← Product Code [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX) · K252319

# Q-Switched Nd: YAG Laser machine (LFS-C13U) (K252319)

_Sanhe Lefis Electronics Co., Ltd. · GEX · Oct 20, 2025 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K252319

## Device Facts

- **Applicant:** Sanhe Lefis Electronics Co., Ltd.
- **Product Code:** [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX.md)
- **Decision Date:** Oct 20, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

The Q-Switched Nd: YAG Laser machine is indicated for use in tattoo removal, treatment of benign vascular lesions, treatment of benign pigmented lesions, incision, excision, ablation, vaporization of soft tissue for general dermatology as follows: 1064nm: -Tattoo Removal Dark ink: blue and black. -Treatment of Benign Pigmented Lesions Nevus of ota. 532nm: -Tattoo Removal Light ink: red, Light ink: sky blue and green. -Treatment of Benign Vascular Lesions Port wine birthmarks; Telangiectasias; Spider angioma; Cherry angioma; Spider nevi. -Treatment of Benign Pigmented Lesions Cafe-au-lait birthmarks; Solar lentiginos; Senile lentiginos; Becker's nevi; Freckles; Nevus spilus.

## Device Story

Q-Switched Nd:YAG laser system; delivers 1064nm or 532nm wavelengths via articulated arm. Neodymium ions in YAG crystal excited by pump source; photons reflected in optical resonant cavity to produce high-energy laser beam; 532nm achieved via frequency doubling. Used in medical institutions by clinicians for dermatological procedures. Operates via photothermal effect on target tissue. Clinician controls pulse output via footswitch. Output parameters (energy, spot size) adjusted to achieve therapeutic effect on skin lesions or tattoos. Benefits include precise tissue interaction for removal or ablation.

## Clinical Evidence

No clinical data. Bench testing only, including electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), laser safety (IEC 60601-2-22, IEC 60825-1), and biocompatibility (ISO 10993-5, -10, -23).

## Technological Characteristics

Nd:YAG laser; 1064nm/532nm wavelengths; Q-switched pulse mode; 1-10Hz repetition rate; 4-16ns pulse width; 2-10mm spot size. Articulated arm delivery. Internal distilled water cooling. Aluminum alloy treatment head. Class 4 laser. Electrical input 110/230VAC. Complies with IEC 60601-1, 60601-1-2, 60601-2-22, 60825-1.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Q-Switched Nd: YAG Laser machine ([K193609](/device/K193609.md))
- CuRAS Nd:YAG Laser ([K173038](/device/K173038.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

October 20, 2025

Sanhe LEFIS Electronics Co., Ltd.
Dandan Wang
Registration Manager
Bldg 11 #1-101, Phase 1, Zhongnan High tech·Yanjiao Science &amp;
Technology Innovation Smart Valley Industrial Park 1
Langfang, Hebei 065201
China

Re: K252319
Trade/Device Name: Q-Switched Nd: YAG Laser machine (LFS-C13U)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: GEX
Dated: July 25, 2025
Received: July 25, 2025

Dear Dandan Wang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K252319 - Dandan Wang
Page 2

Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn

{2}

K252319 - Dandan Wang
Page 3

(https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

TANISHA L. HITHE -S
Digitally signed by
TANISHA L. HITHE -S
Date: 2025.10.20
23:05:05 -04'00'

Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF

|  DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.  |
| --- | --- |
|  510(k) Number (if known) K252319  |   |
|  Device Name Q-Switched Nd: YAG Laser machine (LFS-C13U)  |   |
|  Indications for Use (Describe) The Q-Switched Nd: YAG Laser machine is indicated for use in tattoo removal, treatment of benign vascular lesions, treatment of benign pigmented lesions, incision, excision, ablation, vaporization of soft tissue for general dermatology as follows: 1064nm: -Tattoo Removal Dark ink: blue and black. -Treatment of Benign Pigmented Lesions Nevus of ota. 532nm: -Tattoo Removal Light ink: red, Light ink: sky blue and green. -Treatment of Benign Vascular Lesions Port wine birthmarks; Telangiectasias; Spider angioma; Cherry angioma; Spider nevi. -Treatment of Benign Pigmented Lesions Cafe-au-lait birthmarks; Solar lentiginos; Senile lentiginos; Becker's nevi; Freckles; Nevus spilus.  |   |
|  Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)  |   |
|  CONTINUE ON A SEPARATE PAGE IF NEEDED.  |   |
|  This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*  |   |
|  The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov  |   |
|  "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."  |   |

{4}

510(k) K252319

# 510(k) Number: K252319

# 510(k) Summary

This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of Title 21, CFR Section 807.92.

1. Date of Submission: 2025/7/25
2. Sponsor Identification

**Sanhe LEFIS Electronics Co., Ltd.**

Building 11 #1-101, Phase 1, Zhongnan High tech· Yanjiao Science and Technology Innovation Smart Valley Industrial Park 1, South Side of Liushan Street and West Side of Gushan West Road, Yanjiao High tech Zone, Sanhe City, 065201 Langfang City, Hebei Province, China

Contact Person: Dandan Wang

Position: Registered Manager

Tel: +86-18131561031

Email: 597782121@qq.com

3. Designated Submission Correspondent

Ms. Dandan Wang

Tel: +86-18131561031

Fax: +86-316-3096027

Email: 597782121@qq.com

4. Identification of Proposed Device

Trade Name: Q-Switched Nd: YAG Laser machine

Model Name: LFS-C13U

Common Name: Surgical Laser Device

**Regulatory Information**

Classification Name: Laser Surgical Instrument for Use In General And Plastic Surgery And In Dermatology

Classification: II

Product Code: GEX

Regulation Number: 21CFR 878.4810

Review Panel: General &amp; Plastic Surgery

{5}

510(k) K252319

# 5. Intended Use and Indications for Use

The Q-Switched Nd: YAG Laser machine is indicated for use in tattoo removal, treatment of benign vascular lesions, treatment of benign pigmented lesions, incision, excision, ablation, vaporization of soft tissue for general dermatology as follows:

1064nm:

-Tattoo Removal

Dark ink: blue and black.

-Treatment of Benign Pigmented Lesions

Nevus of ota.

532nm:

-Tattoo Removal

Light ink: red, Light ink: sky blue and green.

- Treatment of Benign Vascular Lesions

Port wine birthmarks; Telangiectasias; Spider angioma; Cherry angioma; Spider nevi.

-Treatment of Benign Pigmented Lesions

Cafe-au-lait birthmarks; Solar lentiginos; Senile lentiginos; Becker's nevi; Freckles; Nevus spilus.

# 6. Environment of Use

Medical institutions, including hospitals, clinics, etc.

# 7. Product Description

The Q-Switched Nd: YAG Laser machine is a laser system which delivers light at a wavelength  $1064\mathrm{nm}$  or  $532\mathrm{nm}$ . The Q-Switched Nd: YAG Laser machine is indicated for use in tattoo removal, treatment of benign vascular lesions, treatment of benign pigmented lesions, incision, excision, ablation, vaporization of soft tissue for general dermatology. The device comprises a main unit, light guide arm and footswitch.

Table 1 Main Components introduction

|  Component name | Function  |
| --- | --- |
|  Main unit | Main Interface  |
|  Light guide arm | Articulated arm for holding of Treatment Probe  |
|  Footswitch. | Control pulse light output  |

Table 2 external dimensions

|  Overall weight | ≤100kg  |
| --- | --- |
|  Volume (without light guide system) | 1000mm×480mm×1050mm  |

{6}

510(k) K252319

Working Principle

The working principle of Q-Switched Nd: YAG Laser machine is that neodymium ions in neodymium doped yttrium aluminum garnet crystals are excited by the energy provided by the pump source and release photons. These photons are repeatedly reflected and stimulated to emit in the optical resonant cavity of the crystal, ultimately producing a specific wavelength (mainly 1064nm) and high energy density laser beam. It can also be converted from a 1064nm wavelength laser beam to a 532nm laser beam through frequency doubling technology; Then, it is effectively transmitted to the treatment site through the beam transmission device and plays a therapeutic role through the photothermal effect.

8. Materials

|  Components | Material | Category | Contact Level | Contact Duration  |
| --- | --- | --- | --- | --- |
|  Treatment Head | Al Alloy | Surface device | Intact skin | Short-term (<24h)  |

The treatment hand piece used in the Machine has passed the Biocompatibility test. For details, please refer to "Biocompatibility Discussion".

9. Identification of Predicate Device(s)

510(k) Number: K193609

Product Name: Q-Switched Nd: YAG Laser machine

Manufacturer: Beijing Lead Beauty S&amp;T Co., Ltd.

510(k) Number: K173038

Product Name: CuRAS Nd:YAG Laser

Manufacturer: Ilooda Co., Ltd

10. Non-Clinical Test Conclusion

Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

&gt; ANSI/AAMI ES60601-1:2005, ES60601-1:2005/AMD1:2012, ES60601-1:2005/AMD2:2021
Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
$\succ$  IEC 60601-1-2 Edition 4.1 2020-09, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -

{7}

510(k) K252319

## Requirements and tests

- IEC 60601-2-22 2019 Medical electrical equipment - Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
- IEC 60825-1 2014 Safety of laser products - Part 1: Equipment classification, and requirements

The body-contacting components of this device are the treatment head. The biocompatibility evaluation for the Q Switched Nd: YAG Laser machine was conducted in accordance with the guidance “Use of International Standard ISO 10993-1, "Biological evaluation of medical devices -Part 1: Evaluation and testing within a risk management process"” as recommended by FDA. The treatment head is considered skin and subcutaneous tissue contacting for a duration of less than 24 hours. The biocompatible testing included In Vitro Cytotoxicity, Skin Sensitization and Intracutaneous Reactivity was conducted in compliance with:

- ISO 10993-5: 2009 Biological Evaluation Of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity
- ISO 10993-10:2021 Biological evaluation of medical devices - Part 10: Tests for skin sensitization.
- ISO 10993-23:2021 Biological evaluation of medical devices — Part 23: Tests for irritation.

In summary, the results of the non-clinical tests indicate that the device complies with relevant international standards in terms of basic safety and essential performance, electromagnetic interference, laser safety, and biocompatibility. This demonstrates the safety and effectiveness of the device, providing a scientific basis for further clinical application and market access. Detailed test methods and results can be found in the test report.

{8}

510(k) K252319

We have also conducted:

&gt; Software verification and validation test according to the requirements of the FDA "Content of Premarket Submissions for Device Software Functions". The software for this device was considered as an "Enhanced" documentation level. Software validation demonstrated that the software functions as specified in the software requirement specifications.
&gt; Bench performance testing to show that the device delivers set laser energy parameters within specifications.

# Sterilization and Shelf-Life

The proposed device is not provided sterile and does not need to be sterilized. The handpiece and the body are cleaned with a soft cloth moistened with isopropyl alcohol or ethanol of  $75\%$  strength or higher. The proposed device is reusable and does not have a restricted shelf-life.

11. Technological characteristics and substantial equivalence:

Table 1 General Comparison

|  ITEM | Proposed Device | Predicate Device1 K193609 | Predicate Device 2 K173038 | Remark  |
| --- | --- | --- | --- | --- |
|  Device Name | Q-Switched Nd:YAG Laser machine | Q-Switched Nd:YAG Laser machine | CuRAS Nd:YAG Laser | SE  |
|  Manufacturer | Sanhe LEFIS Electronics Co., Ltd. | Beijing Lead Beauty S&T Co., Ltd. | Ilooda Co., Ltd | SE  |
|  Product Code | GEX | GEX | GEX | SE  |
|  Regulation No. | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | SE  |
|  Class | Class II | Class II | Class II | SE  |
|  Intended Use | The Q-Switched Nd:YAG Laser machine is indicated for use in | The Q-Switched Nd:YAG Laser Therapy System is indicated | The CuRAS Nd:YAG laser | SE  |
|   | the device is indicated for use in the device is indicated for use in the device is indicated for use in the device is indicated for use in the device is indicated for use in the device is indicated for use in the device is indicated for use in the device is indicated for use in the device is indicated for use in the device is |  |  |   |

{9}

510(k) K252319

|   | tattoo removal, treatment of benign vascular lesions, treatment of benign pigmented lesions, incision, excision, ablation, vaporization of soft tissue for general dermatology as follows: 1064nm: -Tattoo Removal Dark ink: blue and black. -Treatment of Benign Pigmented Lesions Nevus of ota. 532nm: -Tattoo Removal Light ink: red, Light ink: sky blue and green. - Treatment of Benign Vascular Lesions Port wine birthmarks; Telangiectasias; Spider angioma; Cherry angioma; Spider nevi. -Treatment of Benign Pigmented Lesions Cafe-au-lait birthmarks; Solar lentiginos; Senile lentiginos; Becker's nevi; Freckles; Nevus spilus. | for use in tattoo removal, treatment of benign vascular lesions, treatment of benign pigmented lesions, incision, excision, ablation, vaporization of soft tissue for general dermatology as follows: 1064nm: -Tattoo Removal Dark ink: blue and black. -Treatment of Benign Pigmented Lesions Nevus of ota. 532nm: -Tattoo Removal Light ink: red, Light ink: sky blue and green. - Treatment of Benign Vascular Lesions Port wine birthmarks; Telangiectasias; Spider angioma; Cherry angioma; Spider nevi. -Treatment of Benign Pigmented Lesions Cafe-au-lait birthmarks; Solar lentiginos; Senile lentiginos; Becker's nevi; Freckles; Nevus spilus. | system in indicated for: the incision,excision, ablation, vaporization of soft tissues for general dermatology, dermatologic and general surgical procedures for coagulation and hemostasis. 532nm Wavelength:-Tattoo removal:light ink(red,tan,purple,orange,skyblue,green -Removal of Epidermal Pigmented Lesions -Removal of Minor Vascular Lesions including but not limited to telangiectasias -Treatment of Lentigines -Treatment of Cafe-Au-Lait -Treatment of Seborrheic Keratoses -Treatment of Post Inflammatory  |
| --- | --- | --- | --- |

6

{10}

510(k) K252319

Table 2 Technological characteristics

|  ITEM | Proposed Device | Predicate Device1K193609 | Predicate Device 2 K173038 | Remark  |
| --- | --- | --- | --- | --- |
|  Laser | ND:YAG | ND:YAG | ND:YAG | SE  |
|  Wavelength | 1064nm
532nm | 1064nm
532nm | 1064nm
532nm | SE  |

{11}

510(k) K252319

|  ITEM | Proposed Device | Predicate Device1K193609 | Predicate Device 2 K173038 | Remark  |
| --- | --- | --- | --- | --- |
|  Output Power (Maximum) | 1064nm: 800mJ 532nm:250mJ | 500mJ for 1064nm 260mJ for 532nm | Max 1.6J @1064 nm Max 0.4J @532 nm | Different Note 1  |
|  Spot Size | 2-10mm | 2-10mm | 2mm-10mm | SE  |
|  Pulse Width | 4ns-16ns | 6ns±1ns | 5-20ns | SE  |
|  Aiming Beam Wavelength | 630 nm-670 nm | 635nm | 635nm | Different Note 2  |
|  Max.Aiming BeamPower Watts | ≤1mW | 0.1mW-5mW | 5mW | Different Note 2  |
|  Laser Class | Class 4 | Class 4 | Class 4 | SE  |
|  Laser output mode | Q-switched pulse | Q-switched pulse | Q-switched pulse | SE  |
|  Repetition rate | 1-10Hz | 1-10Hz | 1-15Hz | SE  |
|  Beam delivery | articulated arm light guide | articulated arm light guide | Articulated arm | SE  |
|  Cooling System | internal distilled water circulating cooling | internal distilled water circulating cooling | / | SE  |
|  Power Input | 110VAC, 60Hz; 230VAC, 50Hz | / | 220-230VAC, 50-60 Hz | SE  |

Table 3 Safety Comparison

{12}

510(k) K252319

|  ITEM | Proposed Device | Predicate Device1 K193609 | Predicate DeviceK173038 | Remark  |
| --- | --- | --- | --- | --- |
|  Patient contact material | Aluminum alloy | Aluminum alloy | / | SE  |
|  Biocompatibility | Passed the test as per ISO 10993-5, ISO 10993-10andISO10993-23 | Passed the test as per ISO 10993-5 and ISO 10993-10 | / | SE  |
|  Electrical Safety | Comply with IEC 60601-1, IEC 60601-2-22 | Comply with IEC 60601-1, IEC 60601-2-22 | Comply with IEC 60601-1, IEC 60601-2-22 | SE  |
|  EMC | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 | SE  |
|  Laser Safety | Comply with IEC 60601-2-22, IEC 60825 | Comply with IEC 60601-2-22, IEC 60825 | Comply with IEC 60601-2-22, IEC 60825 | SE  |

Note 1:

Different - Output Power (Maximum)

The proposed device has different Maximum pulse energy from the predicate device. Additionally, the reference devices K193609 and K173038 that we have listed have similar intended uses to the proposed device. The maximum pulse energy of the proposed device is greater than that of K193609, but the maximum pulse energy of the reference device K173038 encompasses the pulse energy of the proposed device. Therefore, we believe that a maximum pulse energy of 800mJ for 1064nm is a mature parameter that will not affect the safety and effectiveness of the equipment. The difference in output energy will not impact safety and effectiveness, as the ultimate safety and effectiveness regarding clinical indications will depend on the amount of energy output per unit area, which will produce thermal effects to the irradiated skin area of the patient to achieve the claimed indication for use. Moreover, the proposed device has passed the IEC60601-1 test, IEC60601-1-2 test, IEC60601-2-22 test, IEC60825-1 test, and performance test, ensuring the safety of the product.

Note 2:

{13}

S10(k) K252319

Different - Aiming beam wavelength and Power

The proposed device has different Aiming beam wavelength and Power from the predicate device. For the difference on Aiming beam wavelength and Power between the predicate and proposed device(s), we can see that the proposed device has similar Aiming beam wavelength and Power with predicate device (K193609). They are only with minor difference. And we think this minor difference will not affect the effectiveness and safety.

And the purpose of aiming beam is only for aiming. And at the same power, the difference in the Wavelength and Power of the aiming beam will only cause differences in visual color and brightness, and will not affect the effectiveness and safety.

**Conclusion:**

Q-Switched Nd: YAG Laser machine is substantial equivalent to the predicate device.

**12. Clinical Test Conclusion**

No clinical study is included in this submission.

**13. Substantially Equivalent(SE) Conclusion**

Q-Switched Nd: YAG Laser machine the same intended use, similar indications for use, the same technological characteristics, the same energy used, and the same operating principles as its predicates. The non-clinical data and performance testing reports in this submission demonstrate that Q-Switched Nd: YAG Laser machine meets the expected performance requirements. Any difference between the subject and predicate device do not raise new issues of safety or effectiveness. Based on above analysis, Q-Switched Nd: YAG Laser machine is substantial equivalent to the cited predicate device.

10

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K252319](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K252319)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
