K971814 · Mehl/Biophile Intl., Inc. · GEX · Aug 1, 1997 · General, Plastic Surgery
Device Facts
Record ID
K971814
Device Name
IMPROVEMENTS TO THE MEHL RUBY LASER
Applicant
Mehl/Biophile Intl., Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Aug 1, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Chromos 694 (MEHL) Ruby Laser, Mark I, is indicated for the removal of unwanted body and facial hair.
Device Story
The Chromos 694 Ruby Laser, Mark I, is a medical laser system designed for hair removal. It utilizes a 694 nm ruby laser source to deliver energy to the skin. The device transmits laser light via an optical fiber delivery system, which replaces the articulated arm used in the predecessor model to eliminate Gaussian hot spots. The system features a 7mm spot size. Operation is managed through software-directed controls that include safety safeguards, specifically shutdown provisions for at least seven fault conditions. The device is intended for use by clinicians in a medical practice setting. By delivering controlled laser energy to the hair follicle, the device facilitates the removal of unwanted body and facial hair, offering improved safety and reliability compared to the previous MEHL Ruby Laser model.
Clinical Evidence
No new clinical data was generated for the Mark I model. Equivalence is supported by clinical studies performed on the predecessor MEHL Ruby Laser, which showed no unexpected safety issues, no scarring, and no purpuric reactions. Infrequent hyper- and hypopigmentation were observed but resolved predictably. Substantial equivalence is based on technical similarity and improved safety/control features.
Technological Characteristics
694 nm ruby laser; optical fiber delivery system; 7mm spot size; software-directed controls with seven fault condition shutdown provisions; Class II medical laser (21 CFR 878.4810).
Indications for Use
Indicated for the removal of unwanted body and facial hair in patients.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
MEHL Ruby Laser for Hair Removal
Related Devices
K991634 — FOTONA NOVALIS R RUBY LASER SYSTEM · Fotona D.D. · Aug 10, 1999
K980187 — MLT R694 RUBY LASER SYSTEM · Medical Laser Technologies , Ltd. · Apr 3, 1998
K980517 — EPILASER NORMAL MODE RUBY LASER · Palomar Medical Products, Inc. · Jul 21, 1998
K962446 — SHARPLAN RUBY LASER SYSTEM · Sharplan Lasers, Inc. · Mar 7, 1997
K963947 — EPILASER NORMAL MODE RUBY LASER · Spectrum Medical Technologies, Inc. · Mar 5, 1997
Submission Summary (Full Text)
{0}------------------------------------------------
K971814
## MEHL / Biophile International CORPORATION
AUG - 1 1997
4127 N.W. 27th Lane, Suite A Gainesville, Florida 32606 Phone (352) 373-2565 Fax (352) 373-2481
## 510(k) SUMMARY
## CHROMOS 694 RUBY LASER FOR HAIR REMOVAL
The Mark I Chromos 694 Ruby Laser is a laser instrument used in medical practice for the removal of unwanted body and facial hair. Reviewed by the General and Plastic Surgery Device Classification Panel, medical lasers are assigned the Product Code 79GEX and are regulated under 21 CFR 878.4810 as Class II (Performance Standards) devices.
The Mark I Chromos 694 Ruby Laser is substantially equivalent to its predecessor, the MEHL Ruby Laser for Hair Removal. Both use the same wavelength, deliver the same energies to the skin, and are constructed of similar components for the removal of unwanted body and facial hair.
The MEHL Ruby Laser utilized manual controls and a standard, mirrored articulated arm for transmitting laser light to the skin site. Further. it contained a Q-switched feature which was intended for other purposes.
Safety and reliability are improved with the newer version, the Mark I Chromos 694 Ruby Laser. Those improvements include:
- Elimination of the O-switched option ●
- Replacement of the articulated arm with optical fiber which eliminates the . Gausian hot spot common to the former
- . A 7mm vs. 5mm spot size
- Software directed controls and safeguards which include shutdown provisions for no fewer than seven fault conditions
Establishing substantial equivalence between the two devices were (1) engineering data showing similarity in light delivery characteristics; (2) calculations proving essentially equivalent fluences for the 5mm and 7mm spots; and (3) thermographic profiles, optical fiber vs. articulated arm. Reliability of the newer laser also was assured with a preponderance of quality control and validation information.
{1}------------------------------------------------
510(k) Summary Chromos 694 Ruby Laser for Hair Removal Page 2
In clinical studies with the MEHL Ruby Laser, no unexpected safety issues were identified. No scarring or purpuric reactions were observed. Though hyper- and hypopigmentation were seen infrequently, these effects resolved predictably within a few months.
With the similarity in technical characteristics between the MEHL Ruby Laser and the Chromos 694 Ruby Laser, equivalence in clinical performance can therefore be assured, with more safeguards, better control and increased reliability intrinsic to the newer model.
Thomas Blake
Thomas Blake, R.Ph. Regulatory Strategist
6/9/97
Date
{2}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Thomas Blake, R.Ph. MEHL/Biophile International Corporation 48 Mt. Olive Road …………… Budd Lake, New Jersey 07828
AUG - I 1997
Re: K971814
Trade Name: Chromos 694 (MEHL) Ruby Laser, Model Mark I Regulatory Class: II Product Code: GEX Dated: May 14, 1997 Received: May 15, 1997
Dear Mr. Blake:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements . as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does
{3}------------------------------------------------
Page 2 - Mr. Thomas Blake, R.Ph.
not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
KG711841/A'...
Page _______ of ______________________________________________________________________________________________________________________________________________________________
510(k) Number (if known): K971814
Device Name: CHROMOS 694 (MEHL) RUBY LASER, MODEL MARK I
Indications for Use: . . . . . .
The Chromos 694 (MEHL) Ruby Laser, Mark I, is indicated for the removal of unwanted body and facial hair.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

| (Division Sign-Off) | |
|-----------------------------------------|---------|
| Division of General Restorative Devices | |
| 510(k) Number | K971814 |
Prescription Use_ V (Per 21 CFR 801.109)
OR
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
\$\kappa=37\$
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.