← Product Code [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX) · K960905

# SDL DIODE LASER SYSTEM (K960905)

_Slt-J , Ltd. · GEX · Jun 3, 1996 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K960905

## Device Facts

- **Applicant:** Slt-J , Ltd.
- **Product Code:** [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX.md)
- **Decision Date:** Jun 3, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Device Story

SDL Diode Laser System is a near-infrared surgical laser. Device accepts 110 VAC input power; utilizes touch pad controls for operating parameters; features footswitch or fingerswitch operation. System connects to fiberoptic delivery systems via industry-standard SMA 905 connector. Laser energy is delivered to specialized fiberoptic tips (e.g., NEOS Alloy Scalpel, Fiber Cap, Bipolar Dissector) which absorb laser light and convert it into thermal energy for surgical tissue effects. Device is intended for use by clinicians in surgical settings. By acting as a heat source independent of specific laser wavelength, the system provides thermal coagulation and cutting capabilities equivalent to existing Nd:YAG and diode laser systems.

## Clinical Evidence

No clinical trials performed. Evidence consists of laboratory bench-top studies comparing temperature profiles of fiberoptic delivery tips (NEOS Alloy Scalpel) when powered by the subject diode laser versus a 1.06µm Nd:YAG laser at 3.5W and 6.5W. Results showed equivalent maximum temperatures. Literature review (Judy et al., 1993) compared thermal coagulation depths in rabbit tissue using diode and Nd:YAG lasers in contact mode, supporting equivalent tissue effects.

## Technological Characteristics

Near-infrared diode laser system; 110 VAC power; SMA 905 fiberoptic connector; touch pad interface; footswitch/fingerswitch control. Complies with 21 CFR 1040.10 and 1040.11. System functions by coupling laser energy into specialized fiberoptic delivery tips (e.g., stainless steel hypo tubing or surface-treated alloy scalpel) that convert light to thermal energy.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Diomed 25, 25W surgical diode laser system ([K914520](/device/K914520.md), [K914521](/device/K914521.md))
- ZOE Nd:YAG laser system ([K909128](/device/K909128.md), [K923991](/device/K923991.md))

## Reference Devices

- NEOS Alloy Scalpel ([K914197](/device/K914197.md))
- Fiber Cap ([K924120](/device/K924120.md))
- Hybrid Surgical Device ([K924160](/device/K924160.md))
- Closed End ([K931070](/device/K931070.md))
- Closed End/Electrocautery ([K932272](/device/K932272.md))
- Bipolar Dissector ([K901365](/device/K901365.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

K960905
JUN - 3 1996

V. Summary of Safety and Effectiveness

Summary of 21 CFR 807.87

A. Device Name:

1. Proprietary Name: SDL Diode Laser System and Associated Fiberoptic Delivery Systems

2. Common Name: Near-infrared Diode Laser System

B. Establishment Registration Number:

SLT-J, Ltd. has not yet registered as a medical device establishment with the FDA. The company intends to complete all regulatory filings prior to introducing the SDL diode laser system into interstate commerce.

Address all correspondence to:
Mr. Michael D. Johnson
johnson+associates
5848 Price Road
Milford, OH 45150
513-248-8804

C. Device Classification:

Diode surgical laser systems are currently considered Class II medical devices subject to pre-market notification provisions for many surgical applications.

Although not formally classified, diode surgical laser delivery systems typically have been regulated as Class II devices.

0039

{1}

Electrosurgery handpieces (pencils, electrodes) are considered Class II 79 GEI [21 CFR 878.4400]. SLT-J, Ltd. anticipates these accessory devices would receive the same classification if such energy delivery systems were officially classified.

## D. Compliance With Standards

The SDL diode laser conforms with federal regulations and the performance standards 21 CFR 1040.10 and 1040.11 for medical laser systems. Certification reports will be submitted to CDRH certifying compliance with this standard and are currently in preparation and will be submitted by SLT-J, Ltd. prior to commercial distribution of this product.

## E. Labeling

Product labels comply with 21 CFR 1040.10 and 1040.11 as applicable. An Operator’s Manual for the SDL diode laser system is currently in preparation. A draft Operator’s Manual is included in this submission (See Attachments).

Product labels comply with 21 CFR 801 as applicable for accessory devices. Copies of the proposed text for these package labels are located in Section IV of this submission.

## F. Statement of Equivalence

In the opinion of SLT-J, Ltd., the SDL diode laser system, when used with any of the following fiberoptic delivery systems:

NEOS Alloy Scalpel (K#914197)
LCA, Inc. (Cincinnati, OH)

0040

{2}

Fiber Cap (K#924120)
LCA, Inc. (Cincinnati, OH)

Hybrid Surgical Device (K#924160)
LCA, Inc. (Cincinnati, OH)

Closed End (K#931070)
LCA, Inc. (Cincinnati, OH)

Closed End/Electrocautery (K#932272)
LCA, Inc. (Cincinnati, OH)

Bipolar Dissector (K#901365)
LCA, Inc. (Cincinnati, OH)

is substantially equivalent to the use of any of the above-mentioned devices when used with any of the following medical devices which are used as a laser (energy) source:

Diomed® 25, 25W surgical diode laser system (K#s 914520, 914521), Diomedics Inc. (The Woodlands, TX)

ZOE Nd:YAG laser system (K#s 909128, 923991), LCA, Inc. (Cincinnati, OH)

Section III of this submission provides information in support of this equivalence claim.

This belief of equivalence is based upon the following facts:

1. The SDL diode laser system has the same operating characteristics as the diode laser system offered by Diomedics Inc with respect to power output, operating parameters and operating controls and indicators. The SDL diode laser system incorporates touch pad controls

0041

{3}

for setting of operating parameters, uses identical input power (110 VAC) as the Diomed 25W diode laser system offered by Diomedics Inc. (The Woodlands, TX).

2. The SDL diode laser system uses the industry standard SMA 905 fiberoptic connector system for its fiberoptic delivery systems. This is identical to the connector system used by Diomedics (The Woodlands, TX). The delivery system which will be used with the SDL diode laser system will be able to directly attach to the fiberoptic connector system.

3. Due to the technologies used in the manufacturing of the NEOS family of fiberoptic delivery systems as well as the Bipolar Dissector, the laser system used with these products is not a dependent variable to achieve a desired tissue effect. The particular surface treatment used on the NEOS Alloy Scalpel, Fiber Cap, Hybrid Surgical Device and the Bipolar Dissector, provide a product which completely absorbs the laser energy, thus converting laser light energy into thermal energy, regardless of the wavelength of the laser light source.

The Closed End and Closed End/Electrocautery fiberoptic delivery systems use swaged stainless steel hypo tubing to convert light energy transmitted through the fiberoptics into thermal energy, again regardless of the wavelength of the laser light source.

Laboratory bench top studies were performed to compare temperature profiles of a NEOS Alloy Scalpel tip with respect to the use with a diode laser or a $1.06\mu \mathrm{M}$ Nd:YAG laser system. Temperature plots were made at typical power settings of 3.5 and 6.5 watts. The maximum temperatures obtained using both laser devices were equivalent. These studies support the belief that the NEOS family of fiberoptic delivery systems and the

0042

{4}

Bipolar Dissector can be used with a variety of sources of energy (i.e. - many different laser wavelengths) due to their surface coatings which makes these devices true heat sources independent of input laser wavelength.

In 1993, Judy et al reported a study in which comparisons of the depths of thermal coagulation in rabbit tissues obtained with a diode laser and a $1.06\mu$ Nd:YAG laser using contact fibers. Results of this study support the equivalency of tissue thermal effects obtained with the diode and Nd:YAG lasers in a contact mode; and support the use of the diode laser as an alternative medical device.

4. The SDL diode laser system can be operated by either a footswitch or a fingerswitch. This technology is equivalent to, and has been offered since 1990 on Nd:YAG laser systems offered by Surgical Laser Technologies (The Oaks, PA) and LCA (Cincinnati, OH). This allows the operator to select between either footswitch or fingerswitch control of the fiberoptic delivery system.

{5}

# Bibliography

Absten GT, Joffe SN. *Lasers in Medicine - An Introductory Guide*. 2nd ed. London: Chapman and Hall, Ltd, 1989.

Dixon, JA. *Surgical Applications of Lasers*. 2nd ed. Chicago: Year Book Medical Publishers, Inc, 1987.

Joffe SN, Oguro Y. *Advances in Nd:YAG Laser Surgery*. New York: Springer-Verlag, 1988.

Joffe, SN. *Lasers in General Surgery*. Baltimore, Maryland: Williams &amp; Wilkins, 1989.

Judy MM, Matthews JL, Aronoff BL, Hults DF. Soft Tissue Studies With 805nm Diode Laser Radiation: Thermal Effects With Contact Tips and Comparison With Effects of 1064nm Nd:YAG Laser Radiation. *Lasers in Surgery and Medicine*, 13:528-536 (1993).

0044

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K960905](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K960905)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
