue. A physician's feedback is also used to validate the use of the device in a clinical setting. # IX. Conclusions Drawn from Testing The device is biocompatible, performs as intended, and has acceptable mechanical properties when used according to its labeling. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of a bird-like figure. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUL 12 1999 Ms. Susan H. Gamble Vice President, Regulatory Affairs and Quality Trimedyne, Inc. 2801 Barranca Road P.O. Box 57001 Irvine, California 92619 K992230 Re: > Trade Name: OmniTip™ Side Firing Switchable Tip with Suction/Irrigation Regulatory Class: II Product Code: GEX, GCX, and GBX Dated: July 1, 1999 Received: July 2, 1999 Dear Ms. Gamble: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (OS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3}------------------------------------------------ Page 2 - Ms. Susan H. Gamble This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Celia M. Witten, Ph.D., M.D. Director Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Attachment 2 # Indications for Use Statement | 510(k) Number | K992230 | |---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name | OmniTip™ Side Firing Switchable Tip with Suction/Irrigation | | Indications for Use | The OmniTip™ Side Firing Switchable Tip with Suction/Irrigation is intended for use with any pulsed Holmium: YAG 2.1 micrometer laser with a compatible connector for incision, excision, ablation, vaporization, and coagulation. | # PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED Concurrence of CDRH, Office of Device Evaluation (ODE) | Prescription Use | <span style="text-decoration: overline;">X</span> | OR | Over-The-Counter Use | <span style="text-decoration: overline;"></span> | |------------------|---------------------------------------------------|----|----------------------|--------------------------------------------------| |------------------|---------------------------------------------------|----|----------------------|--------------------------------------------------| (Per 21 CFR 801.109) Signature Title of General Representative | (Ax) Number | K992230 | Page 28 of 32 | |-------------|---------|---------------| |-------------|---------|---------------|
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...