← Product Code [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX) · K963389

# FLEX-LASE 600 SERIES ION LASER SYSTEM (K963389)

_American Laser Medical · GEX · Nov 26, 1996 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K963389

## Device Facts

- **Applicant:** American Laser Medical
- **Product Code:** [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX.md)
- **Decision Date:** Nov 26, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

The Flex-lase 600 Series Ion Laser is intended for use in the treatment of selected vascular and pigmented lesions of the skin. Laser light of a selected wavelength is used to photocalculate tissue based on the absorption characteristic of the targeted chromophore within the tissue.

## Device Story

Flex-Lase 600 Series Ion Laser (Model 620 argon; Model 640 krypton) delivers continuous wave laser energy for dermatological procedures. System inputs include electrical power (208 VAC) and water cooling; operator activates via footswitch. Device emits specific wavelengths (488/514 nm for argon; 568/520-530 nm for krypton) delivered via fiber-optic handpiece to target chromophores in skin lesions. Aiming beam provided by adjustable incoherent white light. Used in clinical settings by trained professionals. Output energy photocoagulates targeted vascular or pigmented tissue, facilitating lesion treatment. System is not computer-controlled.

## Clinical Evidence

No clinical data provided; substantial equivalence is based on bench testing and comparison of technical specifications and intended use with predicate devices.

## Technological Characteristics

Continuous wave ion laser (argon or krypton). Model 620: 488/514 nm; Model 640: 568/520-530 nm. Power output up to 5.5W (argon) or 3.0W (krypton). Beam delivery via fiber-optic handpiece. Cooling: filtered flow-through tap water (2.5 gal/min). Electrical: 208-240 VAC, 40 amp, single phase. Aiming beam: adjustable incoherent white light. Not computer-controlled.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- HGM Spectrum K1 ([K913428](/device/K913428.md), [K913569](/device/K913569.md))
- Metalaser Vasculase ([K903883](/device/K903883.md))
- Coherent Artisan

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

11/08/96 10:55 *000000000000000 + 301 594 2358
NO. 655 008
NOV 26 1996
K963399

Flax-Lase 600 Series Ion Laser
American Laser Medical
November 7, 1996

# Summary of Safety and Effectiveness Information

## Regulatory Authority:

Safe Medical Devices Act of 1990, 21 CFR 807.92

## Company Name/Contact:

**Company:** American Laser Medical
1832 South 3850 West
Salt Lake City, UT 84104
(801) 972-1311, FAX (801) 972-5251

**Contact:** Mr. Daniel Hoefer
Regulatory Affairs
American Laser Medical
1832 South 3850 West
Salt Lake City, UT 84104
(801) 972-1311, FAX (801) 972-5251

## Name of Device:

**Trade Name:** Flax-Lase 600 Series Ion Laser
**Common Name:** Dermatology Laser

## Classification:

Powered surgical laser instruments are assigned the unique device classification code 79GEX. The published physical description of this device is found in 21 CFR, § 878.4810 (a) (2). This device classification section specifically describes argon laser systems. The American Laser Medical device may be configured as either an argon laser system or as a krypton laser system, however American Laser Medical maintains that there is no substantial difference between argon and krypton lasers for the purpose of classification. Presently, these lasers are considered Class II medical devices.

1

{1}

11/25/96 12:23 *000000000000000000000000000000000000000000000000000000000000000000000000000
NO. 674 D03

Flex-Lase 600 Series Ion Laser
American Laser Medical
November 7, 1996

## Description of Device:

The Flex-Lase 600 Series Ion Laser submission covers all of the internal parts, subassemblies and components as well as the completed device. The Flex-Lase Model 640 is a continuous wave krypton laser system emitting light at the 568 nm (yellow) and 520/530 nm (green) wavelengths. The Flex-Lase Model 620 is a continuous wave argon laser system emitting light at the 488 nm (blue) and the 514 nm (green) wavelengths.

Treatment beam output for the Flex-Lase Model 640 (krypton) is 1.0 Watt (yellow) and 3.0 Watts (yellow-green simultaneous). The power output for the Flex-Lase Model 620 (argon) is 5.5 watts blue-green or 2.0 watts green. The aiming beam is provided by an adjustable intensity incoherent white light source. Exposure times for the Flex-Lase 600 Series Ion Laser (in seconds) are 0.02, 0.05, 0.1, 0.25, 0.5, 1.0, and continuous.

Laser activation occurs by footswitch. Overall weight of the system is 112 lbs. (51 kg). The Flex-Lase 600 Series Ion Laser has a footprint of 16 X 24 inches (38 X 60 cm) and is 36 inches (90 cm) high.

The electrical power requirement is 208 VAC 40 amp single phase. Cooling is provided by external filtered flow-through tap water at a minimum rate of 2.5 gallons (9.5 liters) per minute.

Accessories available for use with the Flex-lase 600 Ion Laser include a focusing handpiece with interchangeable guide tips and/or the Hexascan Mark II. The Flex-laser 600 series is not a computer controlled device.

## Intended Use:

The Flex-lase 600 Series Ion Laser is intended for use in the treatment of selected vascular and pigmented lesions of the skin. Laser light of a selected wavelength is used to photocalculate tissue based on the absorption characteristic of the targeted chromophore within the tissue.

## Indications for Use:

The Flex-Lase 600 Ion Laser is intended for use in the treatment of selected pigmented and vascularized lesions of the skin. Representative indications for use are:

2

{2}

11/25/96 12:24 +000000000000000 - 301 594 2358 NO. 674 D04

Flex-Lase 600 Series Ion Laser
American Laser Medical
November 7, 1996

Flex-Lase Model 640 Laser System
For vascular lesions: (Yellow - 568 nm)
1. Capillary Hemangioma (port wine hemangioma)
2. Strawberry hemangioma
3. Telangectasia
4. Rosacea

For pigmented lesions: (Yellow/Green - 568 nm and 520-530 nm).
1. Freckles
2. Age Spots
3. Cafe-au-lait
4. Lentigo

Flex-Lase Model 620 Laser System
For Vascular Lesions: (Blue/Green - 488 nm and 514.5 nm)
1. Capillary Hemangioma (port wine hemangioma)
2. Strawberry Hemangioma
3. Telangectasia
4. Rosacea

For pigmented lesions: (Green - 514.5 nm)
1. Freckles
2. Age Spots
3. Cafe-au-lait
4. Lentigo

Comparison of Device Characteristics:

For vascular lesions:
The Flex-Lase 600 Ion Laser System is substantially equivalent to the HGM Spectrum K1, the Metalaser Vasculase and the Coherent Artisan already in commercial distribution. The materials, design, intended use, method of manufacture, warnings, cautions, precautions and treatment parameters are substantially the same.

For pigmented lesions:
The Flex-Lase 600 Series Ion Laser system is substantially equivalent to the HGM Spectrum K1, the Metalaser Vasculase and the Coherent Artisan already in commercial distribution. The

3

{3}

11/08/96 10:56 *000000000000000 + 301 594 2358 NO. 655 D11

Flex-Lase 600 Series Ion Laser
American Laser Medical, Inc.
November 7, 1996

materials, design, intended use, method of manufacture, warnings, cautions and precautions are substantially the same.

Feature comparison tables which follow illustrate the equivalence of features of the Flex-Lase 600 and equivalent available systems. See also appendix II for marketing materials on substantially equivalent devices.

4

{4}

11/08/96 10:56 *00000000000000 + 301 594 2356 NO. 655 D12

# Laser Equivalency Table
## Vascular Lesions

|   | FLEX-LASE
MODEL 620 | FLEX-LASE
MODEL 640 | HGM K1 | METALASER
VASCULASE | COHERENT
ARTISAN  |
| --- | --- | --- | --- | --- | --- |
|  Type | Continuous.
wave
argon | Continuous
wave
krypton | Continuous
Wave
krypton | Copper
Vapor
Laser | Continuous
wave
Argon/Dye  |
|  Wavelength
(nanometers) | 514, or
488/514 | 568, or
568/520-
530 | 568,
520/530 | 511, 578 | 514 only,
or
488/514,
or 577-630
selectable  |
|  Exposure
Time (sec.) | 0.02,
0.05,
0.1,
0.25,
0.5, 1.0,
and cont. | Same | 0.05, 0.1,
0.2, 0.5,
1.0, 2.0,
5.0, and
continuous | 0.075,
0.1, 0.2,
0.3, and
continuous | 0.01,
0.02,
0.05, 0.1,
0.2, 0.5,
1.0, 2.0,
5.0, cont.  |
|  Output Mode | Multimode | Multimode | Multimode | Multimode | Multimode  |
|  Beam
Delivery | Handpiece
via fiber
optic | Handpiece
via fiber
optic | Handpiece
via fiber
optic | Handpiece
via fiber
optic | Handpiece
via fiber
optic  |
|  Stability
of Output
Energy | +/- 10% | +/- 10% | +/- 10% | +/- 20% | +/- 5%  |
|  Spot Size
(mm) | 0.25,
0.5, 1.0 | 0.25, 0.5,
1.0 | 0.1, 0.2,
1.0 | 0.1, 0.2,
1.0 | 0.1, 0.25,
0.5, 1.0,
fixed,
variable
0.2-2.5  |
|  Energy
density | 2-30
J/cm² | 2-30 J/cm² | 2-30 J/cm² | 4-15 J/cm² | 2-30 J/cm²  |
|  Aiming
beam | Adjust-
able in-
coherent
white
light
source | Adjustable
incoherent
white
light
source | HeNe
(633nm)
adjustable
to 5 mW | HeNe (633
nm) | HeNe (633
nm)  |

5

{5}

11/08/96 10:57 +0000000000000000000000000000000000000000000000000000000000

|  PARAMETER | FLEX-LASE MODEL 620 | FLEX-LASE MODEL 640 | HOM K1 | METALASER VASCULASE | COHERENT ARTISAN  |
| --- | --- | --- | --- | --- | --- |
|  Cooling | Filtered flow-through tap water 2.5 gal/min | Filtered flow-through tap water 2.5 gal/min | Internal liquid cooled | Air-cooled, 40F-80F ambient air temp. | Filtered flow-through tap water 2.5 gal/min  |
|  Electrical Requirements | 208-240 VAC 40 amp single phase 50/60 Hz | 208-240 VAC 40 amp single phase 50/60 Hz | 208-240 VAC 40 amp 50/60 Hz single phase | 208 VAC +/- 10%, 20 amp 60Hz single phase | 208 VAC 50 amp 50/60 Hz 3 phase  |
|  Indications for Vascular lesions | Capillary (port wine) hemangioma, Strawberry hemangioma, Telangec-tasia, Rosacea | Capillary or port wine hemangioma Strawberry hemangioma Telangec-tasia Rosacea | Vascular Lesions (same) | Vascular Lesions (same) | Vascular Lesions (same)  |
|  K Number |  |  | K913428, K913569 | K903883 | na  |

{6}

11/08/96 10:57 +0000000000000000000000000000000000000000000000000000000000

NO.655

Laser Equivalency Table
Pigmented Lesions

|   | FLEX-LASE
MODEL 620 | FLEX-LASE
MODEL 640 | HGM K1 | METALASER
VASCULASE | COHERENT
ARTISAN  |
| --- | --- | --- | --- | --- | --- |
|  Type | Continuous wave
argon | Continuous wave
krypton | Continuous Wave
krypton | Copper Vapor
Laser | Continuous wave
Argon/Dye  |
|  Wavelength (nanometers) | 514, or
488/514 | 568, or
568/520-530 | 568,
520/530 | 511, 578 | 514 only, or
488/514, or
577-630 selectable  |
|  Exposure Time (sec.) | 0.02,
0.05,
0.1,
0.25,
0.5, 1.0,
and cont. | Same | 0.05, 0.1,
0.2, 0.5,
1.0, 2.0,
5.0, and
continuous | 0.075,
0.1, 0.2,
0.3, and
continuous | 0.01,
0.02,
0.05, 0.1,
0.2, 0.5,
1.0, 2.0,
5.0, cont.  |
|  Output Mode | Multimode | Multimode | Multimode | Multimode | Multimode  |
|  Beam Delivery | Handpiece
via fiber
optic | Handpiece
via fiber
optic | Handpiece
via fiber
optic | Handpiece
via fiber
optic | Handpiece
via fiber
optic  |
|  Stability of Output Energy | +/- 10% | +/-10% | +/-10% | +/-20% | +/- 5%  |
|  Spot Size (millimeters) | 0.25,
0.5, 1.0 | 0.25, 0.5,
1.0 | 0.1, 0.2,
1.0 mm | 0.1, 0.2,
1.0 mm | 0.1, 0.25,
0.5, 1.0,
fixed,
variable
0.2-2.5 mm  |
|  Energy density | 2-30 J/cm² | 2-30 J/cm² | 2-30 J/cm² | 4-15 J/cm² | 2-30 J/cm²  |
|  Aiming beam | Adjustable incoherent white light source | Adjustable incoherent white light source | HeNe (633nm)
adjustable to 5 mW | HeNe (633 nm) | HeNe (633 nm)  |

{7}

11/08/96 10:58 +00000000000000 + 301 594 2358 NO. 655 D15

| PARAMETER | FLEX-LASE
MODEL 620 | FLEX-LASE
MODEL 640 | HGM K1 | METALASER
VASCULASE | COHERENT
ARTISAN |
| --- | --- | --- | --- | --- | --- |
| Cooling | Filtered flow-through tap water 2.5 gal/min | Filtered flow-through tap water 2.5 gal/min | Internal liquid cooled | Air-cooled, 40F-80F ambient air temp. | Filtered flow-through tap water 2.5 gal/min |
| Electrical Requirements | 208-240 VAC 40 amp single phase 50/60 Hz | 208-240 VAC 40 amp single phase 50/60 Hz | 208-240 VAC 40 amp 50/60 Hz single phase | 208 VAC +/- 10%, 20 amp 60Hz single phase | 208 VAC 50 amp 50/60 Hz 3 phase |
| Indications for Pigmented lesions | Freckles, Age spots, cafe-au-lait, lentigo | Freckles, Age spots, cafe-au-lait lentigo | Pigmented Lesions (same) | Pigmented Lesions (same) | Pigmented Lesions (same) |
| K Number |  |  | K913428, K913569 | K903883 | na |

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K963389](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K963389)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
