← Product Code [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX) · K240474

# Multi-function Platform Systems   (BL-M10) (K240474)

_Shanghai Bele Medical Technology Co.,Ltd · GEX · May 20, 2024 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K240474

## Device Facts

- **Applicant:** Shanghai Bele Medical Technology Co.,Ltd
- **Product Code:** [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX.md)
- **Decision Date:** May 20, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

The Therapy IPL is intended for medical use in the treatment of the following dermatologic conditions: - Permanent hair reduction- long-terms table reduction in number of hairs re-growing after a treatment regimen; - Moderate inflammatory acne vulgaris; - Benign pigmented epidermal lesions including dyschromia, hyperpigmentation, melasma, ephelides (freckles); - Cutaneous lesions including scars; - Benign cutaneous vascular lesions including port wine stains, hemangiomas, facial truncal and leg telangiectasias, erythema of rosacea, leg veins, spider angiomas and venous malformations. Diode laser treatment (808nm) handpiece handset is indicated for permanent reduction in hair regrowth defined as a long term, stable reduction in the number of hairs re-growing when measured at 6,9 and 12 months after the completion of a treatment regimen. It is suitable for all skin types (Fitzpatrick skin type I-VI), including tanned skin. The device is intended for use in dermatologic and general surgical procedures. The Triple diode laser Module is intended for use in dermatology procedures requiring coagulation. The Triple diode laser Module is indicated for: - Benign vascular and vascular dependent lesions removal. The Picosecond Nd:YAG Laser is intended for use in surgical and aesthetic applications in the medical specialties of dermatology and general and plastic surgery. - The 1064nm wavelength of the Picosecond Nd:YAG Laser is indicated for tattoo removal for dark colored tattoo inks and for multicolored tattoos containing dark colored tattoo inks on patients with all skin types (Fitzpatrick I-VI). - The 532mm wavelength of the Picosecond Nd:YAG Laser is intended for tattoo removal for lighter colored tattoo inks, including red and yellow inks, on patients with Fitzpatrick skin types I-III. The 1064nm Long pulse Nd:YAG is indicated for - Benign vascular lesions. - Superficial and deep telangiectasias (venulectasias). - Benign cutaneous lesions. - Pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments. - The non-ablative treatment of facial wrinkles. - Laser skin resurfacing procedures. - Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar. - Indicated for use on all skin types (Fitzpatrick I-VI), including tanned skin. - Removal of unwanted hair, for stable long-term, or permanent, hair reduction through selective targeting of melanin in hair follicles. - Removal or lightening of unwanted hair (with and without adjuvant preparation). - Treatment of pseudofolliculitis barbae (PFB). The Q-switch Nd:YAG Laser Module is indicated for: - Benign vascular and pigmented lesions, age spots. - Nevus spilus. - Tattoo removal.

## Device Story

Multi-function platform (BL-M10) integrates six therapeutic modalities: IPL, 808nm Diode laser, Triple Diode laser, Picosecond Nd:YAG, 1064nm Long pulse Nd:YAG, and Q-switch Nd:YAG. System comprises AC-powered mobile console with 12.1" display and interchangeable handpieces. Operator (physician/clinician) selects modality and parameters via interface to deliver energy to skin. IPL uses Xenon flash lamp; lasers use specific wavelengths for selective photothermolysis or coagulation. Output affects hair follicles, tattoo pigments, vascular lesions, or collagen/skin tissue. Benefits include versatile dermatologic/surgical treatment options in clinical settings. System includes safety features like emergency shut-off.

## Clinical Evidence

No clinical data. Bench testing only. Compliance verified against IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1, IEC 62471, and IEC 60601-2-57. Performance testing confirmed energy output, wavelength accuracy, pulse duration, and spot size meet specifications.

## Technological Characteristics

Mobile AC-powered platform; 12.1" display. Modalities: IPL (Xenon lamp), Diode (808nm), Triple Diode (755/808/1064nm), Picosecond Nd:YAG (532/1064nm), Long pulse Nd:YAG (1064nm), Q-switch Nd:YAG (532/1064nm). Cooling: Water, forced-air, TEC. Standards: IEC 60601-1, 60601-1-2, 60601-2-22, 60825-1, 62471, 60601-2-57.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- IPL Treatment System ([K200746](/device/K200746.md))
- Diode Laser Therapy Device ([K200118](/device/K200118.md))
- Modified Alma Lasers Soprano XL Family ([K172193](/device/K172193.md))
- PICOSECOND Nd:YAG Laser System ([K233018](/device/K233018.md))
- Platform treatment system ([K203395](/device/K203395.md))
- MT ONE ([K192856](/device/K192856.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in a smaller font.

May 20, 2024

Shanghai Bele Medical Technology Co., Ltd Jeremy Li Room 402, 4F, Building 1, NO.3255 Shengang Road, Songjiang District Shanghai, Shanghai 201600 China

Re: K240474

Trade/Device Name: Multi-function Platform Systems (BL-M10) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX, ONF Dated: April 30, 2024 Received: April 30, 2024

Dear Jeremy Li:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).

{2}------------------------------------------------

Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Tanisha L. Hithe -S Date: 2024 05 20 15.18.25 -04'00

Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical & Infection Control Devices, Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

### Indications for Use

510(k) Number (if known) K240474

Device Name Multi-function Platform Systems (BL-M10)

#### Indications for Use (Describe)

(1) IPL: The Therapy IPL is intended for medical use in the following dermatologic conditions: Permanent hair reduction- long-terms table reduction in number of hairs re-growing after a treatment - Moderate inflammatory acne vulgaris; - Benign pigmented epidermal lesions including dyschromia, hyperpigmentation, melasma, ephelides (freckles); - Cutaneous lesions including scars; - Benign cutaneous vascular lesions including port wine stains, hemangiomas, facial truncal and leg telangiectasias, erythema of rosacea, leg veins, spider angiomas and venous malformations.

(2) Diode laser: Diode laser treatment handset is indicated for permanent reduction in hair regrowth defined as a long term, stable reduction in the number of hairs re-growing when measured at 6,9 and 12 months after the completion of a treatment regimen. It is suitable for all skin type I-VI), including tanned skin. 3) Triple Diode laser: The device is intended for use in dermatologic and general surgical procedures.

The Triple diode laser Module is intended for use in dermatology procedures requiring coagulation. The Triple diode laser Module is indicated for: - Benign vascular and vascular dependent lesions removal.

(4) Picosecond Nd: Y AG Laser: The Picosecond Nd: Y AG Laser is intended for use in surgical and aestheir applications in the medical specialties of dermatology and general and plastic surgery. - The 1064nm wavelength of the Picosecond Nd:YAG Laser is indicated for tattoo removal for dark colored tattoo inks and for multicolored tattoos containing dark colored tattoo inks on patients with all skin types (Fitzpatrick I-VI). - The 532mm wavelength of the Picosecond Nd: Y AG Laser is indicated for tattoo removal for lighter colored tattoo inks, including red and yellow inks, on patients with Fitzpatrick skin types 1-III.

(5) 1064nm Long pulse Nd: YAG: The 1064nm Long pulse Nd: YAG is indicated for - Benign vascular lesions. -

Superficial and deep telangiectasias). - Benign cutaneous lesions. - Pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments. - The non-ablative treatment of facial wrinkles. - Laser skin resurfacing procedures. - Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar. - Indicated for use on all skin types (Fitzpatrick I-VI), including tanned skin. - Removal of unwanted hair, for stable long-term, or permanent, hair reduction through selective targeting of melanin in hair follicles. - Removal or lightening of unwanted hair (with and without adjuvant preparation). - Treatment of pseudofolliculitis barbae (PFB).

(6) Q-switch Nd:YAG laser: The Q-switch Nd:YAG Laseris Module indicated for: -Benign vascular and pigmented lesions, age spots. - Nevus spilus. - Tattoo removal.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

{4}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

#### l. Submission Submission Number K240474 Device submitter Shanghai Bele Medical Technology Co. Ltd. Room 402, 4F, Building 1, NO.3255 Shengang Road, Songjiang District, Shanghai, China Contact Person Jeremy Li +86-21 51012882 Phone Number Email jeremy@belemed.com Submission Date May 16, 2024 II. Device Information Trade Name Multi-function Platform Systems (BL-M10) Laser surgical instrument for use in general and plastic surgery Common Name and in dermatology Classification Name Powered Laser Surgical Instrument Regulation Number 21 CFR 878.4810 Product Code GEX, ONF

### III. Predicate Devices

| Trade Name of Device:   | IPL Treatment System                                                                   |
|-------------------------|----------------------------------------------------------------------------------------|
| Common name:            | Laser surgical instrument for use in general and plastic<br>surgery and in dermatology |
| Classification:         | Class II, 21 CFR 878.4810.                                                             |
| Product Code:           | ONF                                                                                    |
| Premarket Notification: | K200746                                                                                |
| Company name:           | Shanghai Apolo Medical Technology Co., Ltd.                                            |

| Trade Name of Device:   | Diode Laser Therapy Device                  |
|-------------------------|---------------------------------------------|
| Common name:            | Powered Laser Surgical Instrument           |
| Classification:         | Class II, 21 CFR 878.4810                   |
| Product Code:           | GEX                                         |
| Premarket Notification: | K200118                                     |
| Company name:           | Shanghai Apolo Medical Technology Co., Ltd. |

| Trade Name of Device: | Modified Alma Lasers Soprano XL™ Family ofMulti-<br>Application & MultiTechnology Platforms [SopranoXL,<br>SopranoXLi, SopranoICE and Soprano ICE Platinum] with<br>Duo and Trio Diode Laser Modules., Soprano Duo and Trio |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

{6}------------------------------------------------

|                         | Diode Laser Modules                                |
|-------------------------|----------------------------------------------------|
| Common name:            | Laser Powered Surgical Instruments (& Accessories) |
| Classification:         | Class II, 21 CFR 878.4810.                         |
| Product Code:           | GEX, ILY                                           |
| Premarket Notification: | K172193                                            |
| Company name:           | Alma Lasers Inc.                                   |

| Trade Name of Device:   | PICOSECOND Nd:YAG Laser System |
|-------------------------|--------------------------------|
| Common name:            | /                              |
| Classification:         | Class II, 21 CFR 878.4810.     |
| Product Code:           | GEX                            |
| Premarket Notification: | K233018                        |
| Company name:           | IDS, Ltd.                      |

| Trade Name of Device:   | Platform treatment system                   |
|-------------------------|---------------------------------------------|
| Common name:            | Powered Laser Surgical Instrument           |
| Classification:         | Class II, 21 CFR 878.4810.                  |
| Product Code:           | ONF, GEX                                    |
| Premarket Notification: | K203395                                     |
| Company name:           | Shanghai Apolo Medical Technology Co., Ltd. |

| Trade Name of Device:   | MT ONE                                     |
|-------------------------|--------------------------------------------|
| Common name:            | Intense Pulsed Light(IPL) and laser System |
| Classification:         | Class II, 21 CFR 878.4810.                 |
| Product Code:           | GEX, ONG, ONF, ONE                         |
| Premarket Notification: | K192856                                    |
| Company name:           | M&TS.R.L.                                  |

### IV. Device description

The BL-M10 is a new platform treatment system with 12.1" display screen which combines IPL/ Diode Laser/Triple Diode Laser/ Picosecond Nd:YAG Laser/ 1064nm Long pulse Nd:YAG laser and Q-switch Nd:YAG Laser functions. This is 6 in 1 platform machine, that provides different function with attaching different hand pieces. The system is designed with a mains electricity (AC-powered) mobile (on wheels) assembly of devices that uses multiple therapeutic modalities in combination or in isolative and non-ablative treatment of skin surface. The system includes various energy sources and dedicated hand-pieces intended to apply the different energies to the skin for respective indications for use.

K240474 Shanghai Bele Medical Technology Co.,Ltd.

{7}------------------------------------------------

### V. Indications for use

- (1) IPL: The Therapy IPL is intended for medical use in the treatment of the following dermatologic conditions: - Permanent hair reduction- long-terms table reduction in number of hairs re-growing after a treatmen; - Moderate inflammatory acne vulgaris; - Benign pigmented epidermal lesions including dyschromia, hyperpigmentation, melasma, ephelides (freckles); - Cutaneous lesions including scars; - Benign cutaneous vascular lesions including port wine stains, hemangiomas, facial truncal and leg telangiectasias, erythema of rosacea, leg veins, spider angiomas and venous malformations.
- (2) Diode laser: Diode laser treatment handpiece (808nm) handset is indicated for permanent reduction in hair regrowth defined as a long term, stable reduction in the number of hairs re-growing when measured at 6,9 and 12 months after the completion of a treatment regimen. It is suitable for all skin types (Fitzpatrick skin type I-VI), including tanned skin.
- (3) Triple Diode laser: The device is intended for use in dermatologic and general surgical procedures. The Triple diode laser Module is intended for use in dermatology procedures requiring coagulation. The Triple diode laser Module is indicated for:
	- Benign vascular and vascular dependent lesions removal.
- (4) Picosecond Nd:YAG Laser: The Picosecond Nd:YAG Laser is intended for use in surgical and aesthetic applications in the medical specialties of dermatology and general and plastic surgery. - The 1064nm wavelength of the Picosecond Nd:YAG Laser is indicated for tattoo removal for dark colored tattoo inks and for multicolored tattoos containing dark colored tattoo inks on patients with all skin types (Fitzpatrick I-VI). - The 532nm wavelength of the Picosecond Nd:YAG Laser is indicated for tattoo removal for lighter colored tattoo inks, including red and yellow inks, on patients with Fitzpatrick skin types I-III.
- (5) 1064nm Long pulse Nd:YAG: Benign vascular lesions Superficial and deep telangiectasias (venulectasias) - Benign cutaneous lesions to reduce lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments. - The non-ablative treatment of facial wrinkles. - Laser skin resurfacing procedures. - Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar. - Indicated for use on all skin types (Fitzpatrick I-VI), including tanned skin. - Removal of unwanted hair, for stable long-term, or permanent, hair reduction through selective targeting of melanin in hair follicles. - Removal or lightening of unwanted hair (with and without adjuvant preparation). - Treatment of pseudofolliculitis barbae (PFB).
- (6) Q-switch Nd:YAG laser: The Q-switch Nd:YAG Laseris Module indicated for: Benign vascular and pigmented lesions, age spots • Nevus spilus • Tattoo removal

## VI. Comparison of Technological Characteristics with Predicate Devices

The Multi-function Platform Systems (BL-M10) shares the same technological characteristics as its predicate devices, including energy sources (laser/IPL), control mechanisms, and core specifications.

{8}------------------------------------------------

|  | Tabel 1 Substantial equivalence discussion - IPL |  |
|--|--------------------------------------------------|--|
|  |                                                  |  |

| Device feature       | IPL (subject device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | IPULSELIGHT IPL SYSTEM K200746 (predicate device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Discussion  |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Intended use         | The Therapy IPL is<br>intended for medical use in<br>the treatment of the<br>following dermatologic<br>conditions:<br>- Permanent hair reduction-<br>long-terms table reduction<br>in number of hairs re-<br>growing after a treatment<br>regimen;<br>- Moderate inflammatory<br>acne vulgaris;<br>- Benign pigmented<br>epidermal lesions including<br>dyschromia,<br>hyperpigmentation,<br>melasma, ephelides<br>(freckles);<br>- Cutaneous lesions<br>including scars;<br>- Benign cutaneous<br>vascular lesions including<br>port wine stains,<br>hemangiomas, facial<br>truncal and leg<br>telangiectasias, erythema<br>of rosacea, leg veins,<br>spider angiomas and<br>venous malformations. | The IPL treatment<br>systems is intended for<br>medical use in the<br>treatment of the following<br>dermatologic conditions:<br>- Permanent hair<br>reduction- long-terms<br>table reduction in number<br>of hairs re-growing after a<br>treatment regimen;<br>- Moderate<br>inflammatory acne<br>vulgaris;<br>- Benign pigmented<br>epidermal lesions<br>including dyschromia,<br>hyperpigmentation,<br>melasma, ephelides<br>(freckles);<br>- Cutaneous lesions<br>including scars;<br>- Benign cutaneous<br>vascular lesions including<br>port wine stains,<br>hemangiomas, facial<br>truncal and leg<br>telangiectasias, erythema<br>of rosacea, leg veins,<br>spider angiomas and<br>venous malformations. | Identical   |
| Light source         | Intense pulsed light (Xenon<br>Flash Lamp)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Intense pulsed light<br>(Xenon Flash Lamp)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Identical   |
| Wavelength           | Filters:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Filters:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Identical   |
| Device feature       | IPL (subject device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | IPULSELIGHT IPL SYSTEM K200746<br>(predicate device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Discussion  |
| Range                | 420 -1200nm: Acne;<br>510 -1200nm: Acne,<br>vascular, pigment;<br>560 -1200nm: Acne,<br>vascular, pigment;<br>610-1200nm: Hair<br>removal;<br>640-1200nm: Hair<br>removal;<br>690-1200nm: Hair<br>removal;                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | 420 -1200nm: Acne;<br>510 -1200nm: Acne,<br>vascular, pigment;<br>560 -1200nm: Acne,<br>vascular, pigment;<br>610-1200nm: Hair<br>removal;<br>640-1200nm: Hair<br>removal;<br>690-1200nm: Hair<br>removal;                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |             |
| Handpiece<br>Ports   | IPL treatment handpiece                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | HS-650K&HS-660K: 2<br>HS620K, HS-300CK &<br>HS-310K: 1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Different 1 |
| Structure            | Vertical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | HS-650K&HS-660K:<br>Vertical<br>HS620K, HS-300CK &<br>HS-310K: Table top                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Different 2 |
| Energy output        | 10-50 J/cm²                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | 4.1-50.8 J/cm²                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Different 3 |
| Pulse width          | 2-20ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 5-20 ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Different 4 |
| Pulse duration       | 5-50 ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | 5-50 ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Identical   |
| Spot size            | 15mm×50mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 12*35mm, 15*50mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Different 5 |
| Output mode          | Pulse mode                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Pulse mode                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Identical   |
| Deliver<br>materials | Direct sapphire Coupling                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Direct sapphire Coupling                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Identical   |
| Cooling method       | Water cooling, forced-air<br>cooling and TEC water tank<br>cooling system and<br>semiconductor chip                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | HS-650K&HS-660K:<br>Water cooling, forced-air<br>cooling, copper and TEC;<br>HS620K,<br>HS-300CK&HS-310K:<br>Water cooling and forced-air cooling                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Different 7 |
| -                    | Spot Size:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | The subject device offers a spot size of 15mm×50mm, covered by the predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |             |

{9}------------------------------------------------

## Conclusion:

Handpiece Ports: The subject device has one IPL treatment handpiece port, while the ।

{10}------------------------------------------------

predicate device offers multiple handpiece ports depending on the model. However, this difference in handpiece ports does not affect the intended use of the device, as both devices are capable of delivering the required treatments with their respective handpieces.

- Energy Output: The energy output ranges of both devices are within similar ranges, with the subject device ranging from 10-50 J/cm² and the predicate device ranging from 4.1-50.8 J/cm². While there is a difference in the upper limit of the energy output, both devices offer sufficient energy levels for effective treatment across various dermatologic conditions.
- Pulse Width: The subject device offers pulse widths ranging from 2-20ms, whereas the predicate device offers pulse widths ranging from 5-20 ms, which are basically similar. These differences in pulse parameters are minor and do not significantly impact the intended use of the devices for dermatologic treatments.

| Device feature | Diode Laser<br>(subject device)                                                                                                                                                                                                                                                                                                                                                                 | Diode Laser Therapy<br>Device K200118<br>(predicate device)                                                                                                                                                                                                                                                                                                                   | Discussion  |
|----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Intended use   | Diode laser treatment<br>(808nm) handpiece<br>handset is indicated for<br>permanent reduction in hair<br>regrowth defined as a long<br>term, stable reduction in<br>the number of hairs re-<br>growing when measured at<br>6,9 and 12 months after the<br>completion of a treatment<br>regimen. It is suitable for all<br>skin types (Fitzpatrick skin<br>type I-VI), including tanned<br>skin. | The Diode Laser Therapy<br>Device is indicated for<br>permanent reduction in<br>hair regrowth defined as a<br>long term, stable<br>reduction in the number of<br>hairs re-growing when<br>measured at 6,9 and 12<br>months after the<br>completion of a treatment<br>regimen. It is suitable for<br>all skin types (Fitzpatrick<br>skin type I-VI), including<br>tanned skin. | Identical   |
| Spot Size      | 12mm×23mm                                                                                                                                                                                                                                                                                                                                                                                       | 12x16mm;<br>12x20mm                                                                                                                                                                                                                                                                                                                                                           | Different 1 |
| Energy Source  | 808nm diode laser                                                                                                                                                                                                                                                                                                                                                                               | 755nm handpiece<br>810nm handpiece<br>1064nm handpiece                                                                                                                                                                                                                                                                                                                        | Different 2 |
| Wavelength     | 808nm                                                                                                                                                                                                                                                                                                                                                                                           | 755nm/810nm/1064nm                                                                                                                                                                                                                                                                                                                                                            | Different 3 |

Tabel 2 Substantial equivalence discussion - Diode Laser

{11}------------------------------------------------

| Device feature             | Diode Laser<br>(subject device)                                                                                                                                                                                                                                                                                                                                         | Diode Laser Therapy<br>Device K200118<br>(predicate device)                                                                                                                                                                                                                                                                                                                                                  | Discussion  |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
|                            |                                                                                                                                                                                                                                                                                                                                                                         | for option                                                                                                                                                                                                                                                                                                                                                                                                   |             |
| Maximum<br>Repetition Rate | 1,2,3,5,8,10Hz                                                                                                                                                                                                                                                                                                                                                          | 1,2,3,5,8,10Hz                                                                                                                                                                                                                                                                                                                                                                                               | Similar     |
| Pulse Duration             | 10-300ms                                                                                                                                                                                                                                                                                                                                                                | 10-300ms                                                                                                                                                                                                                                                                                                                                                                                                     | Identical   |
| Energy                     | 1-62J/cm²                                                                                                                                                                                                                                                                                                                                                               | 1~64J/cm2<br>1~62J/cm2                                                                                                                                                                                                                                                                                                                                                                                       | Different 4 |
| Device feature             | Triple Diode laser<br>(subject device)                                                                                                                                                                                                                                                                                                                                  | Modified Alma Lasers<br>Soprano XL™ Family<br>ofMulti-Application &<br>MultiTechnology<br>Platforms [SopranoXL,<br>SopranoXLi,<br>SopranoICE and<br>Soprano ICE Platinum]<br>with Duo and Trio Diode<br>Laser Modules.,<br>Soprano Duo and Trio<br>Diode Laser Modules<br>K172193<br>(predicate device)                                                                                                      | Discussion  |
| Intended use               | Intended Use<br>The device is intended for<br>use in dermatologic and<br>general surgical<br>procedures.<br>Indications for Use<br>The triple diode laser<br>Module is intended for use<br>in dermatology procedures<br>requiring coagulation. The<br>Triple diode laser Module is<br>indicated for:<br>● Benign vascular and<br>vascular dependent<br>lesions removal. | Intended Use<br>The device is intended for<br>use in dermatologic and<br>general surgical<br>procedures.<br>Indications for Use<br>The Soprano trio Diode<br>Laser Module is intended<br>for use in dermatology<br>procedures requiring<br>coagulation. The<br>indications for use for the<br>Soprano Trio Diode Laser<br>Module include:<br>● Benign vascular and<br>vascular dependent<br>lesions removal. | Identical   |
| Spot Size                  | 12mm×23mm                                                                                                                                                                                                                                                                                                                                                               | Trio 4 cm² applicator<br>Trio 2 cm² applicator                                                                                                                                                                                                                                                                                                                                                               | Different 1 |
| Wavelength                 | 755/808/1064nm                                                                                                                                                                                                                                                                                                                                                          | 755/808/1064nm                                                                                                                                                                                                                                                                                                                                                                                               | Identical   |
| Maximum<br>Repetition Rate | 1~10Hz                                                                                                                                                                                                                                                                                                                                                                  | 1-10 Hz                                                                                                                                                                                                                                                                                                                                                                                                      | Identical   |
| Device feature             | Triple Diode laser<br>(subject device)                                                                                                                                                                                                                                                                                                                                  | Modified Alma Lasers<br>Soprano XL™ Family<br>ofMulti-Application &<br>MultiTechnology<br>Platforms [SopranoXL,<br>SopranoXLi,<br>SopranoICE and<br>Soprano ICE Platinum]<br>with Duo and Trio Diode<br>Laser Modules.,<br>Soprano Duo and Trio<br>Diode Laser Modules<br>K172193<br>(predicate device)                                                                                                      | Discussion  |
| Pulse Duration             | 10-130ms                                                                                                                                                                                                                                                                                                                                                                | 3.3-280 ms                                                                                                                                                                                                                                                                                                                                                                                                   | Different 2 |
| Energy                     | 1~32J/cm²                                                                                                                                                                                                                                                                                                                                                               | up to 60 J/cm²                                                                                                                                                                                                                                                                                                                                                                                               | Different 3 |

#### Conclusion:

- Different 1: ।
Spot Size: The subject device has a larger spot size (12mm×23mm) compared to the predicate device. A larger spot size can affect the treatment speed and efficiency, reducing treatment time.

- Different 2&3: -
Energy Source (Wavelength): Both the subject device and the HS-817 & HS-819 models of the predicate device operate with diode lasers in the 808nm to 810nm range. This slight difference in wavelength (808nm for the subject device vs. 810nm for the predicate models) is minimal and likely does not affect the efficacy or safety profiles of the devices significantly. Operating in a similar near-infrared spectrum, they are expected to have comparable penetration depths and effects on hair follicles, ensuring a similar mechanism of action for hair removal.

- Different 4
The subject device operates with an energy range of 1-62 J/cm², which is slightly lower than the maximum value of 64 J/cm² for one of the settings of the predicate device. This variation in maximum energy output may affect the ability of the device to adapt to different hair densities and skin types. A lower maximum energy may mean that the emphasis on comfort and minimising skin irritation is beneficial for sensitive skin types or for use on more delicate areas.

Both the subject and the predicate devices share the primary function of hair removal, employing diode lasers to achieve this end. Despite minor variations in their specifications, such as energy output and spot size, the core technology remains consistent, suggesting a substantial equivalence in therapeutic outcome. Therefore, the differences identified do not raise concerns regarding the intended use or safety of the subject device.

{12}------------------------------------------------

Tabel 3 Substantial equivalence discussion - Triple Diode laser

{13}------------------------------------------------

### Conclusion:

- Different 1: -
Spot Size: The subject device's spot size of 12mm×23mm (2.76cm²) is different from the predicate device's options of 4cm² and 2cm², but is within the range of the predicate device.

- Different 2: -
Pulse Duration: The subject device offers pulse duration within the range covered by the predicate.

- Different 3:
Energy: The subject device offers energy level covered by the predicate.

In conclusion, the differences in some parameters between the subject device and the predicate device are minor. The core technology and operational principles remain consistent, ensuring that the subject device is substantially equivalent to the predicate device.

## Tabel 4 Substantial equivalence discussion - Picosecond Nd:YAG Laser

| Device feature          | Picosecond Nd:YAG<br>Laser (subject device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | PICOSECOND Nd: YAG<br>Laser System K233018<br>(predicate device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Discussion  |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| <b>Intended use</b>     | The Picosecond Nd:YAG<br>Laser is intended for use in                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | The PICOSECOND Nd:<br>YAG Laser System (PICO                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Identical   |
| Device feature          | Picosecond Nd:YAG<br>Laser (subject device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | PICOSECOND Nd: YAG<br>Laser System K233018<br>(predicate device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Discussion  |
|                         | surgical and aesthetic<br>applications in the medical<br>specialties of dermatology<br>and general and plastic<br>surgery.<br>- The 1064nm wavelength<br>of the Picosecond Nd:YAG<br>Laser is indicated for tattoo<br>removal for dark-colored<br>tattoo inks and multicolored<br>tattoos containing dark-<br>colored tattoo inks on<br>patients with all skin types<br>(Fitzpatrick I-VI).<br>- The 532nm wavelength of<br>the Picosecond Nd:YAG<br>Laser is indicated for tattoo<br>removal for lighter-colored<br>tattoo inks, including red<br>and yellow inks, on patients<br>with Fitzpatrick skin types I-<br>III. | PREMIUM) is intended for<br>use in surgical and<br>aesthetic applications<br>in the medical<br>specialties of<br>dermatology and general<br>and plastic surgery.<br>- The 1064nm wavelength<br>of the PICOSECOND Nd:<br>YAG Laser System (PICO<br>PREMIUM) is indicated<br>for tattoo removal for<br>dark-colored tattoo inks<br>and multicolored<br>tattoos containing<br>dark-colored tattoo<br>inks on patients with all<br>skin types (Fitzpatrick I-<br>VI).<br>- The 532nm wavelength<br>of the PICOSECOND Nd:<br>YAG Laser System (PICO<br>PREMIUM) is indicated<br>for tattoo removal for<br>lighter-colored tattoo<br>inks, including red<br>and yellow inks, on<br>patients with<br>Fitzpatrick skin types I-III. |             |
| Spot Size               | 1, 2, 3, 4, 5, 7 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | 2mm - 10mm (Adjustable<br>by 1mm) - Both Modes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Different 1 |
| Wavelength              | Double wavelength                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | 532 or 1064nm (+/-10%)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Identical   |
| Device feature          | Picosecond Nd:YAG Laser (subject device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | PICOSECOND Nd: YAG Laser System K233018 (predicate device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Discussion  |
|                         | 1064nm&532nm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |             |
| Maximum Repetition Rate | 1-10Hz                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | 1-10Hz (Adjustable by 1Hz)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Identical   |
| Pulse Duration          | < 980ps                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 300 ps - Both Modes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Different 2 |
| Energy                  | 50mJ to 400mJ(1064nm);<br>50mJ to 200mJ(532nm)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 532nm mode: 50 - 250mJ (Adjustable by 25mJ)<br>1064nm mode:<br>100-500mJ (Adjustable by 25mJ)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Different 3 |

{14}------------------------------------------------

{15}------------------------------------------------

#### Conclusion:

- Spot Size: The subject device offers a spot size range of 1, 2, 3, 4, 5, 7 mm, while the । predicate device provides a range of 2mm to 10mm, adjustable by 1mm. While the subject device's spot size range is narrower, both devices offer adjustable spot sizes suitable for various treatment areas.
- -Pulse Duration: The subject…

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K240474](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K240474)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
