← Product Code [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX) · K234057

# NIRVANA (K234057)

_El.En S.P.A. · GEX · Apr 10, 2024 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K234057

## Device Facts

- **Applicant:** El.En S.P.A.
- **Product Code:** [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX.md)
- **Decision Date:** Apr 10, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

The Nirvana device is indicated for hair removal and permanent hair reduction. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

## Device Story

Nirvana is a diode laser device for hair removal and permanent hair reduction. It utilizes two handpieces to deliver laser energy to the patient's skin. Operated by a clinician in a professional setting, the device transforms electrical energy into laser output at a wavelength of 809 ± 6 nm. The clinician selects parameters such as fluence (up to 40 J/cm²), pulse duration (up to 200 ms), and repetition rate (up to 10 Hz) to target hair follicles. The laser energy targets melanin in the hair, leading to thermal destruction of the follicle. This process provides patients with a long-term, stable reduction in hair regrowth. The device is designed for use in dermatology or aesthetic clinical environments.

## Clinical Evidence

No clinical data provided. Substantial equivalence is supported by non-clinical performance data, including electrical safety (ANSI AAMI ES60601-1), electromagnetic compatibility (IEC 60601-1-2), laser safety (IEC 60825-1), and performance requirements for surgical/cosmetic laser equipment (IEC 60601-2-22), alongside software verification and validation.

## Technological Characteristics

Diode laser system; wavelength 809 ± 6 nm; max fluence 40 J/cm²; max peak power 1800W; pulse duration up to 200 ms; repetition rate up to 10 Hz. Handpiece spot sizes: 12x24 mm and 12x12 mm. Electrical requirements: 100-115V, 50/60Hz, 1600VA. Complies with ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-22, and IEC 60825-1.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- DEKA LUXEA ([K192539](/device/K192539.md))

## Reference Devices

- ASCLEPION MeDioStar ([K192483](/device/K192483.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 10, 2024

El.En. S.p.A. Paolo Peruzzi Regulatory Affairs Manager Via Baldanzese 17 Calenzano, FL 50141 Italy

Re: K234057

Trade/Device Name: Nirvana Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: March 14, 2024 Received: March 14, 2024

Dear Paolo Peruzzi:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/ofdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).

{2}------------------------------------------------

Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Tanisha Digitally signed by Tanisha L L. Hithe - Hithe -s Date: 2024.04.10 2 1 15:21:02 -04'0

Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K234057

Device Name Nirvana

Indications for Use (Describe)

The Nirvana device is indicated for hair removal and permanent hair reduction.

Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

Type of Use (Select one or both, as applicable)

| <div> <div> <span> <svg height="15" width="15"> <rect fill="none" height="15" stroke="black" width="15"></rect> <line stroke="black" x1="2" x2="13" y1="2" y2="13"></line> <line stroke="black" x1="13" x2="2" y1="2" y2="13"></line> </svg> </span>      Prescription Use (Part 21 CFR 801 Subpart D)     </div> </div> |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <div> <span> <svg height="15" width="15"> <rect fill="none" height="15" stroke="black" width="15"></rect> </svg> </span>      Over-The-Counter Use (21 CFR 801 Subpart C)     </div> </div>                                                                                                                        |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

#### 510(k) Summary

#### K234057

#### NIRVANA

#### Submitter:

El.En. S.p.A. Via Baldanzese, 17 50041 Calenzano (FI), Italy

# Contact:

Paolo Peruzzi Regulatory Affairs Manager & Official Correspondent Phone: +39.055.8826807 E-mail: p.peruzzi@elen.it

# Date Summary Prepared:

March 22nd, 2024

#### Device Trade Name:

NIRVANA

#### Common Name:

Medical Laser

# Regulation Number:

21 CFR 878.4810

#### Regulation Name:

Laser Surgical Instrument for use in General and Plastic and in Dermatology

# Regulatory Class:

Class II

#### Product Code:

GEX

# Predicate Devices:

| Primary predicate:   |  |
|----------------------|--|
| Reference predicate: |  |

DEKA LUXEA (K192539) ASCLEPION MeDioStar (K192483)

{5}------------------------------------------------

#### Device Description:

The NIRVANA is a diode laser device indicated for hair removal and permanent hair reduction. The device is provided with two different handpieces that deliver the laser energy to the patient. The NIRVANA electrical specifications are: 100-115V~, 50/60Hz, 1600VA.

#### Indications for Use:

The Nirvana device is indicated for hair removal and permanent hair reduction.

Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

#### Comparison with The Predicate Device:

The NIRVANA is as safe, as effective, and performs as well as the legally marketed predicate devices (K192539 and K192483):

| Device<br>Trade Name             | Proposed Device<br>Nirvana                                                                                                                                                                                                                                                       | Predicate Device<br>DEKA LUXEA<br>VIVID LASER handpiece<br>(K192539)                                                                                                                                                                                                                                                                                         | Reference Device<br>ASCLEPION<br>MeDioStar<br>(K192483)                                                                                                                                                                                                                                                                                                                                                                                                     | Comment                                                                            |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| Indications<br>for Use           | Indicated for hair removal<br>and permanent hair<br>reduction.<br>Permanent hair reduction<br>is defined as the long-<br>term, stable reduction in<br>the number of hairs<br>regrowing when measured<br>at 6, 9, and 12 months<br>after the completion of a<br>treatment regime. | Indicated for the<br>treatment of benign<br>vascular lesions, benign<br>pigmented lesions, hair<br>removal and permanent<br>hair reduction.<br>Permanent hair reduction<br>is defined as the long-<br>term, stable reduction in<br>the number of hairs<br>regrowing when measured<br>at 6, 9, and 12 months<br>after the completion of a<br>treatment regime | Intended for surgical,<br>aesthetic and cosmetic<br>applications in the medical<br>specialties of general and<br>plastic surgery and<br>dermatology.<br>Intended for the treatment<br>of benign vascular lesions.<br>Intended for the treatment<br>of benign pigmented<br>lesions.<br>Intended for hair removal,<br>permanent hair reduction<br>defined as reduced hair<br>growth with or without<br>maintenance when<br>measured at 6, 9 and 12<br>months. | Subset of<br>predicate<br>device and<br>reference<br>device                        |
| Regulation<br>number             | 21 CFR 878.4810                                                                                                                                                                                                                                                                  | 21 CFR 878.4810                                                                                                                                                                                                                                                                                                                                              | 21 CFR 878.4810                                                                                                                                                                                                                                                                                                                                                                                                                                             | Identical                                                                          |
| Product<br>Code                  | GEX                                                                                                                                                                                                                                                                              | GEX                                                                                                                                                                                                                                                                                                                                                          | GEX                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Identical                                                                          |
| Device Type                      | Diode Laser                                                                                                                                                                                                                                                                      | Diode laser                                                                                                                                                                                                                                                                                                                                                  | Diode Laser                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Identical                                                                          |
| Device<br>Trade Name             | Proposed Device<br>Nirvana                                                                                                                                                                                                                                                       | Predicate Device<br>DEKA LUXEA<br>VIVID LASER handpiece<br>(K192539)                                                                                                                                                                                                                                                                                         | Reference Device<br>ASCLEPION<br>MeDioStar<br>(K192483)                                                                                                                                                                                                                                                                                                                                                                                                     | Comment                                                                            |
| Wavelength                       | 809 ± 6 nm                                                                                                                                                                                                                                                                       | 808 ± 10 nm                                                                                                                                                                                                                                                                                                                                                  | 760± 10 nm<br>808± 10 nm<br>940 ± 10 nm                                                                                                                                                                                                                                                                                                                                                                                                                     | Same as<br>predicate<br>device, subset<br>of reference<br>device                   |
| Handpiece<br>Spot Size           | 12x24 mm<br>12x12mm                                                                                                                                                                                                                                                              | 10x12 mm                                                                                                                                                                                                                                                                                                                                                     | Spot size at 808± 10 nm:<br>10 x 10 mm<br>15 x 10 mm<br>30 x 10 mm<br>31.6 x 31.6 mm                                                                                                                                                                                                                                                                                                                                                                        | Larger than<br>predicate<br>device, within<br>the ranges of<br>reference<br>device |
| Max Fluence                      | 40 J/cm²                                                                                                                                                                                                                                                                         | 40 J/cm²                                                                                                                                                                                                                                                                                                                                                     | Max fluence at 808± 10<br>nm:<br>60 J/cm²                                                                                                                                                                                                                                                                                                                                                                                                                   | Identical to<br>predicate<br>device, within<br>the range of<br>reference<br>device |
| Pulse<br>Duration                | Up to 200 ms                                                                                                                                                                                                                                                                     | Up to 200 ms                                                                                                                                                                                                                                                                                                                                                 | Up to 400 ms                                                                                                                                                                                                                                                                                                                                                                                                                                                | Identical to<br>predicate<br>device, within<br>the range of<br>reference<br>device |
| Max output<br>peak power         | 1800W                                                                                                                                                                                                                                                                            | 900W                                                                                                                                                                                                                                                                                                                                                         | 4200W                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Higher than<br>predicate<br>device, within<br>the range of<br>reference<br>device  |
| Max output<br>average<br>power   | 174W                                                                                                                                                                                                                                                                             | 81W                                                                                                                                                                                                                                                                                                                                                          | 250W                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Higher than<br>predicate<br>device, within<br>the range of<br>reference<br>device  |
| Pulse<br>Repetition<br>Rate (Hz) | Single shot to 10 Hz                                                                                                                                                                                                                                                             | Single shot to 10 Hz                                                                                                                                                                                                                                                                                                                                         | Up to 20Hz                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Identical to<br>predicate<br>device, within<br>the range of<br>reference<br>device |

{6}------------------------------------------------

{7}------------------------------------------------

# Clinical Performance Data:

None

# Non-Clinical Performance Data:

# Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the NIRVANA device, according to the following standards:

. ANSI AAMI ES60601-1Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.

. IEC 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.

. IEC 60601-2-22: Medical electrical equipment - Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment

. IEC 60825-1: Safety of laser products - Part 1: Equipment classification and requirements.

# Software Validation and Verification Testing

Software verification and validation testing were conducted and documented as recommended by FDA's Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Device Software Functions".

# Conclusion:

On the basis of the comparison with the predicate device and on the non-clinical performance data, we can conclude that NIRVANA is as safe, as effective, and performs as well as the legally marketed predicate devices (K192539 and K192483).

# Additional Information:

None.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K234057](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K234057)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
