K241642 · Beijing Mega Beauty Technology Co., Ltd. · GEX · Aug 2, 2024 · General, Plastic Surgery
Device Facts
Record ID
K241642
Device Name
Diode Laser Hair Removal System (MBT-Diode Laser)
Applicant
Beijing Mega Beauty Technology Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Aug 2, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Diode Laser Hair Removal System (Model: MBT-Diode Laser) is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term. stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.
Device Story
Diode laser system for hair removal; utilizes semiconductor diode emitting 808nm infrared radiation; energy absorbed by melanin. System comprises main console (CPU, power supply, cooling system), handpiece, and footswitch. Operator activates laser via footswitch or handle button. Used in clinical settings by trained professionals. Laser energy delivered to treatment area via handpiece. Permanent hair reduction defined as stable reduction in hair regrowth at 6, 9, and 12 months post-treatment. Benefits patient through long-term hair reduction.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
808nm diode laser; Class IV; 15mm x 15mm spot size; 1-55.96 J/cm² fluence; 2-240ms pulse duration; 1-10Hz frequency. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1. Biocompatibility per ISO 10993-5, ISO 10993-10, ISO 10993-23.
Indications for Use
Indicated for hair removal and permanent hair reduction on all skin types (Fitzpatrick I-VI), including tanned skin.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration" in a larger font. The word "Administration" is written in a smaller font below the other words.
August 2, 2024
Beijing Mega Beautv Technology Co.,Ltd. % Ray Wang General Manager Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District Beijing, 102401 China
Re: K241642
Trade/Device Name: Diode Laser Hair Removal System (MBT-Diode Laser) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: June 6, 2024 Received: June 7, 2024
Dear Ray Wang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{1}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product: and 21 CFR 820.100. Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations. please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).
{2}------------------------------------------------
Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
TANISHA TANISHA L. HITHE -S L. HITHE -S Date: 2024.08.0
Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
K241642
Device Name
Diode Laser Hair Removal System (MBT-Diode Laser)
Indications for Use (Describe)
The Diode Laser Hair Removal System (Model: MBT-Diode Laser) is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term. stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### K241642 510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
The assigned 510(k) Number: __________________________________________________________________________________________________________________________________________________
- Date of Preparation:2024/06/06 1.
- 2. Sponsor Identification
## Beijing Mega Beauty Technology Co.,Ltd.
Room 1-524, Building No.4, Xinyuan Road No.27, Daxing District, Beijing, 102629 China.
Contact Person: Alex Yu Position: General Manager Tel: +86 13581869234 Fax: +86 58497698 Email: alex@mbt-laser.com
- 3. Designated Submission Correspondent
Mr. Ray Wang
### Beijing Believe-Med Technology Service Co., Ltd.
Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District, Beijing, 102401, China
Tel: +86-18910677558 Fax: +86-10-56335780 Email: information@believe-med.com
{5}------------------------------------------------
- 4. Identification of Proposed Device
Trade Name: Diode Laser Hair Removal System Common Name: Powered Laser Surgical Instrument Model(s): MBT-Diode Laser
Regulatory Information: Classification Name: Powered Laser Surgical Instrument Classification: II; Product Code: GEX; Regulation Number: 21 CFR 878.4810; Review Panel: General & Plastic Surgery;
- ર. Identification of Predicate Device(s) Predicate Device 510(k) Number: K211335 Product Name: Diode Laser Machine Manufacturer: Zhengzhou Bestview St Co., Ltd.
Reference Device 510(k) Number: K200525 Product Name: Medical Diode Laser Hair Removal System Manufacturer:Weifang Mingliang Electronics CO., LTD.
- Device Description: 6.
Diode Laser Hair Removal System (Model: MBT-Diode Laser) is a surgical device, which is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI). The Dermatological Diode Laser Systems utilize a semiconductor diode with invisible infrared radiation as a laser source to emit 808nm wavelength laser which is absorbed by melanin. The laser power is delivered to the treatment area via laser handle. The emission laser is activated by footswitch or handle button.
The Diode Laser Hair Removal System consists of the following major components
- The main console unit that incorporates the main CPU board, power supply modules, laser a. power supply, laser device, cooling system and switching module
- Handpiece b.
- Footswitch C.
- Indication For Use Statement: 7.
The Diode Laser Hair Removal System (Model: MBT-Diode Laser) is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.
{6}------------------------------------------------
#### 8. Substantially Equivalent (SE) Comparison
| Table 7-1 General Comparison | | | | |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Item | Proposed Device | Predicate Device<br>K211335 | Reference Device<br>K200525 | Remark |
| Device Name | Diode Laser Hair<br>Removal System | Diode Laser<br>Machine | Diode Laser<br>Machine | / |
| Classification<br>Regulation | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | SAME |
| Classification<br>Panel | General & Plastic<br>Surgery | General & Plastic<br>Surgery | General & Plastic<br>Surgery | SAME |
| Class | II | II | II | SAME |
| Product Code | GEX | GEX | GEX | SAME |
| Indication for use | The Diode Laser Hair<br>Removal System is intended<br>for hair removal, permanent hair<br>reduction on all skin<br>types (Fitzpatrick skin<br>type I-VI), including<br>tanned skin.<br>Permanent hair<br>reduction is defined<br>as the long-term,<br>stable reduction in<br>the number of hairs<br>regrowing when<br>measured at 6, 9, and<br>12 months after the<br>completion of a<br>treatment regime. | The Diode Laser<br>Machine<br>Model:BM-100<br>is intended<br>for hair removal, permanent hair<br>reduction on all skin<br>types (Fitzpatrick skin<br>type I-VI), including<br>tanned skin.<br>Permanent hair<br>reduction is defined as<br>the long-term, stable<br>reduction in the<br>number of hairs<br>regrowing when<br>measured at 6, 9, and<br>12 months after the<br>completion of a<br>treatment regime. | The Medical Diode<br>Laser Hair<br>Removal System is<br>used for<br>permanent reduction<br>in hair<br>regrowth defined as a<br>long term,<br>stable reduction in the<br>number of<br>hairs re-growing<br>when measured at<br>6, 9 and 12 months<br>after the<br>completion of a<br>treatment regimen.<br>Use on all skin types<br>( Fitzpatrick<br>I-VI), including<br>tanned skin. | SAME |
Table 7-1 General Comparison
# Table 7-2 Performance Comparison
| ITEM | Proposed Device | Predicate Device<br>K211335 | Reference Device<br>K200525 | Remark |
|-------------------------|-----------------|-----------------------------|-----------------------------|--------|
| Laser Type | Diode Laser | Diode Laser | Diode Laser | SAME |
| Laser<br>Classification | Class IV | Class IV | Class IV | SAME |
| Laser | 808nm | 808nm | 808nm | SAME |
{7}------------------------------------------------
| Wavelength | | | | |
|----------------|------------------------|--------------------------|---------------------------|-----------|
| Spot Size | 15mm×15mm | 2.25cm²<br>(15mm×15mm) | 1.44cm2<br>(1.2 x 1.2 cm) | SIMILAR |
| Fluence | 1-55.96J/cm² | 0-120J/cm² | 48 J/cm² | Different |
| Frequency | 1-10Hz | 1-10Hz | 1-10Hz | SAME |
| Pulse Duration | 2-240ms | 10-400ms | 1-300ms | Different |
| Power Supply | 110 VAC/50Hz-60Hz | 220/110<br>VAC/50Hz-60Hz | 220/110<br>VAC/50Hz-60Hz | Different |
| Dimension | 650mm×580mm×<br>1210mm | 112 cm × 42 cm ×<br>60cm | / | Different |
| Weight | 68Kg | 63Kg | / | Different |
### Analysis :
### Different - Fluence
The proposed device has different Fluence from the predicate device.
For the difference on fluence between the predicate and proposed device(s), The Fluence of Predicte Device K211335 is 0-120./cm², while the propose device is 55.96 J/cm². It is lower than 120 J/cm², the safety of the product is no problem. However, the effectiveness cannot be determined, so we selected Reference Device K200525 with the same intended use for Fluence comparison,The Fluence of the Reference Device is 48J/cm², which is similar to the Fluence of the proposed device, so the effectiveness of the product is not affected.And the proposed device has passed the IEC60601-1 test, IEC60601-1-2 test, IEC60601-2-22 test. IEC60825-1 test and performance test, the safety and performance of the product can be ensured.
### Different - Pulse Duration
The proposed device has different pulse duration from the predicate device.
For the difference on pulse duration between the predicate and proposed device(s), we can see that the pulse duration range of proposed device is within the range of the predicate device. The pulse duration between the proposed device and the predicate device only with minor difference. And we think this minor difference will not affect the effectiveness and safety. And the proposed device has passed the IEC60601-1 test, IEC60601-1-2 test, IEC60601-2-22 test, IEC60825-1 test and performance test, the safety and performance of the product can be ensured.
### Different - Power Supply
The power supply for the proposed device is different from the predicate device. However, electrical safety and EMC test has been conducted on the proposed device and the test result show that the device can work normally under this power supply. Therefore, this difference will not affect safety and effectiveness of the proposed device.
### Different - Dimension, Weight
The proposed device is different in dimension and weight from the predicate device. By complying with IEC 60601-1, the mechanical performance of the proposed device is determined to be accepted, therefore,
{8}------------------------------------------------
this difference of dimension and weight have no effect the effectiveness and safety.
| Item | Proposed Device | Predicate Device<br>K211335 | Remark |
|------------------------------------------------|--------------------------------------------|--------------------------------------------|--------|
| EMC, Electrical and Laser Safety | | | |
| Electrical Safety | Comply with IEC 60601-1, IEC<br>60601-2-22 | Comply with IEC 60601-1, IEC<br>60601-2-22 | SAME |
| EMC | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 | SAME |
| Laser Safety | Comply with IEC 60601-2-22, IEC<br>60825 | Comply with IEC 60601-2-22,<br>IEC 60825 | SAME |
| Patient Contact Materials and Biocompatibility | | | |
| Cytotoxicity | No Cytotoxicity | No Cytotoxicity | SAME |
| Sensitization | No evidence of sensitization | No evidence of sensitization | SAME |
| Irritation | No evidence of irritation | No evidence of irritation | SAME |
Table 7-3 Safety Comparison
#### Non-Clinical Test Conclusion 9.
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- A IEC 60601-1 Edition 3.2 2020-08, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- A IEC 60601-1-2 Edition 4.1 2020-09, Medical Electrical Equipment-Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility-Requirements And Tests
- A IEC 60825-1:2014, Safety of laser products - Part 1: Equipment classification, and requirements
- > IEC 60601-2-22: 2012 Medical electrical equipment - Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
- > ISO 10993-5: 2009 Biological Evaluation of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity
ISO 10993-10 Fourth edition 2021, Biological evaluation of medical devices - Part 10: Tests for skin A sensitization
> ISO 10993-23 First edition 2021-01, Biological evaluation of medical devices - Part 23: Tests for irritation
- 10. Clinical Test Conclusion
No clinical study is included in this submission.
{9}------------------------------------------------
- 11. Substantially Equivalent (SE) Conclusion
Based on the nonclinical tests performed, the subject device is as safe, as effective, and performs as well as the legally marketed predicate device, Diode Laser Machine Model:BM-100 cleared under K211335.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.