K954597 · Trimedyne, Inc. · GEX · Mar 11, 1996 · General, Plastic Surgery
Device Facts
Record ID
K954597
Device Name
RIGHT ANGLE LASER FIBERS
Applicant
Trimedyne, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Mar 11, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Right Angle Laser Fiber and Optilase Lasers are indicated for a wide variety of surgical uses including incision, excision, resection, vaporization, ablation, coagulation, and hemostasis for the treatment of benign prostatic hyperplasia (BPH) and other urological conditions. Viewing scopes, which can be either flexible or rigid, may be used for visualization during laser procedures.
Device Story
System comprises Right Angle Laser Fiber (3m buffered quartz fiber, tip, connector) and Optilase Nd:YAG laser (continuous-wave source, aiming beam, adjustable power). Used in urological surgery; operated by physicians. Laser energy delivered via fiber through flexible or rigid viewing scopes to prostate tissue. Enables incision, excision, resection, vaporization, ablation, coagulation, and hemostasis. Clinical benefit: minimally invasive treatment of BPH compared to traditional resection.
Clinical Evidence
Prospective, controlled, well-matched study comparing visual laser ablation of the prostate (VLAP) using the subject devices to transurethral resection of the prostate (TURP). Data concluded the devices are safe and effective for BPH treatment.
Indicated for patients with benign prostatic hyperplasia (BPH) and other urological conditions requiring surgical incision, excision, resection, vaporization, ablation, coagulation, or hemostasis.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}
k954597
Trimedyne
MAR 11 1996
# Summary of 510(k) Safety and Effectiveness Information
## Trimedyne® Right Angle Laser Fibers and
## Optilase® Nd:YAG Laser System Models 1000, 4000, and 1000-100 (PL100)
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
### I. Submitter Information:
Trimedyne, Inc.
P.O. Box 57001
Irvine, CA 92719-7001
714/559-5300
714/559-1330 (facsimile)
Contact Person:
Susan H. Gamble
Director, Regulatory Affairs & Quality Assurance
Summary Date:
October 2, 1995
### II. Device Name
Right Angle Laser Fibers:
Proprietary: Not yet selected
Common: Laser fiber
Classification: Accessory to Laser-powered Instrument (unclassified)
Optilase Lasers:
Proprietary:
- Optilase® Model 1000 Nd:YAG Laser System
- Optilase® Model 4000 Nd:YAG Laser System
- Optilase® PL100 Model 1000-100 Nd:YAG Laser System
Common: Neodymium Yttrium Aluminum Garnet (Nd:YAG) laser system
Classification: Laser-powered Instrument
### III. Predicate Device
The predicate devices for the intended use are various configurations of the Trimedyne Right Angle Laser Fiber cleared under 510(k) number K944204 and the Trimedyne Optilase Nd:YAG Laser Systems cleared under 510(k) numbers K922497 and K932883.
TRIMEDYNE, INC.
2801 BARRANCA ROAD
P.O. BOX 57001, IRVINE, CA 92619-7001
714 559-5300, FAX 714 559-1330
800 33 5271
{1}
## IV. Device Description
The Right Angle Laser Fiber is a fiber optic Nd:YAG energy delivery system consisting of: a 3 meter length of buffered quartz fiber, a tip which allows delivery of laser energy, and a fiber optic connector.
The Optilase laser systems are continuous-wave Nd:YAG energy sources equipped with an aiming beam, adjustable power output, and a fiber optic connector.
## V. Intended Use
The Right Angle Laser Fiber and Optilase Lasers are indicated for a wide variety of surgical uses including incision, excision, resection, vaporization, ablation, coagulation, and hemostasis for the treatment of benign prostatic hyperplasia (BPH) and other urological conditions.
Viewing scopes, which can be either flexible or rigid, may be used for visualization during laser procedures.
## VI. Technological Characteristics
There are no technological characteristics differences between the devices cleared under this 510(k) and the predicate devices with regard to the intended use.
## VII. Nonclinical Tests
No nonclinical tests were submitted in this Premarket Notification.
## VIII. Clinical Tests
A prospective, controlled and well-matched study was undertaken to compare visual laser ablation of the prostate (VLAP) to transurethral resection of the prostate (TURP).
## IX. Conclusions Drawn from Testing
The clinical data show that the Trimedyne Right Angle Laser Fibers and Optilase Nd:YAG Laser Systems are safe and effective for treatment of Benign Prostatic Hyperplasia (BPH).