← Product Code [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX) · K241144

# Picosecond Nd:YAG Laser (PICOCAREMAJESTY) (K241144)

_Wontech Co., Ltd. · GEX · Jun 24, 2024 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K241144

## Device Facts

- **Applicant:** Wontech Co., Ltd.
- **Product Code:** [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX.md)
- **Decision Date:** Jun 24, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

The PICOCAREMAJESTY is indicated for the following at the specified wavelength: · 1064 nm: Removal of tattoos for all skin types (Fitzpatrick I-VI) to treat the following tattoo colors: black, brown, green, blue and purple. · 532 nm: Removal of tattoos for Fitzpatrick skin types I-III to treat the following tattoo colors: red, yellow and orange. · Treatment of benign pigmented lesions for Fitzpatrick skin types I-IV.

## Device Story

Solid-state Nd:YAG laser system delivering 1064 nm or 532 nm wavelengths in picosecond pulses. System comprises main unit, articulated arm, handpiece, and foot switch. User selects fluence and frequency via LCD touch panel on main unit; energy output adjusted automatically based on selected fluence. Laser energy delivered to skin via handpiece for tattoo removal and treatment of benign pigmented lesions. Used in clinical settings by trained professionals. Output provides therapeutic laser energy to target chromophores in skin; benefits include removal of tattoos and pigmented lesions.

## Clinical Evidence

No clinical studies were performed. Substantial equivalence is supported by bench testing, including electrical safety (AAMI ES60601-1), electromagnetic compatibility (IEC 60601-1-2), usability (IEC 60601-1-6), and laser safety (IEC 60825-1).

## Technological Characteristics

Solid-state Nd:YAG laser; 1064 nm and 532 nm wavelengths; 275-300 ps pulse width; max 10 Hz repetition rate; 2-10 mm spot size. Handpiece tip material: Aluminum powder (CAS 7429-90-5). Delivery via articulated arm. Software: Moderate level of concern. Standards: AAMI ES60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-22, IEC 60825-1.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- PICOANDY ([K234104](/device/K234104.md))
- PICOHI ([K201773](/device/K201773.md))

## Reference Devices

- PICOCARE Family ([K181272](/device/K181272.md))
- PICOCAREMAJESTY ([K212127](/device/K212127.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 24, 2024

Won Tech Co., Ltd. Hyun Sik Yoon Team Leader 64 Techno8-ro 64, Yuseong-gu Daejeon, 34028 Korea, South

### Re: K241144

Trade/Device Name: Picosecond Nd:YAG Laser (PICOCAREMAJESTY) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: April 15, 2024 Received: April 25, 2024

Dear Hyun Sik Yoon:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).

{2}------------------------------------------------

Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Tanisha L. Digi Hithe -5 Date: 2024.06.24
15:13:13 -04'00' Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use

Submission Number (if known)

K241144

Device Name

Picosecond Nd:YAG Laser (PICOCAREMAJESTY)

Indications for Use (Describe)

The PICOCAREMAJESTY is indicated for the following at the specified wavelength:

· 1064 nm: Removal of tattoos for all skin types (Fitzpatrick I - VI) to treat the following tattoo colors: black,brown, green, blue and purple.

· 532 nm: Removal of tattoos for Fitzpatrick skin types I · III to treat the following tattoo colors: red, yellow and orange.

· Treatment of benign pigmented lesions for Fitzpatrick skin types I-IV.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for WONTECH. The logo consists of an orange circle with a stylized white "W" inside of it. To the right of the circle is the text "WONTECH", with "WON" in orange and "TECH" in gray. The logo is simple and modern, and the use of orange and gray gives it a professional look.

# 510(k) Summary K241144

[As required by 21 CFR 807.92]

#### 1. Date Prepared [21 CFR 807.92(a)(a)]

June 6th, 2024

#### Submitter's Information & Contact Person [21 CFR 807.92(a)(1)] 2.

- Name of Manufacturer: WON TECH Co., Ltd.
- Address: 64 Techno 8-ro, Yuseong-gu, Daejeon, 34028, Republic of Korea Hyun Sik Yoon - Contact Name: - Telephone No.: +82-10-6750-5346 - Fax No.: +82-70-7836-0110 - Email Address: yoonhs21@wtlaser.com
- 

#### 3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)|

Common name: Nd:YAG Laser System Trade name: PICOCAREMAJESTY

| Classification Description        | 21 CFR Section | Product Code |
|-----------------------------------|----------------|--------------|
| Powered Laser Surgical Instrument | 878.4810       | GEX          |

As stated in 21 CFR, parts 878.4810, this generic types of devices has been classified as Class II.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for WONTECH. The logo consists of an orange circle with a stylized "W" inside, followed by the text "WONTECH". The "WON" part of the text is in orange, while the "TECH" part is in gray.

#### 4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)|

The identified predicate devices within this submission are shown as follow:

#### #1 Predicate device

| • 510(k) Number:       | K234104                           |
|------------------------|-----------------------------------|
| • Applicant:           | WON TECH Co., Ltd.                |
| • Classification Name: | Powered Laser Surgical Instrument |
| • Trade Name:          | PICOANDY                          |

#### #2 Predicate device

| • 510(k) Number:       | K201773                           |
|------------------------|-----------------------------------|
| • Applicant:           | WON TECH Co., Ltd.                |
| • Classification Name: | Powered Laser Surgical Instrument |
| • Trade Name:          | PICOHI                            |

#### ર. Description of the Device [21 CFR 807.92(a)(4)]

The PICOCAREMAJESTY is the solid state laser capable of delivering energy at wavelengths of 1064mm or 532nm at short durations. The device system consists of a main unit, an articulated arm, a handpiece and a foot switch. The laser output is delivered to the skin through the articulated arm delivery system terminated by the handpiece. The fluence (energy density) and frequency are controlled from the LCD display/Touch Pad located on the front of the main unit. The LCD display is used to obtain feedback from the system, such as the number of pulses delivered or spot size selected.

For treatment, the user can select the appropriate fluence value. (The treatmented is provided in the user Manual) The energy is changed automatically in accordance with the selected fluence (down) button and then, the fluence is increased / decreased by 0.1 J/cm2.

The selectable fluence values are 0.2 to 10.0 J/cm2 at 1064nm and 0.1 to 2.5 J/cm2 at 532nm

#### 6. Indications for Use [21 CFR 807.92(a)(5)]

The PICOCAREMAJESTYis indicated for the following at the specified wavelength:

· 1064 nm: Removal of tattoos for all skin types (Fitzpatrick I-VI) to treat the following tattoo colors: black, brown, green, blue and purple.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for WONTECH. The logo consists of an orange circle with a stylized "W" inside, followed by the text "WONTECH" in a sans-serif font. The "WON" part of the text is in orange, while the "TECH" part is in gray.

- · 532 nm: Removal of tattoos for Fitzpatrick skin types I-III to treat the following tattoo colors: red, yellow and orange.
- · Treatment of benign pigmented lesions for Fitzpatrick skin types I-IV.

#### Determination of Substantial Equivalence [21 CFR 807.92(a)(6) and 21 CFR 807.92(b)] 7.

There are no significant differences between the PICOCAREMAJESTY and the predicate devices that would adversely affect the use of the product. It is substantially equivalent to this device in design, function, and technical characteristics.

|                        | Proposed Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Predicate Device #1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Predicate Device #2                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | SE decision |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| K Number               | -                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | K234104                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | K201773                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | -           |
| Manufacturer           | WON TECH Co., Ltd.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | WON TECH Co., Ltd.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Hironic Co., Ltd                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | -           |
| Model                  | PICOCAREMAJESTY                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | PICOANDY                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | PICOHI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | -           |
| Indications<br>for Use | The<br>PICOCAREMAJESTY is<br>indicated for the<br>following at the specified<br>wavelength:<br><br>• 1064 nm: Removal of<br>tattoos for all skin types<br>(Fitzpatrick I-VI) to treat<br>the following tattoo<br>colors: black, brown,<br>green, blue and purple.<br><br>• 532 nm: Removal of<br>tattoos for Fitzpatrick skin<br>types I-III to treat the<br>following tattoo colors:<br>red, yellow and orange.<br><br>• Treatment of benign<br>pigmented lesions for<br>Fitzpatrick skin types I-<br>IV. | The PICOANDY is<br>indicated for the<br>following at the specified<br>wavelength:<br><br>• 1064 nm: Removal of<br>tattoos for all skin types<br>(Fitzpatrick I-VI) to treat<br>the following tattoo<br>colors: black, brown,<br>green, blue and purple.<br><br>• 532 nm: Removal of<br>tattoos for Fitzpatrick<br>skin types I-III to treat<br>the following tattoo<br>colors: red, yellow and<br>orange.<br><br>• Treatment of benign<br>pigmented lesions for<br>Fitzpatrick skin types I-<br>IV. | The PICOHI is indicated<br>for the following at the<br>specified wavelength:<br><br>• 1064 nm: Removal of<br>tattoos for all skin types<br>(Fitzpatrick I-VI) to treat<br>the following tattoo<br>colors: black, brown,<br>green, blue and purple.<br><br>• 532 nm: Removal of<br>tattoos for Fitzpatrick<br>skin types I-III to treat<br>the following tattoo<br>colors: red, yellow and<br>orange.<br><br>• Treatment of benign<br>pigmented lesions for<br>Fitzpatrick skin types I-<br>IV. | Same        |
| Anatomical<br>site     | Skin and subcutaneous<br>tissue                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Skin and subcutaneous<br>tissue                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Skin and subcutaneous<br>tissue                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same        |
| Wavelength             | 1064 & 532nm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | 1064 & 532nm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 1064 & 532nm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Same        |

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for WONTECH. The logo consists of an orange circular icon with a stylized "W" inside, followed by the text "WONTECH". The "WON" part of the text is in orange, while the "TECH" part is in gray.

|                                | Proposed Device                         | Predicate Device #1                     | Predicate Device #2                    | SE decision                                                                                                               |
|--------------------------------|-----------------------------------------|-----------------------------------------|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| Pulse Width                    | 275~300ps                               | 350~550 ps                              | 275~300ps                              | Same with Predicate<br>Device #2                                                                                          |
| Pulse Energy                   | 1. Zoom 1064nm : 30 ~<br>500mJ ±10%     | 1. Zoom 1064nm : 30 ~<br>550mJ ±10%     | 1. Zoom 1064nm : 10~<br>508mJ ±20%     | Diffenence<br>However, the pulse<br>energy of<br>PICOCAREMAJESTY<br>is included in those of<br>the Predicate Device<br>#2 |
|                                | 2. Zoom 532nm : 10 ~<br>250mJ ±10%      | 2. Zoom 532nm : 10 ~<br>200mJ ±10%      | 2. Zoom 532nm : 4 ~<br>255mJ ±20%      |                                                                                                                           |
|                                | 3. Collimation Mode: 38<br>~ 500mJ ±10% | 3. Collimation Mode: 38<br>~ 500mJ ±10% | 3. Collimation Mode: 7 ~<br>471mJ ±20% |                                                                                                                           |
|                                | 4. MLA (1064nm): 30 ~<br>500mJ ±10%     | 4. MLA (1064nm): 30 ~<br>550mJ ±10%     | 4. ZMLA (1064nm): 5 ~<br>508mJ ±20%    |                                                                                                                           |
|                                | 5. MLA (532nm): 10 ~<br>250mJ ±10%      | 5. MLA (532nm): 10 ~<br>200mJ ±10%      | 5. ZMLA (532nm): 2 ~<br>255mJ ±20%     |                                                                                                                           |
|                                | 6. 1064nm DOE : 30 ~<br>450mJ ±10%      | 6. 1064nm DOE : 30 ~<br>550mJ ±10%      | 6. 1064nm DOE : 10 ~<br>500mJ ±20%     |                                                                                                                           |
|                                | 7. 532nm DOE : 20 ~<br>200mJ ±10%       | 7. 532nm DOE : 20 ~<br>200mJ ±10%       | 7. 532nm DOE : 10~<br>250mJ ±20%       |                                                                                                                           |
| Spot Size                      | 2 to 10 mm                              | 2 to 10 mm                              | 1.5 to 13 mm                           | Same                                                                                                                      |
| Pulse<br>Repetition            | Max.10Hz                                | Max.10Hz                                | Max.10Hz                               | Same                                                                                                                      |
| Laser<br>Delivery<br>Type      | Articulated Arm with<br>Handpiece       | Articulated Arm with<br>Handpiece       | Articulated Arm with<br>Handpiece      | Same                                                                                                                      |
| Handpiece                      | Zoom(1064nm, 532nm)                     | Zoom(1064nm, 532nm)                     | Zoom(1064nm, 532nm)                    | Same                                                                                                                      |
|                                | Collimated(1064nm)                      | Collimated(1064nm)                      | Collimated(1064nm)                     |                                                                                                                           |
|                                | MLA (1064nm, 532nm)                     | MLA (1064nm, 532nm)                     | ZMLA (1064nm, 532nm)                   |                                                                                                                           |
|                                | DOE (1064nm, 532nm)                     | DOE (1064nm, 532nm)                     | DOE (1064nm, 532nm)                    |                                                                                                                           |
| Patient<br>Contact<br>Material | Aluminum powder<br>(Handpiece)          | Aluminum powder<br>(Handpiece)          | Aluminum powder<br>(Handpiece)         | Same                                                                                                                      |

### Non-Clinical Test Summary |21 CFR 807.92(b)(1)]

1) Electrical Safety, Electromagnetic Compatibility Testing

Bench tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:

| Standard<br>(Edition) | Standard Title |
|-----------------------|----------------|
|-----------------------|----------------|

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for WONTECH. The logo consists of an orange circle with a stylized white "W" inside of it. To the right of the circle is the text "WONTECH" in a sans-serif font. The "WON" part of the text is orange, while the "TECH" part is gray.

| AASI AAMI ES60601-<br>1:2005/(R)2012 and A1:2012 | Medical electrical equipment – Part 1: General requirements for<br>basic safety and essential performance                                                                              |
|--------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ANSI AAMI IEC 60601-1-<br>2:2014                 | Medical electrical equipment - Part 1-2: General requirements<br>for safety and essential performance - Collateral standard:<br>Electromagnetic compatibility - Requirements and tests |
| IEC 60601-1-6 Edition 3.1<br>2013                | Medical electrical equipment - Part 1-6: General requirements<br>for basic safety and essential performance - Collateral standard:<br>Usability                                        |
| IEC 60601-2-22 Edition 3.1<br>2012-10            | Medical electrical equipment - Part 2-22: Particular<br>requirements for basic safety and essential performance of<br>surgical, cosmetic, therapeutic and diagnostic laser equipment   |
| IEC 60825-1:2014 (Third<br>Edition               | Safety of laser products - Part 1: Equipment classification and<br>requirements                                                                                                        |

### 2) Software Validation

The PICOCAREMAJESTY contains MODERATE level of concern software was designed and developed according to a software development process and was verified and validated.

The software information is provided in accordance with FDA guidance: The content of premarket submissions for software contained in medical devices, on May 11, 2005.

### 3) Biocompatibility

| Part                         | Material                                | Patient<br>Contact | Duration of Contact<br>by ISO 10993-1 | Bio-<br>compatibility |
|------------------------------|-----------------------------------------|--------------------|---------------------------------------|-----------------------|
| Handpiece Tip of<br>PICOANDY | Aluminium Powder<br>(Cas No. 7429-90-5) | Intact Skin        | Limited<br>(< 24 hours)               | Yes                   |

- The material of handpiece for the PICOCAREMAJESTY is same to previous registered material of own ● product (PICOCARE Family <K181272>, ICOCAREMAJESTY <K212127>, and PICOANDY <K234104)) in formulation, processing, and sterilization, and no other chemicals have been added. Thus, biocompatibility is not necessary for this device, as this material has been deemed biocompatible in others previously cleared submissions.
- 4) Performance Testing

performance as The of the PICOCAREMAJESTY has been defined follows. -Laser wavelength: 1064nm±10%, 532nm±10% - Laser output power: 1064nm mode: (30 to 500)mJ, 532nm mode: (20 to 250)mJ - Pulse Width: 1064mm mode: 300-400 ps, 532mm mode : 300-400 ps - Pulse repetition rate: 1064nm

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for WONTECH. The logo consists of an orange circle with a stylized "W" inside it, followed by the text "WONTECH". The "WON" part of the text is in orange, while the "TECH" part is in gray. The logo is simple and modern, and the use of orange and gray gives it a professional look.

64 Techno 8-ro, Yuseong-gu, Daejeon, Republic of Korea TEL: +82 (42) 934 6800 FAX: +82 (42) 934 9491 mode:1-10Hz, 532nm mode : 1-10Hz - Radiation diameter: 1064mm mode: (2 to 10mm) step: 1mm, 532nm mode: (2 to 10mm) step: 1mm

5) Cleaning Validation

Cleaning validation is not needed, because PICOCAREMAJESTY is not in contact with patients

### Clinical Test Summary [21 CFR 807.92(b)(2)]

No clinical studies were considered necessary and performed.

### Conclusion [21 CFR 807.92(b)(3)]

In according with the Federal Food & drug and cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification WON TECH Co., Ltd. concludes that the PICOCAREMAIESTY is substantially equivalent to predicate devices as described herein.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K241144](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K241144)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
