← Product Code [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX) · K231910

# DermRays Revive (K231910)

_Wuhan Lotuxs Technology Co., Ltd. · GEX · Jan 30, 2024 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K231910

## Device Facts

- **Applicant:** Wuhan Lotuxs Technology Co., Ltd.
- **Product Code:** [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX.md)
- **Decision Date:** Jan 30, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

DermRays Revive is intended for hair removal, permanent hair reduction. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. DermRays Revive is also intended for treatment of wrinkles. DermRays Revive is intended for use on all skin types (Fitzpatrick skin types I - VI), including tanned skin.

## Device Story

DermRays Revive is a diode laser surgical instrument (1064nm) for dermatology applications. Device utilizes selective photothermal effect to target melanin in hair follicles for hair removal and to treat wrinkles by acting on dermis. System comprises main unit, power adapter, and power cord; operated via fingerswitch. Used in clinical settings by professionals. Laser energy absorbed by melanin, hemoglobin, and water; destroys hair follicles in growth phase; treats wrinkles without damaging skin. Output parameters (energy density, pulse width) controlled by user to ensure safety and efficacy. Benefits include permanent hair reduction and wrinkle treatment.

## Clinical Evidence

No clinical data. Evidence based on non-clinical bench testing, including performance testing (appearance, function, safety, labeling), shelf-life (accelerated aging), and cleaning/disinfection verification. Biocompatibility testing performed per ISO 10993-1:2018 (cytotoxicity, sensitization, irritation). Electrical safety and EMC testing performed per IEC 60601-1, 60601-1-2, 60601-1-11, 60601-2-22, and IEC 60825-1.

## Technological Characteristics

Diode laser, 1064nm±10nm wavelength, solid-state. 15mm spot size, 1.766cm2 working area. Pulse width 400ms; energy density 5.0-10.0 J/cm2. Fingerswitch activation. Input AC100-240V, 50/60Hz. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-22, IEC 60825-1. Biocompatibility per ISO 10993-5, -10, -23.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Cynosure Elite+ Laser ([K141425](/device/K141425.md))
- Diode Laser Hair Removal ([K232117](/device/K232117.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 30, 2024

Wuhan Lotuxs Technology Co., Ltd. Na Wu, Quality Manager 501/E2, No.999 High-tech Avenue Wuhan, Hubei 430206 China

#### Re: K231910

Trade/Device Name: DermRays Revive Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: December 13, 2023 Received: December 13, 2023

Dear Na Wu:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).

{2}------------------------------------------------

Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Tanisha L. Hithe Tanisha L. s 2024.01.30 Hithe -S 14:57:30 -05'00'

Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use

510(k) Number (if known) K231910

Device Name DermRays Revive (LHR-S5-1064)

Indications for Use (Describe)

DermRays Revive is intended for hair reduction. Permanent hair reduction is defined as the longterm, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

DermRays Revive is also intended for treatment of wrinkles.

DermRays Revive is intended for use on all skin types (Fitzpatrick skin types I - VI), including tanned skin.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------------------|-----------------------------------------------------------|
|------------------------------------------------------------|-----------------------------------------------------------|

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# Section 5 - 510(k) Summary K231910

#### Date of Summary Preparation: June 25, 2023 Date of Summary Modification: January 09, 2024

### 1. Submitter's Identifications

Submitter's Name: Wuhan Lotuxs Technology Co., Ltd. Address: 501/E2, No.999 High-tech Avenue, Wuhan 430206, China Contact Person: Na Wu Contact Title: Quality Manager Contact Email Address: na.wu@lotuxs.com Telephone: +86-27-87619668

#### 2. Correspondent's Identifications

Correspondent's Name: Wuhan Lotuxs Technology Co., Ltd. Address: 501/E2, No.999 High-tech Avenue, Wuhan 430206, China ZIP Code: 430206 Contact Person: Na Wu Contact Title: Quality Manager Contact E-mail Address: na.wu(@)lotuxs.com Telephone: +86-27-87619668

#### 3. Name of the Device

Device Classification Name: Powered laser surgical instrument Product Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Trade Name: DermRays Revive Model: LHR-S5-1064 Classification Panel: General & Plastic Surgery Product Code: GEX Regulation Number: 21 CFR 878.4810 Device Classification: Class II

#### 4. The Predicate Devices

Primary predicate device: K141425 Cynosure Elite+ Laser Secondary predicate device: K232117 Diode Laser Hair Removal

#### 5. Device Description

DermRays Revive uses a diode laser with an output wavelength of 1064nm, to act on the dermis through the epidermis, and the laser energy is absorbed by the target tissues (melanin, hemoglobin, and water).

Due to the selective photothermal effect of the laser, it can treat wrinkles, without damaging the skin. Simutaneously, the laser can act on the melanin in the hair follicles, destroy hair follicles in the growth around the hair follicles, so as to remove hairs.

The device consists of a power adapter, a power cord and a main unit.

{5}------------------------------------------------

## 6. Intended Use of Device

DermRays Revive is intended for hair removal, permanent hair reduction. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. DermRays Revive is also intended for treatment of wrinkles.

DermRays Revive is intended for use on all skin types (Fitzpatrick skin types I - VI), including tanned skin.

#### 7. Summary of Substantial Equivalence

|                                        | Proposed device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Primary predicate<br>device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Secondary predicate<br>device                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Comparison |
|----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| 510k<br>Number                         | K231910                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | K141425                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | K232117                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | —          |
| Product<br>Code                        | GEX                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | GEX                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | GEX                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | —          |
| Proprietary<br>Name                    | DermRays Revive                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Cynosure Elite+ Laser                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Diode Laser Hair<br>Removal                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | —          |
| Model                                  | LHR-S5-1064                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | /                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | LHR-V6S-1064,<br>LHR-V6S/B-1064                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | —          |
| Manufacture                            | Wuhan Lotuxs<br>Technology Co., Ltd.                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Cynosure LLC                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Wuhan Lotuxs<br>Technology Co., Ltd.                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | —          |
| Indications<br>for use                 | DermRays Revive is<br>intended for hair<br>removal, permanent<br>hair reduction.<br>Permanent hair<br>reduction is defined as<br>the long-term, stable<br>reduction in the number<br>of hairs regrowing<br>when measured at 6, 9,<br>and 12 months after the<br>completion of a<br>treatment regime.<br>DermRays Revive is<br>also intended for<br>treatment of wrinkles.<br>DermRays Revive is<br>intended for use on all<br>skin types (Fitzpatrick<br>skin types I - VI),<br>including tanned skin. | 1064 nm:<br>The Cynosure Elite+<br>Laser is intended for<br>the coagulation and<br>hemostasis of benign<br>vascular lesions such<br>as, but not limited to,<br>port wine stains,<br>hemangiomas, warts,<br>telangectasia, rosacea,<br>venus lake, leg veins,<br>spider veins and<br>poikiloderma of civatte;<br>and treatment of benign<br>cutaneous lesions such<br>as warts, scars, striae<br>and psoriasis. The laser<br>is also intended for the<br>treatment of benign<br>pigmented lesions such<br>as, but not limited to,<br>lentigos (age spots),<br>solar lentigos (sun | Diode Laser Hair<br>Removal<br>(LHR-V6S-1064,<br>LHR-V6S/B-1064) is<br>intended for hair<br>removal, permanent<br>hair reduction.<br>Permanent hair<br>reduction is defined as<br>the long-term, stable<br>reduction in the number<br>of hairs regrowing<br>when measured at 6, 9,<br>and 12 months after the<br>completion of a<br>treatment regime.<br>Diode Laser Hair<br>Removal<br>(LHR-V6S-1064,<br>LHR-V6S/B-1064) is<br>intended for use on all<br>skin types (Fitzpatrick<br>skin types I - VI), | Same       |
|                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |            |
|                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | spots), cafe au lait<br>macules, seborrheic<br>keratoses, nevi,<br>chloasma, verrucae,<br>skin tags, keratosis and<br>plaques.<br>The laser is also<br>indicated for the<br>treatment of wrinkles<br>such as, but not limited<br>to, periocular and<br>perioral wrinkles.<br>Additionally, the laser<br>is indicated for the                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |            |
|                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | treatment of<br>pseudofolliculitis<br>barbae (PFB) and for<br>stable long-term, or<br>permanent hair<br>reduction. Permanent<br>hair reduction is<br>defined as long-term<br>stable reduction in the<br>number of hairs<br>regrowing when<br>measured at 6, 9 and 12<br>months after the<br>completion of a<br>treatment regime.                                                                                                                                                                                                                                                 | including tanned skin.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |            |
| Laser Type                             | Solid state                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Solid state                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Solid state                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Same       |
| Wavelength                             | 1064nm±10nm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 1064nm±10nm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 1064nm±10nm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Same       |
| Treatment<br>Activation                | Fingerswitch                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Fingerswitch                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Fingerswitch                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same       |
| Rx/OTC                                 | Prescription                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Prescription                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Prescription                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same       |
| Pulse Width                            | 400ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 0.1ms-300ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | LHR-V6S-1064:<br>350ms-450ms<br>LHR-V6S/B-1064:<br>280ms-400ms                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same1      |
| Single Pulse<br>Max. Energy<br>Density | $5.0J/cm^2$ - $10.0J/cm^2$                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | $4.7J/ cm^2$ (under 24mm<br>handpiece spot size)<br>$12.0J/cm^2$ (under 15mm<br>handpiece spot size)                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | LHR-V6S-1064:<br>$5J/cm^2$ , $6 J /cm^2$ , $7J/cm^2$<br>LHR-V6S/B-1064:<br>$4J/cm^2$ , $5J/cm^2$ , $6J/cm^2$                                                                                                                                                                                                                                                                                                                                                                                              | Similar2   |
|                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |            |
| Handpiece<br>Spot Size                 | 15mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 3mm, 5mm, 7mm,<br>10mm, 12mm, 15mm,<br>18mm, 20mm, 22mm<br>& 24mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | 30mm×10mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same3      |
| Working<br>Area                        | 1.766cm2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | 4.522cm2 (under 24mm<br>handpiece spot size)<br>1.766cm2 (under 15mm<br>handpiece spot size)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | 3cm2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same3      |
| Input<br>Voltage                       | AC100-240V,<br>50/60Hz, 1.6A Max.                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 208/240VAC, 30A,<br>50/60Hz, Single Phase                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | AC100-240V,<br>50/60Hz, 1.6A Max.                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Same4      |

Table 1

{6}------------------------------------------------

{7}------------------------------------------------

Note 1: Pulse Width of the proposed device is same as the secondary predicate device LHR-V6S/B-1064 of K232117.

Note 2: Maximum energy density of a single pulse of the proposed device is similar to those of the primary predicate device K141425 (1064nm) and secondary predicate device K232117. These slight differences do not affect to safety and effectiveness of the proposed device.

Note 3: Handpiece Spot Size and Working Area of the proposed device are same to those of the secondary predicate device K141425 (15mm, 1.766 cm-).

Note 4: Input Voltage of the proposed device is same as the secondary predicate device K232117.

Discussion for Substantially Equivalent (SE):

The proposed device DermRays Revive LHR-S5-1064 has the same indications for use, laser type, wavelength, treatment activation, prescription, pulse width, handpiece spot size and working area, input voltage; has similar single pulse max. energy density. These different items can be controlled within the scope of application. These slight differences between the proposed devices and predicate devices do not cause new safety and effectiveness problems. According to the non clinical test results and use evaluation report, the proposed device is as safe, effective and has good performance as the predicate device.

So the proposed device is Substantially Equivalent (SE) to the predicate device which is US legally market device.

#### 8. Non-Clinical Tests Submitted

The following non-clinical testing was provided in this 510(k) submission:

#### Biocompatibility Testing

{8}------------------------------------------------

According to ISO 10993-1:2018, we performed biocompatibility test for the parts that may directly contact the intact skin of the patient or operator, including in vitro cytotoxicity test per ISO 10993-5:2009, skin sensitization test per ISO 10993-10:2021, irritation test per ISO 10993-23:2021.

### Electrical Safety and Electromagnetic Compatibility Testing

The propoesd device was tested and complicable requirements of the following standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-22 and IEC 60825-1, demonstrated that the basic safety and performance of the device met the requirements.

### Software Verification and Validation

Software verification and validation was performed, and it was demonstrated that the software performs as intended according to the FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.

#### 9. Concise summary for performance testing

From the performance testing - bench, we have a brief summary as follow:

1. Performance Test: Appearance, Function, Safety, Label.

2. Shelf Life Test: Accelerated aging test, and performance test after accelerated aging test).

3. Cleaning and Disinfection Verification: Simulated cleaning and disinfection test, and performance test after simulated cleaning and disinfection test.

All the bench test results are provided in Performance Test Report.

#### 10. Clinical Study

No clinical testing has been performed.

#### 11. Conclusions

After analyzing both bench and external laboratory testing data, the intended use and supporting data can conclude that the device in the submission is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K231910](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K231910)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
